Hantek DSO2000 Series

Contents

Hantek DSO2000 Series Digital Storage Oscilloscope User Manual

Models: DSO2C10, DSO2C15, DSO2D10, DSO2D15

1. Introduction

The Hantek DSO2000 series digital storage oscilloscopes offer a cost-effective solution with bandwidths up to 150MHz and a sampling rate of 1GSa/s. These devices feature an 8M memory depth and, for some models (DSO2D10, DSO2D15), a built-in 1CH 25MHz waveform generator supporting arbitrary waveform output. They include 14 types of trigger modes, 5 kinds of serial protocol triggers and decodes, 32 types of auto measurements with statistics, a 3-digit digital voltmeter, and a 6-digit hardware frequency indicator. The 7-inch color TFT LCD screen with a Windows-style interface ensures ease of use for various applications in communication, aerospace, embedded systems, computers, research, and education.

For a more detailed and comprehensive guide, please refer to the official Hantek DSO2000 Series User Manual (PDF).

Video: Hantek DSO2D15 product overview and demonstration of key features.

2. General Safety Summary

To prevent injury and damage to the product, use it only as specified. Maintenance should only be performed by qualified personnel.

2.1. Avoid Fire or Personal Injury

  • Use suitable power cord: Use only the power cord specified for this product and certified for your country of use.
  • Connect and disconnect properly: Connect a probe to the oscilloscope before connecting to measured circuits; disconnect the probe from the oscilloscope after disconnecting from measured circuits.
  • Ground the product: The product is grounded through the grounding conductor of the power cord. Ensure proper grounding before connecting to input/output terminals.
  • Connect the probe in a right way: The probe ground lead is at ground potential. Do not connect it to an elevated voltage.
  • Check all terminal ratings: Refer to the product manual for detailed information on ratings to avoid fire or shock hazards.
  • Do not operate without covers: Do not operate with covers or panels removed.
  • Avoid exposed circuitry: Do not touch exposed connections and components when power is present.
  • Do not operate with suspected failures: If damage is suspected, have it inspected by qualified service personnel.
  • Assure good ventilation: Ensure adequate ventilation during operation.
  • Do not operate in wet/damp environments: Avoid operating in high humidity conditions.
  • Do not operate in an explosive atmosphere: Do not use the device in potentially explosive environments.
  • Keep product surfaces clean and dry: Maintain cleanliness and dryness of the device.

2.2. Safety Terms and Symbols

The following terms and symbols may appear on the product or in this manual:

  • Danger: Indicates immediate harm if the operation is performed.
  • Warning: Indicates potential harm if the operation is performed.
  • Notice: Indicates possible damage to the product or other properties.
Safety Symbols: Danger, Protective Earth, Test Ground, Shell Ground Terminal

Image: Safety symbols including Danger, Protective Earth, Test Ground, and Shell Ground Terminal.

3. Setup

3.1. General Inspection

  1. Check the shipping container for damage: Retain the container and cushioning material until the contents are verified.
  2. Check the accessories: Verify all accessories listed in 'Appendix B: Accessories' are present and undamaged. Notify the seller if anything is missing or damaged.
  3. Check the instrument: Inspect for any mechanical damage or defects. If the instrument does not operate properly, notify the seller.

3.2. Prepare Instrument for Use

  1. Adjust the Supporting Legs: Tilt the oscilloscope upwards for stable placement and better observation.
  2. Connect the Power Cord: The oscilloscope accepts 100-240 V, 45-440 Hz AC power. Use the supplied power cord.
  3. Power Switch: Turn on the instrument by pressing the power switch in the lower left corner of the front panel. Ensure the power cord is firmly connected to an energized source.
Hantek DSO2D15 Oscilloscope with included accessories: USB cable, probe, signal output lines, power line, certificate.

Image: Hantek DSO2D15 Oscilloscope with included accessories.

3.3. Functional Check

Perform a quick functional check:

  1. Connect the oscilloscope: Set the probe switch to X10. Align the slot on the probe connector with the protuberance on the CH1 BNC and push to connect, then turn to lock. Connect the probe tip and reference lead to the PROBE COMP connectors (~5V@1KHz).
  2. Observe the waveform: Press the [Auto Set] button. A square wave of approximately 5V peak-to-peak at 1KHz should appear on the display.
Diagram showing how to connect the probe to the oscilloscope and the probe compensation terminals.

Image: Probe connection diagram.

3.4. Probe Introduction

  1. Safety: Keep fingers behind the guard on the probe body. Do not touch metallic portions of the probe head when connected to a voltage source. Connect the probe to the oscilloscope and ground terminal to ground before measurements.
  2. Manual Probe Compensation: Adjust the probe compensation to match the input channel. Uncompensated probes can lead to measurement errors. Use a nonmetallic screwdriver to adjust the variable capacity of your probe until the waveform is correctly compensated.
Waveforms showing correctly compensated, overcompensated, and undercompensated probe signals.

Image: Probe compensation waveforms (correctly, over, and under compensated).

  1. Probe Attenuation Setting: Set the Probe option attenuation in the channel menu to 10X and the switch on the probe to 10X for full bandwidth. When set to 1X, bandwidth is reduced to 6MHz.
Diagram of a probe with an attenuation switch.

Image: Probe attenuation switch.

4. Operating Instructions

4.1. Front Panel Overview

The front panel features various controls and connectors for operating the oscilloscope.

Hantek DSO2D15 Digital Storage Oscilloscope front panel with labeled controls.

Image: Hantek DSO2D15 Digital Storage Oscilloscope front panel.

4.2. User Interface

The user interface displays waveforms, measurement parameters, and menu options. Key elements include:

  • Hantek logo: Brand identification.
  • Trigger Status: Indicates current acquisition mode (AUTO, READY, ROLL, STOP, ARM1/ARM).
  • Main time base: Current window's time base.
  • Sample rate: Current sampling rate.
  • Storage depth: Current memory depth.
  • Horizontal Trigger time: Time related to horizontal trigger.
  • Operating Menu: Shows different information for respective function keys.
  • USB icon: Lights up when a USB disk is connected.
  • Signal source information: (For models with function generator)
  • Channel information: Coupling, bandwidth, and volt/div for CH1-CH2.
  • Channel Marker: Indicates channel position.
  • Trigger level: Current trigger level.
Hantek DSO2D15 user interface showing waveforms and various display indicators.

Image: Hantek DSO2D15 user interface display.

4.3. Menu and Control Keys

The oscilloscope features various menu and control keys for comprehensive operation:

  • Menu Keys: [SAVE/RECALL] for saving/recalling files, [MEASURE] for waveform parameters, [ACQUIRE] for acquisition settings, [UTILITY] for system functions, [CURSOR] for cursor measurements, [DISPLAY] for display parameters.
  • Operation Mode Keys: [AUTO SET] for automatic setup, [RUN/STOP] for continuous/stopped acquisition, [SINGLE SEQ] for single trigger acquisition.
  • Shortcut Keys: [DEFAULT SETUP] for factory settings, [HELP] for help messages, [SAVE TO USB] for quick screenshot save, [DECODE] for protocol decoding.
  • Vertical Control System: [CH1 MENU], [CH2 MENU] for channel parameters, [MATH MENU] for math operations, [POSITION] knob for vertical offset, [VOLTS/DIV] knob for vertical scale.
  • Horizontal Control System: [HORIZ MENU] for display mode, [POSITION] knob for horizontal offset, [SEC/DIV] knob for horizontal time base.
  • Trigger Control System: [TRIG MENU] for trigger parameters, [FORCE TRIG] to stabilize waveforms.
  • Signal Source: [EXT TRIG/WAVE GEN] for signal source parameters/external trigger, [BURST/GEN TRIG] for waveform burst output.

4.4. Connectors

The front panel includes BNC connectors for CH1 and CH2 input, and an EXT TRIG/GEN OUT connector for external trigger input or signal generator output.

4.5. Oscilloscope Setup

  • Auto Set: Automatically adjusts horizontal and vertical scales, trigger coupling, type, position, slope, level, and mode for a stable waveform display.
  • Saving a Setup: The oscilloscope saves the current setup before closing. You can save up to 10 settings internally.
  • Recalling a Setup: Recall any saved setup or the default factory setup.
  • Default Setup: Resets the oscilloscope to its factory default settings.

4.6. Horizontal Controls

Use the horizontal controls to change the horizontal scale and position of the waveform. The horizontal position readout shows the time represented by the center of the screen, using the trigger time as zero.

  • Horizontal Offset Knob: Controls the trigger position against the screen center.
  • SEC/DIV Knob: Changes the horizontal time scale to magnify or compress the waveform.
  • HORIZ MENU: Accesses display modes like XY mode (for Lissajous patterns) and Roll mode (for slow-moving waveforms).

4.7. Vertical System

Vertical controls adjust waveform display, vertical scale, position, input parameters, and math calculations.

  • Vertical Offset Knob: Moves the channel waveform up and down.
  • VOLTS/DIV Knob: Controls the oscilloscope to magnify or attenuate the source signal.
  • CH1/CH2 Menu: Displays vertical menu options for coupling (DC, AC, GND), bandwidth (OFF, 20MHz), division (Coarse, Fine), probe (1X, 10X, 100X, 1000X), and invert (OFF, ON).

4.8. Math Operation

The oscilloscope supports various math operations on analog channel waveforms, including addition (+), subtraction (-), multiplication (*), division (/) and FFT.

  • Addition or Subtraction: Performs arithmetic operations point by point on two analog input channels.
  • FFT Operation: Computes the Fast Fourier Transform to convert time domain waveforms to the frequency domain. Used to measure harmonic components, noise characteristics, and analyze vibration.

4.9. Trigger System

The trigger determines when the oscilloscope begins to acquire and display data.

  • Trigger Source: Can be CH1, CH2, or an external trigger channel.
  • Trigger Mode: Auto (acquires freely), Normal (updates only on valid trigger), Single SEQ (acquires one trigger then stops).
  • Trigger Position: Establishes time between trigger position and screen center.
  • Trigger Level: Sets amplitude level for acquisition.
  • Force Trigger: Completes acquisition regardless of trigger signal.
  • Holdoff: Sets time the oscilloscope waits before the next trigger for stable display of complex waveforms.

Trigger Types:

  • Edge Trigger: Triggers on rising, falling, or both edges at a specified level.
  • Pulse Trigger: Triggers on positive or negative pulses of a specified width.
  • Video Trigger: Captures standard analog video signals (NTSC, PAL).
  • Slope Trigger: Triggers on rising or falling transitions between two specified levels within a time range.
  • Overtime Trigger: Triggers when the time interval between edges exceeds a set timeout.
  • Window Trigger: Triggers when the input signal passes through a high or low trigger level.
  • Pattern Trigger: Triggers based on a logical AND/OR combination of channel states (high, low, don't care).
  • Interval Trigger: Triggers when the time difference between neighboring edges meets a time limit.
  • Under Amp Trigger: Triggers on pulses that cross one threshold but not another.
  • UART Trigger: Triggers on UART serial data, with settings for idle level, baud rate, parity, data bits, and trigger conditions (start, stop, data, parity error, COM error).
  • LIN Trigger: Triggers on LIN bus signals, including Sync Break exit, Frame ID, or Frame ID and Data.
  • CAN Trigger: Triggers on CAN bus signals, with settings for baud rate, idle level, identifier, and trigger conditions (start bit, remote frame ID, data frame ID, error frame, etc.).
  • SPI Trigger: Triggers on SPI serial data, with settings for SCL/SDA sources, slope, data width, and timeout.
  • IIC Trigger: Triggers on IIC bus signals, with settings for SCL/SDA sources, signal threshold, and trigger conditions (start bit, stop bit, no ack, address, restart, address and data).

4.10. Protocol Decode

The oscilloscope supports decoding for various serial protocols, displaying data in hexadecimal and providing a visual representation on the waveform.

  • UART Decode: Decodes UART serial communication.
  • LIN Decode: Decodes LIN bus communication, showing Frame ID, Data, and Checksum.
  • CAN Decode: Decodes CAN bus communication, showing Frame ID, Type, DLE, Data, and CRC.
  • SPI Decode: Decodes SPI serial communication.
  • IIC Decode: Decodes IIC bus communication, showing Address, Data, and Acknowledge status.

4.11. Save/Recall

Save and recall oscilloscope setups, waveforms, and reference waveforms.

  • Internal Save and Recall: Save up to 9 setup files, waveforms, or reference waveforms to internal memory.
  • External Save and Recall: Save to a USB storage device. Supports saving setups, waveforms, reference waveforms, CSV files, and BMP pictures.
  • Save Picture: Save the oscilloscope display as a BMP picture to a USB device.
  • File Manager: Create, delete, and rename files/folders on external storage.

4.12. Measure System

The oscilloscope offers several methods for taking measurements:

  • Scale Measurement (Graticule): Visual estimation using graticule divisions and scale factor.
  • Cursor Measurement: Use horizontal (amplitude) and vertical (time) cursors to measure parameters. Includes Manual and Tracking modes.
  • Automatic Measurement: Performs automatic calculations for 32 types of measurements (e.g., frequency, period, peak-to-peak, RMS, rise time, fall time, etc.). Up to 4 measurements can be displayed simultaneously.
  • Statistic Function: Displays current, average, minimum, maximum, RMS, and count values for measurements.
  • Gate Measurement: Measures waveforms between two cursors.
  • DVM: Supports 3-digit voltage and 6-bit frequency measurement for any analog channel.
Hantek DSO2D15 display showing 32 kinds of auto measurements with statistics.

Image: Display showing 32 kinds of auto measurements with statistics.

4.13. Acquire

Control waveform acquisition:

  • Run Control: Use [Run/Stop] to start/stop acquisition or [Single] to capture a single acquisition.
  • Acquisition Modes: Normal (samples evenly), Peak Detect (captures max/min values to detect glitches), Average (reduces random noise), High Resolution (smoother waveform).

4.14. Display

Configure waveform display settings:

  • Waveform Type: Vectors (fills space between points) or Dots (displays sample points).
  • Intensity: Adjustable waveform brightness.
  • Grid: Dotted line or Real line, with adjustable intensity.
  • Persist: Sets the time length for displayed sample points (1s, 5s, 10s, 30s, infinite).

4.15. Utility System

Access system utilities:

  • Language: Set the display language.
  • Sound: Set the buzzer.
  • Update Firmware: Upgrade software via USB flash disk.
  • Self Calibration: Optimizes signal path for accuracy. Perform if ambient temperature changes by >5°C.
  • Pass/Fail: Judges if input signal is within a defined rule range.

4.16. Dual-window Mode

Press the Sec/DIV knob to enter or exit dual-window mode.

4.17. Waveform Generator (DSO2D10/DSO2D15 only)

The waveform generator can output various waveforms.

  • Set Wave Type and Parameters: Select waveform type (Sine, Ramp, Square, Exponent, Noise, DC, Arb), frequency, amplitude, offset, duty cycle, symmetry, and output impedance.
  • Waveform Modulation Setting: Supports Amplitude Modulation (AM) and Frequency Modulation (FM) with adjustable modulation waveform, frequency, and depth/deviation.
  • Burst Setting: Output a burst with a specified number of cycles (1-1024).
  • Edit Arbitrary Waveform: Use the WaveEditor software to create and edit arbitrary waveforms.
  • Output Arbitrary Waveform: Download edited waveforms to the device for output from the GEN OUT BNC port.

4.18. Remote Control

Connect the oscilloscope to a computer via USB (Type-A to Type-B cable) for remote control. Install the Keysight IO Libraries Suite driver for communication. Use Keysight Connection Expert to verify the connection and send SCPI commands.

5. Maintenance

5.1. General Care

Avoid exposing the LCD display to direct sunlight for long periods. Do not expose the oscilloscope or probes to sprays, liquids, or solvents.

5.2. Cleaning

Regularly examine the oscilloscope and probes. To clean:

  1. Use a lint-free cloth to remove dust from the exterior. Avoid scratching the display filter.
  2. Dampen a soft cloth with water (or 75% isopropyl alcohol for more efficient cleaning) to clean the oscilloscope.

Note: Do not use corrosive or chemical cleaning agents.

6. Troubleshooting

6.1. If the oscilloscope does not start up at power on:

  1. Check the power cord connection.
  2. Verify the power on/off button is pushed.
  3. Restart the oscilloscope.

If the issue persists, contact HANTEK Technical Support.

6.2. If there is no display of waveform on the screen:

  1. Check the probe's proper connection to the input BNC.
  2. Ensure the channel switch (CH1~CH2 menu buttons) is turned on.
  3. Verify the input signal is correctly connected to the probe.
  4. Confirm that all measured circuits have signals to output.
  5. Adjust the magnitude for DC signals with large magnitude.
  6. Try pressing the Auto Measure button for automatic signal detection.

If the issue persists, contact HANTEK Technical Support.

6.3. If the waveform of the input signal is distorted seriously:

  1. Check the probe's proper connection to the channel BNC.
  2. Check the probe's proper connection to the measured object.
  3. Verify the probe has been well calibrated.

6.4. If the waveform is rolling continuously but cannot be triggered:

  1. Check that the trigger source is consistent with the input channel.
  2. Verify the trigger level is correctly adjusted. Push the TRIGGER LEVEL knob to reset it to the center.
  3. Confirm the trigger mode is appropriate for the input signal. The default is edge trigger, which may not suit all signals.

7. Specifications

All technical specifications are applicable to DSO2000 series digital storage oscilloscopes. For guaranteed specifications, the oscilloscope must have been operating for twenty minutes under specified operating temperature, and self-calibration must be performed if the temperature changes by more than 5°C. The oscilloscope must also be within its factory calibration interval.

7.1. Model Selection Table

Table showing DSO2000 Series models with channels, bandwidth, sample rate, memory depth, and AWG availability.

Image: DSO2000 Series - Selection Table.

7.2. Oscilloscope Parameters

ParameterDSO2D15DSO2D10DSO2C15DSO2C10
Bandwidth150MHz100MHz150MHz100MHz
Oscilloscope channels2CH2CH2CH2CH
Waveform generator1CH1CH--
Sample rate1GSa/s (single channel) 500MSa/s (two channels)
Maximum record lengthSingle channel maximum 8M, Two channels maximum 4M
Input couplingDC, AC, GND
Input impedance1MΩ±2% ||20pF±3pF
Probe attenuation factor1X, 10X, 100X, 1000X
Voltage rating300V CAT II
Maximum input voltage300VRMS (10X)
Horizontal scale range2ns/div~100s/div 1, 2, 5 step by step
Time base modeY-T, X-Y, Roll
Vertical resolution8 bits resolution, each channel samples simultaneously
Vertical sensitivity2mV/div to 10V/div
Mathematical operation+, -, ×, ÷, FFT
Bandwidth Limit20MHz
Trigger typeEdge, Pulse width, Video, Slope, Overtime, Window, Pattern, Interval, Under Amp, UART, LIN, CAN, SPI, IIC
Trigger level range±5 divisions from the center of the screen
Trigger modeAuto, Normal, single

7.3. Signal Generator Parameters (DSO2D10/DSO2D15 only)

ParameterSpecification
Channel1
Sample rate200MSa/s
Vertical resolution12 bits
Maximum frequency25 MHz
Standard waveformssine, square, ramp, Exp, noise, DC
Arbitrary waveformArb1, Arb2, Arb3, Arb4
Frequency range (Sin)0.1Hz~25MHz
Frequency range (Square/pulse)0.1Hz~10MHz
Frequency range (Triangular wave)0.1Hz~1MHz
Frequency range (Sampling wave)0.1Hz~1MHz
Frequency range (Index)0.1Hz~5MHz
Frequency range (Noise)0.1Hz to 10 MHz
Waveform length4KSa

7.4. General Technical Specifications

ParameterSpecification
Display type7" diagonal TFT liquid crystal
Display resolution800 (horizontal)*480 (vertical) pixels
Display color16 million colors (24 bits true color)
Persistence timeminimum, 1 s, 5 s, 10 s, 30 s, infinite
InterfaceUSB Host, USB Device
Supply Voltage100-120VACRMS(±10%), 45Hz to 440Hz, CAT II; 120-240VACRMS(±10%), 45Hz to 66Hz, CAT II
Power Consumption<15W
Operating Temperature0 to 50 °C (32 to 122 °F)
Storage Temperature-40 to +71 °C (-40 to 159.8 °F)
Humidity≤+40°C: ≤90% relative humidity; +41°C~50°C: ≤60% relative humidity
Altitude3,000m (10,000 feet)
Dimension (L x W x H)318 x 110 x 150mm
Weight1900g

8. Warranty and Support

Thank you for choosing HANTEK. For any inquiries regarding our products, please contact us through the following channels:

  • Your local HANTEK distributor.
  • Your local HANTEK field office.
  • HANTEK headquarters in China.

8.1. HANTEK Headquarters Contact Information:

Qingdao Hantek Electronic Co., Ltd
Website: http://www.hantek.com
Address: 2/F., Zone D2, No. 112 Keyuan Longitude 7th Road, Qingdao City, Shandong Province, China 266101
Tel: +86-532-88703687 / 88703697
Fax: +86-532-88705691
Email: service@hantek.com

8.2. Technical Support:

Tel: +86-532-88703687
Email: support@hantek.com

9. User Tips

When purchasing, ensure you select the correct plug type for your region. The available plug types are US, EU, UK, and AU.