1. Safety Information
Please read this manual carefully before using the instrument and save it for future reference. Pay special attention to all safety warnings and instructions to prevent electric shock, personal injury, or damage to the instrument.
1.1. General Safety Warnings
- Warning: Conditions and operations that may cause danger to users.
- Caution: Conditions and operations that may cause damage to the instrument or equipment.
- Always read the manual carefully before use.
- Strictly follow the operation instructions; otherwise, the protection function of the instrument may be damaged or weakened.
- Exercise caution when measuring voltages above 30V AC (True RMS), 42V AC peak, or 60V DC, as there is a risk of electric shock.
- Ensure the voltage applied between terminals or between each terminal and the grounding point does not exceed the rated value.
- Before each use, verify the meter's functionality by measuring a known voltage. Do not use if it is not working normally or is damaged.
- Inspect the instrument shell and plastic parts for cracks or damage before use. Do not use if damaged.
- Check test probes for cracks or damage. Replace with probes of the same model and electrical specifications if damaged.
- Do not exceed the lowest rated Category of Measurement (CAT) for products, probes, or accessories.
- Do not measure current when a probe is inserted into the input jack.
- Do not work alone when performing electrical measurements.
- Comply with local and national safety codes. Wear appropriate personal protective equipment (e.g., approved rubber gloves, masks, flame-retardant clothing) to prevent electric shock and arc flash.
- Replace batteries promptly when the low battery indicator is shown to avoid measurement errors.
- Do not use the instrument in explosive gas environments or wet conditions.
- When using probes, keep fingers behind the finger protector.
- When measuring, connect the neutral or ground wire first, then the live wire. When disconnecting, disconnect the live wire first, then the neutral/ground wire.
- Before opening the outer cabinet or battery cover, remove all probes from the instrument.
1.2. Safety Symbols
| Symbol | Description |
|---|---|
| High voltage warning | |
| ~ | AC (Alternating current) |
| --- | DC (Direct current) |
| ~--- | AC or DC |
| Warning, important safety signs | |
| Ground | |
| Π | Equipment with double insulation/reinforced insulation protection |
| Low battery | |
| CE | Product complies with all relevant European laws |
| INRUSH | Inrush current measurement |
| The additional product label shows that do not discard this electrical/electronic product into household garbage |
2. Overview
The GVDA Digital Clamp Meter is a new generation of true RMS high-performance intelligent digital clamp meters. It combines a wide range of functions to make electrical work easier, more efficient, and safer. It is capable of measuring AC/DC voltage, AC/DC current, inrush current, frequency, duty cycle, resistance, capacitance, temperature, diode, continuity, NCV (Non-Contact Voltage), and live wire detection.
2.1. Models Available
The GVDA Digital Clamp Meter is available in three models:
- GD166A: AC Clamp Meter (AC current measurement only)
- GD166B: AC/DC Clamp Meter (measures both AC and DC current)
- GD168B: AC/DC Clamp Meter (measures both AC and DC current)

2.2. Instrument Panel Description (GD168B Example)

- NCV probe: For non-contact voltage detection.
- Clamp: For measuring current without breaking the circuit.
- Flashlight: For illuminating the work area.
- Alarm Indicator LED: Provides visual alerts.
- Trigger: Opens the clamp jaws.
- Power button: Turns the device on/off.
- FUNC / AUTO button: Switches between functions or activates auto mode.
- MAX / MIN button: Records maximum and minimum measured values.
- SEL / INRUSH button: Selects measurement modes or activates inrush current measurement.
- Display: HD backlight color LCD for showing measurement results.
- Black meter pen socket (COM): Common input for test leads.
- Red meter pen socket (INPUT): Input for test leads.
2.3. Key Features
- HD Colour LCD Display
- Maximum 6000 Counts display
- Non-contact AC voltage measurement (NCV)
- AC Inrush current measurement
- MAX/MIN value measurement
- Live wire detection
- Auto power off
- Low voltage display: low battery indication
- Over range indication
- DCA zero (only GD166B, GD168B)
- True RMS measurement
- Diode measurement
- Continuity test
- Flashlight
- Data hold
3. Setup
3.1. Package Contents
Before starting, ensure all items are present in the package:

- Digital Clamp Meter
- Test Leads (Original pen)
- Temperature Probe
- Cloth Bag
- User Manual
- Gift Box
3.2. Battery Installation
The clamp meter requires 3x 1.5V AAA batteries (not included).
- Locate the battery compartment on the back of the device.
- Use a screwdriver to open the battery cover.
- Insert three AAA batteries, ensuring correct polarity (+/-).
- Close the battery cover and secure it with the screw.

3.3. Connecting Test Leads
The clamp meter features a design for proper placement of test leads.

- Insert the black test lead into the 'COM' (Common) jack.
- Insert the red test lead into the 'INPUT' jack.
- When not in use, or when using only one probe, the other probe can be stored in the integrated pen rack on the clamp meter.
4. Operating Instructions
4.1. Power On/Off
- To turn on: Press and hold the power button (
) for approximately 2 seconds. - The meter will perform a self-test, displaying "CAL". Do not press the trigger during this time.
- A series of "beep, beep, beep" sounds indicates the self-test is complete, and the meter is ready for measurement.
- To turn off: Press and hold the power button for approximately 2 seconds.
4.2. Smart Measurement Mode (AUTO)
The meter defaults to Smart Mode upon power-on. In this mode, it automatically identifies and measures AC/DC voltage, AC/DC current, resistance, and continuity.

- Turn on the meter. After the self-test, "Auto" will be displayed.
- For current measurement, use the clamp. For other measurements (voltage, resistance, continuity), use the test leads connected to the pen input jacks.
- Read the measurement results directly from the display screen.
4.3. Professional Measurement Mode
To access specific measurement functions, press the FUNC / AUTO button to cycle through available modes.
4.3.1. AC/DC Current Measurement
This function is available on GD166B and GD168B models. GD166A measures AC current only.

- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "A" (Current).
- Press the SEL / INRUSH button to select AC or DC current measurement.
- For DC current, if the display is not zero when not measuring, press and hold the ZERO button to clear it.
- Press the trigger to open the clamp, place it around a single conductor (live line or zero line, NOT both), and slowly release the trigger until the clamp is completely closed. Ensure the conductor is centered in the clamp jaws for accurate readings.
- Read the measurement results. Frequency is also shown for AC current.
4.3.2. AC Inrush Current Measurement

- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "A" (Current).
- Press the SEL / INRUSH button to select inrush current measurement, indicated by the "INRUSH" symbol.
- Press the trigger to open the clamp, place it around a single conductor, and slowly release.
- Turn on the device to be tested (e.g., a motor) to trigger the inrush current measurement.
- Read the measurement results from the display.
4.3.3. AC/DC Voltage Measurement


- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "V" (Voltage).
- Press the SEL / INRUSH button to select AC or DC voltage measurement.
- Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Contact the probes to the measured circuit in parallel.
4.3.4. Resistance Measurement

- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "Ω" (Resistance).
- Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Contact the probes across the component to measure resistance.
4.3.5. Capacitance Measurement

- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "F" (Capacitance).
- Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Contact the probes across the capacitor to measure capacitance.
4.3.6. Frequency/Duty Cycle Measurement
- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "Hz/%" (Frequency/Duty Cycle).
- Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Contact the probes to the circuit to measure frequency or duty cycle.
4.3.7. Temperature Measurement

- Turn on the meter.
- Press the FUNC / AUTO button until the pointer is at "°C/°F" (Temperature).
- Insert the temperature probe into the input jacks.
- Place the temperature probe on or near the object to be measured.
- The display will show the temperature in Celsius or Fahrenheit.
4.3.8. Continuity Test

- Turn on the meter.
- Press the FUNC / AUTO button until the continuity symbol (
) is displayed. - Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Touch the probes to the two points of the circuit or component to be tested. A continuous beep indicates continuity.
4.3.9. Diode Test

- Turn on the meter.
- Press the FUNC / AUTO button until the diode symbol (
) is displayed. - Insert the red probe into the "INPUT" jack and the black probe into the "COM" jack.
- Touch the probes across the diode. The display will show the forward voltage drop. Reverse the probes to check for open circuit.
4.3.10. NCV (Non-Contact Voltage) Measurement

- Turn on the meter.
- Press the FUNC / AUTO button until "NCV" is displayed.
- Bring the NCV probe (top part of the clamp meter) close to the suspected live wire or electrical field.
- The meter will beep and the LED indicator will light up. A slow beep and green LED indicate a weak signal ("--L"), while a fast beep and red LED indicate a strong signal ("--H").
4.3.11. Live Wire Detection

- Turn on the meter.
- Press the FUNC / AUTO button until "Live" is displayed.
- Use the red test lead or the NCV probe to touch the suspected live wire.
- Similar to NCV, the meter will indicate the presence of a live wire with beeps and LED lights.
4.4. Data Hold
Press the H button to freeze the current reading on the display. Press it again to release.
4.5. MAX/MIN Value Measurement
Press the MAX/MIN button to activate maximum and minimum value measurement. The meter will record the highest and lowest readings. Press again to cycle between MAX and MIN values. Press and hold for more than 2 seconds to exit this mode.
4.6. Flashlight

Press and hold the flashlight button for more than 2 seconds to turn the flashlight on or off. This is useful for working in dimly lit conditions.
4.7. Auto Power Off (APO)
The meter will automatically shut down after approximately 15 minutes of inactivity to conserve battery life.
4.7.1. Cancelling Auto Power Off
- While the meter is off, press and hold the FUNC button.
- While holding FUNC, press the power button to turn on the meter.
- Release the FUNC button. The display will show a symbol (e.g.,
) indicating that APO is disabled. - To re-enable APO, restart the meter.
5. Specifications
5.1. General Specifications
- Display: HD Colour LCD, 6000 counts
- MAX. Voltage between terminals and earth ground: 600V
- Over range indication: “OL”
- Low battery indication: Yes
- Input polarity indication: Display “-”
- True RMS: Yes
- Clamp opening size: 36mm
- Power requirement: 3x 1.5V AAA batteries (not included)
- Product size: 199*81*32mm (GD166A/GD166B), 199*76*32mm (GD168B)
- Weight: 0.45KG
- Operating Temperature: 0℃~40℃
- Certifications: CE, FCC
- Safety Rating: EN61010-1 CAT III 600V


5.2. Function Comparison Table
| Function | Range | Accuracy (GD166A) | Accuracy (GD166B) | Accuracy (GD168B) |
|---|---|---|---|---|
| DC Voltage | 600mV/6V/60V/600V | ±(0.5%+5) | ±(0.5%+5) | ±(0.5%+5) |
| AC Voltage | 600mV/6V/60V/600V | ±(0.8%+5) | ±(0.8%+5) | ±(0.8%+5) |
| DC Current | 60A/600A | N/A | ±(2.5%+5) | ±(2.5%+5) |
| AC Current | 60A/600A | ±(2.5%+5) | ±(2.5%+5) | ±(2.5%+5) |
| Resistance | 600Ω/6KΩ/60KΩ/600KΩ/6MΩ/60MΩ | ±(1.0%+5) | ±(1.0%+5) | ±(1.0%+5) |
| Capacitance | 6nF/60nF/600nF/6µF/60µF/600µF/6mF/60mF | ±(4.0%+5) | ±(4.0%+5) | ±(4.0%+5) |
| Frequency | 100Hz/1KHz/10KHz/100KHz/1000KHz/10MHz | ±(1.0%+3) | ±(1.0%+3) | ±(1.0%+3) |
| Duty | 1%~99% | ±(3.0%+3) | ±(3.0%+3) | ±(3.0%+3) |
| Temperature (°C/°F) | -40℃~1000℃ / -40F~1832F | ±(1.0%+3) | ±(1.0%+3) | ±(1.0%+3) |
| Continuity | 600Ω | ✔ | ✔ | ✔ |
| Diode | ✔ | ✔ | ✔ | |
| NCV | ✔ | ✔ | ✔ | |
| Live | ✔ | ✔ | ✔ | |
| Counts | 6000 | 6000 | 6000 | 6000 |
| Analog bar | ✔ | ✔ | ✔ | |
| Data hold | ✔ | ✔ | ✔ | |
| Max/Min | ✔ | ✔ | ✔ | |
| Backlight | ✔ | ✔ | ✔ | |
| Flashlight | ✔ | ✔ | ✔ | |
| DCA ZERO | N/A | ✔ | ✔ | |
| SMART FUNCTION (AUTO) | ✔ | ✔ | ✔ | |
| Low battery indication | ✔ | ✔ | ✔ | |
| Auto power off | ✔ | ✔ | ✔ | |
| True RMS | ✔ | ✔ | ✔ | |
| Power | 3xAAA (1.5V) batteries (not include) | ✔ | ✔ | ✔ |
| Safety Rating | EN61010-1 CAT III 600V | ✔ | ✔ | ✔ |
6. Maintenance
6.1. Cleaning
Wipe the instrument with a dry, soft cloth. Do not use abrasive cleaners or solvents.
6.2. Battery Replacement
Refer to Section 3.2 for battery installation instructions. Replace batteries promptly when the low battery indicator appears on the display.
6.3. Probe Replacement
If test probes are damaged, replace them with new ones of the same model number and electrical specifications to ensure safety and accuracy.
7. Troubleshooting
- Instrument not working or damaged: Do not use the instrument if it is not working normally or appears damaged.
- Low battery indicator: Replace the batteries immediately to avoid measurement errors.
- Inaccurate current measurement: Ensure the conductor is centered within the clamp jaws when measuring current. Off-center clamping can lead to errors.
- DC current display not zero: If the DC current display shows a non-zero value when no measurement is being taken, press and hold the ZERO button to clear it.
- Continuity test fluctuations: For stable continuity readings, ensure firm contact between the test probes and the circuit points. Loose contact can cause fluctuating resistance values and intermittent beeping.
8. User Tips
- The clamp meter offers intuitive operation and good build quality.
- While the auto mode is convenient, it might take a bit longer to settle on a reading compared to manual mode.
- The NCV sensitivity is well-balanced, providing reliable detection without being overly sensitive to distant fields.
- The device includes a sturdy case, which helps prevent accidental power-on when stored in a bag.
- The calibration feature is a valuable addition for maintaining accuracy over time.
- Remember that batteries (3x AAA) are not included in the package.
9. Warranty and Support
For detailed item guides and documents, please refer to the official user manual PDF: User Manual (PDF).
GVDA is committed to providing reliable and accurate measurement tools. For further assistance or inquiries, please contact GVDA customer support.
10. Product Demonstration Video
Watch this video for a visual demonstration of the GVDA Digital Clamp Meter's features and operation: