ASAIR AS200

Contents

ASAIR AS200 Series MEMS Mass Flow Controller Instruction Manual

Model: AS200 | Brand: ASAIR

1. Introduction

The ASAIR AS200 series gas mass flow controllers are designed for precise control and measurement of gas mass flow. Utilizing advanced MEMS technology, these controllers offer high stability, high accuracy, low leakage rates, and high-pressure resistance. They support both digital and analog signals, making them suitable for various working scenarios.

The AS200 series plays a crucial role in scientific research and production across diverse fields. Typical applications include:

  • Photovoltaics: For processes requiring precise gas flow control.
  • Semiconductor Integrated Circuit Technology: Used in integrated circuit process equipment such as epitaxy, diffusion, plasma etching, sputtering, ion implantation, and various CVD equipment.
  • Petrochemical Industry: For gas mixing and distribution systems.
  • Medicine: In various medical applications requiring controlled gas flow.
  • Environmental Protection: For gas sampling devices and analytical instruments like gas chromatographs.
  • Vacuum Technology: In vacuum systems where accurate gas flow is critical.

The product also comes with powerful free client host software to facilitate user debugging operations.

Video: Overview of ASAIR AS200 Mass Flow Controller features and applications.

2. Safety Precautions

Failure to follow these precautions may result in property loss or personal injury. The manufacturer is not responsible for any consequences arising from such failures.

  1. Do not replace or modify the product or parts: Do not replace any parts or make unauthorized modifications. Safety labels must remain intact.
  2. Contact professional staff for service: Any technical support or part replacement must be provided by professional technicians authorized by the company.
  3. Special attention when using hazardous gases: If using hazardous gases, take precautions and completely purge the product if conditions permit. If the hazardous gas is moist, ensure it does not react chemically with the product and sealing materials.
  4. Special attention when blowing and washing products: Before and after installation, purge the entire system and remaining gas residues with dry gas.
  5. Follow appropriate steps for flushing: The product must be blown under a ventilation hood, and operators must wear gloves.
  6. Do not use in explosive environments: This product is not certified for use with flammable or explosive gases unless explicitly stated.
  7. Use appropriate connectors and follow tightening procedures: Connectors must match the product's connectors. Follow the operating instructions for tightening.
  8. Check for leaks at product joints: Carefully check all component joints after installation to ensure no leaks.
  9. Operate under safe inlet pressure: Do not exceed the maximum working pressure (refer to product data sheet).
  10. Keep the entire system away from contamination: Do not use gas containing pollutants (dust, dirt, fiber, glass fragments, iron filings) when the system is running.
  11. Preheat the product before use: Preheat the product, especially for dangerous gases. Use an absolutely closed valve to prevent wrong flow during preheating.

3. Technical Specifications

FeatureSpecification
ModelAS200
Valve TypeNormally Closed
Flow Specifications (Air, O2, N2)1~100, 2~200, 3~300 SCCM (0.03~3, 0.05~5, 0.1~10, 0.2~20, 0.3~30, 0.5~50) SLM
Flow Specifications (Ar, CO2)1~100, 2~200, 3~300 SCCM (0.03~3, 0.05~5, 0.1~10, 0.2~20, 0.3~30) SLM
Accuracy±1.0% S.P. (≥35% F.S.) / ±0.35% F.S. (<35% F.S.)
Linearity±0.5% F.S.
Repeatability±0.2% F.S.
Response Time≤2 sec
Working Pressure Difference Range0.05~0.35MPa (0 <flow specification≤10SLM)
0.1~0.35MPa (10 < flow specification≤30SLM)
0.2~0.45MPa (30<flow specification≤50SLM)
Temperature CoefficientZero point: ≤±0.05% FS/°C; When adjusting: ≤±0.1% FS/°C (Flow specification ≤30SLM); ≤±0.2% FS/°C (Flow specification > 30SLM)
Compressive Strength3MPa
Leak Rate1×10-10 Pa·m3/sec He
Sealing MaterialFluororubber
Working Temperature5~45°C
Input SignalDigital: RS485 or EtherCAT
Analog: (1~5) VDC or (4~20) mA
Output SignalDigital: RS485 or EtherCAT
Analog: (1~5) VDC or (4~20) mA
Zero DriftWithout zero point correction <0.6%FS/year
Power Supply+15~+28VDC
Standard Connector TypeDouble ferrule φ6; Double ferrule 1/4"; Double ferrule 3/8"; VCR1/4" male connector; Double ferrule φ3; Double ferrule 1/8"; VCR 1/4"; φ6 (inner) x 1 hose connector; φ5 (inner) x 1.5 hose connector; φ4 (inner) x 1 hose connector; double ferrule φ10mm; A-type sealing connector
Power Connector TypeD-type 9-pin male connector; EtherCAT connector
WeightAbout 0.95 Kg (for some models), About 2.56Kg (for some models)
InstallationHorizontal installation is the best method. Zero point calibration is recommended for vertical installation.

4. Setup

4.1 Mechanical Installation

Before installation, ensure the air is dry and clean. Avoid external impact or vibration. Do not remove the protective caps on the connectors until ready for installation.

The AS200 series products should be installed in the orientation specified during ordering. Horizontal installation is generally recommended for best accuracy. If vertical installation is necessary, zero point calibration is recommended.

It is recommended to install a filter in the upstream pipeline to prevent contamination. Ensure the flow direction arrow on the product matches the gas flow direction in the pipeline.

ASAIR AS200 Mass Flow Controller

Figure: General appearance of the ASAIR AS200 Mass Flow Controller.

Connector Installation Methods:

  • VCR Connector Installation: For VCR joints, use appropriate VCR washers (non-positioned or position-based). Hand-tighten the nut and joint, then use two wrenches for final tightening. One wrench holds the joint, the other rotates the nut. This method is crucial for proper sealing, especially during disassembly.
  • Double Ferrule Fitting Installation: For double ferrule joints, install the front ferrule, then the rear ferrule and nut. Hand-tighten, then use two wrenches to tighten to the appropriate position to prevent leakage.

Refer to the full user manual (PDF) for detailed diagrams and specific tightening procedures for each connector type (e.g., double ferrule 1/4", VCR 1/4" male connector, D-type 9-pin male connector, EtherCAT connector).

4.2 Electrical Installation

The AS200 series operates with a single power supply of +15V to +28VDC (unipolar). It supports both digital and analog signals.

Connection Interfaces:

  • D-type 9-pin connector: This is a SEMI standard connector supporting 4~20mA input and output.
  • EtherCAT connector: For EtherCAT protocol communication.
  • RS485 port: For RS485 serial communication. Users can connect to a computer using an RS485 cable or adapter.

When selecting a control signal method (e.g., 1~5V analog, 4~20mA analog, or digital), the system automatically shields the other two control signals. The flow detection signal is output digitally, and simultaneously as 1~5V voltage and 4~20mA current, ensuring no interference between signals.

Refer to the full user manual (PDF) for detailed pinout diagrams for the 9-pin D-type connector, RS485, and EtherCAT interfaces.

5. Operating Instructions

5.1 Control Method

The AS200 series products are compatible with both digital and analog control methods. Users can select the control signal method that best suits their needs. The product supports:

  • Digital control (via RS485 or EtherCAT)
  • 1~5V voltage control
  • 4~20mA current control

When one control signal is selected, the other two are automatically shielded by the system. The flow detection signal is output digitally, and simultaneously as 1~5V voltage and 4~20mA current, ensuring no interference between signals.

5.2 Zero Adjustment

Customers can correct the zero drift of the flow meter through a zeroing procedure. Before performing zero adjustment, ensure that no gas is passing through the product.

Zeroing can be done via software commands or by using the zeroing button on the device. If using the button method, press and hold the zeroing button for more than 5 seconds until the green LED indicator flashes. Release the button, and once the zeroing process is complete, the green LED indicator will stop flashing.

5.3 Total Flow Recording

The AS200 series products can record and output the cumulative value of gas flow. The unit for cumulative value is standard milliliters (SCC). For example, a reading of 3000 indicates that 3000 standard milliliters of gas have passed through the MFC.

6. Maintenance

6.1 Overview

The MFC generally does not require routine maintenance, only occasional cleaning and recalibration. The frequency of maintenance depends on the gas quality:

  • Ultra-clean and non-corrosive gases: Clean and recalibrate every 3 to 4 years.
  • Low-cleanliness or corrosive gases: Clean and recalibrate every 1 to 2 years.

6.2 Media Usage Requirements

The gas used must be purified and free from dust, liquid, and oil. If necessary, install a filter in the gas line. If the outlet of the flow controller is connected to a liquid source, a one-way valve should be installed to prevent liquid backflow and product damage.

Warning: For products used with corrosive, toxic, or flammable gases, ensure the air tightness of the installation and connection is strictly guaranteed and inspected before ventilation and use. If removal from the system is necessary, thoroughly clean with dry, harmless conventional gas (e.g., nitrogen, air) or inert gas before disconnecting the gas line. Failure to clean may cause fire, explosion, poisoning, or other accidents, resulting in casualties.

6.3 Valve Port Sealing

The solenoid valve of the MFC is a regulating valve, not a stop valve. It cannot be used as a stop valve. If a stop valve is required, the user should purchase one separately. For dangerous gases, a stop valve should always be added to both the inlet and outlet of the MFC to ensure work safety. The internal leakage rate of the valve port of MFC is within 1% F.S., which is considered normal.

7. Troubleshooting

7.1 Preliminary Inspection

  • Check whether the gas source and the gas line leading to the MFC are open.
  • Ensure that power and control signals are properly delivered to the electrical connectors on the circuit board.
  • Check whether the communication line is connected correctly.

7.2 Common Issues and Solutions

Fault PhenomenonPossible Cause of FailureApproach
When there is no ventilation, the displayed value is not zeroGas is still flowingCheck whether the stop valve is closed
Zero driftUsing the Zero Adjustment Function
Other faultsContact "Aosong Electronics"
Unable to control trafficIncorrect gas connectionCheck whether the MFC is correctly connected to the gas line
The pressure difference is not within the required rangeCheck the pressure
MFC works in other modesCheck and change control mode using software
Power IssuesCheck the power supply and whether the power connection is correct.
Incorrect signal settingCheck the set voltage or current signal
MFC is contaminatedContact "Aosong Electronics"
Sensor damage / Circuit board problem / Valve body damageContact "Aosong Electronics"
Unable to communicatePower IssuesCheck the power supply and whether the power connection is correct.
Communication connection problemCheck whether the communication line is connected correctly
Address Conflict / The baud rate setting is incorrect / Circuit board problemCheck whether the address of MFC conflicts with other MFCs / Check the baud rate setting of the MFC / Contact "Aosong Electronics"

8. User Tips

  • For detailed technical specifications, diagrams, and advanced usage scenarios, always refer to the complete product datasheet.
  • If you encounter issues not covered in the troubleshooting section or require further assistance, contact the manufacturer's authorized technical support.
  • Ensure all gas lines are clean and free of contaminants before connecting the MFC to maintain optimal performance and longevity.

9. Warranty and Support

Guangzhou Aosong Electronics Co., Ltd. provides a 12-month warranty for direct purchasers of its products, calculated from the date of shipment. This warranty covers defects in materials and workmanship according to the technical specifications outlined in the Aosong Electronics product manual. If a product is proven defective during this period, the company will provide free repair or replacement services.

The company is responsible only for defective products used in situations that meet the product's technical requirements. No guarantees are made for applications in special scenarios not recommended or for reliability when applied to other unsupported products or circuits.

This manual is subject to change without prior notice. The final right of interpretation of this product belongs to Guangzhou Aosong Electronics Co., Ltd.

All rights reserved © 2024, ASAIR®