MESTEK IR05A

Contents

MESTEK IR05A Digital Infrared Thermometer User Manual

Non-contact High Temperature Meter Pyrometer with Dual Laser and K-type Liquid Probe

1. Introduction

This manual provides detailed instructions for the safe and effective use of your MESTEK IR05A Digital Infrared Thermometer. Please read this manual carefully before using the product and keep it for future reference.

Note: This device is designed for measuring inanimate objects. Temperature readings for humans or animals will be inaccurate. Batteries are not included. Only IR05A and IR02C models include a K-type thermocouple.

2. Safety Instructions and Attention

2.1 Safety Instructions

  • Please read the instructions carefully before using the instrument.
  • Safety Symbols:
    • ⚠ Important information tips of danger
    • CE Comply with European CE safety standards.
    • EN60825-2014
  • Warning! Do not aim the laser at human eyes or reflective surfaces.

2.2 Attention

  • When the ambient temperature changes greatly or suddenly, please wait for 20 minutes before using the instrument to allow its temperature to reach a stable state.
  • Please avoid electromagnetic fields caused by electric welding and induction heating.
  • Please avoid placing the instrument near or on high-temperature objects.
  • Please keep the instrument clean and avoid dust from entering the lens barrel.
  • Do not use solvents to clean the instrument.

3. Product Overview

The MESTEK IR05A is a high-temperature infrared thermometer featuring dual laser positioning and a K-type liquid probe for versatile temperature measurement. It offers various measurement modes and an adjustable emissivity for accurate readings.

MESTEK IR05A Infrared Thermometer with Dual Laser

Figure 1: MESTEK IR05A Digital Infrared Thermometer

3.1 Key Features

  • High-precision probe
  • Dual laser positioning
  • High and low temperature alarm
  • Adjustable emissivity (0.1-1.0)
  • High temperature measurement (up to 1600°C)
  • VA reverse display color screen
  • Object distance ratio: 50:1
  • Unit conversion (°C/°F)
  • Maximum, Minimum, Average, and Temperature Difference value modes
  • K-type thermocouple temperature measurement
  • Fast response time (<0.5S)
  • Automatic shutdown (approx. 35 seconds)
Multiple functions of the MESTEK IR05A thermometer

Figure 2: Multiple Functions Overview

3.2 LCD Display Components

LCD Display Components Diagram

Figure 3: LCD Display with numbered components

  1. Emissivity
  2. Measurement indication
  3. Laser indication
  4. Battery undervoltage indication
  5. Data locking (HOLD)
  6. Radiation measurement temperature value
  7. Temperature unit (°C/°F)
  8. Maximum lock (Max)
  9. Minimum lock (Min)
  10. Average value (Avg)
  11. Temperature difference value (Diff)
  12. Alarm indicator
  13. Hi high alarm
  14. Low low alarm
  15. K-thermocouple temperature

4. Setup

4.1 Battery Installation

The MESTEK IR05A requires 2 x 1.5V AA batteries (not included in the package). To install the batteries:

  1. Locate the battery cover on the handle of the thermometer.
  2. Open the battery cover.
  3. Insert two 1.5V AA batteries, ensuring correct polarity (+/-).
  4. Close the battery cover securely.

Video 1: Demonstrates battery installation and basic operation modes.

5. Operating Instructions

5.1 Power On/Off and Basic Measurement

Pull the instrument trigger to light up the screen and begin measurement. Release the trigger to hold the reading (Data Locking).

5.2 Setting Alarm Modes (High/Low)

To set the high or low temperature alarm:

  1. Long press the SET button to enter the SET mode. The default is High Alarm setting.
  2. When the value flashes, use the ▲ (Up) or ▼ (Down) keys to adjust the desired alarm value.
  3. Pull the trigger to save the setting. The buzzer will sound if the measured temperature exceeds (High Alarm) or falls below (Low Alarm) the set value.
  4. To switch to Low Alarm mode, gently press the SET button again while in SET mode. Repeat steps 2 and 3.
Thermometer parameter settings for alarm and emissivity

Figure 4: Parameter Setting Flow (High Alarm, Low Alarm, Emissivity, Unit)

5.3 Adjusting Emissivity (ε)

Emissivity is the ability of an object surface to radiate infrared rays. The greater the emissivity, the stronger the radiation ability of the object surface. The default emissivity of this thermometer is 0.95. When measuring, set the emissivity of the thermometer consistent with the emissivity of the measured object.

  1. Long press the SET button to enter the SET mode.
  2. Short press the SET button repeatedly until the emissivity value (ε) flashes.
  3. Use the ▲ (Up) or ▼ (Down) keys to adjust the emissivity value (range 0.10-1.00).
  4. Pull the trigger to save the setting.
Adjustable emissivity feature with comparison table

Figure 5: Adjustable Emissivity and Comparison Table

Emissivity Reference Table

Measured SurfaceEmissivityMeasured SurfaceEmissivity
Oxidate Carbon (unoxidized)0.8-0.9Iron (casting)0.2-0.3
Iron Get rusty0.5-0.7Pottery and porcelain0.95
Graphite0.7-0.8Clay0.95
Oxidate0.6-0.95Passivation of iron (forging)0.9
Carborundum0.9Concrete0.95
Electrical terminal board0.4Lead0.95
Coarse0.2-0.6Cloth0.95
Oxidate0.2-0.4Glass plate0.85
Molybdenum oxidation0.2-0.6Gravel0.95
Metallic aluminium A3003 alloy (oxidation)0.3Platinum black0.9
A3003 alloy (rough)0.1-0.3Plaster0.8-0.95
Brass Polish0.3Cold rolling Steel0.7-0.9
Oxidate0.5Ice0.98
Steel Polished steel plate0.4-0.6Paper0.95
Copper Oxidate0.4-0.8Polished steel plate0.1
Electrical terminal board0.6Zinc (Zn) Oxidate0.1
Asbestos0.95Hastelloy B-20.3-0.8
Soil0.9-0.98Basalt0.7
Water0.93Nickaloy Sand blast0.3-0.6
Wood0.9-0.95Electropolishing0.15

5.4 Temperature Unit Switching (°C/°F)

To switch between Celsius and Fahrenheit:

  1. Long press the SET button to enter the SET mode.
  2. Short press the SET button repeatedly until the unit symbol (°C/°F) flashes.
  3. Use the ▲ (Up) or ▼ (Down) keys to select the desired unit.
  4. Pull the trigger to save the setting.
Celsius and Fahrenheit switching on the display

Figure 6: °C/°F Switching

5.5 Laser ON/OFF

The laser aiming function can be turned on or off by short pressing the laser button (▲ with laser icon) when the instrument screen is on or in the measuring state.

5.6 MAX/MIN/AVG/DIFF Modes

To switch between Maximum, Minimum, Average, and Temperature Difference measurement modes:

  1. When the instrument screen is on or in the measurement state, short press the MIN/MAX button repeatedly.
  2. The display will cycle through MAX, MIN, AVG, and DIFF symbols.
  3. The instrument will then lock and display the corresponding value (Maximum, Minimum, Average, or Temperature Difference).
Measurement value mode display showing Max, Min, Avg, Diff

Figure 7: Measurement Value Modes

5.7 K-type Thermocouple Temperature Measurement

For contact temperature measurement using the K-type probe:

  1. Insert the K-type thermocouple plug into the thermocouple socket of the instrument, ensuring correct positive and negative pole markings.
  2. Pull the trigger to display the K-type symbol on the screen.
  3. During measurement, ensure the thermocouple probe touches the measured object.
  4. The screen will display both the K-type thermocouple temperature and the infrared irradiation temperature.
K-type thermocouple temperature measurement in use

Figure 8: K-type Thermocouple Measurement

5.8 Target Distance Ratio (D:S)

The instrument has a certain visual angle and field of view. The ratio of the measuring distance (D) to the measured target size (S) is D:S = 50:1. It is necessary to ensure that the measured object fills the field of view of the instrument, meaning the thermometer only "sees" the measured object and "cannot see" other objects. The larger the measured object, the farther the measurement distance. The smaller the measured object, the closer the measurement distance.

Safety measures and D:S ratio diagram

Figure 9: Target Distance Ratio (D:S) Explanation

6. Application Scope

The MESTEK IR05A thermometer is suitable for a wide range of industrial and commercial applications, including but not limited to:

  • General Industry
  • Mold/Laser applications
  • Steel/Smelting/Chemical Industry
  • Glass/Ceramic Field
  • Electric/Mechanical temperature measurement
  • Heating/Ventilation/Air Conditioning (HVAC)
  • Food safety (e.g., water/oil temperature, cold storage)
Various application scenarios for the infrared thermometer

Figure 10: Application Examples

7. Specifications

ParameterSpecification (IR05A)
ModelIR05A
Temperature Range-40°C ~ 1600°C (-40°F ~ 2912°F)
Accuracy (IR)-40°C~800°C (±1.5%), 800°C~1600°C (±3%)
K-type Range-10°C ~ 500°C (14°F ~ 932°F)
K-type Accuracy±(1.5% + 2°C/4°F)
DisplayVA reverse display color screen
D:S (Distance to Spot Ratio)50:1
Spectral Response8-14µm
Resolving Power0.1°C/0.1°F
Emissivity0.10-1.00 (Adjustable)
Laser<1mW/620-690nm Level 2
Response Time<0.5S
Auto ShutdownAbout 35 seconds
Operating Environment0-50°C / 10%-90%RH
Storage Temperature-20°C ~ 60°C
Size180x120x60mm
Power Supply2 x 1.5V AA battery (not included)
WeightApproximately 265g (excluding battery)
CertificationCE, FCC, RoHS
Product parameters table for IR05A

Figure 11: Detailed Product Parameters for IR05A

7.1 Other Models (for reference)

ParameterSpecification (IR02C)
ModelIR02C
Range-50° ~ 800°C
Inverted color screen✓
Temperature probe✓
Ambient temperature✓
Environment humidity✓
Mildew✓
Dew point✓
Measurement accuracy0-800°C: ±(1.5% reading + 2°C/4°F) (-58°F-1472°F)
Resolution0.1°C
Emissivity0.1-1.0 Adjustable
Laser spot12 point laser
Max✓
HD backlight✓
Data retention✓
Object distance ratio12:1
Response time<0.5S
Response wavelength8-14µm
Operating temperature0-50°C
Storage temperature-20~60°C
Automatic shut-down30s
Battery1.5V*2 AAA battery (not included)
Size150*103*33mm
Product parameters table for IR02C

Figure 12: Detailed Product Parameters for IR02C

For IR600 and IR600PRO models, the temperature range is -50~600°C (-58°F-1112°F) with an accuracy of +/-1.5°C and a D:S ratio of 12:1. The IR600PRO features a color screen.

IR600 model specifications

Figure 13: IR600 Model Specifications

8. Maintenance

8.1 Cleaning the Instrument

  • Keep the instrument clean and avoid dust from entering the lens barrel.
  • Lens cleaning: Clean the dust on the lens surface with clean compressed air, and then gently wipe the surface with a wet cotton swab.
  • Do not use solvents to clean the instrument.

8.2 Battery Replacement

When the battery power is insufficient, the display will show the battery undervoltage symbol (🔋). The battery must be replaced. Open the battery cover and replace it with two new 1.5V AA batteries.

9. Troubleshooting

If the instrument is not functioning as expected, consider the following:

  • No display after power on: Check if the batteries are installed correctly and have sufficient charge. Replace batteries if necessary.
  • Inaccurate readings:
    • Ensure the emissivity setting matches the material being measured.
    • Allow the instrument to stabilize for 20 minutes if there's a sudden change in ambient temperature.
    • Ensure the target fills the field of view of the instrument (refer to D:S ratio).
    • Clean the lens if it's dusty or dirty.
  • Alarm not working: Verify that the high/low alarm thresholds are set correctly and are within the expected range for your measurement.

10. User Tips

  • For best accuracy, always ensure the emissivity setting is appropriate for the material you are measuring. Refer to the emissivity table.
  • When measuring liquids or oils, ensure the surface is calm and free of steam for more accurate readings. Avoid measuring through transparent surfaces like glass or plastic.
  • Utilize the MAX/MIN/AVG/DIFF modes to get a comprehensive understanding of temperature fluctuations over time or across a surface.
  • The K-type thermocouple provides contact temperature measurement, which can be useful for verifying infrared readings or for materials where infrared measurement is difficult.

11. Additional Resources

A detailed user manual in PDF format is available for download:

User manual (PDF)

12. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the product packaging or contact your retailer. Keep your purchase receipt as proof of purchase.

© 2023 MESTEK. All rights reserved.


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.