1. Introduction
The ASAIR AMC2100 gas mass flow controller is a device designed for precise measurement and control of gas mass flow. It incorporates a self-developed MEMS mass flow chip, offering high precision, stability, high and low temperature resistance, excellent linearity, and fast response time. Each unit undergoes rigorous testing and calibration for its measurement range, accuracy, repeatability, and response time before delivery.
This mass flow controller is suitable for monitoring and controlling dry, clean, non-corrosive gases such as air, nitrogen (N2), oxygen (O2), carbon dioxide (CO2), and argon (Ar). Its applications span various fields including universities, scientific research, fire protection, environmental monitoring, tobacco, smart agriculture, food, and medicine.

2. Key Features
- Digital Display Screen: Provides clear, real-time flow rate readings.
- High Accuracy: Achieves an accuracy of ±3%F.S. (Full Scale).
- Integrated Digital + Analog Outputs: Offers versatile communication options including RS485 and analog voltage (0-5V).
- Good Repeatability: Ensures consistent and reliable measurements with 0.5%F.S. repeatability.
- Fast Response Time: Responds quickly to changes in flow, typically less than 2 seconds.
- Configurable Parameters: Allows users to adjust various settings to suit specific application needs.
- Supports Multiple Gas Monitoring: Compatible with Air, N2, O2, CO2, and Ar.
3. Technical Specifications
| Parameter | Description |
|---|---|
| Model | AMC2100 |
| Measuring Medium | Air, N2, O2, CO2, Ar (standard pressure) |
| Range Options | 0-100ml, 0-300ml, 0-500ml, 0-1L, 0-3L (mL/min) |
| Accuracy | ±3%F.S. |
| Repeatability | 0.5%F.S. |
| Response Time | <2s |
| Power Supply | 24V DC |
| Working Voltage | 24VDC |
| Working Current | <120mA |
| Typical Power Consumption | 2W@24V |
| Output / Communication Mode | RS485 / Analog voltage (0-5V) |
| Operating Pressure | 0.15~0.35MPa |
| Connection Piping Specification | 2-RC 1/4 |
| Pipe Material | Aluminum 6061 |
| Operating Temperature | 0-50°C |
| Connector | DB9 male |
| Weight | 269g |
| Dimensions (L×W×H) | Approx. 70×70×30 mm (refer to dimension diagram) |
For a detailed comparison with other models and comprehensive specifications, please refer to the official user manual (PDF) available at: AMC2100 Datasheet (PDF).

4. Structure and Interface Definition
4.1 Structure Overview
The AMC2100 features a sensor air duct that includes a control valve, flow measurement components, and air inlet and outlet interfaces. An arrow on the housing clearly indicates the required gas flow direction.

4.2 Interface Definition
The communication interface is a DB9 male connector. The pin definitions are crucial for proper wiring and communication.

| Pin | Function | Description |
|---|---|---|
| 1 | Analog voltage input | 0~5V, effective when controller works in analog signal input mode |
| 2 | Analog voltage output | 0~5V |
| 3 | RS485B | RS485 B line |
| 4 | RS485A | RS485 A line |
| 5 | +24V | Power positive |
| 6 | GND | Power common terminal 0V |
| 7, 8, 9 | NC | Test pin, user does not need to connect |
4.3 Connect Cable Definition
The AMC2100 uses a DB9 female single-ended cable for wiring. The cable features specific wire colors corresponding to the pin functions.

5. Setup
- Physical Connection: Identify the gas inlet and outlet ports on the AMC2100, ensuring the gas flow direction matches the arrow on the device housing. Connect your gas lines securely.
- Electrical Connection: Connect the DB9 female single-ended cable to the DB9 male connector on the AMC2100. Ensure the wiring for power (Pin 5: +24V, Pin 6: GND) and communication (Pin 3: RS485B, Pin 4: RS485A) is correct according to the pin definitions provided in Section 4.2.
- Power On: Apply 24V DC power to the device. The digital display should illuminate.
- Initial Gas Selection: If the gas to be measured is not Air (the factory default), proceed to Section 6.5 to set the correct measurement gas.
6. Operation
6.1 Normal Working Interface Display
During normal operation, the controller displays the instantaneous flow rate. This value updates in real-time. Use the Up or Down keys to increase or decrease the set flow rate value. A short press of the S (Setting) key enters the flow setting interface, where the display flashes at 1Hz. Adjust the flow rate with the Up or Down keys. Press S again to exit the setting interface and return to normal working mode.
6.2 Set Modbus Communication Address
To set the device's communication address:
- From the normal working interface, press and hold the S key for 2 seconds to enter the menu page (CH01).
- Press the S key again to enter the sub-page for address setting.
- Use the Up or Down keys to set the desired communication address.
- Press the S key to return to the menu page (CH01).
6.3 Set the Communication Baud Rate
To set the communication baud rate:
- From the menu page (CH01), press the Down key once to navigate to the menu page (CH02).
- Press the S key once to enter the sub-page for baud rate setting.
- Use the Up or Down keys to set the desired baud rate.
- Press the S key to return to the menu page (CH02).
6.4 Check or Clear the Cumulative Flow
To check or clear the cumulative flow:
- From the menu page (CH02), press the Down key once to navigate to the menu page (CH03).
- Press the S key once to enter the sub-page, which displays the cumulative flow.
- Use the Up or Down keys to clear the cumulative flow.
- Press the S key to return to the menu page (CH03).
6.5 Set the Measurement Gas
The factory default measurement gas is Air. To select a different gas:
- From the menu page (CH03), press the Down key once to navigate to the menu page (CH04).
- Press the S key again to enter the sub-page.
- Use the Up or Down keys to select the desired measurement gas from the options: Air, N2, O2, CO2, Ar.
- Press the S key to return to the menu page (CH04).
6.6 Set the Display Brightness Level
To adjust the display brightness:
- From the menu page (CH04), press the Down key once to navigate to the menu page (CH05).
- Press the S key once to enter the sub-page.
- Use the Up or Down keys to set the display brightness level (1-8 levels).
- Press the S key to return to the menu page (CH05).
6.7 Exit Parameter Setting
To exit the parameter setting mode:
- From the menu page (CH05), press the Down key once to navigate to the menu page (End).
- Press the S key once to exit to the normal working interface.
7. Communication Protocol (Modbus RTU)
The AMC2100 utilizes a standard Modbus-RTU communication protocol over an RS485 interface, allowing it to be integrated into various control systems. The device acts as a slave, with a default address of 0x01, which can be modified.
7.1 Communication Parameters
- Protocol: Modbus-RTU
- Baud Rates: 1200, 2400, 4800, 9600, 19200, 38400 (Default: 19200)
- Data Bits: 8 bits
- Stop Bits: 1 bit
- Parity: None
- CRC Polynomial: 0xA001
7.2 Function Codes
The AMC2100 primarily supports the following Modbus function codes:
- 0x03 (Read Holding Register): Used to read the value of a holding register.
- 0x06 (Preset Single Register): Used to write a specific hexadecimal value to a holding register.
7.3 Registers
The AMC2100 contains various registers that provide information such as instantaneous flow, cumulative flow, and set flow. These registers can be accessed via Modbus commands. For example, instantaneous flow and set flow are 32-bit float types, while accumulated flow is a 64-bit double type.
For detailed information on Modbus RTU communication, including specific register addresses, data formats, and data conversion examples (e.g., float and double type conversions), please refer to the official user manual (PDF) for comprehensive programming guidance.
8. Maintenance
- Keep Clean: Regularly clean the exterior of the device with a soft, dry cloth. Avoid using harsh chemicals or abrasive materials.
- Gas Quality: Ensure that only dry, clean, and non-corrosive gases are passed through the controller, as specified in the product's application guidelines. Using unsuitable gases can damage the MEMS sensor.
- Ventilation: Ensure adequate ventilation around the device to prevent overheating.
- Connection Integrity: Periodically check all gas and electrical connections for tightness and integrity to prevent leaks or communication issues.
- Calibration: While the device is factory-calibrated, consider periodic re-calibration by qualified personnel if high accuracy is critical for your application over extended periods.
9. Troubleshooting
- No Display/Power: Check the 24V DC power supply connection and ensure it is providing the correct voltage. Verify the wiring to pins 5 (+24V) and 6 (GND) of the DB9 connector.
- Incorrect Flow Reading: Verify that the correct measurement gas is selected in the device settings (refer to Section 6.5). Ensure the gas flow direction matches the arrow on the device. Check for any blockages in the gas lines.
- Communication Issues (Modbus RTU):
- Check RS485 wiring (Pin 3: RS485B, Pin 4: RS485A) for correct polarity and secure connections.
- Verify that the communication address and baud rate settings on the AMC2100 match those of your host system (refer to Sections 6.2 and 6.3).
- The AMC2100 only supports Modbus function codes 0x03 and 0x06. If an unsupported function code (e.g., 0x07) is sent, the device will return an abnormal response (e.g., 01 87 01 82 30). Ensure your Modbus commands use the correct function codes.
- No Flow Control: Ensure the device is set to the appropriate control mode (digital or analog voltage control) if applicable.
10. User Tips
- CO2 Measurement: The AMC2100 is fully capable of measuring and controlling CO2. Remember to correctly configure the device for CO2 gas type in the settings menu (Section 6.5) for accurate readings.
- Measure and Control: The AMC2100 is a mass flow *controller*, meaning it performs both measurement and active control of gas flow, not just passive measurement. Utilize its control capabilities for precise flow regulation in your applications.
11. Packing List
Upon unpacking your AMC2100, please verify that all the following items are included:
- AMC2100 Mass Flow Controller: 1 unit
- Connection Cable (DB9): 1 unit
- Datasheet: 1 copy
- Certificate: 1 copy
12. Warranty and Support
Guangzhou Aosong Electronics Co., Ltd. provides quality assurance for its products, calculated from the date of shipment, in accordance with the technical specifications outlined in the Aosong Electronics product manual.
12.1 Warranty Periods
- AMC2100 Mass Flow Controller: 12 months
- Cable and other accessories: 6 months
If a product is proven to be defective during its warranty period, the company will provide free repair or replacement services. This warranty covers defective products used under conditions that meet the product's technical specifications. The company does not guarantee applications in special scenarios not recommended or make promises regarding the reliability of products applied to unsupported products or circuits.
This manual is subject to change without prior notice. The final right of interpretation of this product belongs to Guangzhou Aosong Electronic Co., Ltd.
For complete warranty terms and conditions, and for further support, please refer to the official user manual (PDF) or contact Guangzhou Aosong Electronics Co., Ltd. directly.
13. Safety Warnings and Personal Injury
IMPORTANT SAFETY INFORMATION: READ BEFORE USE
This product is not intended for use in safety protection devices, emergency stop devices, or any other application where product failure could result in personal injury, unless specifically authorized for such use. Always refer to the product data sheet and this manual before installing, handling, using, or maintaining this product.
Failure to follow these recommendations can lead to death or serious personal injury. Guangzhou Aosong Electronics Co., Ltd. will not be liable for any compensation for personal injury or death resulting from such misuse, and all claims against the company, its managers, employees, affiliated agents, and distributors will be waived.
Copyright ©2024, ASAIR ®