1. Introduction
Thank you for choosing the TOOLTOP ET828Pro digital storage wave multimeter. This product combines innovative functions and a user-friendly design to provide great convenience for field testing. Before use, please read this manual carefully, especially the "Safety Instructions" section. Keep this manual in a safe place for future reference.
The ET828Pro is a 2-in-1 device, functioning as both a 10MHz bandwidth sight wave meter and a 6000-count auto-range multimeter. It features a 50Msps high-speed sampling rate, a 2.4-inch color screen, and various measurement capabilities for improved work efficiency.

Image: The ET828Pro device displaying both oscilloscope waveform and multimeter readings simultaneously.
2. Safety Instructions
This digital storage visual wave multimeter complies with IEC1010-1 safety specifications, overvoltage measurement category II CAT II - 1000V, and pollution protection level 1.
- Pre-use Check: Always inspect the device for damage, cracks, or missing parts. Pay special attention to the test leads and their insulation. Keep fingers away from the metal part of the probe when in use.
- Environmental Conditions: Do not operate in hot, humid, rainy, or flammable environments, or when the instrument is wet.
- Voltage/Current Limits: Never apply voltage/current exceeding the maximum limits specified for the instrument.
- Probe Connection: Do not connect the meter pen across both ends of a voltage source when measuring current, resistance, on/off test, diode, or capacitance.
- Power Off for Resistance/Diode Tests: Before conducting resistance or diode/on/off tests, ensure the equipment under test is powered off and capacitors are discharged.
- Fuse/Battery Replacement: Before opening the back cover to replace the fuse or battery, power off the instrument and remove test leads from the circuit. Use fuses of the same specifications.
- Unauthorized Modification: Do not modify, disassemble, or use the product for purposes other than its functional design. Do not substitute accessories.
- Children Safety: Keep this equipment away from children; it is not a toy.
Safety Signs
- ! Caution, danger: This symbol indicates that instructions in this manual must be followed to prevent injury or damage.
- 4 Electric shock danger: This symbol indicates dangerous voltages at terminals. Avoid touching test lead ends when these terminals carry voltage.
- Tips: A reminder that special care is needed to avoid incorrect measurements or damage.
- Attention: Extreme care must be taken; errors may damage the product or other property.
- Warning: Operation requires high concentration; incorrect use may cause personal injury or endanger life.
Measurement Function Limits
| Measurement Function | Using the input | Maximum limit |
|---|---|---|
| V DC | V/Ω/Cap, COM | 1000V DC+AC peak, within 10 seconds |
| V AC | V/Ω/Cap, COM | 750V DC+AC RMS, within 10 seconds |
| Hz% | V/Ω/Cap, COM | 250V DC/AC RMS, within 10 seconds |
| mA AC/DC | uA/mA, COM | 600mA DC/AC RMS, 250V/750mA fusible Fuse |
| A AC/DC | 10A, COM | 10A DC/AC RMS, within 30 seconds, 15 minutes of cooling interval 250V/10A fusible Fuse |
| Ω | V/Ω/Cap, COM | 250V DC/AC RMS, within 10 seconds |
3. Setup
3.1. Package Contents
The standard accessories include:
- 1 x Main Unit (Oscilloscope Multimeter)
- 1 x Pair of Test Leads (two-color meter pen)
- 1 x Cloth Bag
- 1 x Packing Box
- 1 x Product Manual (this document)

Image: Contents of the product package, including the ET828Pro unit, test leads, and carrying bag.
3.2. Battery Installation
The meter utilizes 3 AA 1.5V batteries. To install or replace batteries:
- Ensure the knob switch is in the "OFF" position and test leads are disconnected.
- Loosen the set screw on the back cover and remove it.
- Insert 3 AA 1.5V batteries, ensuring correct polarity.
- Replace the back cover and secure it with screws.
Note: Do not mix old and new batteries. Do not charge non-rechargeable batteries. If the screen displays a red battery prompt, replace batteries promptly to maintain measurement accuracy.
3.3. Connecting Test Leads
Always ensure the meter is off before connecting or disconnecting test leads.
- Plug the banana plug of the black test lead into the negative polarity "COM" socket.
- Plug the banana plug of the red test lead into the appropriate positive polarity socket ("VΩCx" for voltage/resistance/capacitance/diode/frequency, "mA" for milliampere current, or "10A" for 10A current).
4. Operating Instructions
4.1. Basic Operations
Power On and Off
- To power on, turn the knob to select any measurement position.
- To power off, turn the knob to the "OFF" position.
- Attention: Ensure test leads are away from test points before powering off. Always turn off the instrument when not in use.
Automatic Sleep Function
- The meter automatically hibernates after approximately 15 minutes of inactivity (no key presses or knob operations).
- To disable auto-sleep, press and hold the "F1" key while turning the knob to power on.
- After hibernation, turn the knob to "OFF" and then back to a measurement position to resume operation.

Image: The auto-sleep function activates after 15 minutes of no operation to conserve battery.
4.2. Oscilloscope Operation
Entering Oscilloscope Mode
Turn the knob switch to the "OSC DC" or "OSC AC" gear to enter oscilloscope test mode. In OSC mode, the amplitude ranges from 500mV/div to 200V/div, and it can measure up to 1600Vpp.
Basic Display in Oscilloscope Mode

Image: Oscilloscope display with labels for Amplitude, Time base, Trigger method, Trigger state, Power, Trigger level, Test Data, and DOS menu.
Function Keys (F1-F4)
The multi-functional keys (F1-F4) at the bottom of the LCD screen change their function based on the current operating mode and range. On-screen menus provide prompts for their current operation.
| Key | Name | Function |
|---|---|---|
| F1~F4 | Multi-functional keys | Functions change with range/mode; LCD shows prompts. |
| R | Multimeter Status / Oscilloscope Status | Manual/automatic range switching (Multimeter) / Automatic capture (Oscilloscope) |
| S | Data locking and storage | Locks current data/waveform for inspection. |
| OFF | Power Off | Meter shutdown and charging gear. |
| OSC DC | DC Oscilloscope Gear | Measures DC waveforms. |
| OSC AC | AC Oscilloscope Gear | Measures AC waveforms. |
| V~ | DC, AC Voltage Gear | Measures DC and AC voltage. |
| Ω | Resistance / Diode / On-Off Test | Measures resistance, tests diodes, and continuity. |
| Cx | Capacitor gear | Measures capacitance. |
| Hz% | Frequency/Duty cycle gear | Measures frequency and duty cycle. |
| mA~ | AC/DC mA Gear | Measures AC and DC milliampere current. |
| 10A~ | AC/DC 10A Gear | Measures AC and DC 10A current. |
Time Base Adjustment
In the oscilloscope main menu, press F1 (TIME) to enter the time base adjustment sub-menu.
- Press F1 (EXIT) to exit.
- Press F2 (◄) and F3 (►) to adjust the time base (t/div).
Range Adjustment (Vertical Amplitude)
In the oscilloscope main menu, press F2 (VOLT) to enter the range adjustment sub-menu.
- Press F1 (EXIT) to exit.
- Press F2 (▲) and F3 (▼) to adjust the vertical range (V/div).
Tip: When measuring signals of unknown frequency, start with the fastest time base and gradually select a slower one until the signal is displayed correctly to avoid aliasing effects. Adjust time base or press "R" key to avoid overlapping.
Trigger Control
In the oscilloscope main menu, press F3 (TRIG) to enter the trigger control sub-menu.
- Press F1 (EXIT) to exit.
- Press F2 (∫ or ㄟ) to select rising edge or falling edge trigger.
- Press F3 (MODE) to select the trigger mode (Auto, Normal, Single).
- Press F4 (LEVEL) to enter the trigger level adjustment sub-menu.
Trigger Level Adjustment
In the trigger setting sub-menu, press F4 (LEVEL) to enter the trigger level adjustment sub-menu.
- Press F1 (EXIT) to exit.
- Press F2 (▲) and F3 (▼) to increase or decrease the trigger level. The display will mark the current trigger level position.
- Press F4 (RESET) to reset the trigger level to zero.
About Trigger Mode
- Auto: Acquires waveforms even without a trigger condition. If no trigger, it self-triggers after a period. Useful for monitoring irregular signals or DC power waveforms.
- Normal: Starts collecting data only after a trigger signal is detected. If no trigger, no new waveform is acquired.
- Single: Acquires waveform data once a trigger condition is detected. The latest waveform is automatically held.

Image: Visual representation of Auto, Normal, and Single trigger modes and their effects on waveform acquisition.
Prompt Message for Scan Status
| Status | Description |
|---|---|
| Auto | Waveforms can be acquired without trigger conditions. |
| Wait | Waiting for the trigger condition. |
| Trig | The trigger condition has been detected. |
| Stop | Keep Locked (waveform held). |
Trigger Operation for Single Scan
- Adjust vertical range (V/div) and horizontal range (s/div) to fit the waveform.
- Select appropriate trigger level, trigger edge, and "Single" trigger mode.
- Once the signal is present, the oscilloscope will trigger and capture it. The latest waveform is automatically held.

Image: Various waveforms automatically captured by the oscilloscope, demonstrating the auto-capture function.
Holding of Signal Waveforms
To hold a waveform for closer inspection, press the "S" key. Pressing "S" starts/stops data acquisition, holding the last waveform on screen. The "STOP" icon appears on the LCD.
Storage and Readout of Waveforms
The instrument can store 10 OSC waveform data sets.
- Press "S" to hold the current waveform.
- Press "S" again and hold for 2 seconds to enter the database.
- Use F1 (◄) / F2 (►) to select a save location.
- Press F4 (SAVE) to save the current waveform.
- Press "S" again and hold for 2 seconds to exit the database.
Tip: Saving a new waveform to an existing location will overwrite the old data. Exercise caution.
4.3. Multimeter Operation
Warning: When changing measurement functions, always detach the probe from the test point first.
Basic Display Contents in Multimeter Mode

Image: Multimeter display with labels for High voltage tips, Auto range, Current range, Data retention, Result, REL value, Battery power, Current gear, Real-time data, AC, Unit, MAX/MIN, and P-H menu.
Switching of Measurement Functions
Turn the knob to select the desired test function. Functions cycle in the following order: DC/AC Voltage → Resistance/Diode/Off → Capacitance → Frequency/Duty Cycle → DC/AC Current (mA) → DC/AC Current (10A).
Manual/Automatic Range Selection
The meter defaults to auto-range. To switch to manual range:
- Press F4 (RNG). The "AUTO" mark disappears.
- Each subsequent press of F4 (RNG) switches to the next range.
- Press R to return to auto-range.
Relative Measurement Mode (REL)
This mode displays the difference between the current measured value and a stored reference value.
- Press F1 (REL) to store the current display as a reference and activate REL mode.
- The screen shows the difference.
- Press F1 (REL) again to exit REL mode.
- REL mode automatically disengages if the measurement function is changed.
Peak Hold (P-H) Mode
Displays maximum and minimum measured values, updated with each new measurement. Activated by pressing F2 (P-H).
- Press F1 (EXIT) to exit P-H mode.
- Press F2 (RUN) to start peak hold measurement.
- Press F3 (STOP) to stop refreshing peak measurements.
- Press F4 (RESET) to reset measured peaks for a new measurement.
AC and DC Voltage Measurement
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "V~" function. "VOLTAGE" appears.
- Press F3 (SEL) to switch between DC/AC measurement.
- Touch test point with meter pen.
- Read voltage value.
Warning: If voltage exceeds 24V, a "4" warning message appears with 3 beeps and flashing red light. Do not apply >700V AC or >1000V DC for more than 10 seconds.
AC and DC Current Measurement (mA, 10A)
- Connect black test lead to "COM", red test lead to "mA" or "10A" socket.
- Turn knob to "mA~" or "10A~" function. "CURRENT" appears.
- Press F3 (SEL) to switch between DC/AC measurement.
- Insert meter pen in series with the circuit under test.
- Read current value.
Warning: Do not perform current measurements on circuits with voltages above 250V. When using 10A gear, measurement time should not exceed 30 seconds in every 15 minutes to prevent damage.
Frequency and Duty Cycle Measurement
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "Hz%" gear.
- Press F3 (SEL) to select frequency or duty cycle.

Image: Display examples for Frequency, Duty Cycle, Diode Test, and Continuity measurements.
Resistance Measurement
Warning: Disconnect power and discharge capacitors before resistance measurements.
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "Ω" function. "RESISTANCE" appears.
- Connect meter pen across circuit/component (ideally disconnected from line).
- Read resistance value.
Diode Detection
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "Ω" function. "RESISTANCE" appears.
- Press F3 (SEL) to select diode detection. "DIODE" appears.
- Use meter pen across diode/PN junction.
Warning: Do not test diodes containing voltage.
On/Off Test (Continuity)
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "Ω" function. "RESISTANCE" appears.
- Press F3 (SEL) to select on-off test. "CONTINUITY" appears.
- Contact circuit with meter pen. Buzzer sounds if resistance < 50Ω.
Warning: Do not perform on-off tests on lines containing voltage.
Capacitance Measurement
Warning: Do not test capacitors containing voltage. Capacitance measurement cannot be used with manual ranges.
- Connect black test lead to "COM", red test lead to "VΩCx".
- Turn knob to "Cx" function. "CAPACITANCE" appears.
- Touch meter pen to capacitor.
- Read capacitance value.
Measurement Data Hold
Press "S" to hold the displayed reading. "STOP" icon appears. Press "S" again to resume. In data hold state, a small dynamic data window refreshes at the top right. Range control switches to manual mode. Exits hold state when switching ranges or functions.
Storage and Readout of Measurement Data
The instrument can save up to 100 DMM measurement data sets.
- Press "S" to hold current data.
- Press "S" again and hold for 2 seconds to enter the database.
- The LCD lists 10 save locations, indicating availability.
- If data exists, its value and unit are listed.
- Press "S" again and hold for 2 seconds to exit.
Tip: Use F1 (▼) to select next save location, F2 (►) to select next page, F3 (ERASE) to delete data, F4 (SAVE) to save data.
5. Maintenance
5.1. Battery Care
- Remove batteries if the meter is not used for a long period to prevent leakage.
- Do not mix old and new batteries.
- Do not attempt to charge non-rechargeable batteries.
5.2. General Care
- Keep Dry: If the meter gets wet, dry it immediately and do not use until completely dry.
- Temperature: Use and store at room temperature. Extreme temperatures can shorten electronic life and deform plastic parts.
- Handling: Handle the instrument carefully to avoid drops that may damage the LCD or internal components.
- Cleaning: Wipe the housing frequently with a damp cloth and a tiny amount of mild detergent. Do not use rough objects, chemical solvents, or alcohol.
5.3. Fuse Replacement
- Remove probes from test points and turn off power.
- Loosen the set screw, remove the back cover. The fuse is on the back of the test socket.
- Remove the burnt fuse and replace with a new fuse of the same specification: 750mA/250V for 600mA circuits, 10A/250V for 10A circuits.
- Install the back cover and secure with screws.
Warning: Disconnect probes from any voltage source before opening the battery back cover. Do not use the meter until the back cover is secured.
6. Troubleshooting
If your instrument malfunctions, check the following before seeking repair:
6.1. No Screen Display, No Key Response
- Check battery power. Ensure batteries are not drained and are properly connected.
- Turn the knob to the "OFF" position, then turn it to a measurement position.
6.2. Unable to Measure
- Check the test leads for continuity (short connect them in the on/off gear).
6.3. Current Cannot Be Measured
- The fuse may be melted. Replace the fuse as described in the Maintenance section.
7. Specifications

Image: Detailed product parameter table for the TOOLTOP ET828Pro, including bandwidth, sampling rate, display, battery, storage, size, weight, and fuse specifications.
7.1. General Features
| Parameter | Value |
|---|---|
| Display | 320 x 240 Color Screen |
| Back Light | Ever Bright |
| Battery | AA 1.5V batteries x 3 |
| Static operating current (Multimeter) | Approximately 75mA |
| Static operating current (Oscilloscope) | Approx. 150mA |
| Usage environment | 0°C ~ +40°C; <75%RH |
| Storage environment | -10°C ~ +60°C; <90%RH |
| Outline dimensions | 83 mm × 160 mm × 32mm |
| Weight | About 190g (without battery) |
| Observation area | 49.0mm × 36.7mm |
| Input resistance | About 10MΩ |
| Automatic sleep | 15 minutes or disabled |
| Shutdown Current | <10μΑ |
| Save Capability | 100 records of DMM data, 10 records of OSC waveforms |
7.2. Oscilloscope Features
| Parameter | Value |
|---|---|
| Simulation bandwidth | 10MHz (500mV/div ~ 20V/div) |
| Maximum Equivalent Sampling Rate | 50Msps |
| Number of channels | 1 |
| Input Resistance | About 5MΩ |
| Vertical sensitivity range | 500mV~200V/div |
| Vertical range Accuracy | ±(5%+ 0.2div) |
| Time base range | 200ns~10s/div |
| Time Base Accuracy | ±(0.01%+ 0.1div) |
| Trigger Mode | Automatic / Single / Conventional Mode |
| Trigger edge selection | Rising/falling edge |
| Automatic settings | Automatic setting of time base and vertical range |
| Automatic measurement function | Vp-p, Vavg, Vrms, Hz |
7.3. Multimeter Features
All range uncertainties are expressed as ± (a% reading + word count). Correction period of one year. Guaranteed uncertainty environmental conditions are: 23°C±5°C, <75%RH.
| Function | Measurement range | Resolution | Uncertainty |
|---|---|---|---|
| DC Voltage | 600mV/6V/60V/600V/1000V | 0.1mV/1mV/10mV/100mV/1V | ± 1.2%rdg + 5dgt |
| AC Voltage | 600mV/6V/60V/600V/750V | 0.1mV/1mV/10mV/100mV/1V | ± 1.2%rdg + 5dgt (50Hz~60Hz) ± 1.5%rdg + 5dgt (100-400Hz) |
| DC Current | 60mA/600mA, 6A/10A | 10μΑ/100μΑ, 1mA/10mA | ± 1.2%rdg + 5dgt ± 1.5%rdg + 5dgt |
| AC Current | 60mA/600mA, 6A/10A | 10μΑ/100μΑ, 1mA/10mA | ± 1.2%rdg + 5dgt (50Hz~1kHz) ± 2%rdg + 10dgt (50Hz~1kHz) |
| Resistance | 600Ω/6kΩ/60kΩ/600kΩ/6MΩ/60MΩ | 0.1Ω/1Ω/10Ω/100Ω/1kΩ/10kΩ | ± (1.0%rdg + 5dgt) ± (3.0%rdg + 5dgt) / 60MΩ |
| Capacitance | 99.99nF/999.9nF/9.99µF/99.99µF/999.9µF, 9.999mF/99.99mF | 10pF/100pF/1nF/10nF/100nF, 1µF/10uF | ± (3.0%rdg + 10dgt) ± (5.0%rdg + 15dgt) |
| Frequency | 1.000Hz~20.00MHz (range > 2Vp-p) | ± (0.5%rdg + 5dgt) (20M or more for reference only) | |
| Diode/on/off test | Open circuit voltage about 2.8V; judgment resistance: about 50Ω | ||
| Fuse specification | 750mA/250V, 10A/250V |
7.4. Display Symbols and Icons
| Symbol | Meaning | Symbol | Meaning |
|---|---|---|---|
| DC | Direct Current | AC | Alternating Current |
| A | Ampere | WAVE | Signal Type |
| mA | Milliampere | BL | Back Light |
| V | Volt | mV | Millivolt |
| F | Farad | mF | Millifarad |
| nF | Nanofarad | μF | Microfarad |
| S | second | ms | Millisecond |
| μs | Microsecond | ns | nanosecond |
| Hz | Hertz | Ω | Ohm |
| AUTO | Auto Range / Auto Scan / Auto Set | P-H | Peak hold |
| REL | Relative Value | REST | Peak reset |
| MAX | Maximum value | MIN | Minimum value |
| STOP | Pause/hold function | EXIT | Exit the current state or menu |
| COM | Measurement reference (near the input reference socket) | avg | Average value |
| ERASE | Record Delete | LEVEL | Trigger Level |
| SAVE | Save current data | RESET | Trigger level zeroing |
| Normal | Normal trigger method | Single | Single trigger method |
| Auto | Auto trigger method | Trig | Trigger in progress |
| Wait | Ready and waiting for trigger | Stop | Hold waveform |
| TIME | Time base | VOLT | Select range |
| ◄► | Option Adjustment | ▲▼ | Option Adjustment |
| Battery level | Diode | ||
| Trigger edge | 4 | Safety Warning |
8. User Tips
- Current Measurement in Oscilloscope View: For the ET828Pro, when measuring current (Amperes), you can display the results in oscilloscope view (Amperes on the Y-axis) by pressing the 'R' button. This allows for visual analysis of current waveforms.
- Optimizing Waveform Capture: If you are having trouble capturing a stable waveform, try adjusting the time base or pressing the 'R' key for automatic capture. Starting with a faster time base and gradually slowing it down can help in correctly displaying signals of unknown frequencies.
9. Warranty and Support
This product is a precision instrument. Unauthorized modification of wiring, replacement of components, or product calibration and repair work are not allowed without authorization from our Product Service Center.
The use of this product under warranty or normal wear and tear, as well as damage caused by the user's own operation or improper storage, and any loss outside the scope of the company's after-sales service, are not covered by compensation.
For further assistance or to view the original user manual, please refer to the User Manual (PDF).