AZ8403 Portable Digital Dissolved Oxygen Meter

User Manual

1. Introduction

The AZ8403 Portable Digital Dissolved Oxygen Meter is a high-precision instrument designed for measuring dissolved oxygen (D.O.) and temperature in various liquid environments. It is widely used in applications such as aquaculture, agriculture, industrial water quality testing, and environmental research. This meter features a large LCD display, easy calibration, and advanced compensation functions for accurate readings.

Overall product display of the AZ8403 Dissolved Oxygen Meter, showing the main unit, probe, and connecting cable.
Figure 1: Overall product display of the AZ8403 Dissolved Oxygen Meter.

2. Features

  • Large LCD with D.O. & Temperature value display and stability indicator.
  • Easy 100% Dissolved Oxygen Value calibration.
  • Utilizes a cost-effective Clark type electrode.
  • DO probe does not require frequent electrolyte refills.
  • Accurate dissolved oxygen measurement.
  • Manual salinity and barometric pressure compensation.
  • Stores and recalls 99 data points with temperature readings.
  • Real-time clock for each memory data.
  • USB interface for PC connectivity and IrDA interface.
  • Automatic temperature compensation.
  • IP67 waterproof rating.

3. Package Contents

Upon unpacking, please verify that all items listed below are present and in good condition:

  • AZ8403 Dissolved Oxygen Meter main unit
  • DO Probe with 3-meter cable
  • Operation Manual
  • Carrying Case
  • Electrolyte Solution
  • Replacement Membrane Pack (O-rings included)
  • Syringe for electrolyte filling
  • Small screwdriver (for maintenance)
  • USB Cable & Software (optional, for PC link)
Contents of the AZ8403 Dissolved Oxygen Meter kit, including the meter, probe, manual, electrolyte, and tools in a carrying case.
Figure 2: AZ8403 Dissolved Oxygen Meter and accessories in carrying case.

4. Setup

4.1 Probe Assembly and Connection

  1. Ensure the probe membrane is properly installed and filled with electrolyte solution (refer to Section 6.1 for membrane change instructions).
  2. Connect the probe cable to the probe interface on the top of the main meter unit.

4.2 Powering On

Press the power button (usually the top-left button) to turn on the meter. The LCD display will light up.

5. Operating Instructions

5.1 Function Key Description

Diagram showing the function of each button on the AZ8403 meter: Probe interface, Large HD screen, SET (Press for switch, Long-press for Settings), MODE (Increment value, Switching mode), CAL/Esc (Correcting key, Exit set mode), MEM (Save data, Reduce values/switch modes), HLD/Rec (Tap for backlight, Long press to view data), M/AV (Confirm key).
Figure 3: Function key description.
  • SET: Press to switch, long-press for settings.
  • MODE: Increment value, switch modes.
  • CAL/Esc: Correcting key, exit set mode.
  • MEM: Save data, reduce values/switch modes.
  • HLD/Rec: Tap for backlight, long-press to view data.
  • M/AV: Confirm key.

5.2 Calibration (100% Saturation Calibration)

The AZ8403 meter features an easy 100% saturation calibration in air, without the need for calibration solutions.

  1. Turn on the meter.
  2. Leave the DO probe in the air.
  3. Ensure the DO unit displayed on the top right corner is in percentage (%).
  4. Wait for the DO reading to stabilize.
  5. Press the CAL/Esc button to start calibration. The 'CAL' icon will flash on the LCD.
  6. Wait for a few seconds until the reading stabilizes again.
  7. Press the M/AV (Enter) button. The meter will automatically calibrate to 100% saturated dissolved oxygen and return to measurement mode.
Video 1: How to Calibrate AZ Dissolved Oxygen meter. This video demonstrates the 100% saturation calibration process in air.
Diagram illustrating the highly sensitive waterproof probe and the ease of saturation calibration in air.
Figure 4: Highly sensitive waterproof probe and air calibration.

5.3 Measurement

After successful calibration, immerse the probe into the liquid sample you wish to measure. The meter will display the dissolved oxygen content and temperature. The meter also supports manual salinity and barometric pressure compensation for improved accuracy.

Image showing the AZ8403 meter measuring dissolved oxygen, temperature, and salinity in a liquid, highlighting its triple compensation function and IP67 waterproof rating.
Figure 5: Triple compensation function (Temperature, Dissolved Oxygen, Salinity) and IP67 waterproof rating.
Diagram showing the AZ8403 meter with features like pressure compensation, self-calibration, and salinity compensation.
Figure 6: Pressure, self-calibration, and salinity compensation features.

The meter can store up to 99 data points with corresponding temperature readings. It also features a USB interface for connecting to a PC to view graphs and record data in real-time. A separate RS232 software data cable is required for this functionality.

Image showing the AZ8403 meter connected to a laptop via a cable, displaying real-time data and graphs on the computer screen.
Figure 7: Connecting to a computer for real-time data and graphs.

5.5 Automatic Shutdown

To conserve battery life and due to the oxygen consumption principle of the DO probe, the meter will automatically shut down after 20 minutes without any operation. The automatic shutdown time can be configured to 20, 30, 40, 60, 90, or 120 minutes.

Diagram showing the automatic temperature compensation and auto-shutdown feature of the AZ8403 meter, with selectable shutdown times.
Figure 8: Automatic temperature compensation and auto-shutdown feature.

6. Maintenance

6.1 Probe Membrane Change

The probe membrane is a consumable part and requires periodic replacement. Follow these steps:

  1. Remove the old membrane from the probe.
  2. Put the new membrane on the top & center of the ring.
  3. Find out the four alignment points on the cap before putting it on the ring.
  4. Put the cap onto the membrane. Make sure the alignment points on the cap and ring are matched.
  5. Put the black O-ring on the top of the sensor. Make sure the O-ring is flat.
  6. Find out the alignment points inside the ring of the semi-assembly part from step 4.
  7. Put the semi-assembly part (finished in step 4) on the top of the O-ring.
  8. Put on the probe guard and screw tight.
  9. Run the calibration. If E17 appears on the meter LCD, slightly loosen the outer cover until a reading appears on the LCD and then run calibration again.
  10. After calibration, turn the probe guard tightly and run the calibration again.
Step-by-step diagram for changing the probe membrane, including visual instructions for each step.
Figure 9: Probe membrane change instructions.

6.2 Electroltye Refill

The Clark type electrode used in the AZ8403 does not require frequent electrolyte refills compared to galvanic electrodes. However, ensure the probe is adequately filled with the provided electrolyte solution during membrane changes or if readings become unstable.

7. Troubleshooting

7.1 E31 Error on Temp. and D.O. Reading

Reason: The inside wire of the probe is broken.

Solution: Cannot be fixed by the user. The probe needs to be replaced. Refer to Section 6.1 for membrane change, as sometimes a faulty membrane installation can mimic this error.

8. Specifications

Parameter Model 8403
D.O. range (in %) 0.0~199.9%
D.O. accuracy +/-1.5% F.S
D.O. resolution 0.1%
DO range (in ppm or mg/l) 0.00~19.99ppm
DO accuracy +/-1.5% F.S
DO resolution 0.01
Temp. range 0~50.0°C
Temp. accuracy +/-0.3°C
Temp. resolution 0.1°C
Manual Compensation Salinity: 0.0~50.0 ppt, Pressure: 500~1499mmHg/66.6~199.9KPA
Memory 99 points
Real Time/Recall Included
IrDA Port Included
LCD size 44(H)x26(W)mm
Operating temp. 0~50°C
Operating RH% Humidity <80%
Storage temp. 0~60°C
Storage RH% Humidity <90%
Sensor life time >6 months (with good maintenance)
Dimension 175(L)x70(W)x33(H)
Weight 150g
Battery AAA x 4pcs
Standard Package Meter/Battery/Probe(840P)/Manual/Carry case Electrolyte/Syringe/membrane/o-ring
Detailed specifications table for AZ Dissolved Oxygen Meter models 8401, 8402, and 8403, covering DO range, accuracy, resolution, temperature, salinity, pressure, memory, and interfaces.
Figure 10: Detailed Specifications for AZ8403.

9. User Tips

  • Non-Sleep Mode: If continuous operation is required without automatic shutdown, you can activate a non-sleep mode. First, turn off the device. Then, press and hold both the POWER and HLD keys simultaneously to power on the device. 'nSL' (non-sleep) will be shown on the display. Release the keys, and the meter will remain on indefinitely until manually turned off.
  • Calibration Frequency: Calibrate the meter in air whenever there is a significant change in weather or ambient temperature to ensure accuracy.
  • Probe Care: The probe is a consumable item. Handle it with care and ensure proper maintenance, including membrane changes and electrolyte levels, to prolong its lifespan.

10. Warranty and Support

This product is manufactured by AZ Instrument. For warranty information, technical support, or to inquire about spare parts (such as replacement probes or membrane packs), please contact AZ Instrument directly through their official channels. Keep your purchase receipt as proof of purchase for any warranty claims.

Ask a question about this manual