1. Introduction
This manual provides detailed instructions for the Hantek DSO5000P Series Digital Storage Oscilloscopes, including models DSO5072P, DSO5102P, and DSO5202P. It covers essential information for safe and efficient use, from initial setup to advanced measurement techniques and troubleshooting.
The DSO5000P Series oscilloscopes offer bandwidths from 70MHz to 200MHz, real-time sample rates up to 1GSa/s, and a 7-inch color TFT LCD display (800x480 pixels) with a user-friendly interface. Key features include a record length up to 24K, various trigger modes (edge, pulse width, video, slope, overtime), USB host and device connectivity, multiple automatic measurements, and four math functions including FFTs.
Product Overview Video: Demonstrates the oscilloscope's features and interface.
2. Safety Tips
To prevent injury and damage to the product, observe the following safety precautions:
- Qualified Personnel: Only qualified personnel should perform maintenance.
- Power Cord: Use only the power cord specified for this product and certified for your country.
- Proper Connection: Connect probes to the oscilloscope before connecting to circuits. Disconnect probes from circuits before disconnecting from the oscilloscope.
- Grounding: Ensure the product is properly grounded through the power cord's grounding conductor.
- Probe Ground Lead: The probe ground lead is at ground potential. Do not connect it to an elevated voltage.
- Terminal Ratings: Check all terminal ratings to avoid fire or shock hazards.
- Operation: Do not operate without covers, with exposed circuitry, or with suspected failures.
- Environment: Assure good ventilation. Do not operate in wet/damp environments or explosive atmospheres.
- Maintenance: Keep product surfaces clean and dry.
2.1 Safety Terms and Symbols
WARNING: Indicates conditions or practices that could result in injury or loss of life.
CAUTION: Identifies conditions or practices that could result in damage to the product or other property.
DANGER: Indicates an injury hazard immediately accessible.
Symbols on Product:
- Protective Ground (Earth) Terminal: Indicates the connection point for protective earth.
- Measurement Ground Terminal: Indicates the measurement ground connection.
- CAUTION Refer to Manual: Advises consulting the manual for specific instructions.
- Mains Disconnected OFF (Power): Indicates the device is disconnected from mains power.
- Mains Connected ON (Power): Indicates the device is connected to mains power.
- High Voltage: Warns of high voltage presence.
3. Setup and Installation
3.1 Installation
To ensure proper ventilation, leave a space of more than 5cm (2 inches) around the top and sides of the oscilloscope.
3.1.1 Power Supply
Use a power supply that delivers 90 to 240 VRMS, 45 to 440 Hz.
3.1.2 Power Cord
Only use power cords designed for this product. Refer to the accessories section for specific standards.
3.2 Functional Check
Perform a quick functional check after installation:
- Power on the oscilloscope: Plug in the oscilloscope and press the ON/OFF button. Then push the DEFAULT SETUP button. The default Probe option attenuation setting is 10X.
- Connect the oscilloscope: Set the switch on the probe to 10X. Connect the probe to Channel 1 (CH1) on the oscilloscope by aligning the slot and turning to lock. Connect the probe tip and reference lead to the PROBE COMP connectors (~5V@1KHz).
- Observe the waveform: Press the AUTOSET button. A square wave of approximately 5V peak-to-peak at 1kHz should appear. Press the CH1 MENU button twice to remove Channel 1. Repeat for Channel 2 if desired.

Front Panel of the DSO5000P Series Oscilloscope.

Side view of the oscilloscope, showing the handle and ventilation.
3.3 Probe Examination
3.3.1 Safety
Always keep fingers behind the guard on the probe body to avoid electric shock. Do not touch metallic parts of the probe head when connected to a voltage source. Connect the probe to the oscilloscope and ground terminal to ground before measurements.
3.3.2 Use of Probe Check Wizard
Every time a probe is connected, use the probe check wizard to verify correct operation. This can be done by:
- Using the vertical menu (e.g., CH1 MENU button) to set the Probe option attenuation factor.
- Pressing the PROBE CHECK button and following menu prompts.
3.4 Manual Probe Compensation
Adjust probe compensation to match the probe to the input channel. Uncompensated probes can lead to measurement errors.
- Set Probe option attenuation to 10X in the channel menu. Set the probe switch to 10X. Connect the probe to Channel 1, attaching the tip to PROBE COMP ~5V@1KHz and the reference lead to PROBE COMP Ground. Display the channel and press AUTOSET.
- Check the displayed waveform shape. It should be a correctly compensated square wave.
- If necessary, use a nonmetallic screwdriver to adjust the variable capacity of your probe until the waveform is correctly shaped.
3.5 Probe Attenuation Setting
Probes have various attenuation factors that affect the vertical scale. Ensure the probe attenuation option matches the probe's attenuation switch setting (1X or 10X). When set to 1X, the probe limits the oscilloscope bandwidth to 6MHz. For full bandwidth, set the switch to 10X.
3.6 Self Calibration
The self-calibration routine optimizes signal path for maximum measurement accuracy. Run it if the ambient temperature changes by 5°C or more. For best accuracy, power on the oscilloscope and wait 20 minutes for it to warm up. Disconnect all probes and cables from front-panel input connectors. Push the UTILITY button, select Do Self Cal, and follow on-screen directions.
4. Operation
4.1 Main Feature Description
The oscilloscope's front panel is divided into several functional areas with control buttons and knobs. The screen displays information and relative testing operations.
4.1.1 Oscilloscope Setup
- Autoset: Automatically adjusts horizontal and vertical scales, trigger coupling, type, position, slope, level, and mode for a stable waveform display.
- Saving a Setup: The oscilloscope saves the current setup automatically before closing and recalls it on power-up. You can permanently save up to 10 settings.
- Recalling a Setup: Recall any saved setup or the default factory setup.
- Default Setup: The oscilloscope's factory preset for normal operations.
4.1.2 Trigger
The trigger determines when the oscilloscope acquires data and displays a waveform. Proper trigger setup converts unstable displays into meaningful waveforms.
- Trigger Source: Can be an input channel (CH1 or CH2) or an external signal.
- Trigger Type: Six types are available: Edge, Video, Pulse Width, Slope, Overtime, and Swap.
- Trigger Mode: Auto or Normal mode defines how data is acquired without a trigger condition. Auto mode acquires freely; Normal mode updates only with a valid trigger.
- Trigger Coupling: Determines which part of the signal is delivered to the trigger circuit for stable display.
- Trigger Position: Horizontal position control sets the time between the trigger position and the screen center.
- Slope and Level: Define the trigger point. Slope determines rising or falling edge; Level controls the trigger point's vertical position.
4.1.3 Data Acquisition
Analog signals are converted into digital data. Two types of acquisition exist: Real-time and Equivalent.
- Real-time Acquisition Modes:
- Normal: Samples at evenly spaced intervals, accurately representing signals. May miss rapid variations or narrow pulses.
- Peak Detect: Captures maximum and minimum values over each sample interval, useful for narrow pulses and glitches. Noise may appear higher.
- Average: Acquires and averages several waveforms to reduce random noise.
- Equivalent Acquisition: Used for periodic signals when real-time acquisition rate is too low. It reconstructs waveforms by acquiring data with tiny delays over multiple frames.
- Time Base: Controls how often values are digitized. Use the SEC/DIV knob to adjust.
4.1.4 Waveform Scaling and Positioning
Adjust the display of waveforms by changing their scale and position.
- Vertical Scale and Position: Move waveforms up/down (position) or expand/contract vertically (scale) using the VOLTS/DIV button.
- Horizontal Scale and Position: Adjust the HORIZONTAL POSITION control to view waveform data before or after the trigger. The SEC/DIV knob changes the horizontal scale.
4.1.5 Waveform Measurement
Measure displayed waveforms using graticules, cursors, or automatic measurements.
- Graticule: Quick visual estimation by counting divisions and multiplying by the scale factor.
- Cursor: Move amplitude (horizontal broken line) and time (vertical broken line) cursors to measure specific points. Push the CURSOR button to use.
- Automatic Measurement: Oscilloscope performs calculations automatically, providing precise readouts that are periodically updated.
4.2 Basic Operation
4.2.1 Display Area
The display area shows waveforms and measurement information. It includes:
- Display Format: YT (voltage vs. time), XY (voltage on CH1 vs. CH2), Vectors (fills space between points), Dots (displays sample points).
- Persistence: Gray indicates auto persistence; Green means persistence display is enabled.
- Acquisition Mode: Normal, Peak Detect, or Average.
- Trigger Status: Icons indicate pretriggered data acquisition, ready for trigger, trigger detected, auto mode acquisition, continuous scan mode, stopped acquisition, or single sequence acquisition.
- Tool Icons: Indicate keyboard lock by host computer (USB), USB disk connected, or USB slave interface connected.
- Readouts: Main time base setting, window position, data length, frequency count, horizontal waveform position, trigger type.
4.2.2 XY Format
Used to analyze phase differences (Lissajous patterns). Plots CH1 voltage (horizontal axis) against CH2 voltage (vertical axis). Uses untriggered Normal acquisition mode, displays data as dots, and has a fixed sampling rate of 1 MS/s. To use, stop acquisition and change display format to XY.
4.2.3 Horizontal Controls
Change horizontal scale and position of waveforms. The horizontal position readout shows time relative to the screen center (trigger time as zero). The SEC/DIV knob expands or contracts the waveform. The HORIZONTAL POSITION knob controls the trigger position.
- Scan Mode Display (Roll Mode): When SEC/DIV is 80ms/div or slower and trigger mode is Auto, waveforms update left to right without trigger or horizontal position control.
4.2.4 Vertical Controls
Display/remove waveforms, adjust vertical scale/position, set input parameters, and perform math calculations. Each channel has a separate vertical menu.
- VERTICAL POSITION Knob: Moves channel waveform up/down. Push to return to vertical center.
- Menu (CH1, CH2): Displays vertical menu options, turns channel waveforms on/off.
- Coupling: DC (passes DC/AC), AC (blocks DC), Ground (disconnects input signal).
- 20MHz Bandwidth Limit: Reduces display noise.
- VOLTS/DIV: Selects resolution (Coarse/Fine).
- Probe Attenuation: Selects value based on probe attenuation factor.
- Invert: Inverts waveform relative to reference level.
- VOLTS/DIV Knob: Magnifies or attenuates source signal.
- MATH MENU: Displays waveform math operations (+, -, x, ÷, FFT).
4.2.5 Math FFT (Fast Fourier Transform)
Converts a time-domain (YT) signal into its frequency components (spectrum) to analyze harmonics, distortion, noise, impulse response, and vibration.
- Setting Time-domain Waveform: Set YT waveform by pressing AUTOSET, centering with VERTICAL POSITION knob, positioning with HORIZONTAL POSITION knob, scaling with VOLTS/DIV knob, and setting resolution with SEC/DIV knob.
- Nyquist Frequency: Highest frequency measurable without errors (half of sample rate). FFT spectrum contains 1024 points from DC to Nyquist frequency.
- Displaying FFT Spectrum: Push MATH MENU, select Source channel, Window algorithm, and FFT Zoom factor.
- Selecting FFT Window: Hanning (better frequency, poorer amplitude), Flattop (better amplitude, poorer frequency), Rectangular (special-purpose, for discontinuous waveforms).
- FFT Aliasing: Occurs when time-domain waveform contains frequency components higher than Nyquist frequency, displayed as lower frequency components.
- Eliminating Aliases: Use faster sample rates (SEC/DIV), FFT Zoom, Bandwidth Limit (20MHz), filter input signal, or identify/ignore aliased frequencies.
- Magnifying and Positioning FFT Spectrum: Use FFT Zoom for horizontal magnification and vertical controls for vertical magnification.
- Using Cursors to Measure FFT Spectrum: Use cursors to measure amplitude (in dB) and frequency (in Hz).
4.2.6 Trigger Controls
Defined through the Trigger Menu and front-panel controls. Six types: Edge, Video, Pulse Width, Swap, Slope, and Overtime.
- Level: Amplitude level signal must cross for Edge or Pulse Width trigger.
- Set to 50%: Sets trigger level to vertical midpoint of trigger signal peaks.
- Force Trigger: Completes acquisition regardless of adequate trigger signal (useless if acquisition is stopped).
- TRIG MENU: Displays trigger menus.
4.2.7 Menu and Option Buttons
Six buttons at the top of the front panel recall relative setup menus:
- SAVE/RECALL: Displays menu for setups and waveforms.
- MEASURE: Displays the Measure menu.
- ACQUIRE: Displays the Acquire menu.
- UTILITY: Displays the Utility menu.
- CURSOR: Displays the Cursor menu.
- DISPLAY: Displays the Display menu.
4.2.8 Fast Action Buttons
- AUTOSET: Automatically sets controls for a usable display of input signals.
- SINGLE SEQ: Acquires a single waveform and then stops.
- RUN/STOP: Continuously acquires or stops waveforms.
- HELP: Displays the Help menu.
- DEFAULT SETUP: Recalls the factory setup.
- Save to USB Disk: Saves all displays on screen to a USB device.
4.2.9 Signal Connectors

Front view of the oscilloscope, showing the BNC input connectors and controls.
- CH1, CH2: Input connectors for waveform display.
- EXT TRIG: Input connector for an external trigger source.
- Probe Compensation: Voltage probe compensation output and ground. Used to electrically match the probe to the oscilloscope input circuit. Do not connect a voltage source to these ground terminals.
5. Application Examples
This section provides examples to help solve common test problems.
5.1 Taking Simple Measurements
To quickly measure frequency, period, and peak-to-peak amplitude of an unknown signal:
- Set probe switch to 10X; push CH1 MENU and set Probe option attenuation to 10X.
- Connect CH1 probe to the circuit test point.
- Press AUTOSET. The oscilloscope will automatically set the waveform for best display.
5.2 Taking Cursor Measurements
To measure time and amplitude of a waveform:
- Push CURSOR button to view Cursor menu.
- Push F1 (Type) and select Time. Push F2 or F3 (Source) and select CH1.
- Push F4 to select a cursor (S or E). Turn V0 knob to move the cursor.
- Place Cursor S on the first peak and Cursor E on the second peak of the ring. Delta will show measured time.
- Push F1 (Type) and select Voltage. Place Cursor S on the highest peak and Cursor E on the lowest point. Delta will show amplitude.
5.3 Analyzing Input Signals to Eliminate Random Noise
To display and analyze noisy signals:
- Press ACQUIRE button, select Real Time (Type) and Peak Detect (Mode).
- If needed, push DISPLAY and set Contrast for clearer noise view.
- To eliminate noise, select Average (Mode) and adjust the number of running averages.
5.4 Capturing Single-shot Signal
To capture aperiodic signals like pulses and glitches:
- Set probe and CH1 attenuation to 10X.
- Adjust VOLTS/DIV and SEC/DIV knobs for proper signal examination.
- Press ACQUIRE, select Peak Detect (Mode).
- Push TRIG MENU, select Rising (Slope) and adjust trigger level.
- Push SINGLE SEQ to start acquisition.
5.5 Using X-Y Mode
To view phase differences between two channel signals:
- Prepare two 10X probes. Push CH1 MENU and CH2 MENU, set Probe option attenuation to 10X for both.
- Connect CH1 probe to network input, CH2 probe to output.
- Push AUTOSET. Adjust VOLTS/DIV knobs for similar amplitude signals.
- Push DISPLAY, select XY (Format). The oscilloscope displays a Lissajous pattern.
- Adjust VOLTS/DIV and VERTICAL POSITION knobs to scale waveform.
5.6 Triggering on Pulse Width
To verify pulse width:
- Set probe attenuation to 10X. Push AUTOSET for stable waveform.
- Push SINGLE CYCLE (Autoset menu) and read pulse width.
- Push TRIG MENU. Push F1 (Type) and select Pulse. Push F2 (Source) and select CH1. Turn TRIGGER LEVEL knob to set trigger level at signal bottom.
- Push F6 (Next Page). Select '=' (When). Push F4 (Set Pulse Width). Turn V0 knob to set pulse width to the read value.
5.7 Triggering on Video Signal
To monitor video signals (e.g., NTSC system):
- Push TRIG MENU. Push F1 (Type) and select Video.
- Push Source and select CH1. Push Polarity and select Normal. Push Standard and select NTSC.
- Push Sync and select Odd Field, Even Field, or All Fields.
- Turn TRIGGER LEVEL knob to adjust and stabilize video signals.
- Adjust horizontal SEC/DIV and vertical VOLTS/DIV knobs to display a complete video signal triggering on a video field.
5.8 Using Slope Trigger to Capture Particular Slope Signal
To observe rise and fall times of a signal:
- Push TRIG MENU. Push F1 (Type) and select Slope.
- Push Source and select CH1. Push Slope and select Rising. Push Mode and select Auto. Push Coupling and select DC.
- Click 'Next Page' and select Vertical. Adjust V1 and V2 with V0 knob. Select '=' (When).
- Select 'Time' and turn V0 to adjust time for stable display.
5.9 Using Overtime Trigger to Measure Long Pulse Signal
To observe parts of a long pulse signal:
- Push TRIG MENU. Push F1 (Type) and select OT.
- Push Polarity and select Normal. Push Mode and select Auto. Push Coupling and select DC.
5.10 Using Math Functions to Analyze Waveforms
To analyze input waveforms using math functions (e.g., instantaneous difference):
- Set Probe option attenuation to 10X.
- Open CH1 and CH2, both with 10X attenuation.
- Push AUTOSET for stable waveform.
- Push MATH MENU. Push Operation and select 'CH1+CH2' (or other operations like -, x, ÷, FFT).
- Adjust horizontal SEC/DIV and vertical VOLTS/DIV knobs for easy check.
Note: Compensate both probes before performing math operations to avoid errors.
5.11 Measuring Data Propagation Delay
To measure propagation delay between enable signal and transfer data:
- Connect two oscilloscope probes to CS (chip-select) and DATA pins on the chip.
- Set Probe option attenuation to 10X for both probes.
- Open CH1 and CH2, both with 10X attenuation.
- Push AUTOSET for stable waveform.
- Adjust horizontal and vertical controls to optimize display.
- Push CURSOR. Push Type and select Time.
- Select Cursor S and turn V0 to place on active edge of enable signal.
- Select Cursor E and turn V0 to place on data output transition.
- Read data propagation delay in Delta readout.
6. Maintenance
6.1 General Care
- Do not expose the LCD display to direct sunlight for long periods.
- Do not expose the oscilloscope or probes to sprays, liquids, or solvents to avoid damage.
6.2 Cleaning
Examine and clean the oscilloscope and probes as often as operating conditions require:
- Use a lint-free cloth to remove dust from the exterior, avoiding scratching the display filter.
- Dampen a soft cloth with water or a 75% isopropyl alcohol aqueous solution to clean.
Note: Do not use corrosive or chemical cleaning agents.
7. Troubleshooting
7.1 Problem Settlement
- Oscilloscope does not start up:
- Check power cord connection.
- Check power on/off button.
- Restart oscilloscope.
- No waveform display:
- Check probe connection to input BNC.
- Check channel switch (e.g., CH1, CH2 menu buttons) is on.
- Check input signal connection to probe.
- Affirm measured circuits have signals.
- Turn up magnitude for large DC signals.
- Press Auto Measure button.
- Distorted input signal waveform:
- Check probe connection to channel BNC.
- Check probe connection to measured object.
- Verify probe calibration; calibrate if necessary.
- Waveform rolling continuously, cannot trigger:
- Check trigger source consistency with input channel.
- Check trigger level adjustment; reset to 50% with TRIGGER LEVEL knob or SET TO 50% button.
- Check trigger mode suitability for input signal (default is edge trigger, not always suitable).
8. Specifications
All specifications apply to DSO5000P series oscilloscopes under specified operating conditions (20 minutes continuous operation, ambient temperature changes less than 5°C since last self-calibration, within factory calibration interval). All specifications are typical unless noted.
8.1 Technical Specifications

Detailed technical specifications for Hantek DSO5000P Series.
Horizontal
- Sample Rate Range: 1GS/s
- Waveform Interpolation: (sin x)/x
- Record Length: Max 40K samples per single-channel; Max 20K samples per dual-channel (4K, 20K optional)
- SEC/DIV Range: 2ns/div to 80s/div, in a 2, 4, 8 sequence
- Sample Rate and Delay Time Accuracy: ±50ppm over any ≥1ms time interval
- Delta Time Measurement Accuracy (Full Bandwidth): Single-shot, Normal mode: ± (1 sample interval +100ppm x reading + 0.6ns); >16 averages: ± (1 sample interval + 100ppm x reading + 0.4ns)
- Position Range: (-4div × s/div) to 20ms for 2ns/div to 8ns/div; (-8div × s/div) to 40ms for 20ns/div to 80µs/div; (-8div × s/div) to 400s for 200µs/div to 80s/div
Vertical
- A/D Converter: 8-bit resolution, each channel sampled simultaneously
- VOLTS/DIV Range: 2mV/div to 10V/div at input BNC
- Offset Range: ±400mV (2mV/div to 20mV/div); ±2V (50mV/div to 200mV/div); ±40V (500mV/div to 2V/div); ±50V (5V/div to 10V/div)
- Selectable Analog Bandwidth Limit, typical: 20MHz
- Low Frequency Response (-3db): ≤10Hz at BNC
- Rise Time at BNC, typical: <5.0ns (DSO5072P), <3.5ns (DSO5102P), <1.8ns (DSO5202P)
- DC Gain Accuracy: ±3% (Normal or Average, 10V/div to 10mV/div); ±4% (Normal or Average, 5mV/div to 2mV/div)
- DC Measurement Accuracy, Average Acquisition Mode: ± (3% × reading + 0.1div + 1mV) for 10mV/div or greater; ± [3% × (reading + vertical position) + 1% of vertical position + 0.2div] for others.
- Volts Measurement Repeatability, Average Acquisition Mode: Delta volts between any two averages of ≥16 waveforms.
Note: Bandwidth reduced to 6MHz when using a 1X probe.
Trigger
- Trigger Sensitivity (Edge Trigger Type): CH1/CH2: 1div from DC to 10MHz; 1.5div from 10MHz to Full (DSO5072P, DSO5102P); 1.5div from 10MHz to 100MHz; 2div from 100MHz to Full (DSO5202P). EXT: 200mV from DC to 100MHz. EXT/5: 1V from DC to 100MHz.
- Trigger Level Range: CH1, CH2: ±8 divisions from center of screen. EXT: ±1.2V. EXT/5: ±6V.
- Trigger Level Accuracy, typical: CH1, CH2: 0.2div × volts/div within ±4 divisions. EXT: ± (6% of setting + 40mV). EXT/5: ± (6% of setting + 200mV).
- Video Trigger Type: Supports NTSC, PAL and SECAM broadcast systems.
- Pulse Width Trigger: Selectable from 20ns to 10s.
- Slope Trigger: Selectable from 20ns to 10s.
Acquisition
- Acquisition Modes: Normal, Peak Detect, and Average.
- Acquisition Rate, typical: Up to 2000 waveforms per second per channel.
Inputs
- Input Coupling: DC, AC or GND.
- Input Impedance, DC coupled: 1MΩ±2% in parallel with 20pF±3pF.
- Probe Attenuation: 1X, 10X.
General Specifications
- Display Type: 7 inch 64K color TFT (diagonal liquid crystal), 800 horizontal by 480 vertical pixels.
- Output Voltage, typical: About 5Vpp into ≥1MΩ load.
- Frequency, typical: 1kHz.
- Power Supply: 100-120VACRMS(±10%), 45Hz to 440Hz, CAT II; 120-240VACRMS(±10%), 45Hz to 66Hz, CAT II.
- Power Consumption: <30W.
- Fuse: 2A, T rating, 250V.
- Environmental Temperature: Operating: 0°C to 50°C; Nonoperating: -40°C to +71°C.
- Humidity: ≤90% RH (≤40°C); ≤60% RH (41°C to 50°C).
- Altitude: Operating and Nonoperating: 3,000m (10,000 feet).
- Mechanical Shock: 50g, 11ms, half sine.
- Size: 313mm (Length) x 142mm (Height) x 108mm (Depth).
- Weight: 2.08Kg (exclusive of packing and accessories).
8.2 Accessories
The following standard accessories are included:
- Passive Probes: Two X1, X10 passive probes. 6MHz bandwidth (100Vrms CAT III) at X1; max 300Vrms CAT II at X10.
- USB A-B Line: For connecting to external devices (e.g., printer) or PC communication.
- Power Cord: Specific power cord for the product.
- Software Installation CD: Contains user manual and descriptions for DSO5000P series oscilloscopes.

Hantek DSO5000P Series Oscilloscope with included probes, power cord, USB cable, and software CD.
9. Warranty and Support
9.1 Harmful and Poisonous Substances or Elements
The product contains certain harmful and poisonous substances or elements as per SJ/T 11363-2006 standard. 'X' indicates content exceeds the limit, '0' indicates content is below the limit.
| Component | Pb | Hg | Cd | Cr(VI) | PBB | PBDE |
|---|---|---|---|---|---|---|
| Shell and Chassis | X | 0 | 0 | X | 0 | 0 |
| Display Module | X | X | 0 | 0 | 0 | 0 |
| Circuit Board | X | 0 | 0 | X | 0 | 0 |
| Power Supply | X | 0 | 0 | X | 0 | 0 |
| Electric Wire and Cable Assembly | X | 0 | 0 | 0 | 0 | 0 |
| Connector | X | 0 | 0 | X | 0 | 0 |
| Fastener and Installed Hardware | X | 0 | X | X | 0 | 0 |
| Other Accessories (including probes) | X | 0 | 0 | X | 0 | 0 |
| Others | 0 | 0 | 0 | 0 | 0 | 0 |
For detailed information, please refer to the official user manual PDF: Hantek DSO5000P Series User Manual (PDF).
10. User Tips
Based on common questions and user feedback, here are some helpful tips:
- General Use: This oscilloscope is an electrical DIY supply, suitable for various electrical projects.
- Display: The device features a large display, typically 7 inches or above, providing clear waveform visualization.
- Channels: It comes with two digital channels, allowing for simultaneous measurement and comparison of two signals.
- Battery: The battery is not designed to be removed by the user.
- Performance: Expect a real-time sampling rate of 1GSa/S and a bandwidth of 100-349MHz, depending on the specific model (DSO5072P, DSO5102P, DSO5202P).