1. Introduction
The Micsig DP Series High-Voltage Differential Probe is designed to convert high-voltage differential signals into single-ended low-voltage signals, which can then be displayed on an oscilloscope. These probes offer a bandwidth of 100-300 MHz and a maximum input voltage of 7000 Vpk. Equipped with a standard BNC interface, they are compatible with most oscilloscope brands.
Key features include one-button calibration, an overload alarm, range power-off memory, dual voltage ranges, and a high-resistance, low-capacitance design to minimize circuit loading. The probe ensures strong amplitude-frequency performance and a high common-mode rejection ratio. Command-based programming enables automated testing, and a 5 MHz bandwidth-limit function helps suppress high-frequency noise, resulting in clearer waveforms.
2. Package Contents
The following items are typically included in the product package:

Figure 1: Package Contents Overview
- DP Series Probe x1 (Main unit)
- Extendable Hook Pliers x1 Pair
- Alligator Clips x1 Pair
- Input Extension Cables x1 Pair
- Power Adapter x1
- BNC Output Line x1
- USB Line x1
- Quick Guide x1

Figure 2: Probe and Accessories

Figure 3: Product Packaging
3. Product Features

Figure 4: Key Features Overview
Command & Programming Control
The Type-C interface supports data communication, enabling connection to a computer for command control.
5MHz Bandwidth Limit
This feature is ideal for measuring switching frequencies of FETs in most power supply applications, effectively filtering out high-frequency noise and interference.
Dual Range Selection
One-button auto zero calibration and flexible dual-range switching improve the signal-to-noise ratio.

Figure 5: Functional Zoning and Controls
Standard BNC Interface
The probe features a standard BNC interface, ensuring compatibility with oscilloscopes from most brands.

Figure 6: BNC Interface Connection
Applications
The DP Series probes cover various application scenarios in high-speed digital signals and high-voltage power electronics.

Figure 7: Typical Applications
4. Setup and Operation
4.1 Powering the Probe
Connect the USB input of the probe to a standard power adapter or directly to a USB port on your oscilloscope.
4.2 Connecting to the Oscilloscope
Connect the BNC output of the probe to an oscilloscope channel input. Ensure that the oscilloscope is properly grounded before making any connections.
4.3 Setting the Attenuation Range
Select the appropriate voltage range (attenuation) according to the signal you intend to measure. The probe offers dual voltage ranges.
4.4 Connecting to the Device Under Test (DUT)
Use the provided clips or hooks to connect the probe to the DUT. To reduce interference, twist the red and black input leads into a pair. If an over-voltage alarm occurs (LED flashes and buzzer beeps rapidly), immediately disconnect the power supply and the circuit.
4.5 Oscilloscope Settings
Set the oscilloscope's input impedance. Adjust the channel attenuation ratio on the oscilloscope to match the probe's settings.
4.6 Calibration
To calibrate the probe, short-circuit the input ends, power on the probe, and then press the "Zero" button. A single "Di" sound indicates successful calibration. If you hear "Di Di Di", the calibration has failed, and you should recalibrate.
4.7 Warm-up Time
For more accurate results, it is recommended to allow the probe and connected equipment to warm up for at least 10 minutes before use.
Video 1: Product Unboxing and Initial Setup
5. Maintenance
To ensure optimal performance and longevity of your Micsig High Voltage Differential Probe, keep product surfaces clean and dry. Avoid exposing the probe to excessive moisture or extreme temperatures.
6. Troubleshooting
Overvoltage Warning
If the range LED indicator flashes and the buzzer beeps rapidly, it signifies an overvoltage condition. In this event, immediately switch the probe to a higher range or disconnect the probe from the circuit and stop testing to prevent damage.
Calibration Failure
If the probe emits a "Di Di Di" sound during calibration, it indicates a calibration failure. Repeat the calibration process by short-circuiting the input ends, powering on, and pressing the "Zero" button.
Excessive Noise or Inaccurate Readings
- Ensure input leads are twisted together to reduce interference.
- Avoid using long extension leads if possible. If necessary, ensure they are of equal length and the input signal frequency does not exceed 5MHz.
- Do not touch the input leads with your hand or other objects during measurement.
- Allow the probe to warm up for 10 minutes for optimal accuracy.
7. Specifications

Figure 8: Detailed Specifications Table
| Parameter | Details |
|---|---|
| Bandwidth | 100MHz (DP700, DP1500, DP3000) / 200MHz (DP702, DP1502, DP3002, DP7002) / 300MHz (DP1503, DP3003) |
| Attenuation | 20X/200X (DP700, DP702) / 50X/500X (DP1500, DP1502, DP1503) / 100X/1000X (DP3000, DP3002, DP3003, DP7000, DP7002) |
| Max. Input Differential Voltage (DC+AC PK) | 700V (DP700, DP702) / 1500V (DP1500, DP1502, DP1503) / 3000V (DP3000, DP3002, DP3003) / 7000V (DP7000, DP7002) |
| Max. Voltage to Ground | CAT II 450V (DP700) / CAT II 1000V (DP1500, DP3000) / CAT II 1500V (DP7000) |
| Max. Common Mode Voltage | 700V (DP700) / 1500V (DP1500) / 3000V (DP3000) / 7000V (DP7000) |
| CMRR | DC: >80dB, 100kHz: >60dB, 10MHz: >30dB, 100MHz: >26dB |
| Output Voltage | <3.5V (DP700) / <3V (DP1500, DP3000, DP7000) |
| Accuracy | ±2% (Customizable 1% accuracy) |
| Output Impedance | 1 MΩ (100/200 MHz bandwidth probes) / 50 Ω (300 MHz bandwidth probes) |
| Power Supply | DC 5V |
| Overrange Indication | LED flashes, Buzzer beeps |
| Dimensions (L*W*H) | 13.5 * 5 * 2.5 cm |
| Cable Length | Approx. 31 cm (Input); Approx. 100 cm (Output) |
| Operating Temperature | 0°C ~ 40°C |
| Non-operating Temperature | -30°C ~ 70°C |
| Operating Humidity | 5%~85% RH (0°C~40°C) |
| Non-operating Humidity | 5%~85% RH (≤40°C); 5%~45% RH (40°C~70°C) |
| LVD Standard | EN 61010-1:2010; EN IEC 61010-2-030:2021; EN 61010-031:2015+A1:2021+A11:2021 |
| EMC Standard | EN IEC 61326-1:2021; EN IEC 61326-2-1:2021; EN61000-3-2:2019+A1:2021; EN61000-3-3:2013+A1:2019+A2:2021 |
8. User Tips
- Always twist the red and black input leads together when connecting to the DUT to minimize interference and improve signal quality.
- Allow the probe to warm up for at least 10 minutes before taking critical measurements to ensure maximum accuracy.
- If using extension leads, ensure they are of equal length and the input frequency is below 5MHz to avoid inaccurate test results.
9. Warranty and Support
Warranty Information
Micsig warrants the main body of this differential probe for a period of one (1) year from the date of shipment. Accessories are warranted for a period of six (6) months from the date of shipment. During the warranty period, Micsig will provide free maintenance for any failure caused by product quality under normal use. Replaced parts, modules, and products used for warranty work may be new or reconditioned to like-new performance and become the property of Micsig.
Conditions for Refusal of Service
Micsig reserves the right to refuse maintenance services or charge a fee under the following circumstances:
- Absence of original packaging or anti-counterfeiting label.
- Altered or blurred anti-counterfeiting label.
- Unauthorized disassembly (e.g., changing wires, dismantling internal components).
- No sales voucher or mismatch between the sales voucher content and the product.
Contact Information
Shenzhen Micsig Instruments Co., Ltd.Add: 6F, Jinhuanyu Building, No. 56, Tiezai Rd, Bao'an District, Shenzhen, Guangdong, China.
Tel: +86-(0)755-88600880
Email: sales@micsig.com
Website: www.micsig.com
For the complete user manual in PDF format, please refer to: Micsig High Voltage Differential Probe User Manual (PDF)
10. Safety Precautions
Failure to follow these safety precautions may result in injury or damage to the product.
- Do not open the product casing unless you are a qualified professional.
- Do not use the probe if the case is open or damaged.
- Avoid touching any bare metal parts of the probe or circuit while testing.
- Immediately disconnect the power supply and the circuit if an over-range condition occurs.
- Do not use the probe in flammable or explosive environments.
- Ensure all devices requiring grounding are properly connected to earth ground before making any connections.
- Always be aware of the voltage ratings of the probe and accessories. Never apply a potential that exceeds these ratings.
- Only qualified personnel should perform service on this product.
- Do not touch exposed connections and components when power is present.
- Do not operate in wet or damp conditions.
- Keep product surfaces clean and dry.