1. Product Overview
The Micsig DP Series High-Voltage Differential Probe is designed for accurate measurement of high-voltage differential signals. These probes offer a bandwidth of 100-300 MHz and can handle a maximum input voltage of 7000 Vpk. Equipped with a standard BNC interface, they are compatible with most oscilloscope brands.
Key features include one-button calibration, an overload alarm system, range power-off memory, dual voltage ranges, and a high-resistance, low-capacitance design to minimize circuit loading. The probes ensure strong amplitude-frequency performance and a high common-mode rejection ratio. Automated testing is supported through command-based programming, and a 5 MHz bandwidth-limit function helps suppress high-frequency noise for clearer waveform delivery.

2. What's in the Box
Upon opening the package, you should find the following items:
- DP Series Probe x1
- Extendable Hook Pliers x1 Pair
- Alligator Clips x1 Pair
- Input Extension Cables x1 Pair
- Power Adapter* x1 (Note: Plug options may vary by region)
- BNC Output Line x1
- USB Line x1
- Quick Guide x1

Video 1: Unboxing the Micsig DP Series Probe
3. Setup Instructions
- Power Supply: Connect the probe's power input port to the USB cable's power output port. The probe operates on USB DC 5V.
- Grounding the Product: To prevent electric shock, ensure the grounding conductor is connected to earth ground. Always verify proper grounding before connecting to the input or output terminals.
- Probe Connection: Connect the probe's BNC port to your oscilloscope or other measuring instrument's BNC port.
- Calibration: Before initial use or if accuracy is in doubt, calibrate the probe. Short-circuit the input ends of the probe, power it on, and then press the