MasTech MS850 Series

Contents

MASTECH MS850 Series Digital Multimeter User Manual

Models: MS850A, MS850B, MS850C, MS850D

1. Introduction

This manual provides detailed instructions for the safe and effective use of the MASTECH MS850 series digital multimeters. These devices are designed for measuring AC/DC voltage, AC/DC current, resistance, diode, continuity, and other electrical parameters, depending on the model. Please read this manual thoroughly before operation and keep it for future reference.

2. Safety Information

WARNING: To avoid electrical shock or personal injury, please read and understand all safety instructions before use.

  • Always use the correct function, range, and terminal for measurements.
  • Do not use the device around explosive gas, vapor, or under high temperature and humidity.
  • Inspect the test leads for damaged insulation or exposed metal before each use. If any defects are found, replace them.
  • Ensure the measurement leads are properly seated in the correct input jacks.
  • Do not apply more than the rated voltage, as marked on the device.
  • Connect the common test lead first before the live test lead, and remove the live test lead before the common test lead.
  • Do not use the device if the battery cover is not in place or if the device is damaged.
  • Replace batteries if the device is not used for an extended period of time. If it is stored in temperatures above 45°C, or if the batteries are not removed, battery leakage can damage the device.
  • Always disconnect power to the circuit and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
Quick Start Guide with safety warnings and operational diagrams for MS850D
Figure 1: Quick Start Guide showing safety warnings and general operational diagrams.

3. Product Overview

The MASTECH MS850 series offers several models, each with specific features and capabilities. All models share a robust design and user-friendly interface.

Play video
Watch video online
Video 1: Overview of the MS850D model demonstrating various measurement functions.

Model Comparison

Comparison table for MS850A, MS850B, MS850C, MS850D multimeters
Figure 2: Comparison of features across MS850A, MS850B, MS850C, and MS850D models.

Key differences include:

  • **MS850A:** Manual ranging, Max Display 2000 counts, Frequency (400Hz), Diode Test, Continuity Check, Transistor Test, NCV, Backlight.
  • **MS850B:** Manual ranging, Max Display 2000 counts, True RMS, MAX/MIN/REL, Buzzer/Continuity/Diode, NCV, Temperature, Backlight/Flashlight.
  • **MS850C:** Auto ranging, Max Display 4000 counts, True RMS, MAX/MIN/REL, Buzzer/Continuity/Diode, NCV, Frequency, Capacitance, Duty Cycle, Backlight/Flashlight.
  • **MS850D:** Auto ranging, Max Display 6000 counts, True RMS, MAX/MIN/REL, Buzzer/Continuity/Diode, NCV, Frequency, Capacitance, Duty Cycle, Temperature, Backlight/Flashlight.

4. Setup

4.1 Battery Installation

The MS850 series multimeters require two 1.5V AAA batteries (not included). To install or replace batteries:

  1. Ensure the multimeter is turned OFF and test leads are disconnected.
  2. Locate the battery compartment on the back of the device.
  3. Use a screwdriver to open the battery compartment cover.
  4. Insert two 1.5V AAA batteries, observing the correct polarity (+/-).
  5. Replace the battery compartment cover and secure it with the screw.
Back view of MS850A showing battery compartment
Figure 3: Rear view of the multimeter with the battery compartment.

4.2 Connecting Test Leads

Connect the red test lead to the "VΩmA" or "10A" jack and the black test lead to the "COM" jack. Ensure a secure connection.

Multimeter with test leads connected
Figure 4: Multimeter with test leads properly connected.

5. Operating Instructions

The MS850 series multimeters feature a rotary switch for function selection and buttons for additional features like SEL (Select), HOLD (Data Hold), MIN/MAX/REL (Minimum/Maximum/Relative), and Backlight.

5.1 General Operation

  1. Turn the rotary switch to the desired measurement function.
  2. Connect the test leads to the circuit or component under test.
  3. Read the measurement value on the LCD display.
  4. Use the 'SEL' button to switch between different measurement modes within a single rotary switch position (e.g., AC/DC voltage, resistance/continuity/diode).
  5. The 'HOLD' button freezes the current display reading.
  6. The 'MIN/MAX/REL' button (on MS850B, C, D) allows for recording minimum/maximum values or relative measurements.
  7. The backlight button (light bulb icon) illuminates the display.

5.2 Specific Measurements

Operational diagrams for Voltage, hFE, Frequency, Temperature, MIN/MAX/REL
Figure 5: Diagrams for measuring Voltage (AC/DC), hFE (Transistor), Frequency, Temperature, and using MIN/MAX/REL functions.
Operational diagrams for Current, Continuity/Diode, NCV/Live
Figure 6: Diagrams for measuring Current (AC/DC), Continuity/Diode, and using NCV/Live functions.

AC/DC Voltage Measurement

  1. Turn the rotary switch to the V~ (AC Voltage) or V= (DC Voltage) position.
  2. Connect the test leads in parallel to the circuit or component.
  3. Read the voltage value on the display.

AC/DC Current Measurement

  1. Turn the rotary switch to the A~ (AC Current) or A= (DC Current) position.
  2. Connect the test leads in series with the circuit. Ensure the circuit is de-energized before connecting.
  3. Read the current value on the display.

Resistance Measurement

  1. Turn the rotary switch to the Ω position.
  2. Connect the test leads across the component to measure resistance.
  3. Read the resistance value on the display.

Diode Test

  1. Turn the rotary switch to the Diode symbol position.
  2. Connect the red test lead to the anode and the black test lead to the cathode of the diode.
  3. Read the forward voltage drop. Reverse the leads to check for open circuit (OL) reading.

Continuity Check

  1. Turn the rotary switch to the Continuity symbol position.
  2. Connect the test leads across the circuit or component.
  3. A continuous beep indicates continuity (low resistance).

Transistor Test (hFE)

For models with this feature (MS850A, MS850C, MS850D), insert the transistor leads into the hFE socket according to the transistor type (NPN or PNP) and pin configuration (EBC). The display will show the hFE value.

Non-Contact Voltage (NCV) Detection / Live Wire Recognition

Turn the rotary switch to the NCV position. Place the top of the multimeter near an AC voltage source. The device will indicate the presence of AC voltage with an audible alarm and/or visual indicator. For Live Wire recognition, follow the specific instructions for your model.

Temperature Measurement (°C/°F)

For models with this feature (MS850B, MS850D), connect the temperature probe to the appropriate input jacks and place the probe tip on or near the object whose temperature is to be measured. Select the °C or °F mode as needed.

Duty Cycle Measurement

For models with this feature (MS850C, MS850D), select the Duty Cycle function and connect the test leads to the signal source. The display will show the duty cycle percentage.

Flashlight

For models with a flashlight (MS850B, MS850C, MS850D), press the flashlight button to turn it on or off, providing illumination in dark working conditions.

6. Specifications

The following tables detail the specifications for each model in the MS850 series.

General Specifications (Applicable to all models)

  • Dimensions: 150mm x 75mm x 47mm
  • Weight: Approximately 284g (without batteries)
  • Power: 2 x 1.5V AAA Batteries
  • Safety Rating: CAT III 600V
  • Operating Mode: Auto Range (MS850C, MS850D), Manual Range (MS850A, MS850B)

Detailed Measurement Specifications

MS850A Specifications
Spec Range Accuracy
Resistance200Ω, 2KΩ, 20KΩ, 200KΩ, 2MΩ, 200MΩ±(1.0%+5) for most, ±(5.0%+10) for 200MΩ
DCV200mV, 2V, 20V, 200V, 600V±(0.5%+3) to ±(0.8%+5)
ACV200mV, 2V, 20V, 200V, 600V±(1.0%+10)
DCA2mA, 20mA, 200mA, 10A±(1.0%+5) to ±(3.0%+5)
ACA2mA, 20mA, 200mA, 10A±(1.8%+5) to ±(3.0%+5)
Frequency400HzNot specified
Battery Test1.5V, 3V, 9V±(1.8%+2)
MS850B Specifications
Spec Range Accuracy
Resistance600Ω, 6KΩ, 60KΩ, 600KΩ, 6MΩ, 60MΩ±(1.0%+5) for most, ±(1.2%+8) for 60MΩ
DCV600mV, 6V, 60V, 600V±(0.5%+3) to ±(0.8%+5)
ACV600mV, 6V, 60V, 600V±(1.0%+10)
DCA600uA, 6000uA, 60mA, 600mA, 6A, 10A±(1.0%+5) to ±(3.0%+5)
ACA600uA, 6000uA, 60mA, 600mA, 6A, 10A±(1.8%+5) to ±(3.0%+5)
Frequency10Hz to 10MHz±(0.5%+4)
Capacitance60nF to 100mF±(4.0%+25) to ±(5.0%+25)
Temperature (°C)-20°C ~ 1000°C±(4°C) to ±(3.0%+3)
Temperature (°F)-4°F ~ 1832°F±(5°F) to ±(3.0%+5)
Battery Test1.5V, 3V, 9VNot specified
MS850C Specifications
Spec Range Accuracy
Resistance400Ω, 4KΩ, 40KΩ, 400KΩ, 4MΩ, 40MΩ±(1.0%+5) for most, ±(1.2%+8) for 40MΩ
DCV400mV, 4V, 40V, 400V, 600V±(0.5%+3) to ±(0.8%+5)
ACV400mV, 4V, 40V, 400V, 600V±(1.0%+10)
DCA400uA, 4000uA, 40mA, 400mA, 4A, 10A±(1.0%+5) to ±(3.0%+5)
ACA400uA, 4000uA, 40mA, 400mA, 4A, 10A±(1.8%+5) to ±(3.0%+5)
Frequency10Hz to 10MHz±(0.5%+4)
Capacitance60nF to 100mF±(4.0%+25) to ±(5.0%+25)
Duty Cycle0.1% ~ 99.9%±(1.2%+3) to ±(2.5%+3)
Battery Test1.5V, 3V, 9VNot specified
MS850D Specifications
Spec Range Accuracy
Resistance600Ω, 6KΩ, 60KΩ, 600KΩ, 6MΩ, 60MΩ±(1.0%+5) for most, ±(1.2%+8) for 60MΩ
DCV600mV, 6V, 60V, 600V±(0.5%+3) to ±(0.8%+5)
ACV600mV, 6V, 60V, 600V±(1.0%+10)
DCA600uA, 6000uA, 60mA, 600mA, 6A, 10A±(1.0%+5) to ±(3.0%+5)
ACA600uA, 6000uA, 60mA, 600mA, 6A, 10A±(1.8%+5) to ±(3.0%+5)
Frequency10Hz to 10MHz±(0.5%+4)
Capacitance60nF to 100mF±(4.0%+25) to ±(5.0%+25)
Duty Cycle0.1% ~ 99.9%±(1.2%+3) to ±(2.5%+3)
Temperature (°C)-20°C ~ 1000°C±(4°C) to ±(3.0%+3)
Temperature (°F)-4°F ~ 1832°F±(5°F) to ±(3.0%+5)
Battery Test1.5V, 3V, 9VNot specified

7. Maintenance

7.1 Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the terminals free from dirt and dust.

7.2 Battery Replacement

When the battery low indicator appears on the display, replace the batteries immediately to ensure accurate readings. Follow the steps in Section 4.1 for battery installation.

7.3 Fuse Replacement

If the current measurement function fails, the fuse may need replacement. Refer to the markings on the back of the device for fuse specifications (e.g., F500mA/250V, F10A/250V). Fuse replacement should only be performed by qualified personnel.

Back view of MS850D showing fuse specifications
Figure 7: Rear view of the multimeter indicating fuse specifications.

8. Troubleshooting

If your multimeter is not functioning as expected, try the following basic troubleshooting steps:

  • **Device does not turn on:** Check if the batteries are correctly installed and have sufficient charge. Replace with new batteries if necessary.
  • **"Err-E" displayed:** This typically indicates an error. Try turning the multimeter off and on again. If the error persists, ensure test leads are correctly connected and the measurement range is appropriate for the input.
  • **No reading or incorrect reading:** Verify that the test leads are properly connected to the correct input jacks and to the circuit under test. Ensure the rotary switch is set to the correct function and range.
  • **Delayed continuity beep:** Some users have noted a slight delay in the continuity beep. This may be a characteristic of the device and not necessarily a fault, especially for basic models.
MS850D displaying 'Err-E'
Figure 8: Multimeter displaying an error code 'Err-E'.

9. User Tips

  • For general electrical tasks and hobbyist use, these multimeters offer good value and functionality.
  • While functional, some users perceive the build quality as basic. Handle with care to ensure longevity.
  • Always double-check battery status, especially if the device fails to power on or provides inconsistent readings.

10. Warranty and Support

For warranty information and technical support, please refer to the documentation provided with your purchase or contact the seller directly. Keep your purchase records for any warranty claims.