WINAPEX 268B

Contents

WINAPEX 268B Clamp Meter User Manual

Model: 268B

1. Overview

This series of instruments is a highly reliable digital clamp meter with stable performance and lithium battery drive. It features a full symbol, 17mm word height LCD display for clear reading. The adjustable screen brightness adapts to external environments, protects the eyes, and allows screen rotation for easy reading. Voice broadcasting and full-function overload protection ensure user convenience and safety.

This instrument can measure the true effective value of DC voltage and AC voltage, DC current and AC current (clamp type), resistance, capacitance, frequency, Celsius, Fahrenheit, diode, on/off, and is suitable for leakage detection in various scenarios. The entire machine features a dual integral A/D conversion as its core, offering superior performance. It is an ideal tool for laboratories, factories, radio enthusiasts, and households.

2. Safety Precautions

The design of this instrument complies with IEC1010 safety standards. Please read the following precautions carefully before use:

  1. Please follow the instructions in this manual. Failure to do so may compromise the safety function of the clamp meter.
  2. Comply with national safety regulations. When working with high voltage and exposed wires, personal protective equipment must be used to prevent injuries from electric shock or arc discharge.
  3. Do not hold the current clamp meter in any position other than the protective shield.
  4. Before each use, check the outer shell of the clamp for cracks or defects, and pay special attention to the insulation around the clamp mouth.
  5. When replacing the battery, ensure the clamp is removed from all live circuits and leads are disconnected.
  6. When changing the function or range, the probe should leave the test point first.
  7. Voltage below 36V is considered a safe voltage. When measuring voltage above 36V DC or 25V AC, ensure the probe contacts reliably, is correctly connected, and has good insulation to avoid electric shock.
  8. When measuring voltage, do not exceed the limit voltage of DC 1000V or AC 750V effective value.

3. Common Symbols

SymbolDescriptionSymbolDescription
Warning promptHigh voltage hazard
Double insulationGrounding
Low battery levelDiode
BuzzerZEROZero clearing
ACDC
VFCVariable frequency conversionEFElectric field measurement
LIVEFire line measurementNCVNon-contact measurement
HOLDData hold

4. Setup

4.1 Package Contents

Please check the package for the following items:

  • 1 x WINAPEX 268B Clamp Meter
  • 1 x Pair of Test Leads (Red and Black)
  • 1 x User Manual (PDF available online)
  • 1 x Thermocouple (TP01)
  • 1 x Type-C Cable
  • 1 x Storage Bag
Package List of WINAPEX 268B Clamp Meter
Figure 4.1: Package Contents

4.2 Operation Panel Description

Familiarize yourself with the components of your WINAPEX 268B Clamp Meter:

Operation Interface of WINAPEX 268B Clamp Meter
Figure 4.2: Operation Interface
  1. Jaw: For clamping wires to measure current. Features a 22mm jaw opening.
  2. Model Sticker: Displays 'WINAPEX 268B'.
  3. LCD Display Screen: Digital display for measurements.
  4. Trigger: Opens the clamp jaw.
  5. On/Off/Function Selection Button: Power on/off and cycle through measurement functions.
  6. HOLD/Backlight Adjustment Button: Short press to hold data, long press to adjust backlight brightness.
  7. EF/LIVE Button: For electric field (NCV) and live wire detection.
  8. ZERO Button: For zero clearing, especially for DC current measurements.
  9. COM Socket: Common input terminal for test leads (black probe).
  10. VΩ Socket: Input terminal for voltage, resistance, capacitance, diode, frequency, and temperature measurements (red probe).
  11. Flashlight: Integrated light for illuminating work areas.
  12. Current Switching Button: For switching between AC and DC current modes.
  13. Battery Compartment: Houses the internal lithium battery.
  14. Charge Up Port: Type-C USB port for charging.

5. Operating Instructions

5.1 General Characteristics

  • Display Method: LCD display
  • Maximum Display: 5999 (3 5/6 digits), automatic polarity display
  • Measurement Method: Double integral A/D conversion
  • Updates per Second: 2-3 times
  • Over Range Display: "OL"
  • Low Voltage Display: Battery symbol appears
  • Working Environment: 0-40°C, relative humidity <80%
  • Power Supply: 3.7V lithium battery
  • Volume (Size): 165 × 68 × 26mm (length × wide × thick)
  • Weight: Approximately 280g (including battery)

5.2 DC Voltage (DCV) Measurement

Precautions:

  1. Insert the black test lead into the "COM" socket and the red test lead into the "V Ω" socket.
  2. When the probe reliably contacts the test point, the screen will display the measured voltage value. The displayed value indicates the voltage and polarity of the point connected to the red probe.
  3. When measuring high voltage, be careful not to touch the high voltage circuit.
  4. Do not exceed the input voltage of DC 1000V or AC 750V, as this may damage the instrument circuit.
DC Voltage Measurement Example
Figure 5.1: DC Voltage Measurement

5.3 AC Voltage (ACV) RMS Measurement

Precautions:

  1. Insert the black test lead into the "COM" socket and the red test lead into the "V Ω" socket.
  2. When the test probe reliably contacts the test point, the screen will display the measured voltage value.
  3. The measured voltage should not exceed AC 750V. The probe should leave the test point first when turning off or changing the functional gear.
  4. When measuring high voltage, be careful not to touch the high voltage circuit.
AC Voltage Measurement Example
Figure 5.2: AC Voltage Measurement

5.4 DC Current (DCA) Measurement (Clamp Type)

Precautions:

  1. Before testing, close the clamp head and press the "ZERO" key to reset it. If it is not zero, you can press it multiple times to make the reading zero.
  2. Record the reading after the test, and repeat the reset step for each measurement.
  3. Minimum measurement current is 5mA, maximum measurement current is 60A. Over range display "OL", overload protection 60A.
  4. Be careful when measuring the current of high voltage electricity to avoid electric shock.
DC Leakage Current Detection for Car Maintenance
Figure 5.3: DC Leakage Current Detection (Car Maintenance)

5.5 AC Current (ACA) Measurement (Clamp Type)

Precautions:

  1. All currents are automatic ranges. When measuring large currents, the range is identified from small to large. The larger the fixed current value, the longer the measurement time will be.
  2. When the displayed value shows "OL", it indicates that the measured value has exceeded the maximum current measurement value of the clamp meter.
  3. The maximum measured current values for 268B is 60A.
  4. The testing frequency is 30Hz to 1kHz for 268B.
  5. The minimum measurement current is AC 2mA for 268B.
AC Leakage Current Detection for Home Appliance
Figure 5.4: AC Leakage Current Detection (Home Appliance)

5.6 Resistance (Ω) Measurement

Warning: [Input voltage prohibited in this range]

Precautions:

  1. Do not input voltage in the resistance range, as it may damage the instrument.
  2. When using a 600 Ω range, the probe should be short-circuited first to measure the lead resistance, and then subtracted from the actual measurement.
  3. When measuring above 1M Ω, it is normal for the reading to react slightly slowly. Please wait for the displayed value to stabilize before reading.
  4. Insert the black test lead into the "COM" socket and the red test lead into the "V Ω" socket.
  5. Press the "SELECT" button to automatically or manually select the functional gear on the resistance page, and connect the two probes across the measured resistance.
  6. If the resistance value exceeds 60M Ω, it will display "OL".
  7. When measuring online resistance, confirm that the power supply of the tested circuit has been turned off and all capacitors have been fully discharged.
  8. When the measured resistance value exceeds 1M Ω or more, it takes several seconds for the reading to stabilize, which is normal.
Resistance Measurement Example
Figure 5.5: Resistance Measurement

5.7 Diode and On/Off Test

Warning: [Input voltage prohibited in this range]

Precautions:

  1. Insert the black test lead into the "COM" socket and the red test lead into the "V Ω" socket (note that the polarity of the red lead is "+").
  2. Press the "SELECT" key to switch to the diode test position. Connect the red probe to the positive pole of the diode, and the black probe to the negative pole. The reading is the positive voltage drop value of the diode, in "V".
  3. For continuity test, connect the probe to two points of the circuit to be tested. If the built-in buzzer sounds, the resistance value between the two points is less than approximately (50 ± 20) Ω.
Diode and On/Off Beep Test Example
Figure 5.6: Diode and On/Off Beep Test

5.8 Capacitance (C) Measurement

Warning: [Input voltage prohibited in this range]

Precautions:

  1. Insert the black test lead into the "COM" socket and the red test lead into the "V Ω" socket.
  2. Press the "SELECT" key to switch to the capacitance page and connect the two probes across the measured capacitance.
  3. If the measured capacitance exceeds the maximum value of the range, the display will only show "OL".
  4. If there is severe leakage or breakdown short circuit of the capacitor, a numerical value will be displayed and unstable.
  5. Before testing capacitance, the capacitor should be fully discharged to prevent damage to the instrument.
  6. When capacitance is not measured, it is allowed to have a certain base number. Before measuring, first press the "ZERO" button to reset it to zero.
Capacitance Measurement Example
Figure 5.7: Capacitance Measurement

5.9 Frequency (Hz) Measurement (268B only)

Precautions:

  1. Insert the probe or shielded cable into the "COM" and "V Ω" sockets.
  2. Press the "SELECT" key to switch to the frequency page, and connect the probe or cable to the signal source or tested load.
  3. When the input exceeds 10Vrms, it can be read, but it may exceed the tolerance.
  4. In noisy environments, it is best to use shielded cables when measuring small signals.
  5. When measuring high voltage circuits, special attention should be paid to avoiding electric shock.
  6. It is prohibited to input voltage values exceeding 250V DC or AC peak to avoid damaging the instrument.
Frequency Measurement Example
Figure 5.8: Frequency Measurement

5.10 Temperature Measurement (°C/°F)

Precautions:

  1. Press the "SELECT" key to switch to the temperature page.
  2. Insert the negative end of the temperature measurement probe (thermocouple) into the "COM" input port and the positive end into the "V Ω" hole.
  3. If the temperature measurement range (-20-1000°C) is too large, it will display "OL".
  4. When the input terminal is open, the ambient temperature is displayed when the operating environment is above 18°C but below 28°C, and is displayed when the operating environment is below 18°C and above 28°C for reference only.
  5. Do not replace the temperature sensor arbitrarily, otherwise measurement accuracy cannot be guaranteed.
  6. It is strictly prohibited to input voltage in the temperature range.
Temperature Measurement Example
Figure 5.9: Temperature Measurement

5.11 Non-Contact Voltage (NCV) / Live Wire Measurement (LIVE)

NCV/LIVE Measurement Precautions:

  1. Click the "EF" key. The screen displays "EF" and enters non-contact electric field measurement mode.
  2. Place the front end of the instrument near the AC voltage. Depending on the strength of the signal, the buzzer will continue to emit sounds at different intervals, and the LCD will display different segments.
  3. Press again in NCV mode. The screen displays "LIVE" to enter the live line measurement mode.
  4. Insert the red probe into the "V Ω" socket (single probe operation is sufficient). If the buzzer emits a continuous sound, it is a live wire. "OL" will simultaneously display on the screen.
  5. This function is applicable to frequency signals ranging from 50Hz to 1kHz.
NCV Non-contact Voltage Test
Figure 5.10: NCV Non-contact Voltage Test

5.12 Brightness Adjustment

Screen brightness can be adjusted according to needs in different environments. The method is as follows: Press and hold the backlight adjustment button for more than 2 seconds until the buzzer emits a short beep, then release it to enter brightness adjustment mode. Press and hold this key again to adjust the brightness, which is divided into five levels. Press and hold this key again until the buzzer emits a short beep, then release it to exit brightness adjustment mode.

Brightness Adjustment
Figure 5.11: Brightness Adjustment

5.13 Data Hold

Short press the "HOLD" key to hold the data. Short press "HOLD" again to unlock the data.

5.14 Flashlight Function

To use the flashlight: Press and hold the flashlight button until the flashlight turns on. Press and hold the button again to turn off the flashlight.

Flashlight Feature
Figure 5.12: Flashlight

5.15 Voice Alarm Value (268A, 268C only)

This model can perform Chinese voice value reporting on measurement results. In situations where lighting is dim, there is no need to look at the LCD display. As long as the reading is stable, it will automatically report the number. Long press the 'LIVE' button to cancel the voice value reporting function, and long press again to restore it.

Note:

  • During measurement, only three consecutive readings of the same data can start reporting values.
  • In manual measurement state, DC6V and AC6V gears may have data changes due to minimal magnetic field interference, and voice will not report a value. AC gear has a base number of 10 words and does not report a value. Other gears have a base number of 2 words and do not report a value. This is to prevent interference.
  • DC voltage, DC current, and resistance gears must change by at least 2 words after the initial measurement report before the second voice report can be made. Otherwise, it will be considered as interference data without reporting values.
  • No value is reported for buzzer measurement, electric field induction, or live wire measurement.

5.16 Leakage Current Detection

(268B measures leakage current, while μA-level leakage current measurement is limited to 268C.)

  1. Detect leakage current on the grounding wire: Open the trigger and clamp the pliers to the ground wire and check the reading displayed on the LCD, which is the leakage current value.
  2. Single-phase AC leakage current detection: Open the trigger and clamp the power supply with the clamp head. The wire, live wire, and zero wire are clamped together, and the reading displayed on the instrument is the leakage current value.
  3. Three-phase three-wire leakage current detection: Open the trigger and clamp three pieces of pliers. The reading displayed by the instrument on the live line is the leakage current value.
  4. Three-phase four-wire leakage current detection: Open the trigger and clamp the three live wires with the clamp head and a zero wire. Do not clamp the ground wire. The instrument display reading is the leakage current value.
  5. AC current measurement: When measuring AC current, only one live wire (phase wire) can be clamped. Keep the wire in the center position of the clamp, and then read the data from the screen.
  6. When measuring current, the two clamps should be tightly connected without gaps. Otherwise, the measurement error will be too large.
Leakage Current Detection Setup
Figure 5.13: Leakage Current Detection Setup
Three-phase Three-wire Leakage Current Detection
Figure 5.14: Three-phase Three-wire Leakage Current Detection
Three-phase Four-wire Leakage Current Detection
Figure 5.15: Three-phase Four-wire Leakage Current Detection
AC Current Measurement
Figure 5.16: AC Current Measurement

5.17 Screen Rotation

The screen can be rotated with one key, allowing for free operation with both hands and improved field of view.

Screen Rotation Feature
Figure 5.17: Screen Rotation

5.18 Automatic Identification

The meter features intelligent recognition for AC voltage, DC voltage, resistance, and buzzer functions, automatically detecting the measurement type.

Intelligent Recognition Feature
Figure 5.18: Intelligent Recognition

5.19 Product Video Demonstration

Video 5.1: Product Demonstration of WINAPEX 268B Clamp Meter

6. Maintenance

6.1 Battery Charging

The WINAPEX 268B is equipped with a 3.7V lithium battery and supports Type-C charging. A full charge provides approximately 20 hours of continuous battery life. Use the provided Type-C cable to charge the device.

Type-C Rechargeable Feature
Figure 6.1: Type-C Rechargeable

6.2 Automatic Shutdown

When the instrument is turned on for 15 minutes, it will automatically power off and enter a sleep state. Before turning off, a five-tone prompt will sound. To restart, press and hold the power button for more than 2 seconds.

7. Troubleshooting

If your instrument cannot function properly, the following methods can help you quickly solve general problems. If the fault still cannot be resolved, please contact the repair center or dealer.

Fault PhenomenonInspection Location and Method
No displayPower not connected
Replacing the battery
Large resistance display errorThe test lead is not in good contact

This manual is subject to change without notice. The content of this manual is considered correct. If users find any errors, omissions, etc., please contact the manufacturer. Our company shall not be responsible for accidents and hazards caused by user error operations. The functions described in this manual are not intended as a reason for using the product for special purposes.

8. Specifications

8.1 General Specifications

  • Operating Mode: Auto Range
  • DIY Supplies: Electrical
  • Origin: Mainland China
  • Battery Included: Yes
  • Semi-finished product: No
  • High-concerned chemical: None
  • Display Type: Digital Display

8.2 Measurement Ranges and Accuracy (Model 268B)

FunctionRangeAccuracyResolution
DC Voltage (DCV)6V± (0.5%+3)1mV
60V10mV
600V100mV
1000V± (0.8%+10)1V
AC Voltage (ACV) RMS6V± (0.8%+3)1mV
60V10mV
600V± (0.8%+3)100mV
750V± (1.2%+10)1V
DC Current (DCA clamp)6A± (2.0%+5)1mA
60A± (2.0%+3)10mA
AC Current (ACA clamp type)600mA± (1.5%+10)0.1mA
6A± (2.5%+5)1mA
60A± (2.0%+5)10mA
Resistance (Ω)600Ω± (0.8%+5)0.1Ω
6kΩ1Ω
60kΩ± (0.8%+3)10Ω
600kΩ100Ω
6MΩ± (2.5%+15)1kΩ
60MΩ10kΩ
Diode and on/off testDiode forward voltage dropForward DC current is about 1mA, Reverse voltage approximately 3V
Buzzer (resistance < 50±20Ω)Open circuit voltage approximately 3V
Capacitance (C)6nF± (3.5%+20)1pF
60nF10pF
600nF100pF
6uF1nF
60uF10nF
600uF100nF
6mF± (5.0%+20)1uF
60mF10uF
Frequency measurement (Hz)6Hz± (0.5%+5)0.001Hz
60Hz0.01Hz
600Hz0.1Hz
6kHz1Hz
60kHz10Hz
600kHz100Hz
6MHz1kHz
10MHz10kHz
Temperature measurement (°C/°F)(-20-1000)°C± (1.0%+5) <400°C;
± (1.5%+15) ≥400°C
1°C
(0-1832)°F± (0.75%+5) <750°F;
± (1.5%+15) ≥750°F
1°F

9. User Tips

  • DC Current Measurement: Always remember to close the clamp head and press the "ZERO" key to clear the reading before measuring DC current. This ensures accurate results.
  • Leakage Detection: For AC leakage current detection, clamp both the neutral and live wires simultaneously. If the display shows 0.00mA, there is no leakage. If it shows a value, it indicates leakage.
  • Automatic Identification: The meter intelligently identifies AC voltage, DC voltage, resistance, and buzzer functions, making it very convenient for general electrical checks.
  • Portability: The compact size and included storage bag make it easy to carry for on-the-go measurements.

10. Warranty and Support

This product is designed and manufactured to high-quality standards. For any issues or inquiries not covered in this manual, please contact the manufacturer or your dealer for assistance. Please refer to the included warranty card for specific warranty terms and conditions.

A detailed user manual in PDF format is available for download. Reference: User manual (PDF)