1. Introduction
The Hantek HTM200 series handheld multi-channel temperature detector is a high-performance device designed for real-time collection of multi-channel temperature data. Controlled by a 32-bit ARM microprocessor, it offers stable and accurate measurements, supporting 11 types of thermocouples and featuring spouse detection. This manual provides essential information for the safe and efficient operation of your HTM208B or HTM208C temperature recorder.
2. Safety Information
Please read and understand all safety instructions before operating the device. Failure to follow these instructions may result in injury or damage to the device.
- Electrical Safety: The digital and analog signals are isolated for live testing. However, always exercise caution when working with electrical circuits.
- Battery Safety: Use only specified 18650 standard lithium batteries. Do not expose batteries to excessive heat or dispose of them in fire.
- Environmental Conditions: Operate the device within the specified operating temperature and humidity ranges to ensure optimal performance and prevent damage.
- Maintenance: Refer to the maintenance section for proper cleaning and care. Do not attempt to disassemble or repair the device yourself.
3. Package Contents
Carefully unpack your Hantek HTM200 series temperature recorder and verify that all items listed below are included:
- Hantek HTM208B/HTM208C Multi-Channel Temperature Recorder
- Power Adapter (EU/US Plug Type, depending on model)
- USB Type-C Cable
- Thermocouples (typically two included)
- Carrying Case

4. Product Overview
The HTM200 series is designed for versatility and ease of use in various temperature monitoring applications.
4.1 Key Features
- Multi-Channel Measurement: Simultaneously collects data from up to 8 channels.
- Display: 320*240 TFT high-definition display supporting list and bar graph views.
- Thermocouple Compatibility: Supports 11 types of thermocouples (J/K/T/E/S/N/B/R/A/C/D).
- Alarm Control: Features temperature overload alarm with buzzer for upper and lower limit breaches.
- Data Storage: Standard U disk storage, with optional Micro SD card support.
- Connectivity: Standard Type-C communication, with optional Bluetooth and 4G modules for remote data acquisition.
- Software Support: Equipped with Windows and Android data acquisition terminals for multi-platform monitoring.
- Power: Uses 18650 standard lithium batteries, supports Type-C charging.

4.2 Device Layout

- TFT Display: Shows real-time temperature data, lists, and bar graphs.
- Push Buttons: For navigation, function selection (F1-F4), utility, list, bar graph view, and saving data.
- Thermocouple Connection Ports (1-8): Terminal blocks for connecting up to 8 thermocouples.
- USB Interface: For data transfer to a computer.
- Type-C Interface: For charging and communication.
- SD Card Slot: For optional Micro SD card storage.
- Power Button: To turn the device on/off.
5. Setup
5.1 Battery Installation and Charging
- The device uses 18650 standard lithium batteries. Ensure batteries are correctly inserted according to polarity markings.
- Connect the provided USB Type-C cable to the device's Type-C port and the power adapter.
- Plug the power adapter into a suitable power outlet (100VAC-240VAC, 50Hz/60Hz). The device supports Type-C charging.
- A full charge provides over 15 hours of working time.
5.2 Connecting Thermocouples
The HTM200 series supports 11 types of thermocouples. Ensure you use the correct type for your application.
- Identify the thermocouple connection ports (1-8) at the bottom of the device.
- Connect your thermocouples to the appropriate terminal blocks. Ensure a secure connection for accurate readings.
- Observe the maximum channel-to-channel voltage of 300V DC and maximum input voltage of ±75V DC.
5.3 Powering On/Off
- To Power On: Press and hold the Power button on the side of the device until the TFT display illuminates.
- To Power Off: Press and hold the Power button until the device shuts down. The device saves the last shutdown settings by default.
6. Operating Instructions
6.1 Basic Operation and Display Modes
Use the push buttons for navigation and function selection.
- F1-F4 Buttons: Context-sensitive function buttons displayed on the screen.
- Utility Button: Accesses general settings and utilities.
- Arrow Buttons: Navigate through menus and adjust settings.
- Enter Button: Confirms selections.
- List Button: Switches the display to a list view of all channel temperatures.
- Bar Button: Switches the display to a bar graph view, visually representing temperature changes across channels.
6.2 Temperature Measurement
Once thermocouples are connected and the device is powered on, it will begin real-time temperature measurement. The display will show readings for each active channel.
6.3 Alarm Settings
The device features an over-temperature alarm function, which is enabled by default.
- Access the alarm settings via the Utility menu.
- Set the upper and lower temperature limits for each channel as required.
- If a measured temperature exceeds the upper limit or falls below the lower limit, the device's buzzer will sound continuously to alert you.

6.4 Data Storage and Export
The device supports data logging to a U disk or an optional Micro SD card.
- U Disk Storage: Connect a U disk to the USB interface to store measurement data.
- Micro SD Card (Optional): Insert a Micro SD card into the dedicated slot for real-time data saving.
- Exporting Data: Data can be exported for further analysis and printing.
6.5 Software for Data Acquisition
The HTM200 series is compatible with Windows and Android data acquisition terminals.
- Install the appropriate software on your computer or Android device.
- Connect the temperature recorder via the Type-C port (or optional Bluetooth/4G).
- Use the software to monitor real-time temperature changes, track abnormal temperatures, and quickly read temperature values from each channel on multiple platforms.


7. Maintenance
7.1 Cleaning
- Wipe the device with a soft, dry cloth.
- Do not use abrasive cleaners or solvents, as they may damage the casing or display.
- Ensure no liquids enter the device through ports or buttons.
7.2 Battery Care
- Charge the battery regularly, even if the device is not in use, to maintain battery health.
- If storing the device for an extended period, charge the battery to approximately 50% and store in a cool, dry place.
- Remove batteries if the device will not be used for a very long time to prevent leakage.
7.3 Storage
- Store the device in its carrying case to protect it from dust, moisture, and physical damage.
- Avoid storing in extreme temperatures or high humidity.
8. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Device does not power on | Low battery or no battery installed. | Charge the device using the Type-C cable and power adapter. Ensure batteries are correctly installed. |
| Inaccurate temperature readings | Incorrect thermocouple type selected or poor connection. | Verify the thermocouple type matches the device settings. Check thermocouple connections for tightness. |
| Alarm not sounding | Alarm limits not set or alarm function disabled. | Check alarm settings in the Utility menu and ensure limits are configured correctly. |
| Data not saving to U disk/SD card | U disk/SD card full, corrupted, or not properly inserted. | Check storage capacity. Try reinserting the U disk/SD card. Format the storage device if necessary (this will erase all data). |
| Software not connecting to device | Incorrect cable, driver issues, or software not running. | Ensure the Type-C cable is securely connected. Install necessary drivers. Restart the software and device. |
9. Specifications
The following table details the technical specifications for the HTM208B and HTM208C models:

| Feature | HTM208B | HTM208C |
|---|---|---|
| Model Number | HTM208B | HTM208C |
| Digital Channels | 8 | |
| U disk | Yes | |
| Bluetooth | Without | Yes |
| SD card | Without | Yes |
| 4G (domestic) | Without | Without |
| 4G (Universal Version) | Without | |
| Graduation number | 11 kinds of thermocouples: J/K/T/E/S/N/B/R/A/C/D type | |
| Type K thermocouple basic accuracy | -200°C to 0°C error is ±1.2°C The error from 0°C to 1370°C is ±0.8°C (excluding thermocouple error) | |
| Temperature measurement range | Type K thermocouple: -200°C to 1370°C | |
| Resolution | 0.1°C | |
| Correction | Each channel can be corrected | |
| Alarm | Support upper and lower limit alarms | |
| Display | Support list display and bar graph display | |
| Cold Junction Compensation Accuracy | 0.5°C | |
| Software support | Android and Windows data acquisition software | |
| Storage | Support U disk (standard) and Micro SD card (optional) to save in real time | |
| Communication Interface | USB-Type C communication (standard), Bluetooth and 4G communication (optional) | |
| Maximum withstand voltage between channels | 300VDC | |
| Channel maximum input voltage | ±75VDC | |
| Plug in terminal type | E0308 Tube Terminal Blocks | |
| Display Resolution | 640*480 Pixels | |
| Display Size | 3.0 - 4.9 Inches | |
| Other | Support Chinese and English, Backlight brightness adjustment, Shutdown time settings | |
| Operating environment | 0-40°C, <80%RH | |
| Storage environment | -10-60°C, <90%RH | |
| Preheat time | >30 minutes | |
| Power Requirements | Input: 100VAC ~ 240VAC, 50Hz/60Hz Output: 5V/2A DC | |
| Battery parameters | Two 3.7V 18650 lithium batteries (Excluding batteries) | |
| Working hours | >15 hours | |
| Power consumption | 1W | |
| Size | 200mm*98mm*40mm | |
| Weight | 600g | |
10. User Tips
- Preheat for Accuracy: Allow the device to preheat for at least 30 minutes before critical measurements to ensure optimal accuracy.
- Check Thermocouple Type: Always double-check that the thermocouple type connected matches the setting in the device for correct readings.
- Regular Data Backup: Periodically back up your measurement data to an external storage device (U disk or SD card) to prevent data loss.
- Software for Analysis: Utilize the Windows or Android data acquisition software for advanced analysis and visualization of your temperature data.
11. Warranty and Support
Hantek products are designed for reliability and performance. For warranty information, technical support, or service inquiries, please refer to the official Hantek website or contact your local distributor. Keep your purchase receipt as proof of purchase for warranty claims.