1. Introduction
Thank you for purchasing our company's digital high-voltage insulation resistance meter. Before you use this instrument for the first time, in order to avoid possible electric shock or personal injury, please be sure to read and strictly observe the safety rules and precautions listed in this manual.
The ES3045 series digital insulation resistance meter, also known as a megohmmeter or high voltage insulation resistance tester, is designed for insulation resistance testing. This instrument features a large LCD screen with gray backlight, data storage, data access, alarm functions, automatic shutdown, and charging capabilities. It also measures DC voltage, AC voltage, absorption ratio, and polarization index of DC voltage. The device is robust, offering a wide range, high resolution, convenient operation, portability, accuracy, reliability, stable performance, and strong anti-interference ability. Its shockproof, dustproof, and moisture-proof structure makes it suitable for harsh working environments.
The ES3045 series is an essential instrument for telecommunications, electricity, meteorology, computer rooms, oil fields, electromechanical installation and maintenance, and industrial enterprises that rely on electricity. It is suitable for measuring the resistance value of various insulating materials and the insulation resistance of transformers, motors, cables, and other electrical equipment. It offers a step-adjusting voltage function, allowing it to output various voltage levels (e.g., 15KV models can generate 12KV levels), expanding its utility.
The digital insulation resistance meter is built with medium and large-scale integrated circuits, providing high output power and multiple output voltage levels (6 voltage levels). The rated insulation test voltage range is 100V ~ 15kV (model-dependent), with an insulation resistance measurement range of 0.01MΩ ~ 50TΩ (model-dependent). It also measures DC voltage from 0 ~ 1000V, AC voltage from 0 ~ 750V, DC current from 0.01nA ~ 6mA, and capacitance from 10nF ~ 50uF.
Product Demonstration Video
Video demonstrating the operation and features of the FUZRR Digital Insulation Resistance Tester.
2. Safety Precautions and Procedures
In any case, the use of this instrument should pay special attention to safety.
- The tester conforms to IEC61010 on design, production, and test.
- Do not use high-frequency signal generators like mobile phones to avoid errors during measuring.
- Pay attention to words and symbols on the Tester.
- Ensure that the tester and accessories are in good condition before use; it can only be used when there are no damaged, naked, or broken parts in testing wires or insulation layer.
- During measurement, it is forbidden to touch bare conductors and circuits under measurement.
- Confirm that the connector plug of the lead has been inserted into the tester interface closely.
- Do not impose over 600V A.C. or D.C. voltage on the part between the testing end and interface, as it may damage the tester.
- Do not measure in an inflammable place. Sparks may cause an explosion.
- If exposed metal is caused by a broken enclosure or testing wires, stop using the tester immediately.
- Do not keep or store the tester in spots with high temperature, high moisture, condensation, or under direct sunlight for a long time.
- When the meter displays the low battery voltage symbol, charge it in time, otherwise it will lead to ground error.
- Do not charge or perform data transmission during the measurement process.
- Pay attention to the measuring range and usage environment stipulated for the Tester.
- This measuring device is only to be used, disassembled, adjusted, and repaired by qualified personnel with authorization.
- If continuous use of the Tester itself causes a hazard, stop using it immediately and deposit it for disposal by an authorized agency.
- For risks indicated by the danger icon in the manual, users must perform safety operations strictly in compliance with the manual content.
- The instrument outputs high voltage. Be sure to connect the test line hand to leave the test line before pressing the test button to test, otherwise there is a danger of electric shock.
- After the test is completed, wait for one minute for the discharge to complete before removing the test line. When removing the test line, first remove the test line of the meter and then collect the test line.
3. Product Overview
Structure of Meter
The diagram below illustrates the various components and interfaces of the ES3045 series digital insulation resistance meter.

Figure: Structure of the ES3045 Series Digital Insulation Tester
- LCD display screen
- LINE interface
- V interface
- GUARD interface
- EARTH interface
- Changer interface
- USB interface
- Voltage select button
- Test button
- Power on/off button
- Security Alligator Clip
- Test line (green, black each 1PC)
- High voltage test terminal banana plug into LINE interface
- Shielded cable connector plugged into GUARD interface (no connection required without this connector)
- High voltage test line
Outline Dimensions
The physical dimensions of the device are shown in the diagram below.

Figure: Outline Dimensions (240mm x 188mm x 85mm)
Performance Characteristics
Key performance characteristics of the ES3045 series are summarized below:

Figure: Performance Characteristics Overview
4. Technical Specifications
Model Specifications
| Model | Maximum Output Voltage | Resistance Limit Range |
|---|---|---|
| ES3045 | 5000V | 10TΩ |
| ES3045E | 10KV | 35TΩ |
| ES3045F | 15KV | 50TΩ |
General Specifications
| Feature | Specification | Accuracy / Resolution |
|---|---|---|
| Functions | Insulation resistance test, voltage test, current test, capacitance test | |
| Baseline Conditions | 23℃ ± 5℃, below 75%RH | |
| Rated Applied Voltage (ES3045) | 100V, 250V, 500V, 1000V, 2500V, 5000V | Accuracy: rated voltage × (1±10%) |
| Rated Applied Voltage (ES3045E) | 250V, 500V, 1000V, 2500V, 5000V, 10KV | Accuracy: rated voltage × (1±10%) |
| Rated Applied Voltage (ES3045F) | 500V, 1000V, 2500V, 5000V, 10KV, 15KV | Accuracy: rated voltage × (1±10%) |
| Insulation Resistance Range (ES3045) | 0.01MΩ ~ 10TΩ | Accuracy: ±3%rdg±5dgt, Resolution: 0.01MΩ |
| Insulation Resistance Range (ES3045E) | 0.01MΩ ~ 35TΩ | Accuracy: ±3%rdg±5dgt, Resolution: 0.01MΩ |
| Insulation Resistance Range (ES3045F) | 0.01MΩ ~ 50TΩ | Accuracy: ±3%rdg±5dgt, Resolution: 0.01MΩ |
| DC Voltage Range | 0.1 ~ 1000V | Accuracy: ±1.5%rdg±3dgt, Resolution: 0.1V |
| AC Voltage Range | 0.1 ~ 750V | Accuracy: ±1.5%rdg±3dgt, Resolution: 0.1V |
| DC Current Measurement | 0.01nA ~ 6mA | Accuracy: ±5.0%rdg±5dgt, Resolution: 0.01nA |
| Capacitance Measurement | 10nF ~ 50uF | Accuracy: ±10%rdg±5dgt, Resolution: 10nF |
| Output Short Circuit Current | ≥6mA | |
| Absorption Ratio & Polarization Index Measurements | Included | |
| Power Supply | 12V rechargeable battery | |
| Backlight | Controllable gray and white screen backlight, suitable for use in dark places | |
| Display Mode | 4-digit large LCD display, gray and white screen backlight | |
| LCD Display Size | 108mm × 65mm | |
| Meter Size | 240mm (L) × 188mm (W) × 85mm (H) | |
| USB Interface | With USB interface, software monitoring, storage data can be uploaded to the computer, save and print | |
| Communication Line | 1 USB communication line | |
| Test Line | High voltage bar red 3 meters, high voltage test line black 1.5 meters, green line 1.5 meters | |
| Data Storage | 500 groups, flashing "FULL" symbol indicates that the storage is full | |
| Data Access | Data access function: "READ" symbol display | |
| Overflow Display | Overrange overflow function: "OL" symbol display | |
| Alarm Function | When the measured value exceeds the alarm setting value, an alarm prompt will be issued | |
| Power Consumption (Standby) | 30mA Max (backlight off) | |
| Power Consumption (Backlight On) | 42mA Max | |
| Power Consumption (Measurement) | 200mA Max (backlight off) | |
| Instrument Weight | 2340g (including battery) | |
| Battery Voltage | Low battery voltage, showing low battery symbol | |
| Automatic Shut-down | The meter shuts down after about 15 minutes of inactivity | |
| Insulation Resistance (between measurement line and housing) | ≥50MΩ | |
| Pressure Resistance | AC3kV/50Hz 1min | |
| Operating Temperature and Humidity | -10℃ ~ +50℃ <85%RH | |
| Storage Temperature and Humidity | -15℃ ~ +55℃ <90%RH | |
| Safety Regulations Compliance | IEC61010-1, IEC1010-2-31, IEC61557-1, 5, IEC60529 (IP54), pollution grade 2, CAT Ⅲ 300V | |
5. Operation Method
5.1 Power On/Off
Press the POWER button to turn the switch on and off. After the power is turned on, "APO" is displayed in the lower corner. When it is not in operation, it will automatically shut down after 15 minutes.
5.2 Battery Voltage Check
After powering on, if the LCD shows a low battery voltage symbol, it indicates that the battery is running low. Please charge in time. Sufficient battery power ensures measurement accuracy.
5.3 DC Voltage Test
⚠️ Input instrument DC voltage cannot exceed 1000V.
When measuring, turn the rotary switch to the V= gear to switch to the DC voltage test mode. Connect the green test lead to the V terminal, the black test lead to the COM, and the LCD displays the real-time DC voltage value.

Figure: DC Voltage Test Setup
5.4 AC Voltage Test
⚠️ Input instrument AC voltage cannot exceed 750V.
During measurement, press the V~ button to switch to AC voltage test mode, connect the green pen to the V terminal, and the black test pen to COM. The LCD displays the real-time AC voltage value.

Figure: AC Voltage Test Setup
5.5 DC Current Test
When measuring, turn the rotary switch to the A= gear to switch to DC current test mode. Connect the green lead to the E terminal, the black lead to the COM, and the LCD displays the real-time current value.

Figure: DC Current Test Setup
During the current test, units of mA, nA, uA, and pA are displayed as shown in the figures below.

Figure: Current Unit Displays
5.6 Capacitance Test
⚠️ Pay attention to standard operation during the capacitor test. Check whether the capacitor is charged before the test. After the test, the capacitor must be discharged. Do not touch the capacitor without discharge.
When measuring, press the C button to switch to the capacitance test mode. One end of the high-voltage rod test lead (red) is connected to the meter LINE and the other end is in contact with the capacitor under test. One end is connected to COM and the other end is connected to the capacitor. Press the TEST key to start the test (the meter will emit an intermittent beep: the prompt sound of "drop ~ drop ~ drop..." during the test). After the test is completed, the LCD displays the correct capacitance value. Note: Pay attention to the standard operation during the capacitance test. Manually discharge after each test. Do not touch the capacitor by hand without discharging.

Figure: Capacitance Test Setup
The capacitance measurement mode has 4 gears and requires manual shift testing. The first range is 10nF-100nF, the second range is 100nF-1000nF, the third range is 1uF-10uF, and the fourth range is 10uF-50uF. During the test, if the value is greater or less than a certain value, it needs to be manually switched to the corresponding gear. The gear can be switched by pressing the < (gear minus 1) or > (gear plus 1) button.

Figure: Capacitance Measurement Displays
5.7 Insulation Resistance Test
⚠️ Insulation resistance test can only be carried out on an uncharged circuit. Before testing, check whether the test circuit wiring is in good condition and whether the circuit under test is energized. If the circuit is live, it may damage the instrument and affect the measurement accuracy.
⚠️ Must wear high-voltage insulating gloves to operate.
In the insulation resistance range, press the TEST switch to generate high voltage in the test line head and in the circuit under test. Be careful to avoid touching.
- Be sure to connect the grounding wire (black) to the grounding port of the circuit under test.
- Do not touch the circuit immediately after testing. The stored charge may cause electric shock.
- Do not remove the test lead immediately. Wait until the discharge is complete before touching the circuit under test.
- In order to ensure the measurement accuracy, do not twist the test lines together.
Notes for testing high insulation resistance:
⚠️ After the high-voltage insulation material is added with DC voltage, the current passing through the sample is very small, and it is very susceptible to the influence of external interference, causing large test errors.
- The higher the measured resistance value, the longer the measurement time.
- As humidity increases, surface leakage increases, and bulk electrical current also increases.
- The resistance value of general materials decreases with the increase of ambient temperature and humidity.
Guaranteed temperature and humidity of the insulation resistance accuracy:
| Insulation Resistance Range | Guaranteed Humidity Value | Guaranteed Temperature |
|---|---|---|
| 0Ω-100MΩ | <85% RH (No condensation) | 23℃ ± 5℃ |
| 100MΩ-20GΩ | <75% RH (No condensation) | |
| 20GΩ-1000GΩ | <65% RH (No condensation) | |
| 1000GΩ-1TΩ | <55% RH (No condensation) | |
| 1TΩ-10TΩ | <45% RH (No condensation) | |
| 10TΩ-50TΩ | <35% RH (No condensation) |
Insulation Resistance Test Operation:
1. Insulation resistance test can only be carried out on an uncharged circuit. Before testing, check that the test leads are good and confirm that the circuit under test is uncharged.
2. Turn the rotary switch to the Ω gear to switch to resistance test mode, then press the voltage selection buttons (100V, 250V, 500V, 1000V, 2500V, 5000V, 10KV, or 15KV) to select the voltage value to be tested.
3. One end of the ground wire (black) is connected to the instrument's EARTH connection, and the other end is connected to the ground end of the circuit under test. One end of the high-voltage rod test line (red) is connected to the instrument's LINE end, and the head is in contact with the circuit under test. If the test has a green branch line as a shielded line, connecting the GUARD port during testing improves accuracy (testing resistance above TΩ must be connected with green shielded wire). If the matching test line does not have this line, it is not necessary to connect it. As shown in the figure, press the TEST button (during the test, the meter will emit an intermittent beep: "Drip ~ Drop ~ Drop..." and "TEST" will flash in the lower left corner of the screen). The LCD shows the measured value. Read the insulation resistance value after the measured value is fixed. After the resistance test is completed, press and hold the Ω key to switch the displayed current value, and then press the Ω key to switch back to the resistance value.

Figure: Insulation Resistance Test Wiring Diagram
5.8 GUARD Use of Protective Wires
When measuring the insulation resistance of a cable, leakage current from the covered surface passes through the interior of the insulator and converges, leading to errors in the insulation resistance value. To avoid this, use a protection wire (any conductive bare wire) to divert the leakage current. After connecting to the protection port, the leakage current does not flow through the indicator, allowing the insulator to accurately measure the insulation resistance. Use the protection test cable from the accessories to connect to the protection port.

Figure: GUARD Connection for Accurate Measurement
5.9 Polarization Index (PI) and Absorption Ratio (DAR)
The Polarization Index (PI) and Absorption Ratio (DAR) tests check whether the leakage current of the insulator has increased. These values indicate changes in insulation resistance over time after the measured object withstands the applied voltage.
Difference between PI and DAR:
For general insulation tests (e.g., housing insulation, tool handles), a short-term test is often sufficient, as leakage current increases with voltage application time. This short-term insulation resistance ratio is called the Absorption Ratio (DAR).
For large-capacity and long-term absorption processes (e.g., transformers, generators, cables, electrical equipment), DAR may not fully reflect the absorption process. In such cases, the Polarization Index (PI) is used, which is the ratio between the insulation resistance at 10 minutes (R10min) and 1 minute (R1min), describing the entire absorption process.
The PI and DAR values are calculated as follows:
- PI (Polarization Index) = R 10 Min / R 1 Min
- DAR (Absorption ratio) = R 60 Sec / R 15 Sec
- DAR (Absorption ratio) = R 60 Sec / R 30 Sec
Remarks:
- R10Min = resistance value measured by the voltage applied for 10 minutes.
- R1Min = R60Sec = the resistance value measured by the voltage applied for 1 minute.
- R30Sec = the resistance value measured by the voltage applied for 30 seconds.
- R15Sec = the resistance value measured by the voltage applied for 15 seconds.
- DAR calculation time can be chosen as 15 seconds or 30 seconds.
PI and DAR Test Procedure:
1. The Polarization Index (PI) and Absorption Ratio Test (DAR) can only be performed on uncharged circuits. Before testing, check that the test leads are good and confirm that the circuit under test is uncharged or not.
2. Turn the rotary switch to the Ω gear, then press the voltage selection buttons (100V, 250V, 500V) to select the voltage value to test.
3. Press the MODE key to set the corresponding mode. The LCD shows "10:01m" for polarization index mode, "60:15S" for absorption mode 15 second, and "60:30S" for absorption mode 30 second. Small numbers do not show anything for the insulation resistance measurement mode.

Figure: PI and DAR Mode Displays
4. One end of the ground wire (black) is connected to the instrument's EARTH connection and the other end to the ground end of the circuit under test. One end of the high-pressure rod test line (red) is connected to the instrument's LINE end and the head is in contact with the circuit under test. Press the TEST key. The LCD displays the measured value. After the measured value is fixed, the absorbance or polarization index can be read.
5. After the test is completed, you can press the TEST button to switch to view the divisor and ratio of the absorption ratio or polarization index value, or press the MODE button to switch to view the ratio or divisor of the absorption ratio or polarization index value. (For example, in "60: 15S" mode, the lower left corner displays "60: 15S" as the ratio, the display "15S" as the dividend, and the display "60S" as the divisor; other modes are similar).

Figure: PI and DAR Result Displays
PI and DAR Applications:
In engineering, insulation resistance and absorption ratio (or polarization index) can reflect the degree of moisture in the insulation of generators, oil-immersed power transformers, and other equipment. The value of the absorption ratio (or polarization index) decreases after the insulation is wet, making it an important indicator of whether the insulation is affected by moisture.
It should be noted that sometimes insulation may have obvious defects (e.g., breakdown under high pressure), yet the absorption ratio or polarization index value may still appear good. Therefore, these values cannot be used to find other local insulation defects beyond moisture and dirt.

Figure: Insulation Resistance vs. Time Graph
Polarization Index Reference Judgment Value:
| Polarization Index | Above 4 | 4 ~ 2 | 2.0 ~ 1.0 | Below 1.0 |
|---|---|---|---|---|
| Judge | The best | Good | Need to pay attention | Bad |
Absorption Ratio Reference Judgment Value:
| Absorption Ratio | Above 1.4 | 1.25 ~ 1.0 | Below 1.0 |
|---|---|---|---|
| Judge | The best | Good | Bad |
5.10 Backlight Control
After power on, press the MODE key to turn the backlight on or off. The backlight function is suitable for dim places. The default backlight turns off every time you turn it on.
5.11 Alarm Value Settings
1. After power on, long press the MODE button to turn on and off the alarm function.
2. Long press the V= key to enter alarm value setting mode. Short press the V= key to select resistance setting, short press the V~ key to select DC voltage setting, short press the A= key to select AC voltage setting. Then press the < (plus 10) key or > (minus 10) key, or the YES key (plus 1) or NO key (minus 1) key to change the current number size. Finally, press the TEST button to save and exit.
3. When the measured voltage value is greater than the alarm critical set value or the insulation resistance value is less than the alarm critical set value and the alarm function is turned on, the instrument flashes the "))" symbol and issues a “beep-beep-beep--” alarm sound. The maximum value of the DC voltage alarm setting is 900V, the maximum value of the AC voltage alarm setting is 700V, and the maximum value of the insulation resistance alarm setting is 9999MΩ. The following figure shows an example ("<" is less than the symbol to indicate an alarm, and ">" is greater than the symbol to indicate an alarm):

Figure: Alarm Setting Displays
5.12 Data Lock/Storage
1. After the measurement is completed, press the MEM key to lock the currently displayed data, and automatically numbered storage. Short press the MEM key again to exit the lock. If the storage is full, the meter displays the "FULL" symbol.
2. As shown in the following figure: the measured data is 1258MΩ, and the "MEM" display indicates it's stored as the fifth set of data. The measured data is 1258MΩ, short press the MEM button to display and store as the fifth group of data.

Figure: Data Lock and Storage Displays
5.13 Data Review/Deletion
1. After booting, if the meter has saved data, long press the MEM button to enter the data lookup, and the data read interface "MR" symbol displays. Press the < or > key to check the corresponding data with a step value of 10, press the YES or NO key to check the corresponding data with a step value of 1, and then short press the MEM button to exit the check.
2. As shown in the figure below: The number 5 in the lower left corner of the screen when checking is the currently stored 5th data. If there is no stored data, the LCD displays "NULL". (The resistance data in the lower right corner shows the voltage, current and capacitance display units used in the test).

Figure: Data Review Displays
3. In the data review state, long press the MEM key to enter the data deletion, press the NO key not to delete and return to the data review state, press the YES key to delete all the stored data. Delete page as shown below:

Figure: Data Deletion Confirmation
5.14 Step Adjustment Resistance Measurement Voltage
After starting up, you can modify the voltage value by pressing the < or > key at a step value of 50V below the 10KV voltage range, or by pressing the YES or NO key at a step value of 5V. For 10KV voltage range and above, you can modify the voltage value by pressing the < or > key with a step value of 500V, or press the YES or NO key with a 50V step value to modify the voltage value. Note: The maximum voltage does not exceed 15KV, and the step accuracy is ± 20%.
6. Battery Description
- The meter uses a 12.6V lithium battery for power supply. When the battery power is low, the power symbol is displayed; please charge it in time. Note: When the battery power is low, the measurement accuracy will be affected.
- The higher the measurement voltage, the higher the battery power requirements.
7. Accessories / Packing List
| Item | Quantity | Description |
|---|---|---|
| Instrument | 1 set | ES3045, ES3045E, ES3045F (Optional) |
| High pressure rod | 1 | Red |
| High Voltage Test Lead | 2 | 1 black, 1 green each |
| Monitoring software CD | 1 | |
| USB communication cable | 1 | |
| Charger | 1 | |
| Manual, Warranty | 1 set | |
| Instrument box | 1 |

Figure: Included Accessories
8. Troubleshooting
- Low Battery: If the LCD shows a low battery voltage symbol, charge the battery immediately to ensure accurate measurements.
- Measurement Errors: High-frequency signal generators (e.g., mobile phones) can cause errors. Ensure such devices are not near the tester during operation.
- Alarm Triggered: If the instrument flashes the "))" symbol and beeps, it indicates that the measured voltage or insulation resistance value is outside the set alarm limits. Check your measurement and alarm settings.
- Overflow Display ("OL"): This indicates that the measured value exceeds the instrument's measurement range. Adjust the test voltage or range if possible, or note that the value is beyond the device's capacity.
- No Stored Data ("NULL"): If the data review shows "NULL", it means no data has been saved. Ensure you press the MEM button after a measurement to save data.
- Inaccurate Readings: Ensure test leads are properly connected and not twisted together. Verify the circuit under test is uncharged before starting insulation resistance tests. High humidity or extreme temperatures can affect accuracy.
9. User Tips
- Always perform a quick battery check before critical measurements to avoid interruptions and ensure accuracy.
- For insulation resistance tests, especially on sensitive equipment, utilize the GUARD connection with a shielded cable to minimize leakage current interference and improve measurement precision.
- When testing high insulation resistance, be aware that the measurement time increases with higher resistance values. Plan accordingly for longer test durations.
- Regularly review stored data to track trends and maintain a record of equipment health.
- Keep the instrument clean and store it in a dry, temperature-controlled environment to prolong its lifespan and maintain accuracy.
10. Warranty and Support
The contents of this user manual cannot be used as a reason for using the product for special purposes. The company is not responsible for other losses caused by use. The company reserves the right to modify the contents of the user manual. Subject to change without notice.