1. Safety Information
Please read this manual carefully before first operation and keep it for future reference.
Warning:
- This device is intended for use by qualified dental professionals familiar with anesthesia procedures in hospitals or dental clinics.
- The device must be used within the scope mentioned in this manual. The manufacturer is not responsible for any damage caused by improper use or use for unintended purposes.
- Before using the device, ensure that the voltage is within the specified operating range. Improper input voltage will damage the device and cause harm to the operator and the patient.
- The power plug is used for disconnection in case of problems, it must be easily accessible at all times.
- To avoid electric shock, do not insert other objects into the device.
- Please use the original cartridge holder, otherwise it will be unusable or even cause injury accidents.
- Please use original power adapter. Other power adapter will result in damage to lithium battery and control circuit.
- Please use sterile injection needles and dental cartridge anesthetic that are approved for sale on the market.
- Use only disposable sterile dental needles compliant with YY/T 0587-2018 and cartridges of anesthetic legally registered and approved.
- Do not modify, disassemble or repair the device yourself.
- Avoid dropping or shocking the device.
- Do not replace the battery yourself. Incorrect replacement of lithium batteries can lead to hazards (e.g. over-temperature, fire or explosion).
- Check the equipment before use to ensure that there are no abnormalities.
- In the event of serious abnormality of the equipment caused by improper use or physical damage, stop using it immediately and switch off the equipment; Only the manufacturer and authorised dealers may open the unit for repair.
- The device has electromagnetic interference. Do not use it around patients with cardiac pacemakers or electronic surgery.
Contraindications:
- Haemophilia patients are prohibited.
- Patients with active periodontal disease are prohibited from intra-ligament injections.
- Patients with heart disease should use with caution.
- Pregnant women and young children should use with caution.
2. Product Components
This product consists of the main unit, cartridge holder, charging case, and power adapter, among other items.



3. Product Structure and Screen Display
The device features an intuitive LCD display and clearly labeled buttons for ease of use.

| Button Name | Icon | Function |
|---|---|---|
| Start/Aspiration button | II► | Short press: Start/stop injection Long press: Aspiration |
| Mode/Reset button | ≡ | Short press: Adjust speed mode Long press: Reset plunger to the initial position |
| Power/Volume button | ⏻ | Short press: Adjust volume Long press: Power on/off |

4. Setup and Initial Preparation
4.1 Charging
The main unit can be charged via the charging case. Press and hold the charging case power button to start charging the main unit. Alternatively, connect the adapter to the charging case to charge both the case and the main unit simultaneously.
Charging Indicator Status:
- Blue light flashing: The charging case is charging.
- Blue light always on: The number of lights indicates the power level of the charging case, and 3 lights indicate full charge.
- Orange light flashing: The main unit is charging.
- Green light always on: The main unit is fully charged or not detected.
- No light on: The charging function is abnormal.
Note: The main unit will automatically shut down when the battery is very low. For wireless charging, place the main unit screen up and ensure it is fully seated in the charging slot. If tilted or not in contact with the slot, charging may fail.
4.2 Installation and Disassembly
Installation Steps:
- Press to open the charging case, take out the main unit and required accessories.
- Insert the dental cartridge anesthetic into the cartridge holder, ensuring the seal head enters first.
- Engage the cartridge holder into the finger grips.
- Align the groove of the finger grips with the connector on the front of the main unit, push it to the bottom, and rotate clockwise until the finger grips lock securely with the main unit.



Video Guide: Cartridge Holder Installation and Button Introduction
This video demonstrates the steps for installing the cartridge holder and introduces the various button functions of the device.
4.3 Pre-Use Check
Before each use, perform an operational check on the product's functions (e.g., injection, aspiration, reset). Ensure there are no abnormalities before starting work.
5. Operation
5.1 Turning On/Off
- Long press the Power button (⏻) to turn the main unit on or off.
- If no operation is performed for 3 minutes, the device will automatically shut down. The plunger will not reset automatically. To restart, press and hold the power button.
- If the plunger is not in the initial position when shutting down, press and hold the power button to reset the plunger before shutting down.
5.2 Injection Speed Mode Selection
- Briefly press the Mode/Reset button (≡) to switch injection modes.
- Three modes are available: High, Low, and PDL.
5.3 Volume Adjustment
- Briefly press the Power/Volume button (⏻) to adjust the main unit's volume.
- Three levels are available: High, Low, and Off.
- When the volume is set to "Off", only the music volume is muted. Operation prompts and button sounds remain active.
5.4 Injection and Aspiration
Injection:
- Briefly press the Start/Aspiration button (II►) or the button on the finger grip to start injection. Press again to stop injection.
Aspiration:
- Long press the Start/Aspiration button (II►) or the button on the finger grip to initiate aspiration. Each aspiration draws approximately 0.1mL.
- Perform aspiration after needle insertion and before injecting anesthetic to avoid intravascular injection.
- Aspiration can only be performed after injecting 0.1ml.
5.5 Reset
- Long press the Mode/Reset button (≡) to reset the plunger to its initial position. The displayed injection volume on the screen will also reset to zero.
- When the injection volume reaches 1.8mL, the device will issue an "anesthetic exhausted" voice prompt. Ten seconds after this prompt, the plunger will automatically reset to its initial position.
Operation Warning:
- Do not apply excessive pressure to the main unit during injection.
- Refer to professional dental anesthesia textbooks to carefully set the injection speed and dosage.
6. Cleaning, Disinfection, and Sterilization
Clean, disinfect, and sterilize patient-contacting parts (cartridge holder) before first use and after each use.
6.1 Cleaning
- Main unit, charging case: Wipe off any visible residue with a clean damp cloth.
- Cartridge holder, finger grip: Rinse under tap water and dry thoroughly with a clean, dry cloth.
6.2 Disinfection
- Main unit, charging case, finger grip: Disinfect the surfaces by wiping with 75% ethanol.
- Cartridge holder: Soak in 75% ethanol for 30 minutes for disinfection.
6.3 Drying
- Dry immediately after cleaning and disinfection. Use compressed air if available.
6.4 Sterilization
- Serializable components: cartridge holder.
- Package with a sterilization bag according to approved by the state before sterilization.
- Sterilization method: Pressure steam sterilization is recommended.
- Sterilization conditions: At 210kPa and 134°C, more than 5 minutes.
6.5 Storage
- Store sterilized instruments in a dry, clean, and dust-free environment.
7. Troubleshooting
| Malfunctions | Possible Cause | Solution |
|---|---|---|
| Main unit does not work | Low battery Battery protection activated Battery damaged | Charging Charging Contact the dealer |
| No charging after inserting the power adapter | Poor power connection Incorrect power adapter Power adapter damaged | Check the connection between the adapter and charging case Check the adapter specifications or contact the dealer Contact the dealer |
| Batteries run out quickly | Battery aging | Contact the dealer |
| Aspiration function malfunction | O-ring worn | Replace the O-ring |
| Plunger does not reset to the initial position | The main unit does not have enough power to support normal operation | Place the main unit in the charging case to charge until it can power on normally |
| E1 error | Reset failure or coupler disconnected | Contact the dealer |
| E2 error | Excessive resistance during reset | Remove the cartridge holder, inspect the plunger to clear any foreign objects |
| E3 error | Plunger damage Encoder failure | Contact the dealer Contact the dealer |
8. Technical Specifications
| Parameter | Specification |
|---|---|
| Power adapter input | 100-240V ~ 50/60Hz |
| Power adapter output | 5V 1.5A |
| Main unit lithium battery | 3.7V 220mAh |
| Charging case lithium battery | 3.7V 1800mAh |
| Degree of protection against electric shock | Type B applied part |
| Maximum injection volume | 1.8mL |
| Injected volume range | 0mL-1.8mL |
| Injected volume accuracy | 0.1mL |
| Classified by security | Non-AP/APG |
| Degree of protection against harmful ingress of water | IPX0 |
| Operating mode | Continuous operating device |
| Application part | Cartridge holder |
| Overvoltage category | Class II |
| Pollution degree | Degree 2 |
| Environmental temperature (Operating) | +5 °C ~ +40 °C |
| Relative humidity (Operating) | 20% ~ 80%RH |
| Atmospheric pressure (Operating) | 80kPa ~ 106kPa |
| Environmental temperature (Transport & Storage) | -10°C ~ +55°C |
| Relative humidity (Transport & Storage) | ≤93%RH |
| Atmospheric pressure (Transport & Storage) | 50kPa ~ 106kPa |
9. Warranty and Support
Main unit and Charging case warranty time for 24 months from the date of purchase. The product with accessories (adapter) warranty for 6 months. The rest of the accessories do not warranty.
Upon request, the supplier will provide circuit diagrams, component lists, legends, calibration details, or other information necessary for qualified technicians to help users repair equipment parts designated by the manufacturer for repair.
For further support, please contact your authorized dealer or visit www.coxotec.com.
10. User Tips
No specific user tips were available from provided reviews or Q&A content.
11. Symbols
| Symbol | Meaning |
|---|---|
| ! | General warning |
| i | Note |
| Type B applied part | |
| Vertical up | |
| SN | Serial number |
| Fragile, handle with care | |
| Date of manufacture | |
| Indoor use only | |
| ~ | Alternating current |
| Caution | |
| Refer to operating instructions | |
| Class II equipment | |
| Sterilizable in a steam sterilizer at 134°C | |
| Keep dry | |
| Do not dispose of the product into the ordinary municipal waste or garbage system | |
| Manufacturer | |
| Direct current |
12. Recycling and Disposal
The device and its packaging are designed to be as environmentally friendly as possible.
- Ensure that the parts are not contaminated on disposal.
- Follow your local and country specific laws, directives, standards and guidelines for disposal.
- This product is classified as a Medical device.
- This product is classified as Waste electrical equipment.
- Packaging should also be disposed of responsibly.
13. Electromagnetic Compatibility (EMC)
The COXO GENI Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
13.1 Electromagnetic Emissions
| Emissions Test | Compliance | Electromagnetic Environment - Guidance |
|---|---|---|
| RF emissions CISPR 11 | Group 1 | GENI Pro uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic devices. |
| RF emissions CISPR 11 | Class B | GENI Pro is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
| Harmonic emissions IEC 61000-3-2 | Class A | |
| Voltage fluctuations/ flicker emissions IEC 61000-3-3 | Complies |
13.2 Electromagnetic Immunity
| Immunity Test | IEC 60601 Test Level | Compliance Level | Electromagnetic Environment - Guidance |
|---|---|---|---|
| Electrostatic discharge (ESD) IEC 61000-4-2 | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. |
| Conducted RF IEC 61000-4-6 | 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM and amateur radio bands | 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM and amateur radio bands | Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. |
| Radiated RF IEC 61000-4-3 | 3 V/m, 10 V/m 80 MHz to 2.7 GHz 385MHz-5785MHz Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 60601-1) | 3 V/m, 10 V/m 80 MHz to 2.7 GHz 385MHz-5785MHz Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC 60601-1) | Recommended separation distance: d= [3.5/V1]x√P (150 kHz to 80 MHz), d= 1.2×√P (80 MHz to 800MHz), d= 2.3×√P (800 MHz to 2.7GHz). Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: |
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not be applicable in every case. The propagation of electromagnetic waves is subject to absorption and reflection by buildings, objects, and people.