Kamoer UIP E

Kamoer UIP E V2 Peristaltic Pump User Manual

Model: UIP E

1. Introduction

This manual provides detailed instructions for the installation, operation, and maintenance of the Kamoer UIP E V2 Peristaltic Pump. Please read this manual thoroughly before operating the device to ensure safe and efficient use.

Kamoer UIP E V2 Peristaltic Pump with touch screen and pump head

Figure 1: Kamoer UIP E V2 Peristaltic Pump

2. Safety Information

  • Ensure the power supply matches the device's specifications (AC 100-240V).
  • Do not operate the pump with damaged power cords or tubing.
  • Keep the device away from water or other liquids to prevent electric shock.
  • Always disconnect power before performing any maintenance or installation.
  • Use only Kamoer-approved accessories and replacement parts.

3. Package Contents

Verify that all items are present in the package:

  • 1 × UIP WIFI E Peristaltic Pump (Main Unit)
  • 1 × Power cable
  • 1 × Foot switch
  • 1 × Tube (2m silicone, 7.9mm ID x 11.1mm OD)
  • Terminals
  • Grease
  • Product Manuals
Kamoer UIP E V2 Peristaltic Pump package contents including main unit, power cord, foot switch, silicone tube, grease, terminals, and product manuals.

Figure 2: Package Contents

4. Product Overview

The Kamoer UIP E V2 is a high-flow peristaltic pump designed for laboratory and filling applications. It features a 4.3-inch touch screen for easy operation and supports various control methods.

Key Features:

  • 4.3-inch touch screen display
  • Quick tube change mechanism
  • Programmable operation
  • Optional photoelectric sensor compatibility
  • Multiple control options: touch screen, foot switch, RS485, 0-5V/10V analog, 4-20mA analog
  • Supports flow rate calibration
  • Continuous, Quantitative (Ration), and Remote operating modes
  • Power-down parameter saving
  • Easy-to-connect quick-connect terminals on the back interface
  • Remote firmware upgrade support

Component Identification:

Front, side, back, and positive side views of the Kamoer UIP E V2 Peristaltic Pump, showing its handheld design and thermal vents.

Figure 3: Overall Product Views

The pump features a handheld design and thermal vents for efficient heat dissipation. The rear panel includes a power switch, power connector, and various interface terminals for control.

5. Specifications

ParameterValue
Brand Namekamoer
Model NumberUIP-E
Power SourceElectric (AC 100-240 V)
ApplicationTransfer, Lab, Filling
TheoryPeristaltic pump
StructureRoller pump (3 rotors)
Standard or NonstandardStandard
PressureLow Pressure
OriginMainland China
CertificationCE
Power (W)Electric (<50W)
Peristaltic Pump Flow Rate0.5~1500ml/min
Speed Range0.1-350 rpm
Speed Adjustment Resolution0.1 rpm
Control modeTouch screen, foot switch, 0-5/10V analog quantity, RS485 communication
Motor TypeStepper Motor
LifetimeMore than 6000 H
Pump HeadSingle Head-KK1800 (1.6mm wall thickness)
Temperature0~60℃
Humidity0~80℃
Overall weight3.15KG
Size (L*W*H)324*164*219mm
Noise≤65dB (Less than 60dB measured at a distance of 50cm)

Reference Flow Rate (ml/min)

Tube codeTube materialID(mm) x OD(mm)Pump headFlow rate(ml/min)
18#Silicone-200H7.9 × 11.1KK1800-3 rotors (1.6mm Wall thickness)0.03~1500
18#BPT-1000H7.9 × 11.1KK1800-3 rotors (1.6mm Wall thickness)0.03~1500

6. Setup and Installation

6.1 Power Connection

  1. Connect the power cord to the rear power connector of the pump.
  2. Plug the power cord into a suitable AC 100-240V power outlet.
  3. Flip the power switch on the rear panel to the 'ON' position. The screen will display immediately.

6.2 Pump Tube Installation

  1. Pull the slide bar on the pump head to open the pump cover.
  2. Carefully put the No. 18 silicone tube into the pump head, ensuring it is properly seated.
  3. Snap the pump tube on both sides into the designated grooves.
  4. Straighten the pump tube to ensure no kinks or twists.
  5. Pull the slide lever back to close the pump cover securely.

Video 1: Official Tutorial for UIP WIFI E-KK1800, demonstrating unboxing, device installation, and operation modes.

6.3 Foot Switch Connection

  1. Locate the connector for the foot switch in the lower right corner of the pump's rear panel.
  2. Connect the leads of the foot switch into the terminals. Note: The red wire is connected to the foot switch terminal, and the yellow wire is connected to the GND terminal.
  3. Insert the terminal block into the corresponding connector on the pump.
Rear panel of the Kamoer pump showing power switch, exhaust fan, and two interface blocks. Interface A for RS485 and 0-5/10V analog control. Interface B for digital foot pedal and direction control. Detailed pinout for Interface B is shown: 5 OUT PUT, 6 DIR, 7 Foot Switch, 8 GND.

Figure 4: Rear Panel Interfaces and Foot Switch Wiring Diagram

7. Operating Modes

The Kamoer UIP E V2 offers three main operating modes: Continuous, Ration (Quantitative), and Remote. These modes can be selected via the touch screen interface.

Touch display showing mode selection for Continuity, Ration, and Remote, along with remote control options for WiFi/Bluetooth, RS485, CAN, and analog quantities (0-5V, 0-10V, 4-20mA).

Figure 5: Touch Display Interface for Mode Selection and Remote Control

7.1 Continuous Mode

This mode is used for non-fixed amount liquid transfer. The pump runs continuously until manually stopped by the user via the touch screen or a connected foot switch.

  1. From the main screen, click "Mode" to enter the mode selection interface.
  2. Select "Continuity" to enter the continuous mode working interface.
  3. Set your desired flow rate by directly entering the value or using the plus (+) and minus (-) signs for fine-tuning. You can also set the variable speed of the plus and minus signs yourself.
  4. Click "Start" to begin continuous pumping. Click "Stop" to halt the operation.

7.2 Ration (Quantitative) Mode

Ideal for filling applications where a specific dosage is required. The pump runs for a set volume and then stops automatically. It supports both semi-automatic and full-automatic cycling.

  1. From the main screen, click "Mode" to enter the mode selection interface.
  2. Select "Ration" to enter the dispensing mode working interface.
  3. Click "Edit" to configure the parameters:
    • Volume (ml): Set the desired liquid volume per run.
    • Flow Rate (ml/min): Set the pumping speed.
    • Pause Time (s): Set the pause duration between cycles (for full-automatic mode).
    • Count: Set the number of cycles (for full-automatic mode).
    • Cycle Mode: Choose "Auto" for full-automatic or "Semi-Auto" for manual start after each dose.
  4. Click "Start" to begin the ration operation. In full-automatic mode, the pump will run for the set number of cycles, pausing for the specified time between each dose. In semi-automatic mode, you must click "Start" for each dose.
  5. Note: Before starting, ensure the pump tube is filled with liquid and free of air to maintain accuracy.

7.3 Remote Mode

This mode allows external control of the pump, suitable for integration into larger systems. It supports WiFi/Bluetooth, RS485, CAN, and various analog control methods.

  1. From the main screen, click "Mode" to enter the mode selection interface.
  2. Select "Remote" to enter the remote control interface.
  3. Choose your desired remote control method (e.g., WiFi or Bluetooth, RS485, CAN, or an analog quantity like 0-5V, 0-10V, 4-20mA).
  4. Connect your external control device according to the selected method's specifications and the pump's interface definitions.
  5. The pump's operation will then be governed by the external control signals.

Analog Control Interface Definition:

  • RS485 A: RS485 data+
  • RS485 B: RS485 data-
  • 0-5V/10V+: Voltage analog quantity positive
  • 0-5V/10V-: Voltage analog quantity negative
  • OUT PUT: 24V 5 mA digital output
  • DIR: Digital directional control, supporting dry contact (NPN) signal
  • Foot Switch: Foot switch, digital input, supporting dry contact (NPN) signal, active (PNP) signal
  • GND: Digital ground

8. Fill & Recycle Function

The Fill & Recycle function assists in preparing the pump for operation and recovering liquid. This function is typically accessed from the "Fill & Rec" tab on the main screen.

  1. From the main screen, click the "Fill & Rec" tab.
  2. Fill: Select "Fill" to quickly fill the pump tube with liquid, expelling any air.
  3. Recovery: Select "Recycle" to suck liquid from the outlet back into the original bottle, useful for clearing the tube or recovering unused liquid.

9. System Settings

The "System" tab on the main screen provides access to various pump settings and information. There are 17 functions that can be set, including:

  • Information (device details)
  • Pedal (foot switch settings)
  • Direction (pump direction)
  • Liquid Level Detection (if optional sensor is used)
  • Pumps and Tubes (tube type and calibration)
  • Suck Back (settings for liquid recovery)
  • Power (power-related settings)
  • Touch Correction (touch screen calibration)
  • Motor Drive (motor parameters)
  • Self-check (diagnostic functions)
  • Unit of Display (measurement units)
  • Network (WiFi/Bluetooth settings)
  • Password (security settings)
  • Motor Drive Current
  • Language
  • Touch Sound
  • Factory Settings Reset

Navigate through these options using the touch screen to customize the pump's behavior and view system information.

10. Optional Accessories

Photoelectric Sensor Switch

An optional photoelectric sensor can be used to detect the presence or absence of solution bottles, enabling automatic filling operations. The sensor has a recognition distance of 0.5 meters.

Photoelectric sensor switch with dimensions and wiring diagram.

Figure 6: Photoelectric Sensor Switch

11. User Tips

  • For accurate dosing in Ration Mode, ensure the pump tube is completely filled with liquid and free of air bubbles before starting operation.
  • When using the foot switch, ensure the red wire is connected to the foot switch terminal and the yellow wire to GND for proper functionality.

12. Troubleshooting

If you encounter any issues with your Kamoer UIP E V2 Peristaltic Pump, please refer to the following common troubleshooting steps:

  • Pump not turning on: Check power cable connection, power switch position, and power outlet.
  • Inaccurate flow rate: Ensure the pump tube is correctly installed and free of air. Perform flow rate calibration in System Settings if necessary.
  • Pump not responding to foot switch: Verify the foot switch wiring (red to foot switch, yellow to GND) and ensure it's securely connected. Check foot switch settings in System Settings.
  • Touch screen unresponsive: Try restarting the device. If the issue persists, perform a Touch Correction from the System Settings.
  • Unusual noise during operation: Check for any obstructions in the pump head or kinks in the tubing. Ensure the pump head cover is securely closed.

For persistent issues, please contact Kamoer customer support.

13. Maintenance

Regular maintenance ensures the longevity and optimal performance of your peristaltic pump.

  • Tube Replacement: Peristaltic tubes are wear-and-tear parts. Replace them regularly based on usage and the type of liquid being pumped to maintain accuracy and prevent leaks.
  • Cleaning: Clean the exterior of the pump with a soft, damp cloth. Do not use abrasive cleaners or immerse the unit in water.
  • Pump Head Inspection: Periodically inspect the pump head for any signs of wear or damage. Ensure the rollers move freely.
  • Storage: When not in use for extended periods, store the pump in a clean, dry environment within the specified storage temperature and humidity ranges (-20℃~80℃, 10%~90% RH non-condensing).

14. Applications

The Kamoer UIP E V2 Peristaltic Pump is suitable for a wide range of applications, including:

  • Food industry
  • Factory production
  • Laboratory and research institutes
  • Research institutions
  • Filling and dispensing tasks
  • Medical equipment

15. Warranty and Support

For warranty information, technical support, or service inquiries, please contact Kamoer Official Store customer service. Please have your product model number (UIP E) and purchase details ready when contacting support.

Kamoer Fluid Tech (Shanghai) Co., Ltd. is committed to providing precise and reliable fluid transfer components. We are always at your service!

Related Documents - UIP E

Preview Kamoer UIP WIFI E Smart Peristaltic Pump: Quick Start Guide and Technical Specifications
Quick start guide and technical specifications for the Kamoer UIP WIFI E Smart Peristaltic Pump, detailing its features, interface, app connectivity, and warranty information.
Preview Kamoer UIP WIFI E Smart Peristaltic Pump User Manual
Comprehensive user manual for the Kamoer UIP WIFI E Smart Peristaltic Pump, detailing its features, operation modes (Continuity, Ration, Remote), system settings, maintenance, and technical specifications.
Preview KCP-KK300-D Adjustable Peristaltic Pump User Manual
User manual for the Kamoer KCP-KK300-D adjustable peristaltic pump, detailing its features, specifications, operating conditions, interface description, and safety precautions.
Preview Kamoer AIP WIFI Smart Peristaltic Pump Quick Start Guide | Operation and Features
Get started with your Kamoer AIP WIFI smart peristaltic pump. This guide covers setup, operation, connectivity options (Wi-Fi, analog, RS485, CAN), technical specifications, and warranty information.
Preview AIP WIFI Smart Peristaltic Pump User Manual
Comprehensive user manual for the Kamoer AIP WIFI smart peristaltic pump, detailing its features, operation modes (continuous and ration), system settings, app control, maintenance, and technical specifications.
Preview Kamoer FX-STP WIFI Peristaltic Pump User Manual
User manual for the Kamoer FX-STP WIFI Peristaltic Pump, detailing installation, setup, operation, advanced settings, and troubleshooting. Learn how to connect and control the pump via the Kamoer Remote app.