KKMOON PH-W218

Contents

Tuya WiFi 8 In 1 Water Quality Tester User Manual

Model: PH-W218

1. Introduction

This manual provides comprehensive instructions for the setup, operation, maintenance, and troubleshooting of your Tuya WiFi 8 In 1 Water Quality Tester (Model: PH-W218). This device is designed for professional measurement of S.G (Specific Gravity), PH (Acidity/Alkalinity), EC (Electrical Conductivity), Salinity, ORP (Oxidation-Reduction Potential), TDS (Total Dissolved Solids), CF (Conductivity Factor), and Temperature values in water. It is ideal for various applications including fish tanks, aquariums, hydroponics, laboratories, spas, and swimming pools, providing real-time monitoring and data management.

For a detailed PDF version of this manual, please refer to the User Manual (PDF).

2. Product Features

  • 8-in-1 Measurement: Simultaneously measures S.G, PH, EC, Salinity, ORP, TDS, CF, and Temperature.
  • WiFi Connectivity: Real-time data transmission and monitoring via the Tuya Smart APP on Apple and Android devices.
  • Data Management: Automatic data storage, curve graph analysis, and export to Excel files for easy saving, managing, and sharing.
  • Multi-parameter Display: LCD backlight display for clear readings, suitable for night monitoring.
  • Wall-Mounted Design: Easy to install and operate, designed to prevent water ingress.
  • Automatic Alarm: Set high/low alarm values for various parameters via the mobile application.

3. Package Contents

The package includes:

  • 1 x Water Quality Tester (Main Unit)
  • 1 x PH Electrode
  • 1 x ORP Electrode
  • 1 x Temperature Electrode
  • 1 x PPM/EC/SALT/S.G/CF Probe
  • 1 x Power Adapter (US-plug or EU-plug, depending on selection)

4. Specifications

Detailed specifications table for the water quality tester

Figure 1: Detailed Specifications

Technical Specifications
ParameterRangeResolutionAccuracy
PH0.00-14.00PH0.01PH±0.03PH
ORP-2000mv-+2000mv1mv±5mv
EC0-19000us/cm, 0-199.0ms/cm1us/cm (<1999us/cm), 10us/cm (>1999us/cm), 0.1ms/cm±2% F.S
CF0-1999.0CF1CF±2% F.S
TDS0-19990ppm, 0-199.0ppt1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt±2% F.S
SALT0-199.9ppt, 0%-20%1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt, 0.01%±2% F.S
S.G0.990-1.4000.001±0.001
Temperature0-50°C (32°F-122°F)0.1°C (0.2°F)±1.0°C
  • Operating Temperature: 0°C-50°C
  • Temperature Compensation: 0°C-50°C
  • Power Supply: DC6V
  • Dimensions: 152mm × 80mm × 26mm
  • Weight: 245g
  • Probe Cable Length: Approximately 1.1m (43.31in)

5. Setup

Detailed view of the water quality tester with labeled components and dimensions

Figure 2: Product Details and Components

Wall-mounted design and installation note

Figure 3: Wall-Mounted Installation

  1. Unpack: Carefully remove the water quality tester and all accessories from the packaging.
  2. Connect Probes: Connect the PH electrode, ORP electrode, Temperature electrode, and PPM/EC/SALT/S.G/CF probe to their respective ports on the main unit. Ensure connections are secure.
  3. Mounting (Optional): The device features a wall-mounted design. Use appropriate fasteners to secure the main unit to a suitable surface near your water system. Note: This tester is not waterproof. Do not let the tester get water or damp during use.
  4. Power Connection: Connect the power adapter to the main unit and plug it into a suitable power outlet (110V or 220V, depending on your adapter type). The device will turn on automatically.
  5. Electrode Preparation: Remove the protective caps from the electrodes. Clean the electrodes with distilled water and dry them well before initial use or calibration.

6. Operation

Instructions for using the MODE button to select measurement parameters and turn WiFi on/off

Figure 4: MODE Button Functionality

Instructions for using the TEMP/CAL button to switch units and calibrate pH

Figure 5: TEMP/CAL Button Functionality

  1. Initial Measurement: Immerse the PH electrode, Temperature electrode, and TDS electrode into the water sample you wish to test. Stir lightly and allow a few seconds for the readings to stabilize on the LCD display.
  2. Switching Units (°C/°F): Short press the "TEMP CAL" key to switch between Celsius (°C) and Fahrenheit (°F) temperature units.
  3. Selecting Measurement Parameters (PPM/EC/SALT/S.G/CF): Short press the "MODE" key to cycle through PPM, EC, SALT, S.G, and CF measurements.
  4. WiFi On/Off: Long press the "MODE" key to turn the WiFi function on or off.

7. Calibration

Regular calibration is crucial for maintaining the accuracy of your water quality tester. The electrodes must be re-calibrated if:

  • The device has not been calibrated for a long time.
  • It has been in regular and continuous use.
  • High accuracy is required for your measurements.

7.1. PH Calibration (3-Point)

  1. Prepare three standard buffer solutions: pH 6.86, pH 4.00, and pH 9.18 (at 25°C). Pour them into separate clean beakers.
  2. For accurate calibration, use two beakers for each buffer solution: one for cleaning the electrode and the other for calibration. This minimizes pollution.
  3. Plug in and turn on the device.
  4. Immerse the PH electrode into the pH 6.86 buffer solution. Stir gently until the value stabilizes.
  5. Long press the "TEMP CAL" key for 5 seconds. The screen will display "06.86" and then enter the pH 6.86 automatic correction mode. Once the display value matches the buffer solution, calibration for pH 6.86 is complete.
  6. Clean the electrode with distilled water.
  7. Immerse the PH electrode into the pH 4.00 buffer solution. Stir gently until the value stabilizes.
  8. Long press the "TEMP CAL" key for 5 seconds. The screen will display "04.00" and then enter the pH 4.00 automatic correction mode. Once the display value matches the buffer solution, calibration for pH 4.00 is complete.
  9. Clean the electrode with distilled water.
  10. Immerse the PH electrode into the pH 9.18 buffer solution. Stir gently until the value stabilizes.
  11. Long press the "TEMP CAL" key for 5 seconds. The screen will display "09.18" and then enter the pH 9.18 automatic correction mode. Once the display value matches the buffer solution, calibration for pH 9.18 is complete.

Note: The factory default calibration uses 6.86, 4.00, and 9.19. Be careful not to use the wrong calibration solution during the process.

7.2. EC Calibration

  1. Turn on the meter. Short press the "MODE" key to select EC-ms measurement.
  2. Immerse the EC electrode into distilled water for about five minutes to activate it.
  3. Immerse the EC electrode into a 12.88ms/cm EC calibration solution (at 25°C). Stir gently.
  4. When the reading is stable, long press the "CAL" key for about 5 seconds. The screen will display "12.80" or similar. Continue pressing "CAL" until the display shows "12.70-12.90".
  5. Clean the electrode with distilled water and dry it with filter paper.
  6. (Optional) Immerse the electrode into a 1.413ms/cm EC calibration solution (at 25°C) and stir gently. The display value should be the same as or close to the calibration solution value.

7.3. ORP Calibration

  1. Plug in the power to turn on the device.
  2. Immerse the ORP electrode into distilled water for five minutes.
  3. Immerse the ORP electrode into a 265mv ORP calibration solution (at 25°C) and stir gently.
  4. When the reading is stable, long press the "CAL" key for about 5 seconds. The screen will display "265" or a similar value. Continue pressing "CAL" until the display shows "265" (or the plus/minus value).
  5. Clean the electrode with distilled water and dry it with filter paper.
  6. (Optional) Immerse the electrode in a 100mv ORP calibration solution (at 25°C) and stir gently until the display value is the same as or close to the calibration solution.

8. APP Connection & Monitoring

WiFi Connection and Data Transfer setup with the device and smartphone

Figure 6: WiFi Connection Overview

Mobile application monitoring interface showing various parameters

Figure 7: Mobile App Monitoring Interface

  1. Download Tuya Smart APP: Search for "Tuya Smart" in your mobile APP store (Apple App Store or Google Play Store) and download it.
  2. Register/Log In: Open the Tuya Smart APP. If you are a new user, register for an account. Otherwise, log in with your existing credentials.
  3. Add Device:
    1. Ensure your mobile phone and the water quality tester are connected to the same WiFi network.
    2. In the Tuya Smart APP, click "Add Device" (usually a '+' icon or similar). The app will start searching for nearby devices.
    3. On the water quality tester, long press the "MODE" key to ensure WiFi is enabled (refer to Section 6, Step 4).
    4. Once the device is discovered, click to add it.
    5. Enter your WiFi password when prompted and click "Next".
    6. The app will connect to the device. Once successful, a confirmation page will appear. Click "Finish".
  4. View Parameters: Enter the program for your water quality tester in the Tuya Smart APP to view various parameters in real-time.
  5. Data Export: The app allows you to export data to Excel. Click the "Export" option within the app, enter your email address, and the data will be sent to your mailbox. You can then download, view, and share the data. Data is automatically stored and saved, with curve graphs available for analysis.
  6. Set Alarms:
    1. In the Tuya Smart APP, navigate to the "Smart" section to create a Smart Scene.
    2. Select "When device status changes" as the condition.
    3. Choose your water quality tester and select the parameter you want to monitor (e.g., pH value).
    4. Set the high or low alarm threshold (e.g., pH value lower than 5.38).
    5. For the action, select "Send notification" and choose your preferred notification method (e.g., Message Notification).
    6. Save the alarm setting. The APP will now send a notification if the parameter value crosses the set threshold.

9. Maintenance

  • Electrode Cleaning: After each use, or before calibration, clean the electrodes thoroughly with distilled water. Gently wipe them dry with a soft cloth or filter paper.
  • Protective Caps: Always replace the protective caps on the electrodes when the device is not in use to prevent damage and keep them moist.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
  • Re-calibration: As noted in the Calibration section, re-calibrate the electrodes regularly, especially if the device has been in continuous use, has not been calibrated for a long time, or if high accuracy is critical.

10. Troubleshooting

If you encounter issues with your water quality tester, consider the following:

  • Inaccurate Readings: Ensure electrodes are clean and properly calibrated. Refer to the calibration section for detailed steps.
  • No Display/Power: Check the power adapter connection and ensure it's plugged into a working outlet.
  • WiFi Connection Issues: Verify that the device's WiFi is enabled (long press MODE button) and that your phone is on the same 2.4GHz WiFi network. Ensure the WiFi signal is strong.
  • App Not Detecting Device: Try restarting the app and the water quality tester. Ensure the device is in pairing mode if required by the app.
  • Data Export Problems: Check your internet connection and ensure the email address entered in the app is correct.

11. User Tips

  • Ammonia Measurement: This device does not measure ammonia. It measures S.G, PH, EC, ORP, TDS, CF, SALT, and TEMP.
  • Probe Cable Length: The probe cables are approximately 1.1 meters (43.31 inches) long, providing flexibility for placement.
  • Power Adapter: The device operates on DC6V power supplied by the included adapter, which is available in US or EU plug types to suit regional power standards.
  • Waterproof Caution: Remember that the main unit is not waterproof. Install it in a location where it will not get wet or damp.

12. Warranty & Support

All defects in materials and manufacture of this device are guaranteed for one year from the date of purchase. Within this period, parts will be repaired or exchanged free of charge if the damage is not caused by the user's negligence or improper operation.

For technical support or warranty claims, please contact your retailer or the manufacturer directly.

Documents and downloads