1. Introduction
Thank you for purchasing this Professional Analogue Multimeter!
1.1 Intended Use
- Measuring and displaying electric quantities in the range of overvoltage category CAT II up to max. 1,000V; CAT III up to max. 500V to ground potential, in accordance with EN 61010-1.
- Measuring alternating and direct voltages of up to 2,500 V Max.
- Measuring alternating and direct currents of up to 10 A Max.
- Measuring resistances from 10 to 200 MΩ easily Without Ω Adjust.
- Checking capacitances Max. 2,000uF.
- Acoustic continuity check.
- AC Live identifying.
The product may only be operated with batteries. The measuring instrument must not be operated when it is open, i.e. with an open battery compartment or when the battery compartment lid is missing. Measuring in damp rooms or under unfavorable ambient conditions is not admissible.
Adverse ambient conditions include:
- Excessive dampness or humidity.
- Dust or combustible gases, vapors or solvents.
- Storms or stormy conditions, strong electrostatic fields, etc.
Any use other than that described above will damage the product and involves other risks, such as short-circuit, fire, or electric shock. Do not change or modify any part of the product!
1.2 Safety Instructions
The safety instructions must be observed! Failure to comply with these operating instructions will void the warranty/guarantee. The user must comply with the safety instructions and warnings contained in these instructions. The following symbols must be observed:
- ! An exclamation mark in a triangle indicates important information in these operating instructions that has to be observed.
- ⚡ The triangle containing a lightning symbol warns of danger of an electric shock or of the impairment of the electrical safety of the device.
- ※ It indicates special information and advice on operating.
- CE This device is CE certified and meets the necessary European regulations.
- ☐ Protection class II (double or reinforced insulation)
- CAT II Overvoltage category II for measurements on electric and electronic devices connected to mains power supply. This category also covers all smaller categories (e.g. CAT I for measuring signal and control voltages).
- CAT III Overvoltage category III for measuring building wiring installation (e.g. outlets or sub-distributions). This category also covers all smaller categories (e.g. CAT II for measuring electronic devices).
- ⏚ Earth potential
For safety and licensing reasons (CE), unauthorized conversion and/or modification of the device is not permitted. Consult an expert when in doubt as to the operation, the safety or the connection of the device. Measuring instruments and accessories are no toys and do not belong in the hands of children! The accident prevention regulations of the relevant trade associations for electrical systems and operating materials are to be observed in commercial institutions. In schools, training centers, hobby and self-help workshops, the operation of measuring instruments must be supervised by trained personnel.
Before measuring voltages, always make sure that the measuring instrument is not set to current range. The voltage between any socket of the measuring instrument and the ground may not be higher than 1,000V in excess voltage category II. Take particular care when dealing with voltages exceeding 25 V AC or 35 V DC! If electrical conductors are touched, even these voltages involve the risk of a fatal electric shock. Prior to each measurement, check your instrument including its measuring lines for damage. Never carry out measurements when the protective insulation is damaged (ripped, torn off etc.). In order to avoid an electric shock, ensure that you do not touch the connections to be measured, even indirectly, during measurements. When measuring, only touch within the tangible grip markings on the test prods. Do not use the multimeter shortly before, during or just after a thunderstorm (lightning high-energy over voltages!). Ensure that your hands, shoes, clothing, the floor, the measuring instrument, the measuring lines, switches and switching parts, etc. are dry. Do not operate the measuring instrument in surroundings or unfavorable conditions where combustible gases, vapors or dust are or may be present. Avoid operation in the immediate vicinity of: strong magnetic or electromagnetic fields, transmitting aerials or HF generators. This might lead to distorted measuring results. For safety reasons, only use measuring lines or accessories in correspondence with the specifications of the multimeter. Only measuring equipment with double or reinforced insulation may be used (e.g. fully insulated BNC adapters, etc.) If you have reason to believe that the device can no longer be operated safely, disconnect it immediately and make sure it is not unintentionally operated. It can be assumed that safe operation is no longer possible if: the device shows visible damage, the device does not function any longer or after it has been stored under unfavorable conditions over a period of time or after it has been exposed to heavy stress during transport. Do not switch on the measuring instrument immediately after it has been taken from a cold to a warm environment. The condensation generated could cause serious damage to the device. Allow device to reach room temperature before switching it on. Do not leave packaging material lying around carelessly. It may become a dangerous toy for children. You should also observe the safety instructions in the individual chapters.
2. Product Description
The FUZRR ES3035E+ ES3035+ is a digital insulation resistance tester, also referred to as a megohmmeter or multimeter. It features double insulation protection and is designed for high voltage measurements. The device is housed in a durable, compact case, making it suitable for various electrical testing applications.

Figure 1: FUZRR ES3035E+ ES3035+ Digital Insulation Resistance Tester and included accessories.

Figure 2: The tester and its accessories are provided in a robust carrying case for protection and portability.

Figure 3: Structure of the Meter. Key components include the LCD display screen, LINE interface, V interface, GUARD interface, EARTH interface, changer interface, USB interface, Ω gear, DC voltage button, AC voltage button, Test button, storage data (MEM), consult data (MR), Absorption Ratio and Polarization Index (MODE), backlight function, direction buttons, power on/off button, voltage select buttons, and data deletion buttons.
2.1 Key Features
- Digital high-voltage insulation resistance meter.
- Functions: Insulation resistance test, voltage test.
- Rated voltage options: ES3035+ (100V, 250V, 500V, 1000V, 2500V, 5000V), ES3035E+ (250V, 500V, 1000V, 2500V, 5000V, 10KV).
- Controllable gray screen backlight for dim places.
- USB interface for data upload to computer.
- Automatic shutdown after 15 minutes of inactivity.

Figure 4: Functions and Rated Voltage Options for ES3035+ and ES3035E+ models.
2.2 Scope of Delivery
- Multimeter with removable compact rubber holster enclosed.
- Safety measuring gold-coated Test Leads (red and black).
- Operating instructions.
- Power adapter.
- USB communication line.
- CD with software.
3. Specifications
Guaranteed for one-year warranty.
| Feature | Specification |
|---|---|
| Over-voltage category | CAT II 1,000V to ground potential |
| Standard Environment | 23°C±3°C, less than 75% RH. at 50/60Hz |
| Temperature Ranges (Operating) | 0°C to 40°C, 32 °F to 104 °F |
| Temperature Ranges (Storage) | -10 °C to 50°C, 14 °F to 122°F |
| Humidity Scope (Operating) | less than 90% RH. |
| Humidity Scope (Storage) | less than 80% RH. |
| Operating altitude | Max. 2,000 m |
| Size (L/W/H) | 277.2mm × 227.5mm × 153mm (Instrument size) |
| Weight | 2720g (including battery) |
| Power Supply | 12V rechargeable battery (Battery Included: Yes) |
| Voltage range | 6-24V DC |
| DIY Supplies | Electrical |
| Backlight | Controllable gray screen backlight |
| Display mode | 4-bit large LCD display, gray screen backlight |
| LCD display size | 108mm × 65mm |
| USB interface | With USB interface, software monitoring, storage data can be uploaded to the computer, save the print |
| Communication Line | USB communication line 1PC |
| Test Line | High voltage test line red 3 meter, black 1.5 meter, green line 1.5 meter |
| Data Storage | 500 groups, “FULL” symbol indicates that storage is full |
| Data Review | Data review function: “MR” symbol display |
| Overflow Display | Exceed measurement range overflow function: “OL” symbol display |
| Alarm Function | Alarm when the measured value exceeds the alarm setting value |
| Power Consumption (Standby) | 30mA Max (backlight off) |
| Power Consumption (Turn on backlight) | 42mA Max |
| Power Consumption (Measure) | 200mA Max (backlight off) |
| Battery voltage | low, low battery symbol “ ” will display |
| Automatic shut-down | automatic shutdown after 15 minutes boot |
3.1 Range Specifications & Tolerance List
| Functions | Range | Accuracy | Remarks |
|---|---|---|---|
| DC V | 0-0.1-2.5-10-50-250V -1000V-2500V | ±3% of FSD. ±4% of FSD. at 0.1V/1000/2500V ranges | Input Impedance: 20KΩ/V But 9KΩ/V > 1000V. Overload Protection: Max. 2,500V AC/DC. BUT Range 0.1V/0.5/10/50V: Max. 250V |
| AC V | 0-10-50-250V -1000V-2500V | ± 4% of FSD. Band width: 20 ~ 20K Hz NOTE #1 | Input Impedance: 9 KΩ/V. Overload Protection: Max. 2500Vrms AC/DC. BUT Ranges 10/50V: Max. 250V |
| DC A | 0-0.05-0.25-2.5-25-250mA, 10A | ±3% of FSD. ±4% of FSD. For 10A range (NOTE #2) | Drop Voltage: 250 mV. Overload protected by a Fuse 0.5A at mA ranges, But 10A No Fuse. |
| AC A | 0-10A | ± 4% of FSD. Band width: 20 ~ 20K Hz (NOTE #1 & #2) | Drop Voltage: 250 mV. No Fuse protected. |
| Ω | X 1: 2 ~ 20KΩ Midscale at 200Ω; X 10: 20~200KΩ Midscale at 2,000Ω; X 100: 200~2,000KΩ Midscale at 20KΩ; X1K: 2000~20MΩ Midscale at 200KΩ; X10K: 20K~ 200MΩ Midscale at 2,000KΩ | ±3% of ARC. (Scale Length) | Overload protected by the Oxide Varactor & Fuse. The Input Voltage limit: Min.50V, Max. 250Vrms AC/DC for 5 seconds. |
| AC Live | AC Live | One LED Red Light flashing >AC 50V and Buzzer sounding >AC100V | Use the R x 100 range via test leads |
| Capacitance | 1~2,000uF (CμF) | Approximate Value | |
| Continuity Check | < 50 Ohm | Beeper sounding & LED | Overload protected by Fuse & Oxide Varactor <250V AC/DC(5s). |
NOTE#1: The extra indication error will occur as per the waveforms of measured ACV/ACA other than the sine wave. Its readings of rms value may be lower or higher than the actual root-mean-square value.
NOTE#2: For 10A range, the big current loaded max. 1min. and pause 5 min. at least before next testing.
4. Setup
4.1 Battery Installation
The operation of the measuring instrument requires batteries, which are included in delivery. Insert new batteries when starting initial operation or when null balancing in the resistance measuring range can no longer be performed.
Proceed as follows to insert or change the batteries:
- Remove all measuring lines and turn off the multimeter.
- Loosen the housing screw on the back and carefully slide upside the battery lid cover.
- Take off the old battery, and insert new 1.5V AA batteries as displayed. Watch out for the polarity in the battery compartment.
- Now, carefully close the housing again.

Figure 5: Battery compartment showing correct polarity for 1.5V AA batteries.
! Never operate the measuring device when it is open. Do not leave flat batteries in the device. Even batteries protected against leaking can corrode and thus release chemicals which may be detrimental to your health or destroy the device. If the device is not used over a period of time, remove the batteries in order to prevent leaking. Leaked or damaged batteries can burn your skin if touched. Therefore, use suitable protective gloves. Make sure that batteries are not short-circuited. Do not throw batteries into fire. ※ It's better to use alkaline batteries. They are powerful and long-lasting.
4.2 Replacing the Fuse
! Always adhere to the safety instructions when replacing fuse! Make sure that only fuse of the type stated and of the rated current specified is used as a replacement. Using unsuitable or repaired fuses or bypassing the fuse holder is inadmissible and can cause fire. Both the measurement input “V/mA” But 10A is protected against over-ranging. If measurements can no longer be made via the "V/mA" socket, the internal fuse must be replaced.
Proceed as follows to change the fuse:
- Remove all measuring lines and turn off this multimeter.
- Pull off the soft Holster from the ABS housing.
- Loosen the housing screw on the back and carefully slide off the battery compartment lid.
- Push the plastic peg between the two battery compartments forward until the case is unlocked. Carefully remove the rear housing.
- Replace the inside microfuse with the new fuse of same type: 0.5A/250V quick-acting Φ5 x 20 mm.
- Close the device in reverse order and carefully screw the battery compartment lid back on, and then recover the holster well.

Figure 6: Steps for fuse replacement, showing the location of the 0.5A fuse.
5. Operating Procedures
Prior to working with the measuring device, you have to insert the batteries into this tester. Insert the batteries as described in chapter "Cleaning and Maintenance”.
5.1 Rotary Switch
The individual measuring functions can be set via the rotary range switch. If the rotary switch is set to “OFF”, the power of this measuring device is turned off and the testing functions of resistance & capacitance are not working. Always turn off the device when it is not in use.
5.2 General Measuring Procedures
! Do not exceed the permitted max. input values. Never touch circuits or parts of circuits with a possible voltage of more than 25V/AC rms. or 35 V/DC! Danger to life! Before measuring, check the connected measuring lines for damage such as, for example, cuts, cracks or pinches. Replace defective measuring lines immediately by new ones, defective measuring lines may not be used any longer! Danger to life! Before changing the measuring range, the test prods have to be removed from the test object.
※ Always start measuring with the largest measuring range. Then gradually switch down to achieve an exact measuring result. Results are most accurate in medium scale range (scale range approx. 70 - 110°). The measuring ranges on the rotary switch correspond to the scale end value. For reading, always select the corresponding value (e.g. scale 50 for the measuring ranges 50 and 500).
5.3 Null Balance
Before measuring, always perform null balancing using the calibration screw (scale value 0V). There must be no measuring signal on the measuring lines.
5.4 Measuring AC and DC Voltage
! Do not exceed the permitted max. input values; this also applies when measuring superimposed DC/AC voltages (e.g. ripple voltages).
- Select the measuring range DCV or ACV using the rotary switch.
- Connect the black measuring line with the COM jack and the red measuring line with +V/mA/Ω input jack.
- Now perform null balancing to set the Needle at Zero position at left side.
For DCV testing, it's necessary to observe the correct polarity (red = + / black = -) and connect the two test prods with the test object (battery, circuit, etc.). If the polarity is not correct, no value is displayed. The built-in protective diode prevents measuring. Then Stop testing and measure again observing the right polarity.
For ACV testing, be sure to place the rotary switch at the proper ACV range. Read off the measured value on the "V" scale.
For DC2,500V testing, it's necessary to observe the correct polarity (red = + / black -) and connect the two test prods with the test object (battery, circuit, etc.). If the polarity is not correct, no value is displayed. The built-in protective diode prevents measuring. Then Stop testing and measure again observing the right polarity.
- Select the measuring range DC2500V or AC2500V using the rotary switch;
- Connect the black measuring line with the -COM jack and the red measuring line with the special +2,500V input jack at the right up corner.
For AC2,500V testing, be sure to place the rotary switch at the AC2500V range. Connect the test lines separately with the -COM jack and the special +2,500V input jack at the right up corner. And then contact the test leads with the tested object. Read off the measured value on the "V" scale.
After measuring, put the rotary switch in the “OFF” position to turn off this device.
※ At the ranges DC0.1V/0.5V, before measuring the needle may be off the Zero position and deflected somehow due to its higher impedance. It's the normal case and doesn't affect any measurement and accuracy while testing.

Figure 7: Test lead connections for voltage measurement. Note the specific input jacks for different voltage ranges.
5.5 Measuring AC and DC Current
! Do not exceed the permitted max. input values in the individual measuring ranges. Current measurement is only admissible in circuits of up to 250V.
Proceed as follows to measure direct currents of up to 10A:
- Select the measuring range DC10A or AC10A, using the rotary switch.
- Connect the black measuring line with the COM jack and the red measuring line with special +10A input jack.
- Now perform null balancing to set the Needle at the left Zero scale position.
For DC testing, it's necessary to observe the correct polarity (red = + / black = -) and connect the two test probes in series with the test object (battery, circuit, etc.). If the polarity is not correct, no value is displayed. The built-in protective diode prevents the measuring. Stop testing and measure again observing the right polarity.
For AC testing, be sure to place the range switch at 10A. Read off the measured value on the "A" scale. After measuring, put the rotary switch in the “OFF” position to turn off this device.
Proceed as follows to measure direct currents of up to 250mA:
- Measuring in DC mA ranges, select the corresponding mA range by using the rotary switch.
- Connect the black measuring line with the COM jack and the red measuring line with the +mA input jack.
- Now perform null balancing to set the Needle at the left Zero scale position.
It's necessary to observe the correct polarity (red = + / black = -) and connect the two test probes in series with the test object (battery, circuit, etc.). If the polarity is not correct, no value is displayed. The built-in protective diode prevents the measuring. Stop testing and measure again observing the right polarity.

Figure 8: Test lead connections for current measurement. Ensure correct polarity and input jacks.
5.6 Resistance Measurement
! Make sure that all circuit parts, switches, components and any other test objects are currentless and discharged.
Proceed as follows to measure the resistance:
- Select the measuring range on the rotary switch Ω.
- Connect the black measuring line with the COM socket and the red measuring line with the +V/Ω input jack.
This device is not equipped with the common Zero OHM calibration adjustor because its special circuit design can always keep the pointer at 0 Ohm even changing the Ω range. Read off the measured value on the "Ω" scale. Multiply the displayed value by the measuring range to obtain the measuring result (e.g. 100Ω (display) x10kΩ (measuring range) = 100 x 10,000 = 1MΩ (measuring result)). After measuring, put the rotary switch in the “OFF” position to turn off the multimeter.
NOTE: If the Needle is not kept at the right 0Ω line before testing, the inner battery must be replaced.

Figure 9: Test lead connections for resistance measurement. Connect leads to COM and +V/Ω jacks.
5.7 Acoustic Continuity Check
! Make sure that all circuit parts, switches, components and any other test objects are currentless and discharged. The continuity check allows a quick check e.g. of a line. If forward resistance is <200 Ohm, a beep sounds with the displayed measuring result.
To measure, proceed as follows:
- Select the measuring •))) range on the rotary switch.
- Connect the black measuring line with the COM jack and the red measuring line with the +V/Ω input jack.
- Now connect the two test probes to the test object (component, switch etc.).
A beeper sounds when the resistance value is <50 Ohm; and the red LED lighting is flashing at same time. After measuring, put the rotary switch in the “OFF” position to turn off the multimeter. Note: If wrongly connected with the voltage, quickly remove the test leads from the circuit and can avoid the damage to this tester. (This tester can afford the voltage <250V DC/AC rms. for the period of 5 seconds max.)
5.8 Capacitance Measurement
WARNING: DO NOT APPLY VOLTAGE TO MEASURING TERMINAL WHILE MAKING ANY CAPACITANCE MEASUREMENTS. BEFORE TESTING ANY CAPACITORS, DISCHARGE THE CAPACITOR COMPLETELY.
Measuring the Capacitor as the following procedures:
- Set the selector switch to the R X100(Cuf) range.
- Connect the BLACK test lead to the "-COM" jack and the RED test lead to the Red "+" jack.
- Connect the test leads to the capacitor to be measured (Note the polarity of capacitor).
- Watch the needle deflection to the right topside, and read the Red CμF scale on the Dial.
After measurement, be sure to set the selector switch of this tester to “OFF” position.
5.9 AC Live Check
This function is used to check and identify the AC Live polarity of wires or outlets.
- Set the rotary selector switch to the range AC Live.
- Connect one test probe to the “+V/Ω” jack of this tester.
- Take this test lead's tip to contact the terminal of wire or outlet.
- Then see the case of Red LED lighting at the right up corner of dial and beeper. If this Red LED light is lit and the buzzer sounds, it means the contacted wire or end of outlet is Live line. Otherwise, it should be the Zero line or Ground wire.
Note: the Red LED light flashing once ACV>50V; and the Buzzer sound once ACV>100V approx.
6. Maintenance
6.1 General Maintenance
To ensure the accuracy of the multimeter over an extended period of time, it should be calibrated once a year.
! Regularly check the technical safety of the instrument and measuring lines, e.g. for damage to the housing or squeezed parts etc. Never operate the measuring device when it is open!
6.2 Cleaning
Always observe the following safety instructions before cleaning the device:
! Live components may be exposed if covers are opened or parts are removed (unless these parts are intended to be removed without tools). Before cleaning or repairing the multimeter, any lines have to be disconnected and the device has to be turned off.
Do not use agents containing carbon, benzine, alcohol or similar substances to clean the device. The surface of the multimeter will be corroded. In addition, the vapors are detrimental to health and explosive. Sharp-edged tools, screwdrivers or metal brushes should not be used for cleaning purposes. To clean the device, the display or the measuring lines, use a clean, lint-free, antistatic and dry cleaning cloth.
6.3 Disposal of Used Batteries/Rechargeable Batteries
The end user is legally obliged (battery regulation) to return all used batteries and rechargeable batteries. Disposal in the household waste is prohibited!
7. Troubleshooting
! By purchasing the multimeter, you have acquired a quality product which operates reliably. Nevertheless, problems or malfunctions may occur. For this reason, the following is a description how you can eliminate possible malfunctions yourself:
Please observe the safety instructions firstly!
| Error | Possible cause |
|---|---|
| The multimeter does not work. | Are the 1.5V battery lower power? Or check the state of the battery and the fuse 0.5A. |
| No measurements possible via V/mA socket. | Is the fuse defective? Check the fuse 0.5A (fuse replacement). |
| No measurements possible via 10A socket. | Is the rotary selector placed correctly or Are the test leads connected right? |
| No change in measured values. | Have you selected the right measuring sockets? Is the measuring mode correct (AC/DC)? |
| Faulty measuring results are displayed. | Has null balancing of the display? Or are the batteries properly located? |
8. User Tips
- The device is designed to be operated with batteries. For optimal performance and longevity, it is recommended to use alkaline batteries.
- When measuring DC0.1V/0.5V ranges, it's normal for the needle to be slightly off the Zero position due to higher impedance. This does not affect measurement accuracy.
- Always refer to the specifications section for details on the supply voltage (6-24V DC) and other electrical parameters.
9. Warranty and Support
This product is provided with a one-year quality guarantee by the manufacturer from the date of purchase. The manufacturer assumes no responsibility for any malfunctions or issues arising from user's failure to follow these instructions or general safety practices.
For more detailed information, including comprehensive diagrams and additional technical specifications, please refer to the official user manual PDF available at: FUZRR ES3035E+ ES3035+ User Manual (PDF).