1. Introduction
The FGHGF HW-210K is a handheld dual-channel VI curve tester designed for online detection and analysis of circuit boards. It functions as an ASA tester, providing 4 gear frequency alternating capabilities. This device is essential for diagnosing electronic components and circuits by displaying their voltage-current (VI) characteristics on an oscilloscope, making fault detection easier and more efficient.
2. Safety Information
- Always ensure the circuit board under test is NOT powered or recharged when performing online detection with the HW-210K.
- If the circuit board contains large capacitors, discharge them completely before connecting the VI tester to prevent damage to the tester or the circuit board.
- Use only the provided BNC cables and test cables.
- Do not attempt to open or modify the device. Refer all servicing to qualified personnel.
- Keep the device away from moisture, extreme temperatures, and direct sunlight.

Important safety warning regarding discharging capacitors before testing.
3. Package Contents
Please check the package for the following items:
- 1 x HW-210K VI Curve Tester Unit
- 1 Pair x BNC Cables
- 1 Pair x Test Cables (probes)
4. Product Overview
The HW-210K features a compact design with intuitive controls for selecting test frequencies and managing display modes. It offers dual signal input and an alternating display function for enhanced usability.

Front view of the HW-210K tester and its accessories.

The tester features dual signal input and alternating display.

Side view showing BNC output and power input.
5. Specifications
| Feature | Description |
|---|---|
| Model Number | HW-210K (Circuit Board Online Detection) |
| Brand Name | FGHGF |
| Certification | CE |
| Origin | Mainland China |
| Channels | 2 channels |
| Frequency | 4 gears (LOW, MED1, MED2, HIGH) |
| Compatible Equipment Style | Display Equipment (Oscilloscopes) |
| Package | Yes |
| High-concerned chemical | None |
| Dimensions (L x W x H) | 21 cm x 5 cm x 2 cm |
| Weight | 0.395 kg |
6. Setup
The HW-210K requires connection to an oscilloscope (analog or digital) to display VI curves.
6.1. Connecting to an Analog Oscilloscope
- Oscilloscope Setup: Adjust your analog oscilloscope to X-Y mode. The method for this varies by oscilloscope model (some use buttons, others use a knob). Consult your oscilloscope's manual if unsure.
- Cable Connection: Connect the BNC cables from the HW-210K to the X and Y input channels of your oscilloscope.
- Power On: Power on both the HW-210K and the oscilloscope.
- Initial Display Adjustment (Horizontal Line): Normally, a horizontal line will be displayed. Adjust the X and Y vertical channel parameters (e.g., around 1V/div) and X-Y attenuation so that the horizontal line is within the display frame. The X attenuation potentiometer on the upper side of the VI test board can assist in adjusting the length of this line.
- Initial Display Adjustment (Vertical Line): Short-circuit the test probes of the HW-210K. A vertical line should appear. If it's too long or too short, adjust the Y channel parameters on your oscilloscope to bring the vertical line within the display frame. Aim for about half a grid or one grid from the edges.
- Once both horizontal and vertical lines are properly adjusted, the system is ready for normal measurement.
6.2. Connecting to a Digital Oscilloscope
- Oscilloscope Setup: Enter X-Y mode via the menu or dedicated keys on your digital oscilloscope.
- Channel Adjustment: Adjust both X and Y channels to 1V/div. For single-track oscilloscopes, only the Y channel needs adjustment.
- Time Base: Set the time base to 1-5 milliseconds.
- Coupling and Attenuation: Select Direct Current (DC) coupling for the X-Y channels, and set attenuation to 1x.
- Initial Display Adjustment (Bright Spot): When the oscilloscope is not connected to the HW-210K, the screen should show a bright spot. Adjust horizontal and vertical controls to center this spot.
- Connect HW-210K: Connect the HW-210K to the oscilloscope. A horizontal line should appear. Adjust the '103 potentiometer' on the VI board to ensure the horizontal line is within the display frame, ideally half a grid or one grid from the edges.
- Initial Display Adjustment (Vertical Line): Short-circuit the test leads. A vertical line should now be displayed. Adjust the Y channel parameters on your oscilloscope to ensure this vertical line is also within the display frame.
- Debugging is complete, and the device is ready for use.

HW-210K connected to an oscilloscope for circuit board testing.
7. Operating Instructions
The HW-210K offers flexible testing options with its dual-channel input and adjustable frequency settings.
- Select Frequency: Use the frequency buttons (LOW, MED1, MED2, HIGH) on the front panel to select the desired test frequency. The 4-gear frequency and alternating speed are adjustable to suit your testing needs.
- Connect Test Probes: Connect the red and black test probes to the A and COM (common) terminals, or B and COM terminals, respectively, on the HW-210K.
- Apply Probes to Circuit: Carefully apply the test probes to the component or circuit points you wish to analyze on the unpowered circuit board.
- Observe VI Curve: Observe the VI curve displayed on your connected oscilloscope. The shape of the curve indicates the characteristics and health of the component or circuit.
- Alternating Display: Utilize the one-click single channel/alternating display function for easier comparison and analysis between channels or different test points.
8. Maintenance
- Keep the device clean and dry. Wipe with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- Store the HW-210K and its accessories in a cool, dry place when not in use.
- Inspect cables and probes regularly for any signs of wear or damage. Replace if necessary.
9. Troubleshooting
- No Display on Oscilloscope: Ensure both the HW-210K and oscilloscope are powered on and correctly connected via BNC cables. Verify the oscilloscope is in X-Y mode.
- Incorrect Horizontal Line: If the horizontal line is not displayed or is too long/short during setup, adjust the X-Y attenuation on the oscilloscope and the X attenuation potentiometer on the HW-210K.
- Incorrect Vertical Line: If the vertical line is not displayed or is too long/short when probes are short-circuited, adjust the Y channel parameters on the oscilloscope.
- Distorted VI Curves: Ensure the circuit board under test is completely unpowered. If testing components with large capacitors, confirm they have been fully discharged.
- No Response from Tester: Check the power connection to the HW-210K. Ensure all cables are securely connected.
10. User Tips
- Capacitor Discharge: Always remember to discharge large capacitors on a circuit board before performing online detection to prevent damage to the VI test board.
- Oscilloscope Compatibility: The HW-210K is compatible with both analog and digital oscilloscopes. Familiarize yourself with your specific oscilloscope's X-Y mode settings for optimal performance.
- Calibration Check: Periodically perform the initial display adjustment steps (horizontal and vertical line checks) to ensure your setup is calibrated correctly.
11. Warranty and Support
For warranty information and technical support, please refer to the seller's policies or contact the point of purchase. Keep your purchase receipt as proof of purchase.