VamsLuna SDZ III

Hwato SDZ III Electro Low-Frequency Acupuncture Stimulator User Manual

Model: SDZ III | Brand: VamsLuna

1. Introduction

The Hwato SDZ III Electro Low-Frequency Acupuncture Stimulator is designed to provide electro-acupuncture treatment, nerve and muscle massage, and relaxation. This device features 6 independent output channels, adjustable frequency, and various working patterns to cater to different therapeutic needs. Please read this manual thoroughly before use to ensure safe and effective operation.

2. Important Safety Information

WARNING: This device is intended for professional use or under the guidance of a qualified practitioner. Improper use can lead to injury. Consult a healthcare professional before using this device, especially if you have pre-existing medical conditions.

  • Do not use on individuals with pacemakers, implanted defibrillators, or other electronic medical devices.
  • Do not use on pregnant women, infants, or individuals with epilepsy, heart disease, or acute infectious diseases.
  • Avoid applying electrodes to areas with skin lesions, open wounds, or sensitive areas like the eyes, mouth, or carotid sinus.
  • Keep the device away from water and moisture.
  • Ensure the power adapter matches your local voltage requirements (available in 110V and 220V versions).
  • Do not disassemble or modify the device.
  • Keep out of reach of children.
  • Discontinue use immediately if you experience discomfort or adverse reactions.

3. Package Contents

Verify that all items are present in your package:

  • Hwato SDZ III Host Unit (1)
  • Output Wires (6)
  • Needle Electrode Metal Clips (6 pairs)
  • Skin Electrodes (6 pairs)
  • Power Adapter (1)
  • User Manual (1)
Hwato SDZ III Stimulator and accessories
Figure 3.1: Complete package contents of the Hwato SDZ III Stimulator.

4. Product Overview

Familiarize yourself with the components and controls of your Hwato SDZ III device.

Diagram of Hwato SDZ III components
Figure 4.1: Labeled diagram of the Hwato SDZ III Stimulator and its accessories.

4.1 Device Components

  • ON/OFF Button: Powers the device on or off.
  • Out-connect DC socket: Power input for the DC9V AC adapter.
  • Wave Showing Window (LCD Display): Displays mode, frequency, time, and output status.
  • Function/Choose Switch: Rotary knob to select treatment modes and functions. Press to confirm selection.
  • Output Strength Adjusting Knobs (1-6): Individual knobs to adjust the intensity for each of the six output channels. Rotate clockwise to increase intensity, counter-clockwise to decrease. Turn fully counter-clockwise to the "off" position to disable a channel.
  • Six-Output Sockets (1-6): Ports for connecting output wires to electrodes.

4.2 Accessories

  • Skin Electrodes: Adhesive pads for surface stimulation.
  • Metal Clamps (Filiform Needle-Electrodes): Used to connect to acupuncture needles for electro-acupuncture.
  • Output Wires: Connect the device's output sockets to the electrodes/clips.
  • DC9V AC Adapter: Power supply for the unit.

5. Setup Instructions

Follow these steps to prepare your Hwato SDZ III for use:

  1. Unpack: Carefully remove all components from the packaging and verify against the package contents list.
  2. Power Connection: Connect the DC9V AC adapter to the "Out-connect DC socket" on the back of the main unit. Plug the adapter into a suitable power outlet. Ensure your adapter matches your local voltage (110V or 220V).
    Back of device with power input
    Figure 5.1: Rear view showing DC9V power input.
    Various power adapter types
    Figure 5.2: Examples of different regional power adapters.
  3. Connect Output Wires: Insert the output wires into the desired "Six-output sockets" (1-6) on the front of the device.
    Output sockets on the device
    Figure 5.3: Close-up of the six output sockets.
  4. Attach Electrodes: Attach either the skin electrodes or the metal clips (for acupuncture needles) to the other end of the output wires. Ensure a secure connection.
  5. Prepare for Application: If using skin electrodes, ensure the skin is clean and dry for optimal adhesion and conductivity. If using metal clips, ensure they are securely attached to the acupuncture needles.

6. Operating Instructions

Operating the Hwato SDZ III is straightforward with its integrated controls.

  1. Power On: Press the "ON/OFF" button to turn on the device. The LCD display will illuminate.
    Product step diagram showing power button
    Figure 6.1: The ON/OFF button (labeled '2' in the diagram).
  2. Restore Default Settings (Optional): If you wish to start with default parameters, long-press the "Function/Choose Switch" (shuttle key) for 3 seconds after startup.
  3. Select Function/Mode: Rotate the "Function/Choose Switch" to cycle through the available treatment modes (e.g., Continuous, Discontinuous, Dilatational). Press the switch to confirm your selection. The selected mode will be shown on the LCD display.
    Function selection switch
    Figure 6.2: The Function/Choose Switch.
  4. Adjust Frequency: Use the "Function/Choose Switch" to adjust the output pulse frequency, which is continuously adjustable from 1 to 100 Hz.
  5. Set Treatment Time: Use the "Function/Choose Switch" to set the desired treatment duration.
  6. Apply Electrodes: Place the skin electrodes on the desired treatment areas or connect the metal clips to inserted acupuncture needles. Refer to anatomical charts for appropriate acupuncture points if needed.
    Fourteen meridians diagram
    Figure 6.3: Diagram of Fourteen Meridians and acupuncture points for guidance.
    Application areas for stimulation
    Figure 6.4: Common application areas for nerve and muscle massage.
  7. Adjust Intensity: Slowly turn the individual "Output Strength Adjusting Knobs" (1-6) clockwise for the active channels until a comfortable sensation is felt. Do not set the intensity too high. Each channel can be adjusted independently.
    Intensity adjustment knobs
    Figure 6.5: The six intensity adjustment knobs.
  8. Treatment Completion: The device will automatically stop after the set time. Turn all intensity knobs to the "off" position and then press the "ON/OFF" button to power off the device.
  9. Parameter Memory: The device automatically saves the last used treatment parameters, allowing for direct use with the same settings upon next power-on.
  10. Sleep Function: If no operation is performed for more than 3 minutes after startup (and output channels are closed), the device will automatically enter sleep mode, and the LCD screen will turn off to save power. Operate the "Function/Choose Switch" to awaken the device.

6.1 Working Patterns

The Hwato SDZ III offers several working patterns:

  • Continuous Wave: Provides a steady, uninterrupted pulse.
  • Discontinuous Wave: Delivers pulses in cycles of 15 seconds ON, followed by 5 seconds OFF.
  • Dilatational Wave: Features a thin wave frequency and a dense wave frequency with a ratio of 1:5. The thin wave operates for 5 seconds, followed by 10 seconds off.

7. Maintenance

Proper care ensures the longevity and optimal performance of your device.

  • Cleaning: Wipe the main unit with a soft, dry cloth. Do not use abrasive cleaners or immerse the device in water.
  • Electrode Care: Clean skin electrodes according to their specific instructions. Replace them when their adhesive properties diminish. Clean metal clips after each use.
  • Storage: Store the device and accessories in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Battery (if applicable): The device uses an internal DC9V power supply. If it has a replaceable battery compartment (not explicitly stated but implied by 'internal power supply'), ensure batteries are removed if the device will not be used for an extended period.

8. Troubleshooting

If you encounter issues with your Hwato SDZ III, try the following:

  • Device Not Powering On:
    • Ensure the power adapter is securely connected to both the device and a working power outlet.
    • Verify that the power outlet is functional.
  • No Output Sensation:
    • Check that the output wires are securely connected to both the device and the electrodes.
    • Ensure the electrodes are properly applied to the skin or needles.
    • Verify that the individual output strength adjusting knobs for the active channels are turned ON and set to a sufficient intensity.
    • Confirm that a treatment mode and time have been selected on the LCD display.
  • LCD Display is Off (Sleep Mode):
    • Operate the "Function/Choose Switch" to awaken the device from sleep mode.

If the problem persists, contact customer support or a qualified technician.

9. Specifications

FeatureDetail
ModelSDZ III
BrandVamsLuna
ApplicationBody (Nerve and Muscle Massage Relaxation, Electro Acupuncture Stimulator)
Power SupplyInternal DC9V; AC Adapter Input: 220V/50Hz or 110V/60Hz, Output: DC9V/150mA
Input Power10.0 VA
Output Pulse ShapeAsymmetric Bidirectional Pulse Wave
Output Pulse Channels6 Channels
Maximum Pulse Power0.3 VA (at 250 Ω load)
Output Pulse Frequency1 ~ 100 Hz (Continuously Adjustable)
Working PatternsContinuous Wave, Discontinuous Wave (15s ON, 5s OFF), Dilatational Wave (Thin wave 5s, 10s OFF, 1:5 ratio)
Output Current Limit≤ 10 mA (at 250 Ω load)
Output DC Component0
Dimensions (L x W x H)210 mm x 100 mm x 300 mm
Weight1.6 kg (including battery)

10. User Tips

Based on common user inquiries, here are some helpful tips:

  • Read the Manual: Many questions can be answered by thoroughly reading this instruction manual, especially regarding setup and operation.
  • Frequency Adjustment: The device offers a wide frequency range from 1 to 100 Hz. Experiment with different frequencies under professional guidance to find what works best for your specific needs and comfort.
  • Voltage Compatibility: Ensure you select the correct voltage version (110V or 220V) for your region to prevent damage to the device.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase receipt as proof of purchase.

Related Documents - SDZ III

Preview Hwato SDZ III Electroacupuncture Device User Manual & Specifications
Comprehensive user manual and technical specifications for the Hwato SDZ III Electroacupuncture Nerve and Muscle Stimulator, covering operation, features, and package contents.
Preview EMS Muscle Stimulator User Manual
Comprehensive user manual for the EMS Muscle Stimulator, including safety warnings, operating instructions, precautions, and key features for effective muscle stimulation.
Preview Medivon KTR-2610 Transcutaneous Electrical Nerve Stimulator User Manual
User manual for the Medivon KTR-2610 Transcutaneous Electrical Nerve Stimulator, detailing its features, functions, safety precautions, and maintenance.
Preview OMRON HV-F021 Electronic Nerve Stimulator Instruction Manual
This instruction manual provides detailed information on the OMRON HV-F021 Electronic Nerve Stimulator, including setup, operation, safety precautions, and maintenance for effective pain relief.
Preview Masimo Bridge™ PNFS Quick Reference Guide: Percutaneous Nerve Field Stimulator Instructions
Quick reference guide for the Masimo Bridge™ Percutaneous Nerve Field Stimulator (PNFS). Learn how to apply and remove the device for nerve stimulation therapy.
Preview Bridge™ Percutaneous Nerve Field Stimulator (PNFS) - Directions for Use
Instructions for use for the Bridge™ Percutaneous Nerve Field Stimulator (PNFS) system, designed to aid in reducing opioid withdrawal symptoms by stimulating cranial and occipital nerves. Includes indications, contraindications, warnings, application steps, removal, technical data, and electromagnetic compatibility information.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.