1. Introduction
The NJTY T1 Digital Multimeter is a compact, auto-ranging smart multimeter designed for professional and DIY electrical measurements. It features a 4000-count display and can measure AC/DC voltage, resistance, and includes functions like Non-Contact Voltage (NCV) detection, continuity testing, data hold, backlight, and a flashlight. Its intelligent recognition system automatically switches to the corresponding measurement range, making it user-friendly for various applications.

2. Safety Information
Please read and understand all safety information before operating this multimeter. Improper use can result in electric shock, injury, or damage to the meter or equipment under test.
- Always ensure the test leads are properly connected and in good condition before use.
- Do not apply voltage or current that exceeds the maximum rated values for the meter.
- Exercise extreme caution when working with live circuits.
- Always disconnect power to the circuit and discharge high-voltage capacitors before measuring resistance or continuity.
- Do not operate the meter if it appears damaged or if the battery cover is not properly closed.
- Replace batteries as soon as the low battery indicator appears to ensure accurate readings.

3. Product Overview
The NJTY T1 Multimeter features a clear HD display and intuitive controls for ease of use.

- HD Display: Shows measurement readings and indicators.
- NCV Non-Contact Voltage Sensing Probe: Detects AC voltage without physical contact.
- Flashlight Lighting Window: Provides illumination in dark environments.
- Power Switch Key: Turns the device ON/OFF.
- HOLD NCV Button: Short press for Data Retention (HOLD), long press for NCV test.
- Backlight/Flashlight Button: Activates the display backlight and flashlight.
- Voltage/Resistance Input Terminal: For connecting the red test lead.
- Negative Insertion Hole: For connecting the black test lead (common).
- NCV/Buzzer Indicator: Visual indicator for NCV detection and continuity.
4. Specifications
The NJTY T1 Multimeter offers the following technical specifications:

| Feature | Specification | Remarks |
|---|---|---|
| DC Voltage | 0.8V ~ 600V | Minimum input voltage 0.8V, Maximum input voltage 600V, Overload protection 600V AC/DC |
| AC Voltage | 1V ~ 600V | Minimum input voltage 1V, Maximum input voltage 600V, Overload protection 600V AC/DC |
| Resistance | 0 ~ 40MΩ | Overload protection 600V AC/DC |
| Maximum Display | 4000 Counts | |
| Automatic Shutdown | Yes | Automatically shuts down after 30 seconds of inactivity. |
| Backlight | Yes | Automatically shuts down after 30 seconds. |
| Flashlight | Yes | Automatically shuts down after 30 seconds. |
| Buzzer | Yes | Internal resistance is less than 50Ω. |
| NCV Indicator | Yes | Corresponds to changes in strong/weak signals. |
| Low Voltage Prompt | Yes | |
| Battery | 1.5V AAA x 2 (not included) | |
| Size | 123*61*24mm | |
| Weight | About 78g |
5. Setup
5.1 Battery Installation
The NJTY T1 Multimeter requires two 1.5V AAA batteries (not included) for operation.
- Ensure the multimeter is turned off and disconnect any test leads.
- Locate the battery compartment on the back of the device.
- Use a screwdriver to open the battery cover.
- Insert two AAA batteries, observing the correct polarity (+ and -).
- Close the battery cover and secure it with the screw.

6. Operating Instructions
The NJTY T1 Multimeter features intelligent recognition for most measurements, automatically selecting the correct range.
6.1 Power On/Off
Press the red power button (U) to turn the multimeter on or off. The display will show "Auto" upon startup.
6.2 AC Voltage Measurement
- Insert the red test lead into the Voltage/Resistance input terminal and the black test lead into the Negative insertion hole (COM).
- Touch the test probes to the AC voltage source (e.g., wall outlet).
- The multimeter will automatically detect and display the AC voltage.

6.3 DC Voltage Measurement
- Insert the red test lead into the Voltage/Resistance input terminal and the black test lead into the Negative insertion hole (COM).
- Touch the test probes to the DC voltage source (e.g., battery terminals).
- The multimeter will automatically detect and display the DC voltage.

6.4 Resistance Measurement
- Ensure the circuit is de-energized before measuring resistance.
- Insert the red test lead into the Voltage/Resistance input terminal and the black test lead into the Negative insertion hole (COM).
- Touch the test probes across the component to measure its resistance.
- The multimeter will automatically detect and display the resistance value.

6.5 Continuity Test
- Ensure the circuit is de-energized.
- Insert the red test lead into the Voltage/Resistance input terminal and the black test lead into the Negative insertion hole (COM).
- Touch the test probes to the two points you want to check for continuity.
- If there is continuity (resistance less than 50Ω), the buzzer will sound.
6.6 Non-Contact Voltage (NCV) Detection
- Long press the "HOLD NCV" button to activate NCV mode. The display will show "EF".
- Move the top of the multimeter (where the NCV sensor is located) close to the wire or outlet you want to test.
- The NCV indicator will light up and the buzzer will sound, with increasing frequency as the voltage strength increases.

6.7 Flashlight and Backlight
Press the "*/" button to toggle the display backlight and the integrated flashlight. This is useful for working in dimly lit areas.

6.8 Data Hold
Short press the "HOLD NCV" button during a measurement to freeze the current reading on the display. Press it again to release the hold function.
7. Maintenance
7.1 Cleaning
Wipe the meter with a dry, soft cloth. Do not use abrasives or solvents. Keep the test leads clean and free of debris.
7.2 Battery Replacement
When the low battery indicator appears on the display, replace the batteries promptly to ensure accurate measurements. Refer to section 5.1 for battery installation instructions.
8. Troubleshooting
- No Display/Meter Not Turning On: Check if the batteries are correctly installed and have sufficient charge. Replace if necessary.
- Inaccurate Readings: Ensure test leads are properly connected and not damaged. Verify the battery level. For resistance measurements, ensure the circuit is de-energized.
- Auto Power Off: The multimeter is designed to automatically shut down after approximately 30 seconds of inactivity to conserve battery life. This is a normal function.
- Constant Buzzer Sound during Continuity Test: If the buzzer sounds continuously even when probes are not touching, check for a short circuit in the test leads or a fault in the meter.
- NCV Not Detecting Voltage: Ensure NCV mode is activated by long-pressing the "HOLD NCV" button. The sensor needs to be close to the AC voltage source.
9. User Tips
- Beeping Sound: Some users find the continuity test beeping sound to be loud. There is no dedicated button to turn off the beeping sound.
- Voltmeter Internal Resistance: The internal resistance of the voltmeter is approximately 0.3 MΩ. While the manual might state 10 MΩ, practical measurements indicate a lower value. This is generally not an issue for most applications but is worth noting for precision work.
- Battery Type: The multimeter uses 2 AAA batteries. It is compatible with rechargeable NiMH batteries.
- Compact Size: The small size makes it very portable and easy to handle with one hand, suitable for home use and quick checks.
10. Warranty and Support
For any issues or questions regarding your NJTY T1 Digital Multimeter, please contact the seller or manufacturer's customer support. Keep your purchase receipt for warranty claims.