SMART SENSOR AS840

Contents

SMART SENSOR AS840 Digital Ultrasonic Thickness Gauge User Manual

1. Introduction

The SMART SENSOR AS840 Digital Ultrasonic Thickness Gauge is an intelligent handheld device designed for quick and precise measurement of material thickness using ultrasonic principles. Controlled by a microprocessor, it offers accurate readings for most industrial materials. This unit is widely used in various precise measurements for hardware and parts in industrial settings, notably for monitoring thickness reduction during the operation of various vessels and pressure containers. It finds broad application in manufacturing, metal processing, and commercial inspection fields.

Video: Overview of the SMART SENSOR AS840 Ultrasonic Thickness Gauge in operation.

2. Product Overview

The AS840 is designed for non-destructive testing, allowing thickness measurement without requiring access to both sides of the test piece. This capability makes it invaluable for a wide range of applications.

Key Features:

  • Automatic Calibration Function
  • Sound Speed Measurement Function
  • Sound Velocity of 12 Materials Pre-stored
  • Backlit Large Screen LCD Display
  • Data Storage and Recall Function
  • Low Battery Indication
  • Auto Power Off
SMART SENSOR AS840 Ultrasonic Thickness Gauge with key features highlighted

Image: SMART SENSOR AS840 Ultrasonic Thickness Gauge highlighting its strong performance and multiple measurement functions.

SMART SENSOR AS840 Ultrasonic Thickness Gauge with probe and accessories

Image: The AS840 unit, probe, and included accessories.

3. Specifications

ParameterValue
Measuring Range1.2~225mm
Accuracy±(1%H+0.1)mm
Resolution(<100mm)0.1mm
Sound Velocity Range1000~9999m/s
Frequency5MHz
Operation Temperature0°C~45°C
Sensor Operation TemperatureNormal 0~60°C, High Temp 0~250°C (with high temp. agent)
Minimum limit for tube measuring10mm probe: 20*3mm (steel)
Data Store/Recall FunctionYes
Sound Velocity MeasuringYes
LCD Backlight DisplayYes
Alarm SetupNo
Low Battery IndicationYes
Auto Power OffYes
Power Supply3 * 1.5V AAA alkaline batteries (not included)
Product Weight / Size210g / 150*80.5*34mm

4. Setup

4.1 Battery Installation

  1. Locate the battery compartment cover on the back of the device.
  2. Slide the cover downwards or unclip it to open.
  3. Insert three (3) 1.5V AAA alkaline batteries, ensuring correct polarity (+/-).
  4. Close the battery compartment cover securely.
Back view of the AS840 with battery compartment open

Image: Battery compartment of the AS840, showing where to insert AAA batteries.

4.2 Probe Connection

  1. Connect the transducer probe cable to the two input jacks located at the top of the AS840 unit.
  2. Ensure the connectors are firmly seated.
Top view of AS840 showing probe connection points

Image: Top view of the AS840, illustrating the two input jacks for the transducer probe.

4.3 Initial Calibration

The device features automatic zero calibration and compensation. The non-linear error of the probe can be corrected by computer software across the entire range, and system errors can be corrected to improve accuracy. For initial setup or when changing probes, perform a calibration:

  1. Apply a small amount of couplant to the included 4mm calibration block.
  2. Press the probe firmly onto the calibration block.
  3. Press the 'CAL' button on the device. The display should show the calibrated thickness (e.g., 4.0mm).
  4. Release the probe. The device is now calibrated.
AS840 with couplant and calibration block

Image: The AS840 with couplant and calibration block, demonstrating the items used for calibration.

5. Operating Instructions

5.1 Power On/Off

  • Press the red power button (POWER) to turn the device on.
  • The device will automatically power off after a period of inactivity to conserve battery.

5.2 Basic Thickness Measurement

  1. Ensure the device is calibrated.
  2. Apply a small amount of ultrasonic couplant to the surface of the material to be measured.
  3. Press the transducer probe firmly and perpendicularly onto the couplant-applied surface.
  4. The thickness reading will be displayed on the LCD screen.
  5. For accurate readings, ensure good contact between the probe and the material, and that no air bubbles are present in the couplant layer.
AS840 performing various thickness measurements

Image: Live demonstration of the AS840 measuring thickness on glass, stainless steel, and plastic.

5.3 Sound Velocity Measurement and Storage

The sound velocity can be directly measured according to the thickness of the sample, eliminating the need to look up tables or convert values. The device can store 12 sound velocity values.

  • To measure sound velocity: Use a material of known thickness. Apply couplant, place the probe, and follow the instructions in the full manual to enter the known thickness and calculate sound velocity.
  • To store sound velocity: The device allows for one-click storage of 12 thickness values. This data will not be lost after shutdown, providing convenience for high-altitude and field work. Use the 'STORE' button to save readings.

5.4 Backlight Control

  • Press the 'LIGHT' button to toggle the LCD backlight on or off for improved visibility in different lighting conditions.

6. Maintenance

  • Cleaning: Wipe the device and probe with a soft, dry cloth after each use. Do not use abrasive cleaners or solvents.
  • Storage: Store the device in its protective carrying case in a cool, dry place when not in use. Avoid extreme temperatures and humidity.
  • Battery: Remove batteries if the device will not be used for an extended period to prevent leakage.
  • Probe Care: Ensure the probe surface is clean and free from scratches or damage, as this can affect measurement accuracy.

7. Troubleshooting

  • No Reading/Inconsistent Readings:
    • Ensure sufficient couplant is applied and there are no air gaps between the probe and the material.
    • Verify the probe is making firm, perpendicular contact with the surface.
    • Check if the material's sound velocity is correctly set or calibrated for the material being measured.
    • Ensure the material surface is clean and smooth.
  • Low Battery Indication: Replace the AAA batteries promptly when the low battery indicator appears on the display.
  • Device Not Powering On: Check battery installation and ensure batteries are fresh.

8. User Tips

  • Always use a suitable ultrasonic couplant for accurate measurements. Water, oil, or specialized gels can be used depending on the application and temperature.
  • For very thin materials or materials with rough surfaces, multiple readings and careful probe placement may be necessary to achieve consistent results.
  • When measuring curved surfaces, ensure the probe is positioned to make the best possible contact.
  • Keep the calibration block handy for quick checks of device accuracy.

9. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the documentation included with your product or contact the seller directly. Keep your purchase receipt as proof of purchase.