yieryi PH-W218

Contents

Yieryi PH-W218 WiFi Tuya Smart 8-in-1 Water Quality Monitor User Manual

Model: PH-W218

1. Product Overview

The Yieryi PH-W218 is an advanced 8-in-1 WiFi-enabled water quality monitor designed for real-time measurement and data logging of various water parameters. It integrates with the Tuya Smart mobile application, allowing users to monitor pH, ORP, EC, TDS, SALT, S.G (Specific Gravity), CF (Conductivity Factor), and Temperature from anywhere. This device is ideal for applications such as aquariums, mariculture, drinking water analysis, water sources, soilless cultivation, hydroponics, laboratories, spas, and swimming pools.

Yieryi PH-W218 8-in-1 Water Quality Monitor with probes and app interface

Figure 1: Yieryi PH-W218 8-in-1 Water Quality Monitor with probes and app interface.

Key Features:

  • 8-in-1 Measurement: Simultaneously measures pH, ORP, EC, TDS, SALT, S.G, CF, and Temperature.
  • WiFi Connectivity & Tuya Smart App: Real-time monitoring, data transfer, and remote control via the Tuya Smart app (compatible with Apple and Android).
  • Data Logging & Export: Records historical data (daily, weekly, monthly) and allows export to Excel via email for analysis.
  • Alarm Functions: Customizable high and low alarm settings for various parameters with app notifications.
  • Automatic Correction & Calibration: Supports multi-point calibration for pH, ORP, and EC.
  • Backlight Display: Clear digital LCD display for easy reading in various lighting conditions.
  • Wall-Mounted Design: Convenient installation with included suction cups and clip.

2. Safety Information

  • This tester is not waterproof. Do not let the tester get wet or damp during use. Otherwise, the tester may not work normally.
  • Always use the provided power adapter. Do not use a mobile phone USB cable to connect the power supply, as it is not compatible with this product.
  • Keep the device and probes out of reach of children.
  • Handle calibration solutions with care and follow their specific safety instructions.

3. Package Contents

There are two main types of packages available, differing by the inclusion of calibration liquids:

Standard Package (e.g., PH-W218EU/US/UK/AU):

  • 1x Monitor Unit
  • 1x PH Probe
  • 1x ORP Probe
  • 1x PPM/EC/Salt/S.G./CF Probe
  • 1x Temp Probe
  • 1x Charger (EU/US/UK/AU plug type based on selection)
  • 1x English Manual
  • 2x Suction Cups
  • 1x Clip
  • 1x 4.0, 6.86 & 9.18 PH Calibration Powder

Package with Calibration Liquids (e.g., PH-W218EU-P/US-P/AU-P/UK-P):

  • 1x Monitor Unit
  • 1x PH Probe
  • 1x ORP Probe
  • 1x PPM/EC/Salt/S.G./CF Probe
  • 1x Temp Probe
  • 1x Charger (EU/US/UK/AU plug type based on selection)
  • 1x English Manual
  • 2x Suction Cups
  • 1x Clip
  • 1x 4.0, 6.86 & 9.18 PH Calibration Powder
  • 1x ORP Calibration Powder
  • 1x 1413us/cm/25ml Calibration Solution
  • 1x 12.88ms/cm/25ml Calibration Solution
Package contents including monitor, probes, charger, and calibration solutions

Figure 2: Example package contents with calibration liquids and powders.

4. Setup

4.1 Device Connection

  1. Remove the protective caps from the pH, ORP, and PPM/EC/Salt/S.G./CF probes.
  2. Connect the pH electrode, ORP electrode, and temperature electrode to their respective ports on the monitor unit.
  3. Connect the power adapter to the monitor unit and plug it into a suitable power outlet.
Diagram of the PH-W218 monitor with labeled ports for probes and power adapter

Figure 3: Device diagram with labeled components and connection points.

4.2 Mobile App Installation and Pairing

The PH-W218 connects to your smartphone via the Tuya Smart app for real-time monitoring and data management.

  1. Download the App: Search for "Tuya Smart" in your mobile app store (Apple App Store or Google Play Store) and download it.
  2. Register/Log In: Open the Tuya Smart app and register for a new account or log in if you already have one.
  3. Add Device:
    • On the app homepage, click "Add Device" or the "+" icon.
    • Select "Add Manually" and choose the appropriate device type (e.g., "Socket (Wi-Fi)" or "Power Strip (Wi-Fi)" if specific categories are not available, or search for "Water Quality Tester").
    • On the PH-W218 monitor, long press the "MODE" key to turn WiFi on/off. The WiFi indicator on the device should start blinking, indicating pairing mode.
    • In the app, confirm the device is in pairing mode and proceed.
    • Enter your WiFi network name and password. Ensure your phone and the device are connected to the same 2.4GHz WiFi network.
    • The app will discover and add the device. Once successfully added, click "Done".

Video 1: Step-by-step guide for WiFi connection and data transfer using the Tuya Smart app.

Tuya Smart app showing WiFi connection and real-time monitoring of an aquarium

Figure 4: WiFi connection and data transfer, showing the monitor connected to an aquarium and a phone displaying the Tuya Smart app.

5. Operation

5.1 Basic Measurement

  1. Clean the electrodes with distilled water and dry them well.
  2. Connect the power adapter to turn on the device.
  3. Immerse the pH, ORP, and PPM/EC/Salt/S.G./CF electrodes, along with the temperature probe, into the testing solution.
  4. Stir lightly and wait a few seconds for the readings to stabilize on the LCD screen.
Various applications of the water quality tester including drinking water, pool, fish tank, lab, and hydroponics

Figure 5: The PH-W218 is suitable for various water quality testing applications.

5.2 Switching Display Modes

  • Temperature Unit: Short press the "TEMP CAL" key to switch between °C and °F.
  • Parameter Display: Short press the "MODE" key to cycle through PPM, EC, SALT, S.G, and CF measurements on the rightmost display.
Diagram of the PH-W218 monitor with labeled buttons for mode and calibration

Figure 6: Control buttons for mode switching and calibration.

5.3 Mobile App Monitoring

Once connected to the Tuya Smart app, you can:

  • View current readings for all 8 parameters in real-time.
  • Access historical data (daily, weekly, monthly) in graph format for comparative analysis.
  • Set high and low alarm thresholds for each parameter. The app will send notifications if readings fall outside the set range.
  • Export data to Excel via email for detailed record-keeping and sharing.
Screenshots of the Tuya Smart app showing various monitoring interfaces for temperature, pH, EC, and TDS

Figure 7: Mobile application monitoring interfaces.

Laptop displaying an Excel spreadsheet with exported water quality data from the monitor

Figure 8: Data export to Excel via email.

6. Calibration

Regular calibration is essential for maintaining the accuracy of your water quality monitor. The electrodes must be re-calibrated if there has been no calibration for a long time, after continual use, or if high accuracy is required.

6.1 PH Calibration (3-point)

  1. Prepare standard buffer solutions of pH 6.86, pH 4.00, and pH 9.18 (at 25°C) in three separate clean beakers. For accurate calibration, use two beakers for each solution: one for cleaning the electrode and one for calibration.
  2. Plug in and turn on the device.
  3. Immerse the pH electrode into the pH 6.86 standard buffer solution. Stir gently until the value stabilizes.
  4. Long press the "TEMP CAL" key for 5 seconds. When the screen displays "06.86", the meter enters pH 6.86 automatic correction mode. The calibration is finished when the display value corresponds to the standard buffer solution.
  5. Clean the electrode with distilled water.
  6. Immerse the electrode into the pH 4.00 standard buffer solution. Stir gently until the value stabilizes.
  7. Long press the "TEMP CAL" key for 5 seconds. When the screen displays "04.00", the meter enters pH 4.00 automatic correction mode. The calibration is finished when the display value corresponds to the standard buffer solution.
  8. Clean the electrode with distilled water.
  9. Immerse the electrode into the pH 9.18 standard buffer solution. Stir gently until the value stabilizes.
  10. Long press the "TEMP CAL" key for 5 seconds. When the screen displays "09.18", the meter enters pH 9.18 automatic correction mode. The calibration is finished when the display value corresponds to the standard buffer solution.

Note: Users can choose calibration solutions of 6.86, 4.00, 9.18 or 7.00, 4.00, 10.00. The factory default calibration is 6.86, 4.00, 9.18. Be careful not to use the wrong calibration solution.

6.2 EC Calibration

  1. Turn on the meter. Press the "MODE" key to select EC-ms.
  2. Immerse the electrode into distilled water for about five minutes to activate it.
  3. Immerse the electrode into the 12.88ms/cm (at 25°C) EC calibration solution. Stir gently.
  4. When the reading is stable, long press the "CAL" key for about 5 seconds. The display will show "12.80" or "12.70-12.90". The calibration is complete when the display shows the correct value.
  5. Clean the electrode with distilled water and dry it with filter paper.
  6. (Optional) Immerse the electrode into the 1.413ms/cm (at 25°C) EC calibration solution. Stir gently. The display value should be the same as or close to the calibration solution value.

6.3 ORP Calibration

  1. Plug in the power to turn on.
  2. Immerse the ORP electrode into distilled water for five minutes.
  3. Immerse the ORP electrode in 265mV (25°C) ORP calibration solution. Stir gently.
  4. When the reading is stable, long press the "CAL" key for about 5 seconds. The display will show "265" or a plus/minus value until it shows "265".
  5. Clean the electrode with distilled water and dry it with filter paper.
  6. (Optional) Immerse the electrode in 100mV (25°C) ORP calibration solution. Stir gently until the display value is the same as or close to the calibration solution value.
Calibration powders and solutions for pH, ORP, and conductivity

Figure 9: Calibration powders and solutions.

7. Specifications

ParameterRangeResolutionAccuracy
pH0.00 - 14.00 pH0.01 pH±0.03 pH
ORP-2000mv - +2000mv1mv±5mv
EC0-19000us/cm, 0-199.0 ms/cm1us/cm (<1999us/cm), 10us/cm (>1999us/cm), 0.1ms/cm±2% F.S
TDS0-19990ppm, 0-199.0ppt1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt±2% F.S
SALT0-199.9ppt, 0%-20%1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt, 0.01%±2% F.S
S.G0.990 - 1.4000.001±2% F.S
CF0 - 1999.0 CF1 CF±2% F.S
Temperature0 - 50°C (32 - 122°F)0.1°C (0.2°F)±1.0°C (±2°F)

General Specifications:

  • Power Supply: DC6V
  • Temperature Compensation: 0°C - 50°C
  • Operating Temperature: 0°C - 50°C
  • Dimensions: 152mm × 80mm × 26mm
  • Weight: 245g
Detailed specifications table for the water quality monitor

Figure 10: Comprehensive specifications of the PH-W218 monitor.

8. Maintenance

  • After each use, clean the electrodes thoroughly with distilled water.
  • Always replace the protective caps on the electrodes when not in use to prevent damage and drying out.
  • Store the device in a cool, dry place.
  • Regularly check the probes for any signs of wear or damage. Replace damaged probes promptly to ensure accurate readings.

9. Troubleshooting

  • Inaccurate Readings:
    • Ensure electrodes are clean and free from debris.
    • Perform a full calibration for pH, ORP, and EC as described in Section 6.
    • Check if the electrodes are properly immersed in the solution.
    • Verify that the temperature compensation is correctly applied (if applicable).
  • Device Not Powering On:
    • Ensure the power adapter is securely connected to the device and a working power outlet.
    • Confirm you are using the original power adapter provided with the device.
  • WiFi Connection Issues:
    • Ensure your device is within range of a 2.4GHz WiFi network.
    • Verify that the WiFi password entered in the Tuya Smart app is correct.
    • Restart both the monitor and your router, then attempt pairing again.
  • "The device rusts, cables too short":
    • The device is not waterproof; ensure it is kept dry to prevent rust.
    • The standard sensor cable length is 1.1m. If longer cables are needed, contact customer support for potential extension options or replacement probes with longer cables.
  • Lost Power Supply/Damaged Probe:
    • Replacement power supplies and individual probes may be available. Contact customer support for assistance.

10. User Tips

  • Initial Setup: Many users find the initial setup and app pairing straightforward. Follow the video guide (Video 1) for a visual walkthrough.
  • App Integration: The Tuya Smart app works reliably for real-time monitoring and data logging. While direct integration with other smart home platforms like Home Assistant might require advanced configurations (e.g., via Tuya's developer API), the native app provides comprehensive functionality.
  • Cable Length: The standard sensor cable length is 1.1m. If your setup requires longer reach, consider purchasing extension cables or longer probes if available from the manufacturer or compatible third-party suppliers.
  • Calibration Solutions: Always have fresh calibration solutions on hand for accurate measurements, especially for pH, ORP, and EC. Store them properly to maintain their integrity.
  • Waterproof Note: Remember the device itself is not waterproof. Mount it in a dry location, using the provided hook design, to protect it from splashes and moisture.

11. Warranty and Support

All defects in materials and manufacture of this device are guaranteed for one year from the date of purchase. Within this period, parts will be repaired or replaced free of charge if the damage is not caused by user negligence or improper operation.

For technical support, replacement parts (e.g., probes, power supply), or warranty claims, please contact the seller or the manufacturer's customer service.

For additional information and documents, you may refer to the User Manual (PDF).