1. Introduction
The Hantek HT8050 and HT8100 are high-voltage differential probes designed for use with oscilloscopes. These probes offer high stability, reliability, and precision for various electrical measurements. They utilize a differential input mode, making them ideal for high-voltage floating measurements where safety and accuracy are paramount. The probes feature an over-range alarm function indicated by a red LED for enhanced user awareness.
2. Package Contents
Please check the package contents upon receipt to ensure all items are present:
- 1 x Oscilloscope Probe (HT8050 or HT8100 model)
- 2 x Test Clips (Gator Clips)
- 2 x Probe Hooks (Finger Pins)
- 1 x Power Adapter (DC9V, 1000mA)
- 1 x English User Manual
3. Specifications
3.1. Electrical Performance
| Parameter | HT8050 | HT8100 |
|---|---|---|
| Model Number | HT8050 | HT8100 |
| Bandwidth (-3dB) | 50MHz | 100MHz |
| Rise time | 7ns | 3.5ns |
| Accuracy | ±2% | |
| Attenuation Ratio | 1/50, 1/500 | |
| Input Impedance | Single-ended to ground: 4MΩ Between two input ends: 8MΩ | |
| Input Capacitance | Single-ended to ground: 7pF Between two input ends: 3.5pF | |
| Maximum Difference Measurement Voltage | 1/50: 130V (DC+peakAC) 1/500: 1300V (DC+peakAC) | |
| Maximum Common Mode Input Voltage | 1000Vrms, CATII 600Vrms, CATIII | |
| Noise | 1/50: ≤1.5mVrms 1/500: ≤1mVrms | |
| Linearity | ±1% | |
| Delay Time | 20ns±1ns | |
| CMRR | DC: ≥80dB 100Hz: ≥60dB 1MHz: ≥50dB | |
| Over Range Warning Voltage Threshold | 1/50: 140V±4V (DC+peakAC) 1/500: 1400V±40V (DC+peakAC) | |
| Power Adapter Output Power | DC9V, 1000mA | |
| Power Adapter Input Power | AC100-240V, 50-60Hz | |
3.2. Mechanical Specifications
| Parameter | HT8050 | HT8100 |
|---|---|---|
| Length of Input Cable | About 60cm | About 30cm |
| Length of Output Cable | About 90cm | |
| Gator Clip Dimensions | About 85 x 40 x 13 mm | |
| Finger Pin Dimensions | About 160 x 45 x 13 mm | |
| Probe Body Dimensions | 120 x 55 x 24 mm | |
| Weight | 160g | |
3.3. Environmental Characteristics
| Parameter | Value |
|---|---|
| Working Temperature | 0-40℃ |
| Storage Temperature | -10-45℃ |
| Working Humidity | 85%RH |
| Storage Humidity | 90%RH |
| Working Altitude | 3000m |
| Storage Altitude | 12000m |
4. Setup
- Power Connection: Connect the provided DC 9V power adapter to the "EXT POWER DC 9V" input port on the probe body. Ensure the adapter is plugged into a suitable AC power source (100-240V, 50-60Hz). The green "POWER" LED indicator should illuminate.
- Output Connection: Connect the probe's output cable (BNC connector) to the input channel of your oscilloscope. The probe outputs signals with a maximum voltage of 2.6V into a 1MΩ impedance.
- Input Connection: Attach the red and black input cables from the probe to the test points of the circuit you wish to measure. Use the provided test clips (gator clips) or probe hooks (finger pins) as appropriate for secure connection.
- Select Attenuation: Use the switch on the probe body to select the desired attenuation ratio: 1/50 or 1/500. This depends on the voltage levels you expect to measure.
5. Operating Instructions
- Power On: Ensure the probe is correctly connected to its power adapter and the oscilloscope. The green "POWER" LED should be lit.
- Select Attenuation: Choose the appropriate attenuation ratio (1/50 or 1/500) based on the expected voltage of the signal you are measuring. For lower voltages (up to 130V DC+peakAC), use 1/50. For higher voltages (up to 1300V DC+peakAC), use 1/500.
- Connect to Circuit: Carefully connect the red and black input leads to the points in the circuit where you want to measure the differential voltage. Ensure secure connections to prevent accidental disconnections during measurement.
- Observe Waveform: View the waveform on your connected oscilloscope. Adjust the oscilloscope's vertical and horizontal settings as needed to get a clear and stable display.
- Over-Range Indication: If the red "OVER RANGE" LED illuminates, it indicates that the input voltage exceeds the selected attenuation range. Immediately disconnect the probe from the circuit and switch to a higher attenuation ratio (e.g., from 1/50 to 1/500) or ensure the voltage is within the probe's limits.
5.1. Example Application Video
The following video demonstrates an application of an automotive oscilloscope, which can be used in conjunction with this differential probe for certain measurements. While this video uses a different Hantek device (1008C), the principles of connecting probes and observing waveforms are similar.
6. Maintenance
- Cleaning: Use a soft, dry cloth to clean the probe body. Do not use abrasive cleaners or solvents.
- Storage: Store the probe and its accessories in a dry environment, away from extreme temperatures and humidity, within the specified storage conditions (-10-45℃, 90%RH).
- Cable Care: Avoid sharp bends or kinks in the cables to prevent internal damage.
- Inspection: Periodically inspect the cables and connectors for any signs of wear or damage.
7. Troubleshooting
- No Power Indication: If the green "POWER" LED does not light up, ensure the power adapter is securely connected to both the probe and a live power outlet. Check the power adapter for any visible damage.
- "OVER RANGE" LED Constantly On: This indicates the input voltage is too high for the selected attenuation. Switch to the 1/500 attenuation setting. If it remains on, the voltage may exceed the probe's maximum rating (1300V DC+peakAC), and the probe should not be used for that measurement.
- No Signal on Oscilloscope:
- Verify the probe's output cable is securely connected to the oscilloscope input.
- Check the oscilloscope's input settings (e.g., channel enabled, correct voltage scale).
- Ensure the probe's input leads are making good contact with the circuit under test.
- Confirm the circuit under test is powered and functioning correctly.
- Noisy or Unstable Waveform:
- Ensure all connections are secure and free from interference.
- Check for proper grounding of the oscilloscope and the circuit.
- Minimize electromagnetic interference from other devices.
8. User Tips
No specific user tips were available from the provided customer reviews or Q&A data.
9. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the official Hantek website or contact your authorized dealer. Keep your purchase receipt as proof of purchase for any warranty claims.