Hantek HT8050 HT8100 Oscilloscope High Voltage Differential Probe User Manual

Models: HT8050 (50MHz), HT8100 (100MHz)

1. Introduction

The Hantek HT8050 and HT8100 are high-voltage differential probes designed for use with oscilloscopes. These probes offer high stability, reliability, and precision for various electrical measurements. They utilize a differential input mode, making them ideal for high-voltage floating measurements where safety and accuracy are paramount. The probes feature an over-range alarm function indicated by a red LED for enhanced user awareness.

Front view of the Hantek HT8050 probe
Figure 1: Front view of the Hantek HT8050 High Voltage Differential Probe.

2. Package Contents

Please check the package contents upon receipt to ensure all items are present:

  • 1 x Oscilloscope Probe (HT8050 or HT8100 model)
  • 2 x Test Clips (Gator Clips)
  • 2 x Probe Hooks (Finger Pins)
  • 1 x Power Adapter (DC9V, 1000mA)
  • 1 x English User Manual
Hantek HT8050 probe with all included accessories
Figure 2: Hantek HT8050 probe with included accessories.

3. Specifications

3.1. Electrical Performance

Parameter HT8050 HT8100
Model NumberHT8050HT8100
Bandwidth (-3dB)50MHz100MHz
Rise time7ns3.5ns
Accuracy±2%
Attenuation Ratio1/50, 1/500
Input ImpedanceSingle-ended to ground: 4MΩ
Between two input ends: 8MΩ
Input CapacitanceSingle-ended to ground: 7pF
Between two input ends: 3.5pF
Maximum Difference Measurement Voltage1/50: 130V (DC+peakAC)
1/500: 1300V (DC+peakAC)
Maximum Common Mode Input Voltage1000Vrms, CATII
600Vrms, CATIII
Noise1/50: ≤1.5mVrms
1/500: ≤1mVrms
Linearity±1%
Delay Time20ns±1ns
CMRRDC: ≥80dB
100Hz: ≥60dB
1MHz: ≥50dB
Over Range Warning Voltage Threshold1/50: 140V±4V (DC+peakAC)
1/500: 1400V±40V (DC+peakAC)
Power Adapter Output PowerDC9V, 1000mA
Power Adapter Input PowerAC100-240V, 50-60Hz

3.2. Mechanical Specifications

Parameter HT8050 HT8100
Length of Input CableAbout 60cmAbout 30cm
Length of Output CableAbout 90cm
Gator Clip DimensionsAbout 85 x 40 x 13 mm
Finger Pin DimensionsAbout 160 x 45 x 13 mm
Probe Body Dimensions120 x 55 x 24 mm
Weight160g

3.3. Environmental Characteristics

Parameter Value
Working Temperature0-40℃
Storage Temperature-10-45℃
Working Humidity85%RH
Storage Humidity90%RH
Working Altitude3000m
Storage Altitude12000m
Dimensions of the Hantek HT8050 probe body
Figure 3: Hantek HT8050 Probe Body Dimensions (120mm x 55mm x 24mm).

4. Setup

  1. Power Connection: Connect the provided DC 9V power adapter to the "EXT POWER DC 9V" input port on the probe body. Ensure the adapter is plugged into a suitable AC power source (100-240V, 50-60Hz). The green "POWER" LED indicator should illuminate.
  2. Output Connection: Connect the probe's output cable (BNC connector) to the input channel of your oscilloscope. The probe outputs signals with a maximum voltage of 2.6V into a 1MΩ impedance.
  3. Input Connection: Attach the red and black input cables from the probe to the test points of the circuit you wish to measure. Use the provided test clips (gator clips) or probe hooks (finger pins) as appropriate for secure connection.
  4. Select Attenuation: Use the switch on the probe body to select the desired attenuation ratio: 1/50 or 1/500. This depends on the voltage levels you expect to measure.
Hantek HT8050 probe with input and output cables
Figure 4: Hantek HT8050 probe showing input and output connections.
Close-up view of the Hantek HT8050 probe's controls and indicators
Figure 5: Close-up of probe controls, including POWER LED, OVER RANGE LED, and 1/50, 1/500 attenuation switch.

5. Operating Instructions

  1. Power On: Ensure the probe is correctly connected to its power adapter and the oscilloscope. The green "POWER" LED should be lit.
  2. Select Attenuation: Choose the appropriate attenuation ratio (1/50 or 1/500) based on the expected voltage of the signal you are measuring. For lower voltages (up to 130V DC+peakAC), use 1/50. For higher voltages (up to 1300V DC+peakAC), use 1/500.
  3. Connect to Circuit: Carefully connect the red and black input leads to the points in the circuit where you want to measure the differential voltage. Ensure secure connections to prevent accidental disconnections during measurement.
  4. Observe Waveform: View the waveform on your connected oscilloscope. Adjust the oscilloscope's vertical and horizontal settings as needed to get a clear and stable display.
  5. Over-Range Indication: If the red "OVER RANGE" LED illuminates, it indicates that the input voltage exceeds the selected attenuation range. Immediately disconnect the probe from the circuit and switch to a higher attenuation ratio (e.g., from 1/50 to 1/500) or ensure the voltage is within the probe's limits.
Hantek HT8050 probe connected to an oscilloscope and in use
Figure 6: Hantek HT8050 probe in use, connected to an oscilloscope for measurement.

5.1. Example Application Video

The following video demonstrates an application of an automotive oscilloscope, which can be used in conjunction with this differential probe for certain measurements. While this video uses a different Hantek device (1008C), the principles of connecting probes and observing waveforms are similar.

Video 1: Automotive Oscilloscope testing of Valve EGR. This video illustrates general oscilloscope probe connection and usage principles.

6. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the probe body. Do not use abrasive cleaners or solvents.
  • Storage: Store the probe and its accessories in a dry environment, away from extreme temperatures and humidity, within the specified storage conditions (-10-45℃, 90%RH).
  • Cable Care: Avoid sharp bends or kinks in the cables to prevent internal damage.
  • Inspection: Periodically inspect the cables and connectors for any signs of wear or damage.

7. Troubleshooting

  • No Power Indication: If the green "POWER" LED does not light up, ensure the power adapter is securely connected to both the probe and a live power outlet. Check the power adapter for any visible damage.
  • "OVER RANGE" LED Constantly On: This indicates the input voltage is too high for the selected attenuation. Switch to the 1/500 attenuation setting. If it remains on, the voltage may exceed the probe's maximum rating (1300V DC+peakAC), and the probe should not be used for that measurement.
  • No Signal on Oscilloscope:
    • Verify the probe's output cable is securely connected to the oscilloscope input.
    • Check the oscilloscope's input settings (e.g., channel enabled, correct voltage scale).
    • Ensure the probe's input leads are making good contact with the circuit under test.
    • Confirm the circuit under test is powered and functioning correctly.
  • Noisy or Unstable Waveform:
    • Ensure all connections are secure and free from interference.
    • Check for proper grounding of the oscilloscope and the circuit.
    • Minimize electromagnetic interference from other devices.

8. User Tips

No specific user tips were available from the provided customer reviews or Q&A data.

9. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the official Hantek website or contact your authorized dealer. Keep your purchase receipt as proof of purchase for any warranty claims.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.