1. Overview
The YR1035+ and YR1030+ battery internal resistance testers are true four-wire internal resistance meters designed to simultaneously measure internal resistance and voltage. They can also be used for general resistance measurement. The measurement results are clearly displayed on the same screen.
These instruments are ideal for battery configuration, performance testing, and screening applications. With their exquisite structure and portability, they are perfect tools for electricians and electronic testing.
The YR1030+ model features true 4-wire AC sine measurement, effectively avoiding the impact of test leads and contact resistance. It includes two function interfaces, an old battery sorting function, low power consumption, and various power-saving options. The test range is up to 200 ohms with a minimum resolution of 0.01 milliohms, capable of testing large capacity, high discharge rate, and low resistance batteries, as well as button batteries and 9V laminated batteries.
The YR1035+ model also offers true 4-wire AC sinusoidal measurements, effectively refraining from the effects of test leads and contact resistance. It includes two function interfaces for battery sorting and features a built-in battery with phone recharger charging, low power consumption, and multiple power-saving options. The test range is up to 200 ohms with a minimum resolution of 0.01 milliohms. Pressing the HOLD button enables the ZR function for measuring internal resistance as low as 0.30-0.50 milliohms with enhanced accuracy.

2. Safety Instructions
- The machine must not be connected to a DC voltage greater than 100V.
- No AC voltage can be input under any conditions.
- The MicroUSB port is exclusively for charging, and the input voltage should be between 4.5V-5.5V.
3. Setup
3.1. Button Functions
- 【POWER】 Key (Power / Set / Cancel):
- Short press to turn on from shutdown state.
- In menu options, click to cancel settings and return to the menu list.
- In normal and sorting modes, click to enter the menu list.
- Under the menu list, click to enter the selected menu options.
- In normal and sorting modes, long press to enter the shutdown interface. Use 【HOLD】 and 【RANGE R】 to confirm shutdown.
- 【HOLD】 Key (Save / Confirm):
- Under the "Power Off?" interface, click to confirm shutdown.
- In normal and sorting modes, short press will display "HOLD" on the left side of the screen, indicating the screen is locked. Click again to unlock.
- Under the menu list, short press to exit directly from the menu list.
- From menu options, click to save settings and return to the menu list. Click again to save and return to the measurement interface.
- In parameter setting, click to switch between 【RANGE R】 and 【RANGE U】.
- 【RANGE R】 Key (Move up / Add):
- Under the "Power Off?" interface, click to cancel shutdown.
- In normal and sorting modes, click to switch the internal resistance range.
- Under the menu list, click to move up menu options. Long press for more than 3 seconds for continuous motion.
- From menu options, click to move up to select or change parameter values. Long press for more than 3 seconds for continuous motion.
- 【RANGE U】 Key (Move down / Reduce):
- In normal and sorting modes, click to switch the voltage range.
- Under the menu list, click to move down menu options. Long press for more than 3 seconds for continuous motion.
- From menu options, click to move down to select or change parameter values. Long press for more than 3 seconds for continuous motion.
3.2. Initial Connection
The device comes with various test leads and probes. Ensure you select the appropriate connection method for your battery type.


4. Operating Instructions
4.1. General Measurement Mode
In the measurement interface, click the 【POWER】 key to enter the menu list. Use the 【RANGE R】 and 【RANGE U】 keys to navigate through the menu options. Press 【POWER】 again to select an option and enter its settings. Press the 【HOLD】 key to exit settings and return to the measurement interface.
The display shows internal resistance and voltage simultaneously. The upper left part indicates the internal resistance range (AUTO for automatic ranging, 20mR, 200mR, 2R, 20R, 200R). The lower part shows the voltage range (AUTO, 1V, 10V, 100V).

4.2. Sorting Mode
This mode is used for battery screening. Before using, set the grading parameters in the menu (Section 5. Setting of Grading). After setting, enter sorting mode to measure. The device will display "YES" or "NO" and indicate which parameter is unqualified if the battery meets or fails the set criteria.
4.3. Test Considerations
- The device is factory calibrated; no further calibration or reset is typically needed.
- Ensure internal resistance and voltage ranges are set correctly (e.g., AUTO mode).
- The twisted test wires should not be disassembled. Keep them wound in red and black to enhance anti-interference and accuracy.
- Each test lead must make good contact with the battery, with no connection between individual test leads.
- When using Kelvin test leads, the clamps must be open and not directly touching each other or the battery electrode.
- For 18650 batteries, use the battery stand. Adjust the movable stand to the battery length and ensure negative contact compresses the battery. Positive and negative poles should not expose the middle.
- When using probe test leads, press them vertically onto the battery terminals so both probes touch independently. Ensure battery surfaces are clean for good contact.
5. Maintenance
5.1. Host Maintenance
- Avoid high temperature and humid environments to prevent moisture oxidation of interfaces and circuits.
- Avoid prolonged direct sunlight exposure to the LCD screen to prevent aging.
- If not used for a long time, keep the battery charged to 50% (or between 3.7V-3.9V) to extend its lifespan.
5.2. External Test Line Maintenance
- Regularly check the continuity of each signal line of the test line to avoid measurement errors due to broken lines.
- Avoid high temperature and humid environments to prevent moisture oxidation of interfaces and probes.
6. Troubleshooting
6.1. Common Problems and Solutions
- Voltage value jitter when not measured after starting up: This is normal. The value is random and meaningless, and does not affect measurement results.
- Adjusting YR1035+ to zero / Resistance between Kelvin clips: The device is factory calibrated. The minimum value is generally not zero but does not affect accuracy. No processing is required.
- Measuring large single lithium iron phosphate battery: Recommended for batteries within 100Ah. For batteries above 100Ah, internal resistance is generally lower than 0.3mΩ. This device can measure batteries with internal resistance below 100V and above 0.3-0.5mΩ.
- Measuring internal resistance below 0.3mΩ: It is possible, but reading error is large. For example, 1mΩ error is ±0.03mΩ. More than 0.3mΩ is recommended for reliable results.
- Measuring 18650 battery internal resistance (fully charged or not): For a general idea, test directly. For precise results, test new batteries fully charged, and old batteries after they are drained. The internal resistance change due to charge level is usually small for good batteries.
- Measuring time positive and negative / Connection damage: Connecting in reverse has no effect; the device does not differentiate positive and negative poles. It will not be damaged as long as the voltage does not exceed 28V.
- Figures change quickly and cannot be measured: This indicates poor contact with the battery, strong external interference, or the battery is charging/discharging. Normal use should not show this.
- Testing one of multiple connected batteries in electric cars: You can directly test one section as long as the car is not starting or charging.
- Measuring battery packs: Possible if total voltage is less than 28V. However, connecting line resistance often exceeds battery internal resistance, so series measurement may not be accurate. It's best to measure individual battery poles.
- Measuring capacitance internal resistance: The device measures the approximate internal resistance value of capacitance.
7. Specifications
7.1. General Technical Indicators
- Maximum Voltage between any terminals: 100V
- Display: Internal resistance: 2,000 counts; Voltage: 99,999 counts. Updates 4 times per second.
- Temperature:
- Working: 10℃-40℃
- Storage: -20℃-60℃
- Internal resistance temperature coefficient: 0.1 * (specified accuracy)/℃ (<18℃ or >28℃)
- Battery Type: 3.7V 2000mAh 18650 (included)
- Power Consumption:
- <60mA (200mΩ and above range, 3.7V power, backlight off)
- <120mA (20mΩ range, 3.7V power, backlight off)
- <10mA (low power mode)
- 0mA (power off)
7.2. Precision Specifications
Accuracy format: ±([accuracy percentage] + [least significant digit])
| Range | Maximum Resolution | Accuracy | Display Mode |
|---|---|---|---|
| 20mΩ | 0.01mΩ | 0.7%+7 | 22.00mΩ |
| 200mΩ | 0.1mΩ | 0.5%+5 | 220.0mΩ |
| 2Ω | 1mΩ | 0.5%+5 | 2.200Ω |
| 20Ω | 10mΩ | 0.5%+5 | 22.00Ω |
| 200Ω | 0.1Ω | 0.6%+5 | 220.0Ω |
Note: If the measurement is outside the 5%-95% range of the gear, the error may be greater than the stated accuracy.
| Range | Maximum Resolution | Accuracy | Display Mode |
|---|---|---|---|
| 1V | 0.00001V | 0.15%+0.015 | .99999V |
| 10V | 0.0001V | 0.15%+0.010 | 9.9999V |
| 100V | 0.001V | 0.15%+0.015 | 99.999V |
- Positive and negative symmetry error: ±(0.012% + 5 words)
- Temperature Drift within normal use temperature (10℃-40℃): 100ppm/℃
7.3. Physical Dimensions and Weight
- Package Size: 120 * 60 * 40mm / 4.7 * 2.4 * 1.6in
- YR1030+ Package Weight: 300g / 10.6ounce
- YR1035+ Package Weight: 320g / 11.3ounce


8. User Tips
- For accurate resistance measurements, ensure the battery is fully charged if it's a new battery, or drained if it's an old battery.
- The device is versatile and can be used for various battery types, including NiMH, lithium, and LiFePo4.
- The included replacement tips for the probes are a useful addition for prolonged use.
9. Warranty and Support
For customer service, after-sales support, or to inquire about new products, please scan the QR code below to connect via WeChat.