Victor 14+ Temperature Calibrator User Manual

Model: 14+ | Brand: RuoShui

1. Introduction

This temperature calibrator (referred to as 'the calibrator' in this manual) is a handheld, battery-operated instrument designed to measure and source various electrical and physical parameters. It offers high precision and a range of functionalities for temperature verification and calibration tasks.

Key Features:

  • Measurement Capabilities: DC-voltage, ohm, Thermocouple (Tc), Resistance Temperature Detector (RTD), and continuity.
  • Sourcing Capabilities: DC-voltage, ohm, Thermocouple (Tc), and RTD.
  • 2-wire, 3-wire, and 4-wire connection methods for ohm and RTD measurements.
  • Large LCD display showing TC/RTD measurement values and corresponding voltage/resistance simultaneously.
  • TC measurement/source terminals with built-in lead connector for automatic reference junction (RJ) compensation.
  • Room temperature monitoring during operation.
  • Measurement wave-filter function for stable readings.
  • Measurement manual-holding function for data retention.
Front view of the Victor 14+ Temperature Calibrator highlighting its digital display and control buttons.
Figure 1: Front view of the Victor 14+ Temperature Calibrator.
Dimensions of the Victor 14+ Temperature Calibrator, showing its protective case, USB interface, and bracket.
Figure 2: Dimensions and key components of the calibrator.

2. Safety Information

For the correct and safe use of the instrument, please follow the cautionary notes provided in this manual. The manufacturer is not liable for any damage resulting from improper use of the instrument.

⚠ Warning

  • To avoid possible electric shock or personal injury, do not apply more than the rated voltage (as marked on the calibrator) between terminals or between any terminal and earth ground.
  • Before use, verify the meter's operation by measuring a known voltage.
  • Follow all equipment safety procedures.
  • Do not use the meter if it is damaged. Inspect the case for cracks or missing plastic, especially around connectors.
  • Select the proper function and range for the measurement.
  • Ensure the battery door is closed and latched before operating.
  • Remove test leads before opening the battery door.
  • Inspect test leads for damaged insulation or exposed metal. Replace damaged leads.
  • When using probes, keep fingers behind the finger guards.
  • Connect the common test lead before the live test lead. Disconnect the live test lead first.
  • Do not use the meter if it operates abnormally. Protection may be impaired.
  • Do not operate in areas with inflammable or explosive gases or vapor.
  • Do not operate the meter around explosive gas, vapor, or dust.
  • Use only type 4 AAA batteries, properly installed.
  • Disconnect test leads before shifting functions.
  • When servicing, use only specified replacement parts.
  • Replace batteries when the low battery indicator appears to avoid false readings and potential electric shock.

▲ Caution

  • To avoid possible damage to the meter or equipment under test:
  • Disconnect power and discharge high-voltage capacitors before testing resistance or continuity.
  • Use the proper jacks, functions, and ranges for measurement or source operation.

International Electrical Symbols:

Symbol Description
⏚ Earth Ground
▲ Warning Information

3. Familiarization with the Calibrator

Panel Introduction

Detailed panel introduction of the Victor 14+ Temperature Calibrator with numbered labels for each component.
Figure 3: Detailed Panel Introduction.

The calibrator features a clear LCD display, a numeric keypad, and dedicated function buttons for various measurements and sourcing tasks. The input/output terminals are clearly marked for easy connection.

Keys and Functions

No. Name Function
1~5Source value set keyIncrement of source set point
6~10Source value set keyDecrement of source set point
11MEASURE keyTurn on or off measurement function
12SOURCE keyTurn on or off source function
13Power buttonTurn on or off the power
14Backlight keyTurn on or off the backlight
15mV KeySelect measurement/source DC-voltage function
16RTD KeySelect measurement/source RTD function
17OHM KeySelect measurement/source OHM function
18Tc KeySelect measurement/source Tc function
19))) KeySelect measurement continuity function
20RANGE KeySelect measurement/source range
21RJ-ON KeyIn TC measurement/source function, turn on or off the RJ compensation function.
22T.DISP KeyIn TC/RTD measurement/source function, pressing the key converts the assistance display between room temperature or mV/Ω value; on other measurement/source functions, pressing the key turns on or off room temperature display in assistance display.
23HOLD KeyMeasured value holding
244/3/2W KeyIn ohm/RTD measurement function, select 2W, 3W or 4W method.
25AVG KeyMeasuring average value
26ZERO KeySet the source value to default value

Display Screen

The LCD display provides comprehensive information about the current operation. A typical display screen includes:

  • Measurement/Source indicators (e.g., DCV, TC, RTD, OHM)
  • Battery level indicator
  • Measurement/Source value and unit
  • Reference Junction Compensation (RJ) status
  • Beeper status for continuity
  • Hold and Average indicators
  • Room temperature or mV/Ω subsection value
Detailed view of the calibrator's LCD display and function buttons, showing various indicators and controls.
Figure 4: Typical LCD Display and Function Buttons.

4. Setup

Installing or Replacing the Batteries

The calibrator uses four AAA alkaline batteries. Always ensure the calibrator is turned off and disconnected from all leads and the charger before installing or replacing batteries.

  1. Remove the test leads and charger. Turn off the calibrator.
  2. Remove the protector (if attached).
  3. Using a standard blade screwdriver, turn each battery door screw a quarter counterclockwise to remove the battery door.
  4. Install four new AAA alkaline batteries, ensuring correct polarity as indicated on the holder.
  5. Reinstall and tighten the battery door, then reattach the protector.

Battery Level Indication:

The battery replacement indicator shows the battery level in five steps. Replace batteries when the low battery indicator appears (flashing icon) to avoid false readings.

Diagram showing the process of replacing batteries and fuses in the calibrator.
Figure 5: Replacing Batteries and Fuses.

Connecting the Charger

⚠ Warning

  • The charger is for this specified product only.
  • Ensure AC power voltage matches the charger's rated supply voltage.
  • Do not short-circuit the charger's output plug.
  • Do not charge non-Ni-Cd, non-Ni-MH batteries or wasted batteries.
  1. Turn off the calibrator.
  2. Connect the charger plug into the charging jack of the terminal.
  3. Plug the charger into the AC power outlet.

Note: The indicating light on the charger will be on during normal charging and turn dark when charging is complete. A blinking light indicates the charger is not connected or no battery is inside. Do not use the calibrator while charging, as this may prolong charging time.

5. Operation

5.1 Sourcing Functions

The calibrator can source DC voltage, resistance, thermocouple (TC), and RTD signals.

⚠ Warning

To avoid electrical shock, do not apply more than the rated voltage (as marked on the calibrator) between terminals or between any terminal and earth ground. Always use the calibrator in locations with a voltage to ground below 30 Vpk.

▲ Caution

The instrument is calibrated without accounting for voltage drop due to lead cable resistance. Exercise care when drawing load current. For precise resistance sourcing, use three-wire or four-wire connections.

Connecting Cables for Sourcing

For DC Voltage, Thermocouple (TC):

  1. Connect the black lead cable for source to the COM output terminal.
  2. Connect the red lead cable to the "mVTc" output terminal.
  3. Connect the other ends of the cables to the input of the equipment under test, ensuring correct polarities.
Wiring diagram for sourcing DC voltage and thermocouple signals.
Figure 6: Sourcing DC Voltage, TC Wiring.

For Resistance and RTD Signal:

  1. Connect black lead cables for source to the black terminal of Ω/RTD.
  2. Connect the red lead cable to the red terminal of Ω/RTD.
  3. Connect the other ends of the cables to the input of the equipment under test, ensuring correct polarities.
Wiring diagram for sourcing resistance and RTD signals.
Figure 7: Sourcing Resistances and RTD Wiring.

Sourcing DC Voltage:

  1. Press the (mV) key to select DC voltage source function.
  2. Press the (RANGE) key to select the desired range (100mV, 1000mV).
  3. Set the output value digit by digit using the (▲)/(▼) keys. Pressing (ZERO) initializes the output to 0.

Simulate Sourcing TC:

The calibrator uses an internal temperature sensor for reference junction compensation. The "RJ-ON" mark will display on the screen.

  1. Press the (TC) key to select simulate TC source function.
  2. Press the (RANGE) key to select the desired TC type (K, E, J, T, B, N, R, S, L, U).
  3. Set the output value digit by digit using the (▲)/(▼) keys. Pressing (ZERO) initializes the output to 0.
Diagram showing the simulation of 10 different thermocouple types with their respective temperature ranges.
Figure 8: Simulating 10 Thermocouples.

Sourcing RTD:

  1. Press the (RTD) key to select RTD function.
  2. Press the (RANGE) key to select the desired RTD type (PT100, PT200, PT500, PT1000, Cu10, Cu50).
  3. Set the output value digit by digit using the (▲)/(▼) keys. Pressing (ZERO) initializes the output to 0.
Diagram showing the simulation of 6 different thermal resistance types (RTD signals) with their respective temperature ranges.
Figure 9: Simulating 6 Thermal Resistance Types (RTD).

5.2 Measurement Functions

The calibrator can measure DC voltage, resistance, thermocouple (TC), RTD, and continuity.

⚠ Warning

When using the calibrator with supplied lead cables for measurement, the allowable voltage to ground of the input terminals is 30 Vpeak maximum. Do NOT use the calibrator at any voltage exceeding this maximum. If using a thermocouple convertor, the allowable voltage to ground is also 30V peak maximum.

Connecting Cables for Measurement

For DC Voltage Measurement:

  1. Connect the black lead cable to the "COM" input terminal.
  2. Connect the red lead cable to the "mVTc3W" input terminal.
  3. Connect the other end of the cable to the measuring terminals of the equipment under test, ensuring correct polarities.
Wiring diagram for measuring DC voltage.
Figure 10: Measuring DC Voltage Wiring.

For Thermocouple (TC) Signal Measurement:

  1. Connect the thermocouple convertor to the input terminals.
  2. Connect the positive output lead wire of the thermocouple to the H terminal of the convertor and the negative lead wire to the L terminal.
Wiring diagram for measuring thermocouple signals.
Figure 11: Measuring Thermocouple Wiring.

For Continuity, Ohm/RTD Signal Measurement (2-wire method):

  1. Connect one black lead cable to the "COM" input terminal.
  2. Connect the red lead cable to the "ΩRTD" input terminal.
  3. Connect the two clips of the cables to the measuring terminals of the equipment under test, ensuring correct polarities.
Wiring diagram for 2-wire continuity, ohm, or RTD measurement.
Figure 12: 2-wire Continuity, Ohm/RTD Measurement Wiring.

For Ohm/RTD Signal Measurement (3-wire method):

  1. Connect one black lead cable to the "COM" input terminal.
  2. Connect another black lead to the "mVTc3W" terminal.
  3. Connect the red lead cable to the "ΩRTD" input terminal.
  4. Connect the three clips of the cables to the measuring terminals of the equipment under test, ensuring correct polarities.
Wiring diagram for 3-wire ohm or RTD measurement.
Figure 13: 3-wire Ohm/RTD Measurement Wiring.

For Ohm/RTD Signal Measurement (4-wire method):

  1. Connect one black lead cable to the "COM" input terminal.
  2. Connect another black lead to the "mVTc3W" terminal.
  3. Connect one red lead cable to the "ΩRTD" input terminal.
  4. Connect another red lead to the "4W" terminal.
  5. Connect the four clips of the cables to the measuring terminals of the equipment under test, ensuring correct polarities.
Wiring diagram for 4-wire ohm or RTD measurement.
Figure 14: 4-wire Ohm/RTD Measurement Wiring.

Measuring DC Voltage:

  1. Ensure lead cables are not connected to the measuring instrument under test.
  2. Press the (mV) key to select DC Voltage measurement function.
  3. Connect the lead cables to the measuring terminals of the instrument under test.
  4. Press the (RANGE) key to select the desired range (50mV, 500mV).

Measuring Resistance:

  1. Ensure lead cables are not connected to the measuring instrument under test.
  2. Press the (OHM) key to select resistance measurement function.
  3. Connect the lead cables as shown in Figure 12, 13, or 14.
  4. Press the (RANGE) key to select the desired range (500Ω, 5KΩ).

Measuring Temperature with Thermocouple (TC):

  1. Ensure lead cables are not connected to the measuring instrument under test.
  2. Press the (TC) key to select TC measurement function.
  3. Press the (RANGE) key to select the desired range (K, E, J, T, B, N, R, S, L, U).
  4. Connect the thermocouple convertor to the jack under test as shown in Figure 11.

Measuring Temperature with RTD:

  1. Ensure lead cables are not connected to the measuring instrument under test.
  2. Press the (RTD) key to select RTD measurement function.
  3. Connect the lead cables as shown in Figure 12, 13, or 14.
  4. Press the (RANGE) key to select the desired range (PT100, PT200, PT500, PT1000, Cu10, Cu50).

Measuring Continuity:

Continuity measurement detects the intactness of a circuit (e.g., resistance lower than 50Ω).

  1. Press the ())) key to select continuity measurement function. The LCD will display the continuity symbol ")))".
  2. Connect the devices as shown in Figure 12. The beeper will sound continuously if the loop circuit resistance is less than 50Ω, and the LCD will show the measured resistance value.

Measurement-filtering function:

Pressing the (AVG) key in DCV, OHM, TC, or RTD function calculates the average of samples, stabilizing the displayed value. Pressing it again cancels the calculation.

Measured Value Holding Function:

Press the (HOLD) key to retain the current measured value on the LCD. Pressing it again releases the hold.

5.3 Environmental Temperature Test

The calibrator can measure the surrounding environmental temperature.

  1. After turning on the calibrator, press the (T.DISP) key.
  2. The LCD will display the room temperature value and unit in the assistance districts.
  3. Pressing (T.DISP) again cancels the room temperature display.

5.4 Factory Default Settings

You can reset the calibrator to factory defaults. When turning on the calibrator, press the (HOLD) key immediately to enter the default set mode.

Setting Auto Power-off Time:

  1. In default set mode, the LCD displays "AP.OFF".
  2. Use the (▲)/(▼) keys to set the time (0-60 minutes). 0 means no auto power-off.
  3. Press the (SOURCE) key to save.

Setting Backlight Time:

  1. In default set mode, press the (MEASURE) key. The LCD displays "BL.OFF".
  2. Use the (▲)/(▼) keys to set the time (0-3600 seconds). 0 means backlight stays on.
  3. Press the (SOURCE) key to save.

Setting Temperature Unit:

  1. In default set mode, press the (MEASURE) key. The LCD displays "TEM.U".
  2. Use the (▲)/(▼) keys to switch between °C and °F.
  3. Press the (SOURCE) key to save.

Setting Frequency:

  1. In default set mode, press the (MEASURE) key. The LCD displays "FRSET".
  2. Use the (▲)/(▼) keys to switch between 50Hz and 60Hz.
  3. Press the (SOURCE) key to save.

6. Calibration and Adjustment

Environmental Requirements for Adjustment:

  • Ambient temperature: 23 ± 2°C
  • Relative humidity: 35% to 75% RH
  • Warm-up: Place the meter in the standard environment for 24 hours, then power on and warm up for one hour in non-automatic power-off state.

▲ Caution

For adjustment, new alkaline AAA batteries (type 7) are recommended.

Adjusting Measurement Functions (DC Voltage, Ohms)

To enter measurement adjustment state: Turn on the meter, press (SOURCE) while holding (MEASURE). Then press (MEASURE) again. The LCD will show "CAL" and the reference value.

▲ Caution

If battery level is below 25%, adjustment cannot be performed, and "ERR" will be displayed.

Adjusting DC Voltage Measurement:

  1. Ensure lead cables are disconnected.
  2. Press the (mV) key to select DC voltage adjust function.
  3. Connect lead cables to the output terminals of a standard source (e.g., 5520A) as shown in Figure 15.
  4. Press the (RANGE) key to select the range.
  5. Press the (HOLD) key to enter measurement CAL mode. LCD shows "CA:0" and the reference voltage.
  6. Press the (AVG) key to save the adjusted value. LCD shows "SAVE".
  7. Repeat steps 4-6 for all ranges.
  8. Press the (HOLD) key to exit CAL mode.
Wiring diagram for calibrating DC voltage using a standard source.
Figure 15: Calibrating DC Voltage.

Adjusting Ohms Measurement:

  1. Ensure lead cables are disconnected.
  2. Press the (OHM) key to select ohm adjust function.
  3. Connect lead cables to the output terminals of a standard source as shown in Figure 16 (2-wire compensation of 5520A must be open).
  4. Press the (RANGE) key to select the range.
  5. Press the (HOLD) key to enter ohm CAL mode. LCD shows "CAL:0" and the reference resistance.
  6. Press the (AVG) key to save the adjusted value. LCD shows "SAVE".
  7. Press the (RJ-ON) key to shift between CAL:0 and CAL:F.
  8. Repeat steps 4-7 for all ranges.
  9. Press the (HOLD) key to exit CAL mode.
Wiring diagram for calibrating all ranges of ohms using a standard source.
Figure 16: Calibrating All Ranges of Ohms.

Adjusting Source Functions (DC Voltage, Resistance)

To enter source calibration state: Turn on the meter, press (SOURCE) while holding (MEASURE). The LCD shows "CA-0" or "CA-F" and the highest five digits of the sourced value.

▲ Caution

If battery level is below 25%, adjustment cannot be performed, and "ERR" will be displayed.

Adjusting Voltage Source:

  1. Press the (mV) key to select DC voltage function.
  2. Connect lead cables to a standard digital meter as shown in Figure 17.
  3. Press the (RANGE) key to select the range.
  4. Read the calibrator output on the standard meter. Use (▲)/(▼) keys to adjust the reading to match the CAL adjustment setpoint.
  5. Press the (ZERO) key to save the CAL adjustment reading.
  6. Press the (4/3/2W) key to shift to the next setpoint.
  7. Repeat steps 3-6 to adjust all points for the range.
  8. Repeat steps 2-7 to adjust all ranges.
Wiring diagram for adjusting voltage source using a digital meter.
Figure 17: Adjusting Voltage Source.

Adjusting Resistance Source:

  1. Press the (OHM) key to select resistance function.
  2. Connect lead cables to a standard digital meter as shown in Figure 18.
  3. Press the (RANGE) key to select the range.
  4. Read the calibrator output on the standard meter. Use (▲)/(▼) keys to adjust the reading to match the CAL adjustment setpoint.
  5. Press the (ZERO) key to save the CAL adjustment reading.
  6. Press the (4/3/2W) key to shift to the next setpoint.
  7. Repeat steps 3-6 to adjust all points for the range.
  8. Repeat steps 2-7 to adjust all ranges.
Wiring diagram for adjusting resistance source using a digital meter.
Figure 18: Adjusting Resistance Source.

7. Maintenance

7.1 Cleaning the Calibrator

To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers. Clean the calibrator with a soft cloth dampened with water and mild soap.

7.2 Replacing Batteries or Fuse

⚠ Warning

To avoid electric shock, remove test leads before opening the battery door. Ensure the battery door is tightly closed before turning on the calibrator.

▲ Caution

  • To avoid fluid leakage or battery explosion, ensure correct battery polarity.
  • Do not short-circuit, disassemble, or heat batteries, or throw them into fire.
  • When replacing, use four new batteries of the same specified type.
  • Remove batteries if the meter will not be used for a long time.

Refer to the battery replacement steps in the Setup section (Section 4) and Figure 5 for visual guidance.

To replace a blown fuse, use the same type: F1 (50mA/250V).

8. Troubleshooting

No specific troubleshooting information is available in the provided documentation. If the meter operates abnormally, first inspect the batteries and replace them if necessary. If issues persist, contact the vendor or service center.

9. User Tips

  • For stable readings, utilize the measurement wave-filter function (AVG key).
  • When performing precise measurements or sourcing, ensure the calibrator has warmed up for at least one hour in the specified environmental conditions.
  • Always double-check wiring connections, especially for 3-wire and 4-wire resistance measurements, to ensure accuracy.
  • The manual-holding function (HOLD key) is useful for recording intermittent readings.
  • If the calibrator is exposed to sudden temperature changes, allow it to stabilize for at least one hour before use to prevent condensation and ensure accuracy.

10. Specifications

General Specifications for Measurement:

  • Calibration Cycle: 1 year
  • Operating Temperature: 18°C to 28°C
  • Relative Humidity: 35% to 70% (non-condensing)
  • Accuracy: ± (percentage of reading + percentage of range)
  • Common Mode Rejection Ratio (CMRR): ≥120dB (at 50Hz or 60Hz)
  • Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C for 5°C to 18°C and 28°C to 50°C
  • Internal Temperature Compensation Sensor Range: -0°C to 50°C, compensation error ≤ ±0.5°C
  • Temperature Probe Accuracy: ±0.2°C, measured temperature range: -20°C to 100°C
  • Maximum voltage between VΩ Hz terminal and COM terminal: 30 Vpk

Measurement Technical Indicators:

Measurement Range Measuring Range Resolution Accuracy
DCV50mV-5.000mV~55.000mV1μV0.02+0.02
500mV-50.00mV~550.00mV10μV0.02+0.01
OHM500Ω0.00Ω~550.00Ω0.01Ω0.05+0.02
5KΩ0.0000 KΩ~5.5000KΩ0.1Ω0.05+0.02
TCR0°C~1767°C1°C0~500°C: 1.8°C
500~1767°C: 1.5°C
S0°C~1767°C
K-100.0°C~1372.0°C0.1°C-100.0~0.0°C: 1.2°C
0.0~1372.0°C: 0.8°C
E-50.0°C~850.0°C-50.0~0.0°C: 0.9°C
0.0~850.0°C: 1.5°C
J-60.0°C~1120.0°C-60.0~0.0°C: 1.0°C
0.0~1120.0°C: 0.7°C
T-100.0°C~400.0°C-100.0~0.0°C: 1.0°C
0.0~400.0°C: 0.7°C
N-200.0°C~1300.0°C-200.0~0.0°C: 1.5°C
0.0~1300.0°C: 0.9°C
B600°C~1820°C600~800°C: 2.2°C
800~1000°C: 1.8°C
1000~1820°C: 1.4°C
L-60.0°C~900.0°C-60.0~0.0°C: 0.7°C
0.0~900.0°C: 0.5°C
U-100.0°C~600.0°C-100.0~0.0°C: 0.7°C
0.0~600.0°C: 0.5°C
RTDPt100(385)-200.0°C~800.0°C0.1°C-200.0~0.0°C: 0.5°C
0.0~400.0°C: 0.7°C
400.0~800.0°C: 0.8°C
Pt200(385)-200.0°C~630.0°C-200.0~0.0°C: 0.8°C
0.0~300.0°C: 0.9°C
300.0~630.0°C: 1.0°C
Pt500(385)-200.0°C~630.0°C-200.0~0.0°C: 0.4°C
0.0~300.0°C: 0.5°C
300.0~630.0°C: 0.7°C
Pt1000(385)-200.0°C~630.0°C-200.0~100.0°C: 0.2°C
100.0~300.0°C: 0.5°C
300.0~630.0°C: 0.7°C
Cu10-100.0°C~260.0°C1.8°C
Cu50-50.0°C~150.0°C0.7°C
Continuity500Ω≤50Ω sound0.01Ω

Output Basic Technical Indicators:

  • Temperature coefficient: 0.1 × specified accuracy / °C (temperature range 5 to 18 °C or 28 to 50 °C)
  • Internal temperature compensation sensor RJC, temperature measurement range 0~50C, compensation error ±0.5°C
  • Temperature probe accuracy: ±0.2°C, probe temperature range: -20C~100°C
  • Maximum voltage applied between output terminals and ground: 30Vpk

General Specifications for Source:

  • Calibration Cycle: 1 year
  • Operating Temperature: 18°C to 28°C (64.4°F~82.4°F)
  • Relative Humidity: 35% to 70% (non-condensing)
  • Accuracy: ± (percentage of set value + percentage of range)

Source Technical Indicators:

Function Reference Range Resolution Accuracy Remark
DC voltage100mV-10.000mV~110.000mV1μV0.02+0.01Maximum output current: 0.5mA
1000mV-100mV~1100mV10μV0.02+0.01Maximum output current: 2mA
Resistance400Ω0.00Ω~400.00Ω0.01Ω0.02+0.02Excitation current: ±0.5-3mA; if ±0.1-0.5, add 0.1Ω; Accuracy does not include lead resistance;
4KΩ0.0000 KΩ~4.0000 KΩ0.1Ω0.05+0.025Excitation current: ±0.05-0.3mA; Does not include lead resistance;
TCR0°C~1767°C1°C0~100°C: 1.5°C
100~1767°C: 1.2°C
By using ITS-90 temperature scale; The accuracy does not include the error of internal temperature compensation caused by a sensor;
S0°C~1767°C
K-200.0°C~1372.0°C0.1°C-200.0~-100.0°C: 0.6°C
-100.0~400.0°C: 0.5°C
400.0~1200.0°C: 0.7°C
1200.0~1372.0°C: 0.9°C
E-200.0°C~1000.0°C-200.0~-100.0°C: 0.6°C
-100.0~600.0°C: 0.5°C
600.0~1000.0°C: 0.4°C
J-200.0°C~1200.0°C-200.0~-100.0°C: 0.6°C
-100.0~800.0°C: 0.5°C
800.0~1200.0°C: 0.7°C
T-250.0°C~400.0°C-250.0~400.0°C: 0.6°C
N-200.0°C~1300.0°C-200.0~-100.0°C: 1.0°C
-100.0~900.0°C: 0.7°C
900.0~1300.0°C: 0.8°C
B600°C~1820°C600~800°C: 1.5°C
800~1820°C: 1.1°C
L-60.0°C~900.0°C-60.0~0.0°C: 0.7°C
0.0~900.0°C: 0.5°C
U-100.0°C~600.0°C-100.0~0.0°C: 0.7°C
0.0~600.0°C: 0.5°C
RTDPt100(385)-200.0°C~800.0°C0.1°C-200.0~0.0°C: 0.3°C
0.0~400.0°C: 0.5°C
400.0~850.0°C: 0.8°C
By using ITS-90 temperature scale. Excitation current: ±0.5~±3mA for Pt100, Cu10, Cu50 add 0.5°C when excitation current is ±0.1mA-0.5mA;
Pt200(385)-200.0°C~630.0°C-200.0~100.0°C: 0.8°C
100.0~300.0°C: 0.9°C
300.0~630.0°C: 1.0°C
Pt500(385)-200.0°C~630.0°C-200.0~100.0°C: 0.4°C
100.0~300.0°C: 0.5°C
300.0~630.0°C: 0.7°C
Excitation current: ±0.05mA~±0.3mA for PT200, PT500, PT1000;
Pt1000(385)-200.0°C~630.0°C-200.0~100.0°C: 0.2°C
100.0~300.0°C: 0.5°C
300.0~630.0°C: 0.7°C
Does not include lead resistance.
Cu10-100.0°C~260.0°C1.8°C
Cu50-50.0°C~150.0°C0.7°C

11. Warranty and Support

Contact Us

To purchase parts, obtain operation help, or find the vendor or service center nearest to you, please contact the vendor from which you purchased the product.

If the meter operates abnormally, first inspect the batteries and replace them if necessary. If the meter still fails to operate properly, pack it securely (in its original container if available) and forward it, postage paid, to the nearest Service Center. The company assumes no responsibility for damage in transit.

The company guarantees rapid repair and maintenance. If the warranty is due, you will be billed for the maintenance and repair work. If the calibrator is not within the warranty range, you can contact the warranted service center for enquiring about the expenditure.

PDF User Manual: A detailed user manual in PDF format is available for download. View User Manual (PDF)

12. Operation Demo Video

Watch this video for a demonstration of the calibrator's key functions, including resistance testing, thermocouple testing, RTD testing, output resistance testing, and output DC voltage testing.

Video: Operation Demo of Victor 14+ Temperature Calibrator.

© 2023 RuoShui. All rights reserved.