1. General Safety Summary
Review the following safety precautions carefully before operating the device to avoid any personal injuries or damages to the device and any products connected to it. Use the device as specified by this user's guide only.
To Avoid Fire or Personal Injury:
- Use Proper Power Cord: Use only the power cord specified for this product and certified for the country of use.
- Connect and Disconnect Properly: Do not connect or disconnect probes or test leads while they are connected to a voltage source. Connect the probe output to the measurement device before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement device.
- Observe All Terminal Ratings: To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
- Use Proper Probe: To avoid shock hazard, use a properly rated probe for your measurement.
- Avoid Circuit or Wire Exposure: Do not touch exposed connections and components when power is on.
- Do Not Operate With Suspected Failures: If suspected damage occurs with the device, have it inspected by qualified service personnel before further operations.
- Provide Proper Ventilation: Refer to the installation instructions for proper ventilation of the device.
- Do not operate in Wet/Damp Conditions.
- Do not operate in an Explosive Atmosphere.
- Keep Product Surfaces Clean and Dry.
2. Product Overview and Features
The Hantek 6022BE/BL is a compact, lightweight, and portable digital oscilloscope that requires no external power source, operating via USB. It is designed for production testing, research and development, analog circuit testing, troubleshooting, and educational purposes.
Key Features:
- Model 6022BE: Cost-effective, 20MHz bandwidth, 2 channels, 48MSa/s sample rate.
- Model 6022BL: Includes all 6022BE features plus a 16-channel Logic Analyzer.
- Design: Anodized aluminum casing (similar to iPad) for durability, heat resistance, and wearability.
- Interface: Standard USBXI™ interface for easy integration into combination instruments. USB2.0 for power and data.
- Portability: Compact dimensions (205mm L x 120mm W x 35mm H) for easy transport.
- Performance: High performance with 48MSa/s real-time sampling and 20MHz bandwidth.
- Operating System Compatibility: Windows 7, Windows NT, Windows 2000, Windows XP, VISTA.
- Measurement Functions: 23 automatic measurement functions, PASS/FAIL Check.
- Waveform Analysis: Supports waveform average, persistence, intensity, invert, addition, subtraction, multiplication, division, and X-Y plot.
- Data Saving: Save waveforms as text files, JPG/BMP graphic files, or MS Excel/Word files.
- FFT: Fast Fourier Transform function included.
- Expandability: Multiple DSOs can be connected to one computer to extend channel count.
- Second Design Support: Supply DEMO code (VC, VB, LABVIEW) and technical support.
Package Contents:

The package typically includes:
- Hantek 6022BE or 6022BL Digital Oscilloscope unit
- Two (2) Passive Probes (1:1, 10:1)
- USB Cable
- Software CD (containing drivers and application software)
- User Manual (PDF version available for download)
3. System Requirements
To run the oscilloscope software, your computer must meet the following specifications:
Minimum Requirements:
- Operating System: Windows XP/Vista/Win7/Win8/Win 10
- Processor: Upwards of 1.00GHz processor
- Memory: 128MB RAM
- Disk Space: 500MB free space
- Screen Resolution: 800 x 600
Recommended Configuration:
- Operating System: Windows XP SP2 System
- Processor: 2.4GHz Processor
- Memory: 512MB RAM
- Disk Space: 80GB
- Screen Resolution: 1024 x 768 or 1280 x 1024
- DPI Setting: Normal Size (96DPI)
4. Setup and Installation
4.1. Hardware Connection

- Connect USB Cable: Connect the A-Type plug of the USB cable to your PC's USB port and the B-Type plug to the oscilloscope's USB port. The USB cable is provided with the Hantek unit.
- Connect Probes: Connect the two oscilloscope probes to the CH1 and CH2 BNC connectors on the device.
- Logic Analyzer Connection (6022BL only): If you have the 6022BL model, connect the 16-channel logic analyzer pins and 4 ground pins to the dedicated logic analyzer port.

4.2. Software and Driver Installation
Important: Install the software before using the oscilloscope.
- Insert Software CD: Insert the installation CD into your computer's CD-ROM drive. The installation should start automatically. If not, navigate to the CD-ROM drive in Windows Explorer and run
Setup.exe. - Follow Installation Wizard: Follow the on-screen instructions of the Hantek6022BE/BL Setup program. Click 'Next' to proceed through the steps.
- Driver Installation: After software installation, connect the oscilloscope via USB. Your system should detect new hardware.
- Update Driver: If the device appears as 'Unknown Device' in Device Manager:
- Right-click the 'Unknown Device'.
- Select 'Update Driver Software'.
- Choose 'Browse my computer for driver software'.
- Select 'Let me pick from a list of device drivers on my computer'.
- Browse to the 'Hantek software in CD/Driver' folder (or the location where you copied the CD contents).
- Select the appropriate driver (e.g., Hantek6022BL DRIVER 1 or DRIVER 2).
- Click 'Next' and confirm any Windows security warnings to install the driver.
- Verify Installation: Once the driver is successfully updated, the device should appear correctly in Device Manager (e.g., under 'Hantek6022BL DRIVER 1'). A blinking red LED on the oscilloscope indicates successful recognition by the computer.
5. Operating Basics
5.1. User Interface Overview
After launching the Hantek software, the main user interface will be displayed. It consists of:
- Main Menu: Contains all settings (File, View, Horizontal, Vertical, Trigger, Display, Cursor, Utility, Help).
- Toolbar: Quick access icons for common functions.
- Control Panel: Adjusts horizontal and vertical settings, trigger mode, and source.
- Measure Window: Displays measurement results (e.g., Maximum, Minimum, Peak to Peak, Frequency, Period).
- Output Window: Shows current output status.
5.2. Probe Compensation
This function matches the characteristics of the probe to the channel input and should be performed when attaching a probe to any input channel for the first time.
- Set the attenuation switch on the probe to 10X.
- Connect the probe to CH1 of the oscilloscope.
- Attach the probe tip to the Probe Compensator (CAL output) and the reference lead to the ground connector.
- Select CH1 and press the "AUTOSET" button in the menu or toolbar.
- Check the shape of the displayed waveform. It should be a clean square wave.
- If the waveform is not a perfect square wave, use a non-metallic tool to adjust the trimmer capacitor on the probe until a flat-top square wave is displayed.
5.3. Self-Calibration
The self-calibration routine optimizes the oscilloscope signal path for maximum measurement accuracy. Run this routine if the ambient temperature changes by 5°C or more, or if you suspect accuracy issues.
- Power on the oscilloscope and allow it to warm up for 20 minutes.
- Disconnect any probes or cables from the input connectors.
- Go to "Menu -> Utility -> Calibration" and follow the prompts.
- For zero calibration, connect both channel terminals (CH1 and CH2) to the ground pin on the oscilloscope and wait for some time before calibrating.
5.4. Measuring Audio Output (Example)
This example demonstrates how to measure audio output from a PC using the oscilloscope.
- Prepare Audio Jack Adapter: Obtain a 3.5mm audio jack. Cut female-to-female jumper wires into halves. Solder the wires to the audio jack to create a custom adapter for connecting to the oscilloscope probes.
- Connect to PC Audio Out: Plug the 3.5mm audio jack into the PC's audio output port.
- Connect to Oscilloscope: Connect the jumper wires from the audio jack to the oscilloscope probes (CH1 and CH2). Note: The ground of the Hantek DSO is common with the PC's audio ground via the USB connection, so you may not need to connect a separate ground wire from the audio jack to the probe ground.
- Launch Hantek Software: Open the Hantek oscilloscope software.
- Play Audio: Play an audio file on your PC.
- Observe Waveform: Adjust the vertical and horizontal settings in the software to view the audio waveform. You should see a dancing waveform if the trigger is not set correctly.
- Set Trigger: To stabilize the waveform, adjust the trigger source (e.g., to CH1 or CH2) and trigger mode (e.g., Edge) in the Control Panel.
6. Specifications
| Attribute | Specification |
|---|---|
| Brand Name | Hantek |
| Model Number | 6022BE, 6022BL |
| Digital Channels | 2 |
| Band Width | 20MHz |
| Real Time Sampling Rate | 48MSa/s |
| Max Waveform Capture Rate | 30,000wfm/s |
| Record Length | 1M/CH |
| Vertical Resolution | 8Bit |
| Input Impendence | 1MΩ 25pF |
| DC Accuracy | ±3% |
| Timebase Range | 1ns/div~5000s/div |
| Trigger Mode | Auto, Normal, Single |
| Trigger Source | CH1, CH2 |
| Logic Analyzer (6022BL only) | 16 Channels, 10MHz Bandwidth |
| Power | Bus-powered from USB |
| Dimensions | 200mm x 100mm x 35mm |
| Weight | 0.3KG |
| Certification | CE |
7. Maintenance
7.1. General Care
- Do not store or leave the oscilloscope where the device will be exposed to direct sunlight for long periods of time.
- To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents.
- To avoid damages to the surface of the device or probes, do not use any abrasive or chemical cleaning agents.
7.2. Cleaning
Inspect the device and probes as often as operating conditions require. Make sure the device is disconnected from all power sources before cleaning.
- Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth. Use care to avoid scratching the clear glass display filter.
- Use a soft cloth dampened with water to clean the device.
8. Troubleshooting
8.1. Driver Not Recognized
If your Hantek oscilloscope is not recognized by your computer after connecting, or appears as an 'Unknown Device' in Device Manager:
- Ensure the Hantek software has been installed from the provided CD or downloaded from the official Hantek website.
- Manually update the driver through Device Manager, pointing to the 'Driver' folder within the installed Hantek software directory.
- Try unplugging and replugging the USB cable to force a re-detection.
8.2. Unstable or 'Dancing' Waveforms
If your waveforms are not stable or appear to be 'dancing' on the screen:
- Check Trigger Settings: The most common cause is an incorrectly set trigger. Adjust the 'Trigger Source' to the channel you are observing (CH1 or CH2).
- Adjust Trigger Mode: Set the 'Trigger Mode' to 'Edge' and 'Trigger Sweep' to 'Auto'.
- Adjust Trigger Level: Drag the trigger level indicator on the screen to a stable point on your waveform.
8.3. Incorrect Waveform Shape During Probe Compensation
If the square wave displayed during probe compensation is not ideal (e.g., rounded corners, overshoot):
- Use the provided non-metallic screwdriver to adjust the trimmer capacitor on the probe.
- Turn the screw gently until the waveform appears as a clean, flat-top square wave.
9. User Tips
- DIY Audio Jack Adapter: For measuring audio signals from a PC, consider creating a custom adapter by soldering female-to-female jumper wires to a 3.5mm audio jack. This allows for easy connection to the oscilloscope probes.
- Grounding: When connecting to a PC via USB, the oscilloscope's ground is often common with the PC's audio ground. This means you might not need to connect a separate ground wire from your audio source to the probe ground, simplifying connections.
- Software Availability: If you lose your software CD, check the official Hantek website for downloads.
10. Warranty and Support
For detailed warranty information, technical support, and further documentation, please refer to the official Hantek website or the comprehensive user manual PDF available at: User manual (PDF).