Introduction
The Yieryi Bluetooth Water Quality Tester is a multi-functional device designed for precise measurement of various water parameters including pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), and Temperature (TEMP). It features intelligent control via a mobile application, automatic temperature compensation (ATC), and cloud-based data management, making it suitable for a wide range of applications such as drinking water analysis, aquariums, fish tanks, swimming pools, farmland irrigation, and laboratory use.
Package Contents
Please check the box for the following items:
- 1 x Yieryi Bluetooth Water Quality Pen (Specific model: BLE-9908, BLE-9902, or BLE-9901)
- 1 x English Instruction Manual
- 5 x pH Calibration Powders (pH 4.00, 6.86, 7.00, 9.18, 10.01)

Product Overview
The meter features a backlit LCD display for clear readings. Control buttons include HOLD/TEMP (for data hold and temperature unit conversion), MODE/CAL (for function conversion and calibration), and ON/OFF. It is powered by 3*1.5V button batteries (not included). The device is equipped with a high-precision, highly sensitive glass electrode and an anti-rust platinum electrode, which are removable for maintenance. The device has a waterproof grade of IP67.

Setup
Battery Installation
- Unscrew the top cap of the meter.
- Insert 3*1.5V button batteries into the battery compartment, ensuring correct polarity.
- Replace the cap securely.
App Installation
- Scan the appropriate QR code (Android or iOS) below to download the mobile application.
- Install the application on your smartphone.

Bluetooth Connection
- Turn on the water quality meter. Initially, the Bluetooth logo may not be visible.
- Press and hold the ON/OFF button for 5 seconds until the Bluetooth logo appears on the meter's display. This indicates Bluetooth is active.
- On your smartphone, enable Bluetooth and open the installed water quality app.
- The app should automatically detect and connect to the meter. If prompted, confirm the pairing.
- To turn off Bluetooth, press and hold the ON/OFF button for 5 seconds again.
Operation
Basic Measurement
- Ensure the meter is clean and properly calibrated.
- Immerse the electrode head into the water sample you wish to test.
- The readings (PH, EC, TDS, TEMP) will be displayed on the meter's screen and simultaneously synchronized to the mobile app.
- Press the "HOLD/TEMP" button to hold the current reading on the meter. Press again to release.
- To switch temperature units (Celsius/Fahrenheit), press and hold the "HOLD/TEMP" button.

App Features
- Data Synchronization: Real-time data from the meter is displayed on the mobile app, allowing for convenient viewing of multiple parameters simultaneously.
- Data Saving & Sharing: Data can be saved automatically (with a configurable time interval) or manually within the app. Saved data can be exported to Excel and shared. Notes can be added to manual saves.
- Parameter View: The app provides a comprehensive view of all measured parameters.
- Setting Limits: You can set maximum and minimum limits for pH and EC values in the app. If a measurement exceeds these limits, the value will be displayed in red on the app interface.
- Backlight Control: The app allows you to turn the meter's screen backlight on or off.



Calibration Procedure (pH)
- Prepare the pH buffer solutions (e.g., 4.00, 6.86, 9.18 for Asia standard or 4.00, 7.00, 10.01 for EU standard).
- Clean the electrode with distilled water and dry it gently.
- Immerse the electrode into the first buffer solution (e.g., pH 6.86).
- In the app, navigate to the calibration section. Select the appropriate standard (Asia or EU).
- Follow the on-screen instructions to perform the 5-point calibration. The app will guide you through immersing the electrode in each buffer solution.
- The app will record the calibration data.

Maintenance
Electrode Care
The highly sensitive glass and platinum electrodes are removable. After each use, rinse the electrode thoroughly with distilled water to prevent contamination. Do not wipe the glass bulb, as this can damage it.

Storage
Always replace the protective cap on the electrode when not in use to keep it moist. Store the meter in a cool, dry place. If the electrode becomes dry, soak it in pH 4.00 buffer solution for at least 30 minutes before use.
Troubleshooting
- Cannot connect to Bluetooth:
- Ensure the meter's Bluetooth is activated (press and hold ON/OFF for 5 seconds until the Bluetooth icon appears).
- Check your phone's Bluetooth settings and ensure it's enabled and visible.
- Restart both the meter and your phone.
- Ensure the app has necessary Bluetooth permissions.
- Inaccurate readings:
- Perform a full 5-point calibration using fresh buffer solutions.
- Ensure the electrode is clean and not damaged.
- Check battery levels; low battery can affect accuracy.
- Ensure the automatic temperature compensation (ATC) is functioning correctly.
- Readings outside set limits are not red:
- Verify that the maximum and minimum limits are correctly set in the app's function settings.
- Ensure the app is updated to the latest version.
Specifications
| Feature | Description |
|---|---|
| Models | BLE-9908 (PH/EC/TDS/TEMP), BLE-9902 (PH/EC/TEMP), BLE-9901 (PH/TDS/TEMP) |
| Temperature Range | 0.1-60.0°C (32.1-140.0°F) |
| pH Range | 0.00-14.00 |
| TDS Range (BLE-9908/BLE-9901) | 0-9999ppm |
| EC Range (BLE-9908/BLE-9902) | 0-9999us/cm; 10.01-19.99mS/cm |
| Automatic Temperature Compensation (ATC) | 0-60.0°C (32.0-140.0°F) |
| Calibration Points | Two sets optional: 4.0+6.86+9.18 (Asia Standard) / 4.0+7.00+10.00 (EU Standard) |
| Work Environment | 0.1℃-80.0℃; RH: max 90% |
| Waterproof Grade | IP67 |
| Power Supply | 3*1.5V button battery (not included), low battery warning |
| Dimensions | 187mm x 37mm (7.36 x 1.46 x 1.46 inches) |
| Weight | 95g |
| Smart Device | Yes |
| Backlight | Yes |
| Display Type | Digital |
| Certification | CE |
User Tips
- Electrode Replacement: While the electrodes are designed for durability, they may need replacement over time. Contact the seller or manufacturer for information on replacement electrodes and their cost.
- App Compatibility: Ensure your smartphone's operating system (Android or iOS) is compatible with the latest version of the Yieryi app. If experiencing connection issues with specific phone models, try updating the app or checking for known compatibility issues with the manufacturer.
- Calibration Powder: Always use fresh pH buffer powders for calibration to ensure the most accurate readings.
Warranty and Support
For warranty information, technical support, or inquiries regarding replacement parts, please contact the manufacturer or your point of purchase. Keep your purchase receipt as proof of purchase.