OWON HDS2102S

OWON HDS2102S Digital Oscilloscope Instruction Manual

Model: HDS2102S

Brand: OWON

1. Introduction

The OWON HDS2102S is a versatile 3-in-1 handheld device combining the functionalities of a digital oscilloscope, a multimeter, and a waveform generator. Designed for portability and ease of use, it is ideal for outdoor maintenance, rapid on-site measurements, automobile maintenance, and power detection. It features a 3.5-inch high-resolution, high-contrast color LCD display for clear data visualization.

Video: Overview of the OWON HDS200 series, highlighting its 3-in-1 functionality, portability, and key features.

2. Key Features

  • Multifunctionality: Integrates Oscilloscope, Multimeter, and Waveform Generator (HDS2102S model).
  • Display: 3.5-inch high-resolution, high-contrast color LCD, optimized for outdoor visibility.
  • Power Management: Powered by an 18650 lithium battery, offering approximately 6 hours of continuous operation with a low power consumption of ≤3W.
  • Connectivity: USB Type-C interface supports charging via power bank and PC software connection for data transfer and control.
  • Durability: Integrated rubber covering design for a comfortable grip and enhanced durability.
  • Calibration: Self-calibration function ensures measurement accuracy.
  • Programmability: SCPI (Standard Commands for Programmable Instruments) support facilitates secondary development and automation.
OWON HDS2102S with hangable bracket and oscilloscope performance examples

Image: Device with a hangable bracket for convenient placement and examples of oscilloscope waveform displays.

3. Package Contents

Ensure all items listed below are present in your package:

  • 1 x Main Device (OWON HDS2102S Digital Oscilloscope)
  • 1 x Carrying Bag
  • 1 x Oscilloscope Probe
  • 1 x Power Plug (EU type)
  • 1 x Probe Adjust Tool
  • 1 x USB Cable (Type-C)
  • 2 x Multimeter Leads (Red and Black)
  • 1 x Quick Guide (Instructions Manual)
  • 1 x BNC to Alligator Clip Cable
Contents of the OWON HDS2102S package including the device, bag, probes, cables, and manual.

Image: Visual representation of all included accessories and the main device.

4. Device Overview

4.1 Front Panel

The front panel features the 3.5-inch LCD display, function buttons (F1-F4), mode selection, channel selection (CH1/2), horizontal control (HOR), system settings, save function, measure range, auto-set, trigger controls (Trig/Δ), and play/pause button. The multimeter input connectors are located at the bottom.

Front view of the OWON HDS2102S showing the display, control buttons, and multimeter inputs.

Image: Front view of the OWON HDS2102S handheld oscilloscope.

4.2 Top Panel

The top panel includes the BNC connectors for oscilloscope channels (CH1, CH2) and the waveform generator output (GEN OUT).

Top view of the OWON HDS2102S showing BNC connectors for CH1, CH2, and GEN OUT.

Image: Top view of the device with BNC connectors.

4.3 Side Panel

The side panel houses the USB Type-C interface for charging and communication, and the oscilloscope probe calibration signal output (1kHz square wave).

Side view of the OWON HDS2102S showing the USB Type-C port and calibration signal output.

Image: Side view of the device with USB-C port and calibration signal.

4.4 Rear Panel

The rear panel features the battery compartment and a hangable bracket for convenient positioning.

Rear view of the OWON HDS2102S showing the battery compartment and kickstand.

Image: Rear view of the device with battery cover open.

4.5 Dimensions

Approx. 198mm (Height) x 96mm (Width) x 38mm (Depth). Weight is approximately 0.6 kg (main device, without battery).

Diagram showing the dimensions of the OWON HDS2102S: 198mm height, 96mm width, 38mm depth.

Image: Device dimensions and key interface locations.

5. Setup

5.1 Battery Installation and Charging

The device is powered by an 18650 lithium battery. To install or replace the battery, remove the screw on the rear cover to access the battery compartment. The device can be charged via the USB Type-C interface using the provided power plug or an external power bank (5V/2A).

Diagram showing battery installation and USB-C charging options for the OWON HDS2102S.

Image: Battery compartment and USB Type-C charging port.

5.2 Probe and Lead Connection

Connect the oscilloscope probes to the CH1 and CH2 BNC inputs on the top panel. Use the multimeter leads with the VΩ+C, COM, and mA/A inputs on the front panel for multimeter functions. For waveform generation, connect to the GEN OUT BNC connector.

5.3 Power On/Off

Press and hold the power button located on the front panel to turn the device on or off.

6. Operating Modes

The OWON HDS2102S integrates three main functions: Oscilloscope, Multimeter, and Waveform Generator. You can switch between these functions with just one key.

Diagram illustrating the three main functions: Oscilloscope, Multimeter, and Waveform Generator, with a 'One key switch' indicator.

Image: Instrument Transformers - Function overview.

6.1 Oscilloscope Function

The oscilloscope mode allows for waveform observation and analysis. It features a 100MHz bandwidth and a real-time sampling rate of 500MSa/s. Key features include:

  • Waveform Capture Rate: Max 10,000 wfrms/s.
  • Record Length: Max 8K.
  • Measurement Functions: Cursor measurement and multiple auto measurement functions (Frequency, Period, Amplitude, Max, Min, Mean, PK-PK).
  • Acquisition Modes: Peak detect for jamming burr detection, high refresh, and deep storage for capturing exceptional and low probability events.
  • XY Function: For phase comparison.
  • Auto Set Function: One-key auto set for easy detection and debugging.
  • Save Function: Save up to 4 settings, 4 reference waveforms, and 4 CSV waveform files. Connect to a computer via USB Type-C to recognize the oscilloscope as a USB flash drive for easy access to saved data.
Screenshots of oscilloscope auto measurement, cursor measurement, and save function interfaces.

Image: Oscilloscope functions including auto measurement, cursor measurement, and save options.

6.2 Multimeter Function

The integrated multimeter provides 20,000 counts with true RMS measurement. It supports auto-range functionality for ease of use. Measurements include:

  • Voltage (DC 1000V input voltage)
  • Current
  • Resistance
  • Diode
  • Capacitance
  • Continuity Test

It features a reading hold function and relative measurement function.

Image showing the multimeter function measuring DC voltage, resistance, and capacitance.

Image: Multimeter function demonstrating voltage, resistance, and capacitance measurements.

6.3 Waveform Generator Function (HDS2102S only)

The HDS2102S model includes a waveform generator with a maximum output frequency of 25MHz. It offers a 14-bit vertical resolution and 125MSa/s high sampling rate. Available waveform types include:

  • Sine (0.1Hz - 25MHz)
  • Square (0.1Hz - 5MHz)
  • Ramp (0.1Hz - 1MHz)
  • Pulse (0.1Hz - 5MHz)
  • Arbitrary (0.1Hz - 5MHz) - including Sinc, Bessel, StairUp, StairUpDown, StairDown, AttAlt, AmpAlt.
Screenshots of the waveform generator interface showing sine, square, ramp, pulse, and arbitrary waveforms.

Image: Waveform Generator functions and output types.

7. Specifications

7.1 Oscilloscope Specifications

ParameterHDS2102HDS2102S
Bandwidth100MHz
Channels2-CH Oscilloscope2-CH Oscilloscope + 1-CH Generator
Sample Rate500MSa/s
Acquisition ModelNormal, Peak detect
Record Length8K
Display3.5-inch LCD
Waveform Refresh Rate10,000 wfrms/s
Input CouplingDC, AC, and Ground
Input Impedance1 MΩ±2%, in parallel with 16pF±10pF
Probe Attenuation Factors1X, 10X, 100X, 1000X, 10000X
Max. Input Voltage400V (DC+AC, PK PK, 1MΩ input impedance) (10:1 probe attenuation)
Bandwidth Limit (typical)20MHz / Full bandwidth
Horizontal Scale2ns/div - 1000s/div, step by 1-2-5
Vertical Sensitivity10mV/div - 10V/div
Vertical Resolution8 bits
Trigger TypeEdge
Trigger ModesAuto, Normal, Single
Automatic MeasurementFrequency, Period, Amplitude, Max, Min, Mean, PK-PK
Cursor MeasurementΔV, ΔT, ΔT&ΔV between cursors
Communication InterfaceUSB Type-C

7.2 Digital Multimeter Specifications

ParameterValue
Max. Resolution20,000 counts
Testing ModeVoltage, Current, Resistance, Capacitance, Diode, and Continuity test
Input Impedance10MΩ
Max Input VoltageAC 750V, DC 1000V
Max Input CurrentDC: 10A AC: 10A
Diode0-2V

7.3 Waveform Generator Specifications (HDS2102S only)

ParameterTypeFrequency Range
Frequency OutputSine0.1Hz - 25MHz
Square0.1Hz - 5MHz
Ramp0.1Hz - 1MHz
Pulse0.1Hz - 5MHz
Arbitrary0.1Hz - 5MHz
Sampling Rate125MSa/s
Channel1-CH
Amplitude Range20mVpp - 5Vpp
Waveform Length8K
Vertical Resolution14bits
Output Impedance50Ω

8. User Tips

  • The device is highly portable and functions as an excellent multimeter, making it suitable for various field applications.
  • For extended use, the battery can be quickly replaced by removing the screw on the rear cover.
  • The integrated rubber covering provides a comfortable and secure grip, preventing easy peeling off even after long-term use.
  • The USB Type-C interface is versatile, allowing charging from an external power bank (5V/2A) for prolonged outdoor operation.

9. Troubleshooting

If you encounter any issues with your OWON HDS2102S, please consider the following basic troubleshooting steps:

  • Device not powering on: Ensure the battery is correctly installed and sufficiently charged. Connect the device to a power source via the USB-C cable.
  • Incorrect measurements: Verify that probes and leads are securely connected to the correct input terminals for the selected function. Perform a self-calibration if available.
  • Display issues: Check for any physical damage to the screen. If the display is unresponsive or shows artifacts, try restarting the device.
  • Connectivity problems with PC: Ensure the USB-C cable is properly connected and the correct drivers are installed on your computer.

For persistent problems, refer to the comprehensive user manual (if a PDF link was provided, it would be referenced here) or contact customer support.

10. Maintenance

  • Cleaning: Use a soft, dry cloth to clean the device. Do not use abrasive cleaners or solvents.
  • Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures. If storing for an extended period, ensure the battery is partially charged (around 50%) to prolong its lifespan.
  • Battery Care: Avoid fully discharging the battery frequently. Charge the battery regularly even if the device is not in use.
  • Probe Care: Handle probes and leads carefully to prevent damage. Store them neatly to avoid tangling.

11. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the official OWON website or contact your authorized dealer. Keep your purchase receipt as proof of purchase for warranty claims.


OWON HDS2102S Handheld Oscilloscope Multimeter Waveform Generator Demonstration

OWON HDS2102S Handheld Oscilloscope Multimeter Waveform Generator Demonstration

0:37 • 854×480 • demonstration

Related Documents - HDS2102S

Preview OWON HDS200 Series Handheld Oscilloscope User Manual
Comprehensive user manual for the OWON HDS200 Dual Channel Series handheld oscilloscopes, detailing operation, safety, multimeter, and waveform generator functions. Support available at www.owon.com.hk/download.
Preview OWON HDS200 Dual Channel Series Handheld Oscilloscope Quick Guide
A quick guide for the OWON HDS200 Dual Channel Series Handheld Oscilloscope, covering safety information, general inspection, oscilloscope and multimeter usage, waveform generator functionality, troubleshooting, maintenance, and accessories. Includes model numbers HDS272(S), HDS242(S), and HDS2102(S).
Preview Oscilloscope Portatif Série HDS200 - Guide Rapide
Ce guide rapide présente l'oscilloscope portatif OWON série HDS200 à deux canaux. Il couvre les informations de sécurité, l'utilisation de l'oscilloscope et du multimètre, le générateur de signaux, la communication avec un ordinateur, le dépannage, et les accessoires. Idéal pour une prise en main rapide.
Preview OWON HDS200 Series SCPI Command Protocol Reference
This document details the Standard Commands for Programmable Instruments (SCPI) protocol for the OWON HDS200 Series oscilloscopes, covering horizontal settings, acquisition, channel configuration, triggering, measurements, and function generator control.
Preview OWON HDS200 Series Handheld Oscilloscope User Manual - Operation and Specifications
This comprehensive user manual provides detailed instructions for operating the OWON HDS200 Dual Channel Series handheld oscilloscopes, including models HDS272(S) and HDS242(S). It covers safety information, general inspection, oscilloscope, multimeter, and optional waveform generator functions, along with technical specifications and maintenance guidelines.
Preview Guía Rápida del Osciloscopio Portátil Doble Canal Serie OWON HDS200
Esta guía rápida proporciona información esencial para la configuración y el uso de los osciloscopios portátiles de doble canal OWON Serie HDS200, incluyendo precauciones de seguridad, operaciones básicas, y solución de problemas.