Hantek HT8000 Series High Voltage Differential Probe Instruction Manual

Models: HT8050, HT8100

Please read this instruction manual carefully to avoid injury and prevent damage to this product.

1. Introduction

Thank you for choosing the Hantek HT8000 Series high voltage isolating probes. These probes utilize a differential input mode and are designed for applications requiring high-voltage floating measurements or isolation from ground. The HT8000 Series offers high stability, reliability, and precision, making it an essential tool for various electrical and electronic measurements.

The probes provide two selectable attenuation ranges: 1/50 and 1/500. The maximum measurable differential voltage is 130V (DC + peak AC) for the 1/50 range and 1300V (DC + peak AC) for the 1/500 range. An integrated red LED indicates an over-range condition, enhancing user safety and ease of use.

2. Safety Summary

To prevent injury and avoid damage to this product or any connected devices, please observe the following safety precautions:

  • **Use proper AC adapter:** Only use the AC adapter specified for this product.
  • **Connect and disconnect properly:** Connect the probe output to the measurement instrument *before* connecting the probe to the circuit under test. Disconnect the probe input from the circuit under test *before* disconnecting it from the measurement instrument. Do not connect or disconnect probes while they are connected to a voltage source.
  • **Ground the product:** This product is indirectly grounded through the grounding conductor of the mainframe power cord. To avoid electric shock, ensure the grounding conductor is connected to earth ground. Properly ground the product before making any connections to its input or output terminals.
  • **Do not operate without covers:** Do not operate this product with covers or panels removed.
  • **Power disconnect:** The power cord serves as the product's disconnect from the power source. Ensure the power cord remains accessible at all times and is not blocked.
  • **Environmental conditions:** Do not operate in an explosive atmosphere or in wet/damp conditions.

3. Unpacking and Preparation for Use

3.1 Unpacking

This product has undergone quality checks before leaving the factory. Inspect the packaging and contents for any damage incurred during transportation. If damage is found, contact the transportation company and your local agent immediately.

Packing List:

  • High voltage isolating probe: 1 unit
  • Instruction manual: 1 unit
  • AC adapter: 1 unit
  • Red alligator clip: 1 unit
  • Black alligator clip: 1 unit
  • Red pincer clip: 1 unit
  • Black pincer clip: 1 unit

3.2 Preparation for Use

Before connecting the AC adapter to an electric outlet, verify that the line voltage matches the specifications below. Using an incorrect line voltage can damage the product and the AC adapter.

AC Adapter Input Voltage Frequency
AC100V~240V 50~60Hz

Warning: Use only the AC adapter specified for this product to avoid injury and prevent damage.

4. Product Overview

Hantek HT8000 Series High Voltage Differential Probe with labeled components
Figure 1: General View of the Hantek HT8000 Series Probe. Key components include: Input cable, Maximum differential input voltage, Maximum input voltage to earth, Over range indicator, Power indicator, Attenuation switch, Probe body, Power input connector, and Output cable.
Hantek HT8050 probe dimensions: 120mm length, 55mm width, 24mm height
Figure 2: Dimensions of the HT8050 Probe Body (120mm x 55mm x 24mm).

5. Accessories

The HT8000 Series high voltage isolating probe comes with the following accessories:

  1. AC Adapter: Used to power the probe.

    AC adapter for Hantek probe
    Figure 3: AC Adapter.
  2. Alligator Clips: For secure connections to test points.

    Red and black alligator clips
    Figure 4: Alligator Clips.
  3. Pincer Clips: For fine-point connections.

    Red and black pincer clips
    Figure 5: Pincer Clips.

6. Operating Basics

The HT8000 Series high voltage isolating probe is designed for convenient operation. New users are encouraged to thoroughly read this manual before beginning operation.

6.1 Operation Steps

  1. Verify the AC adapter input voltage is correct before use.
  2. Estimate the voltage to be measured and set the appropriate attenuation switch (1/50 or 1/500). Incorrect attenuation for an over-range input voltage may damage the probe.
  3. Connect the BNC output terminal of the probe to an oscilloscope or other measuring instrument. Ensure input cables are securely connected to the adapt clips.
  4. Connect the AC adapter output to the probe's power input connector. The green POWER indicator should illuminate.
  5. Select the appropriate attenuation range based on the measured voltage. If the measured voltage exceeds the selected range, the red OVER RANGE indicator will light up. In this case, increase the attenuation or disconnect the power and the input measured voltage.
  6. Set the measuring oscilloscope or instrument sensitivity according to the desired displayed amplitude of the signal.
  7. Connect the adapt clips to the input cables.
  8. Connect the clips to the target circuit to begin measuring. Position the probe body away from high voltage pulse circuits to minimize noise and errors.
  9. Upon completion of testing, first disconnect the power, then disconnect the input from the measured target, and finally disconnect the BNC output terminal from the oscilloscope or other measuring instrument.

6.2 Attention to Use

  • Twisting the input leads can help cancel noise induced into the leads and improve the high-frequency response.
  • Avoid extending input cables unnecessarily, as this can increase noise. If extension is required, ensure all extended cables are of the same length. For extended cables, the measured signal frequency should ideally be below 10MHz to prevent significant errors.
  • For high accuracy measurements, allow the probe and measuring instrument to warm up for approximately 20 minutes.

7. Troubleshooting

Problem Possible Solution
The Power LED does not remain lit. Check that the AC adapter is properly plugged in and functional.
The measured waveform cannot be displayed correctly or is unstable.
  • Check that all probe accessories are fully mated.
  • Check the probe connection to your circuit.
  • Perform a functional check on the probe.
  • Try a different channel on the oscilloscope or use a different oscilloscope.

8. Specifications

Electrical Performance

Parameter HT8050 HT8100
Bandwidth (-3dB)50MHz100MHz
Rise time≤7ns≤3.5ns
Accuracy±2%
Attenuation Ratio1/50, 1/500
Input ImpedanceBetween each input to ground: 4MΩ±1%
Between inputs: 8MΩ±1%
Input CapacitanceBetween each input to ground: ≤8pF
Between inputs: ≤4pF
Maximum measurable differential voltage1/50: ±130V (DC + peak AC)
1/500: ±1300V (DC + peak AC)
Maximum common mode input voltage1000Vrms, CAT II
600Vrms, CAT III
Noise1/50: ≤ 2mVrms
1/500: ≤ 1mVrms
Linearity±1%
Delay Time≤10ns
CMRRDC: ≥80dB
100Hz: ≥ 60dB
1MHz: ≥ 50dB
Over Range Warning Voltage Threshold1/50: 140V±4V (DC+peak AC)
1/500: 1400V±40V (DC+peak AC)
Power Adapter Output PowerDC9V, 1000mA
Power Adapter Input PowerAC100-240V, 50-60Hz

Machinery Specification

Parameter HT8050 HT8100
Length of Input CableAbout 60cmAbout 30cm
Length of Output CableAbout 90cm
Gator Clip DimensionsAbout 85×40×13 (mm)
Finger Pin DimensionsAbout 160×45×13 (mm)
Probe Body Dimensions120×55×24 (mm)
Weight (probe only)≈156g

Environmental Characteristics

Parameter Value
Working Temperature0~40°C
Storing Temperature-10~45°C
Working Humidity85%RH
Storing Humidity90%RH
Working Altitude3000m
Storing Altitude12000m

9. User Tips

  • **Minimize Noise:** Some users have noted high noise levels during measurements. Ensure proper grounding and keep the probe body and cables away from strong electromagnetic interference sources, especially high voltage pulse circuits, to achieve more reliable readings.
  • **Cable Length Considerations:** While the provided leads are generally sufficient, be mindful that extending input cables can introduce additional noise. If longer leads are necessary, try to use high-quality, shielded cables of equal length to minimize signal degradation, particularly for frequencies above 10MHz.

10. Warranty and Support

Warranty

Hantek warrants this product to be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If the product proves defective during this period, Hantek will, at its option, repair or replace the defective product without charge for parts and labor. Replaced parts, modules, and products may be new or reconditioned and become the property of Hantek.

This warranty does not cover defects, failures, or damage caused by improper use, inadequate maintenance, or care.

HANTEK AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. HANTEK AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER HANTEK OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Additional Resources

For further detailed information and support, you may refer to the official user manual PDF: Hantek HT8000 Series User Manual (PDF).