1. Introduction
This manual provides detailed instructions for the safe and efficient operation of your A-BF Variable Lab DC Power Supply. This series includes models such as SS-4020HP, SS-8010HP, SS-1605HP, and KPS3020D, designed for laboratory and bench applications requiring adjustable voltage and current regulation. Please read this manual thoroughly before use and keep it for future reference.
2. Technical Specifications
2.1 Model Variants
| Model | Output Voltage (0~Vmax) | Output Current (0~Amax) | Power (0~Cmax) |
|---|---|---|---|
| SS-4020HP | 0~40V | 0~20A | 300W |
| SS-8010HP | 0~80V | 0~10A | 300W |
| SS-1605HP | 0~160V | 0~5A | 300W |
| KPS3020D | 0~30V | 0~20A | 600W |
2.2 General Specifications
- Input Voltage: 110V (US Version) / 220V (EU, UK, AU Version)
- Voltage Set Resolution: 10mV
- Voltage Set Precision: ≤0.1%+2 digits
- Voltage Readback Resolution: 10mV
- Voltage Readback Precision: ≤0.1%+2 digits
- Voltage Ripple: ≤18mVrms
- Voltage Load Regulation: ≤0.1% + 5mV
- Voltage Line Regulation: ≤0.01% + 5mV
- Current Set Resolution: 1mA
- Current Set Precision: ≤0.2%+3 digits
- Current Readback Resolution: 1mA
- Current Readback Precision: ≤0.2%+3 digits
- Current Ripple: ≤18mArms
- Current Load Regulation: ≤0.2% + 3mA
- Current Line Regulation: ≤0.2% + 3mA
- Maximum Output Voltage: Vmax ± 0.2%FS
- Maximum Output Current: Cmax ± 0.2%FS
- Cooling Method: Air-cooled
- Size (bare metal): 80 x 150 x 230 mm
- Weight: 1.95kg
- Operating Conditions: Temperature: 0~40℃, Relative humidity: ≤80%RH
- Suggested Operating Range: 23℃ ± 5°
- Storage Conditions: Temperature: -20~60℃, Relative humidity: ≤80%RH
3. Package Contents
Upon opening the package, please verify that all items are present and in good condition:
- 1 x A-BF Variable Lab DC Power Supply Unit
- 1 x Power Cord (appropriate for your region: US, EU, UK, or AU)
- 1 x User Manual (this document)
- 1 x Test Line (with alligator clips)
- 1 x Serial Communication Line (optional, for software control)

4. Panel Layout and Functions

4.1 Front Panel
- Output Voltage Display: Shows the current output voltage.
- Output Current Display: Shows the current output current.
- Power Display: Shows the current output power.
- OCP Settings Button: Used to set Overcurrent Protection.
- Store/Recall Buttons (M1, M2, M3): Used to save and recall preset voltage/current values.
- Output/Close Button (ON/OFF): Toggles the power output on or off.
- OVP Settings Button: Used to set Overvoltage Protection.
- Adjusting Knob: Rotates to adjust values, press to confirm.
- Voltage Value Setting Button (U): Enters voltage adjustment mode.
- Current Value Setting Button (I): Enters current adjustment mode.
- USB Output Port: Provides power for charging devices.
- Output Terminal Negative Pole (-): Connects to the negative terminal of the load.
- Positive Output Terminal (+): Connects to the positive terminal of the load.
4.2 Back Panel
- Power Switch: Main power switch for the unit.
- Communication Port: For optional serial communication line.
- Cooling Fan: Dissipates heat from the unit.
- Power Interface: Connects the power cord.
- Fuse Holder: Contains the protective fuse.
5. Initial Setup
- Placement: Place the power supply on a stable, level surface with adequate ventilation. Ensure the cooling fan (16) at the back is not obstructed.
- Power Connection: Connect the provided power cord to the power interface (17) on the back panel and then to a suitable AC power outlet (110V or 220V depending on your model and region).
- Load Connection: Connect your load to the positive output terminal (13) and negative output terminal (12) using the test leads. Ensure correct polarity.
- Power On: Flip the power switch (14) on the back panel to the 'ON' position. The display will light up.
6. Operating Instructions
6.1 Adjusting Voltage and Current
- Adjusting Voltage:
a. Press the 'U' button (9) to enter voltage adjustment mode. The voltage display will blink.
b. Rotate the adjusting knob (8) to change the voltage value.
c. Press the adjusting knob (8) to confirm the setting. - Adjusting Current:
a. Press the 'I' button (10) to enter current adjustment mode. The current display will blink.
b. Rotate the adjusting knob (8) to change the current value.
c. Press the adjusting knob (8) to confirm the setting.
6.2 Output Control
Press the 'ON/OFF' button (6) to toggle the power output. When the output is off, the display will show "OFF".
6.3 Memory Functions (M1, M2, M3)
The unit allows you to save and recall three sets of voltage and current settings:
- Saving Settings: Set your desired voltage and current. Long press one of the 'M1', 'M2', or 'M3' buttons (5) until the display indicates the setting has been saved.
- Recalling Settings: Short press one of the 'M1', 'M2', or 'M3' buttons (5) to recall the saved settings.
6.4 Lock Function
To prevent accidental changes to settings, you can lock the buttons:
- Locking: Long press the adjusting knob (8) until a lock icon appears on the display.
- Unlocking: Long press the adjusting knob (8) again until the lock icon disappears.
6.5 USB Fast Charge
The USB output port (11) on the front panel can be used to fast charge compatible devices.
7. Overload Protection
The A-BF DC Power Supply is equipped with multiple protection mechanisms to ensure stable, reliable, and safe operation:
- OVP (Overvoltage Protection): Protects the load from excessive voltage.
- OCP (Overcurrent Protection): Protects the load from excessive current.
- OTP (Over Temperature Protection): Shuts down the unit if it overheats.
- SCP (Short-circuit Protection): Protects against short circuits at the output.
- OPP (Overpower Protection): Limits the output power to prevent damage.
- Software Setting Protection: (Requires optional serial port) Allows setting protected cut-off values via computer software.

8. Maintenance
- Cleaning: Regularly clean the exterior of the unit with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- Ventilation: Ensure the cooling fan and vents are free from dust and obstructions to maintain proper airflow.
- Storage: When not in use for extended periods, store the power supply in a cool, dry place, away from direct sunlight and extreme temperatures.
- Fuse Replacement: If the unit fails to power on, check the fuse in the fuse holder (18) on the back panel. Replace with a fuse of the same type and rating if necessary.
9. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Unit does not power on | No power from outlet; Power cord loose; Blown fuse; Power switch off | Check power outlet; Secure power cord; Replace fuse; Ensure power switch is ON |
| No output voltage/current | Output is OFF; OVP/OCP triggered; Incorrect settings; Loose connections | Press ON/OFF button; Check OVP/OCP settings; Verify voltage/current settings; Secure load connections |
| Output voltage/current unstable | Poor load connection; Overheating; Faulty component | Check load connections; Ensure proper ventilation; Contact support if problem persists |
| Buttons/knob unresponsive | Lock function engaged | Long press the adjusting knob to unlock |
10. User Tips
- For precise measurements, especially with very low current loads (e.g., below 20mA), consider using an external multimeter for verification, as the internal display might have limitations at the lowest ranges.
- Always double-check your voltage and current settings before connecting sensitive equipment to prevent damage.
- Utilize the memory functions (M1, M2, M3) to quickly switch between frequently used voltage/current configurations.
11. Warranty and Support
For any technical assistance, warranty claims, or further inquiries, please contact the manufacturer or your point of purchase. Refer to your purchase documentation for specific warranty terms and contact details.