Hantek 2C42/2D42/2C72/2D72

Contents

Hantek 2000 Series Handheld Oscilloscope User Manual

Models: 2C42, 2D42, 2C72, 2D72

1. Introduction

The Hantek 2000 series is an all-new handheld oscilloscope that combines three universal test instrument functions into one compact, lightweight device: a digital oscilloscope, a signal source (waveform generator), and a multimeter. Designed for single-hand operation, it features a 45-degree suspension bracket for multi-angle adjustment, ultra-low power consumption, and a Type C charging interface.

Each 2-hour battery charge allows for a full day of continuous operation or two months of standby time. The high-definition color LCD screen provides clear and delicate waveform displays. Its minimalist keyboard design simplifies operation, making it easy to use.

Video: Overview of the Hantek 2000 Series Handheld Oscilloscope.

2. Package Contents

The device comes in a branded carrying case. The standard accessories include:

  • Hantek 2000 Series Handheld Oscilloscope
  • Clip Probe (1 or 2, depending on model)
  • Type C Charging Cable
  • Power Adapter
  • Multimeter Probe
  • User's Guide (PDF reference available online)
  • Certification Card

Note: The product does not come with a battery. If you need a battery, please contact customer service.

Hantek 2000 Series with accessories in carrying case

Image: Hantek 2000 Series device and included accessories.

3. Setup

3.1 Battery Installation and Charging

The device is powered by two 18650 replaceable lithium batteries. Batteries are installed from the factory. The device features intelligent charging control, allowing it to be recharged via a high-current 5V/2A charger or a standard computer USB interface. You can use the device while it is charging.

The Type C power and data transfer interface allows you to share the charger and data cable with your mobile phone for convenience.

Back view of Hantek 2000 Series showing battery compartment and USB-C port

Image: Back view of the device showing the battery compartment and USB-C port.

3.2 Connecting Probes

Multimeter probes connect via standard banana connectors. Oscilloscope and waveform generator connections use BNC connectors.

Bottom view of Hantek 2000 Series showing DMM input connectors

Image: Bottom view of the device with DMM input connectors.

Top view of Hantek 2000 Series showing CH1, CH2, and Gen Out BNC connectors

Image: Top view of the device with CH1, CH2 (measured signal input) and Gen Out (waveform generator output) BNC connectors.

4. Operating Instructions

4.1 General Controls

The device features a minimalist keyboard design for ease of use. Buttons are soft and pleasant to the touch. The casing is made of durable plastic with a solid rubber bumper for protection.

Front view of Hantek 2000 Series with labeled buttons and DMM inputs

Image: Front view of the Hantek 2000 Series with labeled menu keys, function buttons (Scope, DMM, AWG, Menu), navigation keys, and DMM input connectors.

4.2 Oscilloscope Mode

The oscilloscope function offers a 250MSa/s sampling rate and 40/70MHz bandwidth (depending on model). It has a 10mV~10V vertical resolution and an auto-automatic measuring function.

Key Features:

  • Adjustable backlight brightness and duration.
  • Auto power-off time setting.
  • Digital voltmeter function.
  • Support for Chinese and English menu switching.
  • Cursor measurement for time and voltage.
  • Reference waveform and storage functions.
Oscilloscope display showing a waveform with channel information, trigger status, and time base

Image: Oscilloscope display showing a waveform with channel information (CH1, CH2), trigger status (TRID), main time base window, trigger level, and battery level.

4.3 Waveform Generator Mode

The waveform generator offers a 250MSa/s sampling rate and 12-bit vertical resolution. It can output various waveforms:

  • Sine: 1Hz~25MHz
  • Square: 1Hz~10MHz
  • Triangle: 1Hz~1MHz
  • Trapezoid: 1Hz~5MHz

It has a frequency resolution of 1Hz and an output impedance of 50Ω. It supports 4 types of arbitrary waveforms.

Waveform generator display showing a sine wave output

Image: Waveform generator display showing a sine wave output.

For models without a built-in signal generator, the "Gen Out" terminal always outputs a 2V @ 1KHz square wave, which can be used for calibrating probe attenuators.

4.4 Multimeter Mode

All series models include a standard multi-functional digital multimeter for measuring AC/DC voltage/current, resistance, diode, capacitance, and on-off tests. It offers high sensitivity and accuracy, a clear display, strong overload ability, and ultra-portability.

Multimeter display showing DC voltage measurement

Image: Multimeter display showing DC voltage measurement.

5. Maintenance

The device features an unibody casing with fine touch feeling, favorable impact resistance, and outstanding heat resistance. It is easy to disassemble for maintenance.

The 18650 replaceable lithium battery power system allows for easy battery replacement when needed.

The self-calibration routine helps optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but should always run it if the ambient temperature changes by 5°C or more. For a more accurate calibration, please power on the oscilloscope and wait for 20 minutes until it has adequately warmed up.

To do self-calibration, ensure that no input signal is connected, otherwise it may damage the instrument.

6. Troubleshooting

If you encounter issues with the device, consider the following:

  • Firmware Update: The device may have bugs or errors in the stock firmware. It is recommended to update the firmware to the latest version available on the manufacturer's website. The firmware and programming tool can be downloaded from the Hantek website under the "Hantek2000 Series" product details.
  • DFU Mode for Firmware Update: To update the firmware, the device needs to be started in DFU (Device Firmware Update) mode. Hold down the F1 button while powering on the device. Then, use the DfuSe Demo software to select the firmware file and click "Upgrade". After a successful update, remember to "Leave DFU mode".
  • FPGA Update: The FPGA (Field-Programmable Gate Array) firmware can also be updated via a USB utility while the device is powered on. Press Alt+Ctrl+Shift+F to open the update dialog, select the FPGA file, and click "Update".
  • Probe Attenuator Switch: The probe attenuator switch (1X/10X) can be easily switched accidentally, leading to incorrect readings. Always verify the probe setting matches the device setting.
  • Multimeter Buzzer: The multimeter buzzer is located deep inside the device and may be difficult to hear. When shorting, it might produce a double beep instead of a single continuous tone.
  • Multimeter Resistance Measurement: Resistance measurements can take over 10 seconds. The device measures actual resistance, not voltage drop, making it unsuitable for quick diode/transistor checks.
  • Multimeter Millivolt Mode: The millivolt mode can sometimes display phantom microamperes.
  • Oscilloscope Menu Overlap: Menu overlays can obscure parts of the screen. These menus disappear automatically after 10 seconds.
  • Device Glitches: The device may occasionally shut down unexpectedly or reset to factory settings. Ensure firmware is up-to-date as this may resolve such issues.
  • Phantom Signals: Sometimes, phantom signals may appear on the display. The trigger should not activate on these.
  • Scan Mode Limitations: In Scan Mode, the buffer is not fully utilized, and signals shorter than a few milliseconds may not be visible. Moving the trigger to the far left may not help much as only a quarter of the memory is displayed.
  • Single Mode for Irregular Pulses: For irregular short pulses, Normal mode may not trigger correctly. Use Single mode to capture and then scroll, zoom, and examine the signal in detail.

7. Specifications

FeatureSpecification
Brand NameHantek
Model Number2D72, 2C72, 2C42, 2D42
DIY SuppliesElectrical
Digital Channels2
Band Width60-99MHz (40MHz for 2C42/2D42, 70MHz for 2C72/2D72)
Real Time Sampling Rate250MSa/s (single channel), 125MSa/s (dual channel)
Record Length16k (Max. 6K for single-channel, 3K samples per dual-channel)
Max Waveform Capture Rate30,000wfm/s
Display Size3.0 - 4.9 Inches
Display Resolution640*480 Pixels (2.8 inch 64K color TFT, 320 horizontal by 240 vertical pixels)
CertificationCE
OriginMainland China
Battery IncludedYes (18650 replaceable lithium battery)
IS Battery DemountableYes
Power Supply100V-240VAC, 50Hz-60Hz; DC INPUT: 5VDC, 2A
Power Consumption<2.5W
FuseT, 3A
Battery Capacity2600mA*2
Operating Temperature0~50 °C (32~122 °F)
Storage Temperature-40~+71 °C (-40~159.8 °F)
Humidity≤+104°F (≤+40°C): ≤90% relative humidity; 106°F~122°F (+41°C~50°C): ≤60% relative humidity
Cooling MethodConvection
Altitude (Operating)3,000m (10,000 feet)
Random Vibration (Nonoperating)0.310RMS from 50Hz to 500Hz, 10 minutes on each axis
Mechanical Shock (Nonoperating)2.46GRMS from 5Hz to 500Hz, 10 minutes on each axis
Mechanical Shock (Operating)50g, 11ms, half sine
Dimension (L x W x H)199 x 98 x 40mm
Weight624g

Image: Detailed specifications table for the Hantek 2000 Series.

8. Warranty and Support

For detailed warranty information and technical support, please refer to the official Hantek website. You can find user manuals, software, and firmware updates in the download section of the product page.

The device has a good potential due to its powerful microcontroller and partial data processing via FPGA. Many of the described shortcomings are purely software-related and may be addressed in future firmware updates.

Function of video help: Scan the QR code on the back of the machine to access video courses for assistance.

9. User Tips

  • Always update the device firmware and FPGA to the latest versions from the Hantek website to ensure optimal performance and fix known bugs.
  • Be mindful of the probe attenuator switch (1X/10X) as it can be easily changed accidentally, affecting measurement accuracy.
  • When performing self-calibration, ensure no input signals are connected to prevent potential damage to the instrument. Allow the device to warm up for about 20 minutes before calibration for best results.
  • For capturing irregular or short-duration pulses, utilize the "Single" trigger mode to freeze the waveform for detailed analysis.
  • Be aware that the multimeter's buzzer might be quiet and its resistance measurement for diodes/transistors might not provide voltage drop readings.