1. Product Overview
The Yieryi TPH02159 is a versatile 7-in-1 digital multifunction water quality monitor designed for precise measurement of various parameters in liquids and air. It can simultaneously measure pH, ORP, TDS (Total Dissolved Solids), EC (Electrical Conductivity), CF (Conductivity Factor), Temperature, and Humidity. Featuring a high-precision LCD digital display and quick 1-second measurement, this device is easy to use and suitable for continuous online monitoring.
Its applications include monitoring water quality in fish tanks, aquariums, mariculture, swimming pools, drinking water, and laboratory settings.

2. Key Features
- Measures pH, ORP, Humidity, Temperature, EC, CF, and TDS (PPM).
- Switchable temperature units: Centigrade (°C) / Fahrenheit (°F).
- Easy calibration for pH and TDS using a mini-screwdriver.
- Instant display of results with continuous online monitoring capability.
- Long-lasting operation: Can work for over 10 hours during power outages or outdoors after a 3-hour battery charge.
- Detachable electrode for convenience and maintenance.
3. Specifications
| Parameter | Measuring Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.01 ~ 14.00pH | 0.01pH | ±0.5pH |
| EC | 0.00~20.00mS/cm | 0.01mS/cm | ±2%F.S |
| CF | 0.1 ~ 199.9CF | 0.1CF | ±2%F.S |
| TDS | 0 ~ 9999ppm | 1ppm | ±2%F.S |
| Temperature | 0.1°C-60.0°C, 32.1°F-140.0°F | 0.1°C/0.1°F | ±0.5°C |
| ORP | -1200mV~+1200mV | 1mV | ±5 mV |
| Humidity (air) | 10%RH-99%RH | N/A | N/A |
Additional Specifications:
- Automatic Temperature Compensation (ATC): 0.1℃-60.0℃
- Working Temperature: 0℃ ~ 50℃ (32℉ ~ 122℉)
- Main Unit Dimensions: 150mm x 80mm
- Probe Cable Length: 120cm
- Power Cable Length: 95cm




4. Package Contents
Upon opening the package, please verify that all the following items are included:
- 1x Digital Multifunction Monitor
- 3x pH Buffer Powder (for calibration)
- 1x English Instruction Manual
- 1x Power Adapter (EU or US plug, depending on region)
- 1x Detachable Electrode
- 1x Package Box

5. Setup and Initial Use
- Connect the Electrode: Carefully connect the detachable electrode to the corresponding port on the main meter unit.
- Remove Protective Cap: Remove the protective cap from the electrode.
- Clean Electrode: Rinse the electrode with distilled water and gently wipe off any excess water.
- Power On: Press the "ON/OFF" button to turn on the meter.
- Immerse Electrode: Immerse the electrode into the solution you wish to test.
- Read Measurements: Lightly shake the electrode and wait for the numeric display to stabilize before taking a reading.
- Switch Units/Modes:
- Press "C/°F" to switch between Celsius and Fahrenheit temperature units.
- Press the "MODE" button under the left screen (EC/PPM/CF) to cycle through EC, PPM, and CF modes.
- Press the "MODE" button under the middle screen (pH/ORP) to switch between pH and ORP modes.
- Power Off: After use, press the "ON/OFF" button to shut off the instrument.
6. pH Calibration Procedure
The pH meter must be re-calibrated before initial use and periodically thereafter for accurate readings.
- Turn on the meter by pressing the "ON/OFF" button.
- Prepare Buffer Solutions: Dissolve each pH buffer powder (e.g., pH 6.86, pH 4.00) into separate cups, each containing 250ml of distilled water.
- Switch to pH Mode: Press the "MODE" button on the "pH/ORP" screen to ensure the meter is in pH mode.
- Calibrate with pH 6.86: Immerse the electrode into the pH 6.86 solution (at approximately 25℃). Press and hold the "CAL" (calibration) button for 5 seconds, then release. The display will start flashing "6.86". Wait until the display stops flashing and stabilizes.
- Clean Electrode: Rinse the electrode with distilled water and dry it gently with filter paper.
- Calibrate with pH 4.00: Immerse the electrode into the pH 4.00 solution. Press and hold the "CAL" button for 5 seconds (the display may flash "6.86" again). Immediately short-press the "CAL" button again. The display will flash "4.00". Wait until the display stops flashing and stabilizes.
- Clean Electrode: Rinse and dry the electrode with distilled water as before.
- Verify Calibration: Test the calibration by inserting the electrode into a standard buffer solution. Gently stir the electrode until the screen display stabilizes. If necessary, repeat the calibration procedure.
- Replace Protective Cap: Clean the electrode and replace the protective cap after calibration.
7. Maintenance and Safety
7.1 General Maintenance
- Always replace the protective cap on the electrode after use to prevent it from drying out. A dried-out electrode can lead to slow and unstable readings. If the electrode has dried out, immerse it in a 3.3mol KCL standard solution for a few hours to recondition it.
- Clean the electrode with pure water after each use.
7.2 Battery Information
- When the battery is low, a battery symbol will appear on the display. Connect the power supply to charge the battery.
- A full charge typically takes about 3 hours for continuous operation of over 10 hours.
7.3 Important Safety Warning

Please keep the instrument's main unit dry at all times. The host unit is NOT waterproof and must not get wet or be immersed in water. Water exposure will cause abnormal temperature readings and other functions to become unusable. Ensure the environment is ventilated and dry when using the device.
8. Troubleshooting
- Device not functioning normally upon receipt: Refer to this instruction manual for setup and operation. If issues persist, please contact customer service. For non-artificial quality problems encountered within one week of receiving the goods, you may be eligible for replacement or refund.
- Missing pH buffer powders: Please check the package contents carefully against the 'Package Contents' section. If items are genuinely missing, contact the seller immediately.
- Calibration difficulties: Ensure you follow the pH calibration steps precisely, using fresh distilled water for buffer solutions and proper cleaning of the electrode between steps. If calibration remains problematic, the electrode might need reconditioning (see Maintenance) or replacement.
9. User Tips
- Optimal Readings: For the most accurate readings, ensure the electrode is fully immersed in the solution and allow a few moments for the readings to stabilize before recording.
- RS485 Connectivity: If you have a model equipped with RS485, ensure you have the correct drivers and software installed for seamless connection to a computer or Raspberry Pi. Consult specific RS485 documentation if available for detailed setup.
- Environmental Control: Always operate the main unit in a dry, well-ventilated area to prevent moisture damage and ensure longevity.
10. Warranty and Support
For any product-related inquiries, technical support, or warranty claims, please contact the seller directly. Non-artificial quality problems reported within one week of receiving the goods may qualify for replacement or refund.