Hantek 1008C Automotive Oscilloscope User Manual

DIGITAL OSCILLOSCOPE V1.0.5

1. General Safety Summary

Review the following safety precautions carefully before operating the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards, use the device as specified by this user's guide only.

To Avoid Fire or Personal Injury:

  • Connect and Disconnect Properly: Do not connect or disconnect probes or test leads while they are connected to a voltage source. Connect the probe output to the measurement device before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement device.
  • Observe All Terminal Ratings: To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
  • Use Proper Probe: To avoid shock hazard, use a properly rated probe for your measurement.
  • Avoid Circuit or Wire Exposure: Do not touch exposed connections and components when power is on.
  • Do Not Operate With Suspected Failures: If suspected damage occurs with the device, have it inspected by qualified service personnel before further operations.
  • Provide Proper Ventilation: Refer to the installation instructions for proper ventilation of the device.
  • Do not operate in Wet/Damp Conditions.
  • Do not operate in an Explosive Atmosphere.
  • Keep Product Surfaces Clean and Dry.

2. Getting Started

The Hantek 1008C is a small, lightweight, portable oscilloscope that requires no external power. It is ideal for production testing, research and design, and troubleshooting analog circuits, as well as for educational purposes.

2.1 System Requirements

  • Operating System: Windows 7, Windows 8, or Windows 10 (32-bit and 64-bit supported).

2.2 Software Installation

To download the software and user guide, scan the QR code below or visit the official Hantek product page:

https://e.hantek.com/Products/hantek1008c

QR code for software download

Text description: QR code for software download and technical support.

Double-click Setup.exe and follow the on-screen instructions to install the software. A "Hantek1008" icon will appear on your desktop.

2.3 Connecting the Device to PC

Connect the Hantek 1008C device to your PC using the provided USB cable.

Hantek 1008C connected to a laptop via USB

Text description: The Hantek 1008C oscilloscope connected to a laptop, illustrating the USB connection.

2.4 Probe Compensation

This function matches the characteristics of the probe and the channel input. Perform this whenever attaching a probe to any input channel for the first time.

  1. From the "Probe" menu, select attenuation to 1:10.
  2. Set the switch to "X10" on the probe and connect it to CH1 of the oscilloscope. Ensure a firm connection if using a probe hook-tip.
  3. Attach the probe tip to the Probe Compensator and the reference lead to the ground connector. Select CH1.
  4. Adjust the vertical and horizontal controls to optimize the waveform display.
  5. Check the shape of the displayed waveform.
Waveform examples for probe compensation: Correctly Compensated, Over Compensation, Under Compensation

Text description: Three waveform diagrams illustrating correctly compensated, over compensated, and under compensated probe signals.

If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the flattest square wave. Repeat if needed.

2.5 Self-Calibration

The self-calibration routine optimizes the oscilloscope signal path for maximum measurement accuracy. Run this routine if the ambient temperature changes by 5℃ or more. For accurate calibration, power on the oscilloscope and wait twenty minutes for it to warm up. Disconnect any probes or cables from the input connectors. Access the "More -> Utility -> Calibration" option and follow the on-screen directions. The routine takes several minutes.

3. Understanding Oscilloscope Functions

The Hantek 1008C offers a range of functions for automotive diagnostics and general oscilloscope use.

3.1 User Interface Overview

After connecting the device and running the software, the main user interface will be displayed.

Hantek 1008C software user interface screenshot

Text description: Screenshot of the Hantek 1008C software interface, showing waveform display, toolbar, and functional panels.

Key elements of the user interface include:

  1. Toolbar: Contains various controls and settings.
  2. Current Time Base (Time/Div): Displays the current horizontal time division.
  3. Trigger Information: Shows the edge trigger slope, source, and level.
  4. Functional Panel: Contains buttons for various functions:
    • Return: Returns to the function select interface.
    • Vehicle Diagnosis Reference Waveform: Toggles on/off reference waveform for vehicle diagnosis.
    • Default: Resets to default settings in vehicle diagnosis.
    • More: Accesses additional functional menus.
    • QR Code: Click or scan to get measurement help.
    • Screenshot: Saves the current screen as a .bmp or .jpg file.
  5. Acquisition Mode: Displays the current data acquisition mode (Auto, Normal, Single).
  6. Horizontal Setting Panel: Allows selection of horizontal Time/Div.
  7. Vertical Setting: Allows turning on/off channels (CH1-CH8) and changing Volt/Div and probe attenuation.
  8. Marker of Trigger Level: Indicates the trigger level on the waveform.
  9. System Time: Displays the current system time.
  10. CH1-CH8 Information: Shows vertical scale factors for each channel.
  11. Software Status: Indicates connection status (Connected, Searching Device, Demo).
  12. CH1-CH8 Reference Points: Markers showing reference points for displayed waveforms.
  13. Marker of Horizontal Trigger Position: Indicates the horizontal trigger position.
  14. Display Waveform in Buffer Position: A window showing the waveform in buffer.
  15. Trigger Status: Indicates the current trigger state (Trig'D, WAIT, STOP, RUN, PLAY).

3.2 Set Vertical System

Access the "More -> Setup -> Vertical" menu to configure vertical settings.

Vertical Setup menu screenshot

Text description: Screenshot of the Vertical Setup menu, showing options for channel selection, ON/OFF, VOLTS/DIV, Coupling, Probe attenuation, and Invert.

  • Select Channel: Choose from CH1-CH8.
  • Channel Setting:
    • ON/OFF: Turn the channel on or off.
    • Volt/DIV: Select the channel voltage per division.
    • Coupling: Set to DC coupling.
    • Probe: Select the channel probe attenuation.
  • Change Volt/DIV: Click "Volt/Div" to select the voltage range.
  • Probe Attenuation Setting: Select the attenuation factor for the probe. This setting remains in effect until changed. The attenuation factor changes the vertical scale to reflect actual voltage levels.

3.3 Set Horizontal System

Access the "More -> Setup -> Horizontal" menu to configure horizontal settings.

Horizontal Setup menu screenshot

Text description: Screenshot of the Horizontal Setup menu, showing options for Time/DIV, Format (Y-T, X-Y), and X-Y settings.

  • Change Time/DIV: Select the horizontal Time/DIV (scale factor). Use left/right arrows or the combobox in the sidebar.
  • Format: Set the waveform display format (Y-T, X-Y).
    • Y-T: Shows the relative relation between vertical voltage and horizontal time.
    • X-Y: Shows CH1 value at X-axis and CH2 value at Y-axis, used for phase difference analysis (Lissajous patterns).
  • Y-T Format:
    • Scan: Waveform updates from left to right.
    • Roll: Waveform updates from right to left. No trigger or horizontal offset control is available in Roll Mode, only for 500ms/div or slower.
  • Change Horizontal Position: Adjusts the displayed waveform position relative to the trigger point. Drag the 'T' icon on screen to change the position.

3.4 Set Trigger System

Access the "More -> Setup -> Trigger" menu to configure trigger settings.

Trigger Setup menu screenshot

Text description: Screenshot of the Trigger Setup menu, showing options for Trigger Mode, Trigger Sweep, Trigger Source, Trigger Slope, and HF Rejection.

  • Edge Trigger: Determines when the oscilloscope acquires data. It finds the trigger point on the rising or falling edge of a signal.
    • Mode: Edge trigger.
    • Sweep: Set to Auto, Normal, or Single.
      • Auto: Acquires waveform even if no trigger occurred.
      • Normal: Acquires waveform when trigger occurred.
      • Single: Acquires waveform once, then stops.
    • Source: Select the signal source for triggering (any channel BNC).
    • Slope: Set to Rising (+) or Falling (-).
      • Rising: Triggers on a rising edge.
      • Falling: Triggers on a falling edge.

3.5 Set MATH Functions

Access the "More -> Setup -> Math" menu to perform mathematical operations on waveforms.

MATH Setup menu screenshot

Text description: Screenshot of the MATH Setup menu, showing options for ON/OFF, Source A/B, Operate type (A+B, A-B, AxB, A/B, FFT), Volt/DIV, and Invert.

  • ON/OFF: Turn the MATH Channel on or off.
  • Source A/B: Set the sources for the math channel.
  • Operate: Set the type of math operation.
  • Volt/DIV: Set the resolution of the math channel.
  • Invert: Turn on/off the invert function.

Mathematical functions include addition, subtraction, multiplication, and FFT.

Operate Types:

  • A + B: Add source A and source B.
  • A - B: Subtract source B from source A.
  • A x B: Multiply source A by source B.
  • A / B: Divide source A by source B.
  • FFT: Converts a time-domain signal into its frequency components (spectrum).

To use the FFT mode, click the FFT icon in the Toolbar to open the FFT setup window.

FFT Setup menu screenshot

Text description: Screenshot of the FFT Setup menu, showing options for Source channel, Window algorithm, Scale, Vertical Scale, and Horizontal Scale.

You can select the Source channel, Window algorithm, FFT number, and FFT Zoom factor. It displays only one FFT spectrum at a time.

3.6 Set Reference

Access the "More -> Setup -> REF" menu to manage reference waveforms.

REF Setup menu screenshot

Text description: Screenshot of the REF Setup menu, showing options for ON/OFF, Volt/DIV, Time/DIV, Load, Save, Ref Data Length, and Move Step.

  • ON/OFF: Turn the reference channel on or off.
  • Volt/DIV: Set the resolution of the reference channel.
  • Time/Div: Select the horizontal Time/DIV of the reference channel.
  • Load: Load a reference waveform from a ".rfc" file.
  • Save: Save the current reference waveform as a ".rfc" file.

3.7 Record and Playback

Access the "More -> Setup -> Record" menu to record and play back waveforms.

Recorder sidebar menu screenshot

Text description: Screenshot of the Recorder sidebar menu, showing options for Storage, Play Back, Interval, Start, and Stop.

  • Storage: Records waveform and saves it as a ".dfr" file.
  • Start: Starts recording waveforms.
  • Stop: Stops recording waveforms.
  • Play Back: Plays back the saved waveform.

3.8 Utility Menu

The Utility Menu provides access to various system settings.

  • Factory Setup: Loads default oscilloscope settings. This resets CH1-CH8 waveforms and removes others. It does not reset Language or Date and Time.
  • Calibration: Refer to section 2.5 Self-Calibration.
  • Language: Selects the display language (English, German, Chinese).

3.9 Measure Signal

The oscilloscope provides various measurement tools.

3.9.1 Cursor Menu

Access the "More -> Cursor" menu to use cursors for measurements.

Cross Cursor measurement screenshot

Text description: Screenshot showing a waveform with cross cursors for measuring vertical and horizontal parameters.

  • Source: Set the source to CH1-CH8 or MATH.
  • Type: Four types of cursors are available: Cross, Trace, Vertical, and Horizontal.
    • Cross: Appears as cross lines, measuring vertical and horizontal parameters. Results are displayed in the status bar (Freq, Time, Volt).
    • Trace: Appears as vertical lines, measuring waveform amplitude at the point the waveform crosses the cursor.
    • Vertical: Appears as vertical lines, measuring vertical parameters.
    • Horizontal: Appears as horizontal lines, measuring horizontal parameters.

3.9.2 Measure Menu

Access the "More -> Measure" menu for automatic parametric measurements.

Automatic measurement display screenshot

Text description: Screenshot showing a waveform with automatic measurements (e.g., +Duty, Freq, Vrms, Vamp, Vpp, Vmax) displayed at the bottom of the screen.

The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time measurements).

  • Source: Select the measurement source.
  • Vertical Measurements: Peak To Peak, Top, Base, Middle, RMS, Amplitude, Mean, Cycle Mean, Preshoot, Overshoot.
  • Horizontal Measurements: Period, Frequency, Rise Time, Fall Time, +Duty Cycle, -Duty Cycle, +Pulse Width, -Pulse Width.
  • Clear Measure: Clears all measure items from the display screen.

3.10 Display System

Access the "More -> Display -> Type" menu to configure display settings.

Waveform display in Vectors mode screenshot

Text description: Screenshot of a waveform displayed in Vectors mode.

  • Display Type:
    • Vectors: Displays the waveform as vectors.
    • Dots: Displays the waveform as dots.
  • Display Grid: Access "More -> Display -> Grid" to toggle the grid display.
  • Intensity: Access "Display -> Intensity" to adjust grid and waveform intensity.

3.11 Zoom In/Out

When the waveform acquisition is stopped (after clicking "Stop"), you can zoom in/out and adjust the waveform display.

  • Click "Zoom In/Out" in the "Acquire" menu, then left or right click the mouse button on the display screen.
  • Alternatively, change Time/Div in the Horizontal menu or Horizontal panel.

3.12 Print and Print Preview

To print or preview reports:

Print Report preview screenshot

Text description: Screenshot of a "Test Report" preview, showing multiple channel waveforms and a notes section.

  1. Click "More -> File -> Print" to set up the printer.
  2. Click "More -> File -> PrintPreview" to enter the preview window.
  3. In the "PrintPreview" window, use "Zoom In" and "Zoom Out" buttons to change the size.
  4. Click "Close" to exit preview, then "Print" to print the report.

3.13 Generate Square Waveform

Click the "Generator" icon to access the signal generator settings.

Generator function menu screenshot

Text description: Screenshot of the Generator function menu, showing icons for Vehicle Diagnosis, Oscilloscope, and Generator.

Square waveform generator display screenshot

Text description: Screenshot of the square waveform generator display, showing multiple channels with adjustable pulse width and level.

  1. Edit Setting: Click "Edit -> Edit On" to open the edit settings.
    • Wave Num: Set the number of waveform channels (3, 5, or 8).
    • Pulse: Set the pulse number.
    • StartLevel: Set the start level (High or Low) of channels.
  2. Output Waveform: Click "Operate -> Output -> ON" to enable output.
  3. Connect D0-D7 digital channels to output waveform.

3.14 Application Examples

4.1 Simple Measurement

To acquire and display a signal:

  1. Connect the signal to CH1 using a probe.
  2. Adjust vertical and horizontal controls to optimize the waveform display.

To measure Frequency and Vpp:

  1. Click "More -> Measure -> Horizontal -> Frequency" to display frequency.
  2. Click "More -> Measure -> Vertical -> Peak-to-Peak" to display Vpp.
Simple measurement waveform screenshot

Text description: Screenshot of a waveform with Vpp and Frequency measurements displayed at the bottom.

To clear measurements, click "More -> Measure -> Clear Measure".

4.2 Capturing a Single-Shot Signal

To capture a single event, set up the trigger level and slope correctly.

  1. Set probe and channel attenuations to X10.
  2. Set up the trigger in the Trigger Menu:
    • Adjust Trigger Mode to Edge.
    • Set Trigger Sweep to Single.
    • Set Trigger Source to CH1.
    • Set Trigger Slope to "+" (rising edge).
    • Adjust Volts/Div and time base to a proper range.
    • Drag the trigger level sign to the proper position (slightly above normal level).
  3. Click the Run button to start capturing. Data appears when trigger conditions are met.

4.3 The Application of the X-Y Operation

X-Y Plot analyzes the correlation of data from two channels, displaying a Lissajous diagram. This allows comparison of frequencies, amplitudes, and phases of waveforms.

  1. Set probe attenuation to "x10" on the probes.
  2. Connect CH1 probe to the input and CH2 probe to the output of the circuit.
  3. Adjust vertical scale and offset for similar amplitude signals.
  4. Select X-Y format in the Horizontal window.
  5. Adjust scale and offset for a desirable waveform display.
  6. Apply the Ellipse Method to observe phase difference.
Lissajous diagram in X-Y format screenshot

Text description: Screenshot of a Lissajous diagram displayed in X-Y format.

Instruction of the Ellipse Method:

Ellipse Method diagram for phase shift calculation

Text description: Diagram illustrating the Ellipse Method with labels A, B, C, D for calculating phase shift (Sine = A/B or C/D).

Sine = A/B or C/D, where θ = phase shift (in degrees) between the two signals. From the formula above: θ = ±arcsine (A/B) or ±arcsine (C/D). θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I and III quadrant. If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π) or (π~3π/2).

4.4 Taking Cursor Measurements

Use cursors for quick time and amplitude measurements.

Measure the Peak Frequency or Time of the First Sine Waveform:

  1. Click "More -> Cursor -> Source", select CH1 (or CH2-CH8).
  2. Click "More -> Cursor -> Type", select Vertical.
  3. Push left mouse button to make vertical lines appear.
  4. Drag the mouse button to the desired measurement point.
  5. Release the left mouse button; frequency and time differences will show in the status bar.
Vertical cursor measurement screenshot

Text description: Screenshot of a waveform with vertical cursors for measuring frequency and time.

Measure the Amplitude:

  1. Click "More -> Cursor -> Source", select CH1 (or other CH).
  2. Click "More -> Cursor -> Type", select Horizontal.
  3. Push left mouse button to make horizontal lines appear.
  4. Drag the mouse button to the desired measurement point.
  5. Release the left mouse button; voltage difference will show in the status bar.
Horizontal cursor measurement screenshot

Text description: Screenshot of a waveform with horizontal cursors for measuring amplitude.

Trace the Amplitude of a fixed position on X-axis in a Waveform:

  1. Click "More -> Cursor -> Source", select CH1 (or other CH).
  2. Click "More -> Cursor -> Type", select Trace.
  3. Click the cursor at the desired position in the waveform window.
Trace cursor measurement screenshot

Text description: Screenshot of a waveform with a trace cursor indicating amplitude at a specific point.

Note: Click "More -> Cursor -> Type", select "Cross" to measure time and amplitude simultaneously.

4. Specifications

CategoryParameterValue
AcquisitionAcquisition ModeReal-Time Sample
Acquisition Rate2.4MSa/s (Single Channel)
InputBandwidth100K (Single Channel)
Input CouplingDC
Input ImpedanceResistance: 1MΩ
Supported Voltage1X, 10X, 100X, 1000X, 10000X, 20:1
Attenuation Factors
Maximum Input Voltage400Vpk (DC + peak)
HorizontalScanning Speed Range (Sec/Div)1ns/div ~ 20000s/div (1-2-5 sequences)
Sample Rate and Delay Time Accuracy±50ppm (any interval ≥1ms)
Memory Depth (Sample Points)Max. 4K (Single Channel)
Time Base Range1ns/div to 20000s/div (1-2-5 sequences)
Time Base Precision±50ppm
VerticalAnalog Channel8
A/D converter12 bit
Vertical Scale (Volt/div) Range10mV ~ 5V/div @ x1 probe (1,2,5 sequence)
100mV ~ 50V/div @ x10 probe
1V ~ 500V/div @ x100 probe
10V ~ 5000V/div @ x1000 probe
100V ~ 50000V/div @ x10000 probe
200mV ~ 100V/div 20:1 probe
DC Gain Accuracy±3%
TriggerTrigger SourceCH1-CH8
Trigger ModeAuto, Normal and Single
Trigger TypeEdge
Trigger Sensitivity0.02 div increments
Trigger Level Accuracy±4 divisions
X-Y ModeX-Axis InputCH1
Y-Axis InputCH2
Cursors & MeasurementVoltage MeasurementVpp, Vamp, Vmax, Vmin, Vtop, Vmid, Vbase, Vavg, Vrms, Vcrms, Preshoot, Overshoot
Time MeasurementFrequency, Period, Rise Time, Fall Time, Positive Width, Negative Width, Duty Cycle
Cursors MeasurementHorizontal, Vertical, Track, Auto Measure Modes
Waveform Signal Process+,-, x,÷, FFT, Invert
Voltage Range10mV to 5V/div @ x 1 probe
100mV to 50V/div @ x 10 probe
10V to 5000V/div @ x 1000 probe
100V to 50000V/div @ x 10000 probe
200mV to 100V/div @ 20:1
Current Range100mA to 50.0A/div @ CC65 (20A)
1000mA to 500.0A/div @ CC65 (65A)
1A to 100.0A/div @ CC650 (60A)
10A to 1000.0A/div @ CC650 (650A)
1A to 200.0A/div @ CC1100 (100A)
10A to 2000.0A/div @ CC1100 (1100A)
CursorTime/frequency difference, voltage difference
FFTRectangular, Hanning, Hamming, Blackman Window
MathAddition, subtraction, multiplication, division
Programmable Square GeneratorChannel8 (D0-D7)
Output LevelLVTTL
Frequency Range0-250KHz
InterfaceUSB 2.0
PowerNo need extra power supply
PhysicalSize185 x 150 x 27 (mm)
Weight0.35KG

5. Maintenance

5.1 General Care

Do not store or leave the oscilloscope where the device will be exposed to direct sunlight for long periods of time.

5.2 Caution

To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents. Do not use any abrasive or chemical cleaning agents on the surface of the device or probes.

5.3 Cleaning

Inspect the device and probes as often as operating conditions require. Make sure the device is disconnected from all power sources before cleaning.

To clean the exterior surface, perform the following steps:

  1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth. Use care to avoid scratching the clear glass display filter.
  2. Use a soft cloth dampened with water to clean the device.

6. Troubleshooting

This section addresses common issues you might encounter with your Hantek 1008C Automotive Oscilloscope.

  • Software Compatibility: The device is compatible with Windows NT, 2000, XP, VISTA, 7, 8, and 10 (32-bit and 64-bit). Ensure your operating system meets these requirements.
  • Connection Issues: If the software shows "Searching Device" or "Demo" mode, ensure the USB cable is securely connected to both the oscilloscope and the PC. Try a different USB port or cable if necessary.
  • Driver Installation: The software installation includes necessary drivers. If you experience connection problems, verify that the drivers are correctly installed.
  • Waveform Display Issues: If waveforms are unstable or blank, check your trigger settings (Mode, Sweep, Source, Slope) and probe compensation (Section 2.4). Perform a self-calibration (Section 2.5) if ambient temperature has changed.
  • Language Settings: The device language supports English, German, and Chinese. If the language is incorrect, navigate to "More -> Utility -> Language" to change it.

7. User Tips

  • Software Installation: Some users have noted that the software is provided on a CD. If your computer lacks a CD drive, you will need to download the software from the official Hantek website (link provided in Section 2.2).
  • Testing Functionality: Always test the equipment upon arrival to confirm proper functioning and ensure all accessories are included as described.
  • Communication: If you encounter any issues or delays, contacting the seller for support can be helpful, as they often provide good communication.

8. Warranty and Support

For technical support, refer to the official Hantek website or the provided user manual PDF for detailed instructions and contact information. The user manual (PDF) can be viewed here.