1. Introduction
The Kamoer KDS series peristaltic pumps are designed for precise liquid transfer in various applications, including aquariums, laboratories, and analytical instruments. These pumps are capable of handling both viscous and non-viscous liquids, as well as liquids containing particles, thanks to their robust design and resilient tubing. Key features include low noise operation, durable construction, and a range of motor options to suit different voltage and flow requirements.

2. Specifications
2.1. Basic Parameters
- Motor Types: DC brush motor (FC/FE), Stepper motor (FB), AC synchronous gear motor (FA), AC shaded-pole motor (FD)
- Rotor: 2 rotors
- Tube Types: Silicone tube, PharMed BPT, Norprene tube
- Operating Temperature: 0~40℃
- Operating Humidity: Less than 80%
DC Motor (FC/FE) Specifics:
- Voltage: 24V/12V
- Current: 0.28A/0.56A
- Flow: 340-380 ml/min
- Net Weight: Approximately 0.75 KG
Stepper Motor (FB) Specifics:
- Voltage: 24V
- Current: 1.8A max
- Flow: 5~600 ml/min
- Net Weight: Approximately 1.03 KG
2.2. Flow Rate & Tube Selection
The following tables detail flow rates for different motor types and tube materials, along with tube characteristics and expected lifespan.
| Material | Code | ID×OD (mm) | KDS-FC/FE (24V/12V; 0.28A/0.56A) Flow Rate (ml/min) @ 300rpm |
|---|---|---|---|
| BPT | B16 | 3.2×6.4 | 340 |
| Norprene | N16 | 3.2×6.4 | 340 |
| Silicone | S16 | 3.2×6.4 | 340 |
| Material | Code | ID×OD (mm) | KDS-FB (24V; 1.8A) Flow Rate (ml/min) @ 1-400rpm |
|---|---|---|---|
| BPT | B16 | 3.2×6.4 | 5-440 (at 400rpm) |
| Norprene | N16 | 3.2×6.4 | 5-440 (at 400rpm) |
| Silicone | S16 | 3.2×6.4 | 5-600 (at 400rpm) |
| Tube Material | Features | Working Temperature | Tubing Life |
|---|---|---|---|
| Silicone (S) | Low adsorption, good temperature resistance, not easy to aging, low precipitation, chemical resistance. Corrosion performance decreases with increasing temperature. Suitable for transmission of various low corrosion (10%) liquids. | -60℃~200℃ | ≥200H |
| PharMed BPT (B) | Very good resistance to general chemicals, as well as excellent resistance to acids, alkalis and oxides. Transparent and resistant to ultraviolet radiation, helping protect sensitive liquids. | -51℃~132℃ | ≥1000H |
| Imported Norprene (N) | FDA certificate, perfect for food and dairy product application. High temperature and corrosion resistance, aging and oxidation resistance, long-term service time. | -59℃~-135℃ | ≥1000H |
3. Setup and Installation
3.1. Motor Identification
The KDS series offers various motor types. Identify your pump's motor type for correct installation and operation.

3.2. Physical Installation
The pump head can be fixed in two ways: by supporting it with its integrated stand or by using a through-hole mounting method. The pipe joints can be oriented in four different directions to accommodate various working environments.

DC Motor Pump Installation Drawings:

Stepper Motor Pump Installation Drawings:


3.3. Electrical Connection (Stepper Motor)
For stepper motor models, ensure correct wiring as per the diagram below. The stepper motor has a step angle of 1.8°.

Note: A dedicated driver board for speed control is not included by default with the stepper motor pump. If required, please ensure this need is communicated at the time of purchase.
4. Operating Instructions
The KDS peristaltic pump operates by compressing and releasing a flexible tube, moving fluid through it. This design ensures that the fluid only contacts the tubing, preventing contamination and making it suitable for a wide range of liquids.
4.1. General Operation
- Ensure the pump is securely installed and the tubing is correctly routed and connected to your fluid source and destination.
- Verify that the power supply matches the pump's voltage requirements (e.g., 12V, 24V).
- Turn on the power to the pump. The rotor will begin to turn, initiating fluid transfer.
- Monitor the flow rate and adjust as necessary. For stepper motor pumps, speed control typically requires an external driver board.
5. Maintenance
5.1. Tubing Replacement
The tubing is a consumable part and should be replaced periodically based on its material and usage. Refer to the 'Tubing Life' in the specifications section. The video demonstrates the process of replacing the tubing.
- Disconnect power to the pump.
- Open the pump head cover.
- Carefully remove the old tubing from the rotor and pump head channels.
- Insert the new tubing, ensuring it is properly seated around the rollers and within the pump head.
- Close the pump head cover securely.

5.2. Cleaning
Regular cleaning of the exterior of the pump is recommended. If the pump is used for corrosive liquids, ensure any spills are immediately cleaned to prevent damage to the pump housing. Since the fluid only contacts the tubing, internal cleaning of the pump mechanism is generally not required unless there is a tube rupture.
5.3. Storage
Store the pump in a dry environment within the specified temperature and humidity ranges (0~40℃, less than 80% humidity).
6. Troubleshooting
This section provides basic troubleshooting steps for common issues. For more complex problems, please contact customer support.
| Problem | Possible Cause | Solution |
|---|---|---|
| Pump not running | No power, incorrect voltage, motor fault | Check power connection and supply voltage. Ensure correct motor type for voltage. |
| No fluid flow or low flow | Tubing kinked, clogged, worn out, or improperly installed; air in line; insufficient fluid source | Inspect tubing for kinks or clogs. Replace worn tubing. Ensure tubing is correctly seated. Prime the pump if necessary. Check fluid level. |
| Excessive noise | Worn tubing, motor issue, improper installation | Replace tubing. Ensure pump is securely mounted. If noise persists, contact support. |
7. User Tips
- For applications requiring precise flow control, especially with stepper motor models, consider investing in a compatible driver board.
- Regularly inspect the tubing for signs of wear or degradation, particularly when transferring aggressive chemicals or operating continuously.
- Take advantage of the flexible installation options (stand or through-hole mounting) and adjustable pipe joint directions to optimize the pump's placement in your setup.
- When working with liquids containing particles, select tubing materials like PharMed BPT or Norprene for extended life and better performance.
8. Warranty and Support
For any questions, technical assistance, or warranty claims, please contact the seller or manufacturer directly. Provide your product model and purchase details for efficient support.