1. Introduction
The ADS8688 is an advanced 8-channel integrated data acquisition system, featuring a 16-bit successive approximation (SAR) analog-to-digital converter (ADC). It boasts a high throughput of up to 500kSPS, making it suitable for demanding data acquisition applications. This module integrates an analog front-end circuit with overvoltage protection up to ±20V for each input channel, an 8-channel multiplexer supporting both automatic and manual scan modes, and an on-chip 4.096V reference voltage with low temperature drift. Designed for flexibility, it operates from a single 5V analog power supply and supports various bipolar (±10.24V, ±5.12V, ±2.56V) and unipolar (0V to 10.24V, 0V to 5.12V) input ranges, configurable via software.
The gain of the analog front end is precisely fine-tuned across all input ranges to ensure high DC accuracy. Each channel's input range can be independently programmed. The device maintains a constant resistive input impedance of 1MΩ, irrespective of the selected input range.
2. Specifications
| Parameter | Value |
|---|---|
| Module Model | ADS8688 |
| Module Type | Analog-to-Digital Conversion Module (ADC) |
| Module Power Supply | 5V |
| Static Power Consumption | <2.2mA (for reference only) |
| Communication Protocol | SPI Serial |
| Input Impedance | 1M ohm |
| Number of Input Channels | 8 |
| Reference Voltage | Default internal 4.069V (weldable SOT-23 package external input reference) |
| Input Range | Bipolar: ±10.24V, ±5.12V, ±2.56V; Unipolar: 0V to 10.24V, 0V to 5.12V (software switching) |
| Resolution | 16 bits |
| Chip Maximum Sampling Rate | 500kSPS (total throughput) |
| Provided Routines | STM32F103RBT6 (analog SPI) |
| Module Features | Integrated Programmable Gain Amplifier (PGA), on-chip low temperature drift reference voltage, constant resistive input impedance |
| Typical Application Fields | Electric power automation, protection of middle-dimensional devices, factory automation, process control, education and learning |
| Module Weight | 13g |
| PCB Size | 50mm x 50mm (Length x Width) |
| Module Interface | 3.81-8P socket, XH2.54 double row pin |
| Battery Included | No |
| High-concerned chemical | None |
3. Setup and Installation
This section provides guidance on connecting and preparing your ADS8688 module for operation.
3.1. Power Supply
- The module is a low-power device; it is recommended that the power supply does not exceed 5.25V.
- For optimal performance and to avoid unnecessary interference, especially given the module's high precision, it is recommended to use a linear power supply.
3.2. Wiring and Connections
The module features a 3.81-8P socket for AD acquisition channels and XH2.54 double row pins for power supply and data interface. Ensure all connections are secure and correct before powering on the module.
For detailed wiring instructions and schematic diagrams, please refer to the User Manual (PDF).
4. Operating Instructions
4.1. Initial Test and Control
To perform a simple functional test of the module, it is recommended to use it in conjunction with a compatible control panel. After ensuring correct wiring, power the control panel to initiate signal acquisition and display. The provided routines are specifically for the STM32F103RBT6 (analog SPI) microcontroller.
4.2. Input Range Configuration
The ADS8688 allows for software-based switching of input ranges. Each channel's input range and polarity can be configured independently due to the integrated Programmable Gain Amplifier (PGA).
4.3. ADC Value to Voltage Conversion
The 16-bit resolution provides a digital output range. To convert the ADC value back to voltage:
- For Bipolar Range: 0V corresponds to 32767. The formula is:
V = (ADC_value - 32767) * Range / 65535 - For Unipolar Range: There is no need to subtract 32767. The formula is:
V = ADC_value * Range / 65535
Where 'Range' refers to the full-scale voltage range selected (e.g., 20.48V for ±10.24V, 10.24V for 0V to 10.24V).
4.4. Sampling Rate
The ADS8688 has a maximum total throughput of 500kSPS. The effective sampling rate per channel depends on the number of channels selected in the multiplexer scan sequence. For example:
- If 2 channels are selected, each channel's throughput is 250kSPS.
- If 4 channels are selected, each channel's throughput is 125kSPS.
- If all 8 channels are selected, each channel's throughput is approximately 62.5kSPS.
5. User Tips
- Power Supply Quality: For high-precision measurements, always use a stable linear power supply to minimize noise and ripple.
- Signal Integrity: Keep input and output signal lines as short as possible. Long or poor-quality cables can introduce noise or signal attenuation.
- Optimizing for Small Signals: If you are collecting small voltage signals and find the accuracy lacking, ensure that the input range (PGA setting) is appropriately configured for the expected signal amplitude. A range that is too large for a small signal will reduce effective resolution.
- Handling Unused Channels: If pins are not connected to a voltage source, they may show floating voltage readings. To ensure a true zero reading for unused channels, connect them directly to ground.
6. Troubleshooting
- Q: Can the module measure negative voltage?
- A: Yes, negative voltages can be measured when the input range is configured for bipolar operation.
- Q: The collected data fluctuates or jumps, even with a stable input. Is this normal?
- A: Data fluctuation can occur due to several factors, including power supply ripple noise, long signal wires, and other environmental interference. Ensure a stable power supply and short, shielded signal lines.
- Q: Why is the ADS8688 not accurate when collecting small voltage signals?
- A: This often happens if the configured input range (PGA setting) is too wide for the small signal. Adjust the chip's measuring range (PGA) to match the expected signal amplitude for better accuracy.
- Q: After the module is driven normally, pins not connected to a voltage source show a voltage reading. Is this normal?
- A: Yes, this is normal for floating pins. The module defaults to continuously collecting from all 8 channels. If a pin is not connected, it will pick up ambient electrical noise, resulting in a non-zero reading. To ensure a zero reading for an unused channel, it should be directly grounded.
6.1. Test Charts for Reference
7. Maintenance
- Power Considerations: Always adhere to the recommended power supply voltage (not exceeding 5.25V) to prevent damage to the module.
- Handling: As a high-precision device, handle the module with care to avoid physical damage or electrostatic discharge.
- Cleanliness: Keep the module clean and free from dust and debris, which can affect performance.
- Storage: Store the module in an anti-static bag in a dry environment when not in use.
8. Warranty and Support
This product comes with after-sales technical support. For any operational problems, technical inquiries, or support needs, please contact the seller directly. The module information PDF file and routine source code are provided to assist with development and functional verification.
Please note that the provided code is for the main control board and is fully commented. Functions beyond the basic display may require self-development.