YIERYI DO9100

Contents

Digital Dissolved Oxygen Meter

Model: DO9100 - Brand: YIERYI

Introduction

The YIERYI DO9100 is a precision instrument designed for measuring dissolved oxygen and temperature in various water environments. It features a built-in microcomputer processor that calculates and compensates all measurement data related to dissolved oxygen and temperature, ensuring accurate readings. The instrument's case is waterproof, making it suitable for field use. The dissolved oxygen probe utilizes a polarographic sensor with a permeable membrane and a thermistor for automatic temperature measurement and compensation. This membrane allows oxygen to pass through to react with the sensing electrode, enabling precise concentration measurement.

Package Contents

Please check the package carefully to ensure all items are present:

  • 1 x DO9100 Dissolved Oxygen Meter
  • 1 x Filling Liquid (for electrode)
  • 6 x Replaceable Membranes
  • 1 x User Manual
DO9100 Dissolved Oxygen Meter package contents
Image: Contents of the DO9100 Dissolved Oxygen Meter package, including the meter, probe, filling liquid, membranes, and manual.

Specifications

FeatureDescription
Model NumberDO9100
Brand NameYIERYI
Display TypeDIGITAL with Backlight
Is Smart DeviceNo
CertificationCE
OriginMainland China
Oxygen Concentration Range0.0 - 40.0 mg/L
Dissolved Oxygen Saturation Range0.0 - 300.0 %
Temperature Range0.1 - 40.0°C (32.0 - 104.0°F)
Oxygen Concentration Resolution0.1 mg/L
Dissolved Oxygen Saturation Resolution0.1 %
Temperature Resolution0.1°C / 0.1°F
Oxygen Concentration Precision±1.5% F.S.
Dissolved Oxygen Saturation Precision±1.5% F.S.
Temperature Precision±0.5℃
Calibration ModeSingle-point or dual-point automatic calibration (zero point or 100% in air)
Response Time<45s (fast response at 20℃)
Temperature Compensation RangeAutomatic (0-40℃)
Battery PowerLow battery warning
BacklightAutomatic backlight
Auto Shut-offApproximately 15 minutes
Power Supply3 x 1.5V button batteries (not included)
Operating Environment0.1 - 40.0℃, RH max 90%
Dimensions183 x 37 x 37 mm (7.2 x 1.5 x 1.5 in)
Weight80g
DO9100 Dissolved Oxygen Meter dimensions and parts
Image: Diagram showing the dimensions of the DO9100 meter and probe, along with labeled components like battery compartment, LCD display, power button, lock/temperature switch key, mode/calibration key, and electrode interface.
DO9100 Dissolved Oxygen Meter features
Image: Key features of the DO9100 meter, including response time (≤45s), low power warning, backlight display, and automatic shutdown (approx. 8 minutes).
Table of oxygen saturation concentration at different temperatures
Image: Table illustrating the saturation concentration of oxygen (mg/L and %) at various temperatures (°C and °F).

Setup

  1. Install Batteries: Open the battery compartment (located at the top of the meter, under the cap) and insert 3 x 1.5V button batteries, ensuring correct polarity. Close the compartment securely.

    DO9100 Dissolved Oxygen Meter battery compartment
    Image: Close-up of the DO9100 meter's battery compartment, showing where to insert 3x1.5V button batteries.
  2. Install Electrode: Connect the dissolved oxygen electrode probe to the meter's electrode interface at the bottom. Ensure the metal knobs are tightened securely to prevent it from falling off.

    DO9100 Dissolved Oxygen Meter probe connection
    Image: The highly sensitive probe of the DO9100 meter, showing the metal knob connector for secure attachment to the main unit.

    Video: Demonstration of installing the electrode onto the dissolved oxygen meter.

Operation

  1. Power On: Press the power button to turn on the instrument. The screen will display the temperature and the dissolved oxygen reading (initially in mg/L).

  2. Taking Measurements: Ensure the instrument has been calibrated. Immerse the probe directly into the solution to be tested. Wait for the value on the screen to stabilize before recording the reading.

    DO9100 Dissolved Oxygen Meter measurement process
    Image: Illustration of the DO measurement process, showing the meter and probe immersed in a liquid, with the display showing dissolved oxygen in mg/L and temperature.
  3. Switching Units (mg/L or %): Short press the 'MODE/CAL' button to switch the dissolved oxygen display between mg/L (oxygen concentration) and % (dissolved oxygen saturation).

  4. Hold Function: Short press the 'HOLD/TEMP' button to lock the current reading on the display.

  5. Switching Temperature Units (°C or °F): Long press the 'HOLD/TEMP' button to switch the temperature display between Celsius (°C) and Fahrenheit (°F).

  6. Automatic Temperature Compensation (ATC): The instrument automatically compensates for temperature effects on dissolved oxygen measurements within the range of 0-40°C. This ensures accuracy despite temperature fluctuations in the sample.

    DO9100 Dissolved Oxygen Meter automatic temperature compensation
    Image: Diagram explaining the automatic temperature compensation feature of the DO9100 meter, showing how it adjusts for temperatures between 0-40°C.
  7. Automatic Shutdown: The meter will automatically shut down after approximately 15 minutes of inactivity to conserve battery life.

Calibration

The DO9100 supports one-point or two-point automatic calibration. Ensure the probe is installed correctly before calibration.

Correct and incorrect probe usage for DO9100
Image: Visual guide showing the correct way to measure (probe directly in solution without protection cover) versus an incorrect way (with protection cover on).

Important Note: Do NOT open the electrode protection cover during measurement and calibration. Put the electrode directly into the solution to measure.

  1. Air Calibration (100% Saturation):

    • With the instrument powered on and the probe exposed to air (not in water), wait for the value to stabilize.
    • Long press the 'MODE/CAL' button for 5 seconds.
    • Release the button. The screen will flash '100.0%' three times.
    • Calibration in air is complete.
    DO9100 Dissolved Oxygen Meter air calibration steps
    Image: Step-by-step guide for air calibration (100% saturation) of the DO9100 meter. It shows pressing the CAL button for 5 seconds until the screen flashes 100%.

    Video: Demonstration of air calibration for the dissolved oxygen meter.

  2. Oxygen-Free Water Calibration (0% Saturation):

    • With the instrument powered on, immerse the probe into oxygen-free water (anaerobic water). Wait for the value to stabilize.
    • Long press the 'MODE/CAL' button for 5 seconds.
    • Release the button. The screen will flash '0.0%' three times.
    • Calibration in oxygen-free water is complete.
    DO9100 Dissolved Oxygen Meter oxygen-free water calibration steps
    Image: Step-by-step guide for oxygen-free water calibration (0% saturation) of the DO9100 meter. It shows immersing the probe in anaerobic water and pressing the CAL button for 5 seconds until the screen flashes 0%.

    Video: Demonstration of oxygen-free water calibration for the dissolved oxygen meter.

Maintenance

  1. Electrode Care: Always keep the electrode clean. After each use, rinse the probe with distilled water and gently wipe it dry. Store the probe with its protective cap on.

  2. Membrane Replacement: The dissolved oxygen probe uses a permeable membrane that may need replacement over time due to wear or damage. The package includes 6 replaceable membranes. Refer to the detailed instructions in the included user manual for the specific procedure to replace the membrane.

    DO9100 Dissolved Oxygen Meter replaceable probe covers
    Image: The detachable probe of the DO9100 meter, showing the probe and several protective covers, indicating the replaceable nature of the probe components.
  3. Filling Liquid: The electrode requires a specific filling liquid for optimal performance. The package includes one bottle of filling liquid. Ensure the electrode is properly filled according to the instructions in the user manual. If you need more filling liquid, ensure you purchase the correct type for dissolved oxygen meters.

    Bottle of electrode filling liquid
    Image: A bottle labeled 'pH Probe' (likely intended as filling liquid for the DO meter electrode), indicating the type of solution used for electrode maintenance.
  4. Battery Replacement: When the low battery warning appears on the display, replace all three 1.5V button batteries promptly to ensure continuous accurate operation.

Troubleshooting

  • No Display/Power On Issue: Check if the batteries are correctly installed and have sufficient charge. Replace batteries if necessary.

  • Unstable Readings: Ensure the probe is clean and properly immersed in the solution. Verify that the electrode membrane is intact and the filling liquid is sufficient. Recalibrate the instrument if readings remain unstable.

  • Calibration Failure: Ensure the calibration environment (air or oxygen-free water) is correct. Make sure the probe is clean and the membrane is in good condition. Repeat the calibration steps carefully.

  • Inaccurate Readings: Perform a calibration. Check the operating temperature of the environment and the sample; while the device has automatic temperature compensation, extreme temperatures outside the specified range (0-40°C) can affect accuracy.

User Tips

  • Filling Liquid Clarification: The 'Filling Liquid' option is an accessory for the Dissolved Oxygen Meter, not a complete product. If you ordered 'Filling Liquid' expecting a full meter, please note it is a consumable for electrode maintenance.

  • Standalone Device: This device is designed as a standalone dissolved oxygen meter and cannot be directly connected to a microcontroller for data output or control.

  • Environmental Conditions: While the meter has automatic temperature compensation, it is designed for use within a specified operating environment of 0.1-40.0℃. Using it in freezing conditions (e.g., frost in winter) may affect its performance and accuracy.

  • Wide Application: The DO9100 is suitable for a wide range of water quality testing applications, including sewage disposal, farming, aquaculture, aquarium maintenance, swimming pool monitoring, and laboratory use.

    DO9100 Dissolved Oxygen Meter in aquarium
    Image: The DO9100 meter being used to measure oxygen concentration and saturation in an aquarium, demonstrating a common application.
    DO9100 Dissolved Oxygen Meter wide applications
    Image: Various applications of the DO9100 meter, including sewage disposal, farming, aquaculture, aquariums, swimming pools, and laboratories.

Warranty and Support

This product comes with a standard manufacturer's warranty. For specific warranty details, please refer to the documentation provided with your purchase or contact the seller directly. For technical support or inquiries, please reach out to the customer service department of the retailer or manufacturer.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.