1. Introduction
The yieryi DO9100 Digital Dissolved Oxygen Meter is a precision instrument designed for measuring dissolved oxygen levels and temperature in various aquatic environments. It features a wide measurement range, automatic temperature compensation (ATC), and a clear LCD backlight display, making it suitable for applications in aquariums, swimming pools, ponds, aquaculture, and research.
2. Safety Information
- Always handle the electrode with care to prevent damage.
- Keep the device away from strong magnetic fields and high temperatures.
- Do not attempt to disassemble or repair the device yourself. Contact qualified personnel for service.
- Ensure the battery compartment is properly sealed to prevent water ingress.
- Dispose of batteries and the device according to local regulations.
3. Package Contents
Verify that all items are present and in good condition upon unboxing:
- 1 x Dissolved Oxygen Meter (DO9100)
- 1 x Dissolved Oxygen Probe
- 6 x Probe Protective Covers
- 1 x Electrode Filling Fluid (30ml bottle)
- 1 x English Instruction Manual
4. Product Overview
The DO9100 meter consists of a main unit with an LCD display and control buttons, and a detachable probe connected via a cable.
Key Features:
- HD Backlight Display: Provides clear readings even in low light conditions.
- Highly Sensitive Probe: Detachable and easy to replace, with metal knobs for secure attachment.
- Automatic Temperature Compensation (ATC): Ensures accurate measurements across a temperature range of 0-40°C.
- Power Supply: Powered by 3*1.5V button batteries for long life.
- Automatic Shutdown: Device automatically turns off after approximately 15 minutes of inactivity to conserve battery.
- Low Power Warning: Indicates when batteries need replacement.
5. Specifications
| Parameter | Value |
|---|---|
| Model Number | DO9100 DO tester |
| Brand Name | yieryi |
| Origin | Mainland China |
| Certification | CE, RoHS |
| Display Type | DIGITAL (with backlight) |
| Dissolved Oxygen Range | 0.0-40.0 mg/L |
| Dissolved Oxygen Saturation Range | 0.0-300.0% |
| Temperature Range | 0.1-40.0°C (32.0-104.0°F) |
| Dissolved Oxygen Resolution | 0.1 mg/L |
| Dissolved Oxygen Saturation Resolution | 0.1% |
| Temperature Resolution | 0.1°C, 0.1°F |
| Dissolved Oxygen Accuracy | ±1.5% F.S. |
| Temperature Accuracy | ±0.5°C |
| Response Time | ≤45s (at 20°C) |
| Temperature Compensation | ATC (0-50°C) |
| Power Supply | 3*1.5V button battery |
| Automatic Shutdown | Approx. 15 minutes |
| Meter Size | 183*37*37mm |
6. Setup
6.1 Battery Installation
The DO9100 meter is powered by 3*1.5V button batteries. To install or replace them, open the battery compartment located at the top of the main unit (refer to Figure 4.2).
6.2 Electrode Installation
Connect the dissolved oxygen probe to the main meter unit via the electrode interface. Ensure the metal knobs are securely tightened to prevent accidental disconnection during use.
7. Operation
7.1 Power On/Off
Press the power button (labeled with a circle and line icon) to turn the device on or off.
7.2 Measurement Procedure
- Ensure the instrument has been calibrated (refer to Section 8).
- Immerse the probe directly into the solution to be tested. Do not open the electrode protection cover during measurement or calibration.
- Wait for the reading on the LCD display to stabilize.
- The dissolved oxygen value will be displayed in mg/L or %. The temperature will be displayed below it.
- Short press the "MODE/CAL" button to switch between mg/L and % display modes.
7.3 Automatic Temperature Compensation (ATC)
The instrument is equipped with ATC, which automatically adjusts for temperature variations between 0-40°C to ensure accurate dissolved oxygen readings. Oxygen solubility is affected by temperature, atmospheric pressure, and salinity, and the ATC feature helps maintain measurement accuracy under varying conditions.
8. Calibration
The DO9100 meter supports one-point or two-point calibration. Ensure the probe is correctly installed before calibration.
8.1 Air Calibration (100.0%)
This calibrates the meter to 100% dissolved oxygen saturation in air.
- Turn on the instrument.
- Ensure the probe is exposed to air (not immersed in water).
- Wait for the value to stabilize.
- Long press the "MODE/CAL" button for 5 seconds and then release it.
- The screen will flash "100.0%" three times, indicating that the air calibration is complete.
8.2 Oxygen-Free Water Calibration (0.0%)
This calibrates the meter to 0% dissolved oxygen saturation using oxygen-free water.
- Prepare oxygen-free water (e.g., by adding sodium sulfite to distilled water).
- Immerse the probe into the oxygen-free water.
- Wait for the value to stabilize.
- Long press the "MODE/CAL" button for 5 seconds and then release it.
- The screen will flash "0.0%" three times, indicating that the oxygen-free water calibration is complete.
For two-point calibration, perform both air calibration and oxygen-free water calibration sequentially.
9. Maintenance
9.1 Probe Care
- After each use, rinse the probe with distilled water to remove any residue.
- Store the probe with its protective cover on and ensure it is kept moist with electrode filling fluid or a suitable storage solution. Do not allow the membrane to dry out.
- If readings become inaccurate or calibration fails, consider replacing the membrane sleeve and refilling the electrode with fresh fluid. The kit includes 6 probe protective covers for replacement.
9.2 Battery Replacement
When the low battery indicator appears on the display, replace the 3*1.5V button batteries. Ensure correct polarity during installation.
9.3 Storage
Store the meter and probe in a cool, dry place, away from direct sunlight and extreme temperatures. Ensure the probe is properly maintained as described above.
10. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Inaccurate readings | Probe not calibrated; dirty probe membrane; depleted electrode filling fluid; damaged probe. | Perform calibration (Section 8). Clean the probe. Replace electrode filling fluid. Replace the probe membrane or the entire probe if damaged. |
| Display not turning on | Dead batteries; incorrect battery installation. | Replace batteries. Check battery polarity. |
| Calibration failure | Probe not stable; incorrect calibration procedure; faulty probe. | Ensure probe is stable during calibration. Follow calibration steps precisely. Replace probe if necessary. |
| "----" displayed on screen | Factory reset initiated. | This indicates a successful factory reset. Proceed with calibration. |
10.1 Restore Factory Settings
To restore the instrument to factory default settings:
- When the instrument is turned on, press and hold the power button for 5 seconds.
- Release the button. The screen will show "----" to confirm that factory settings have been successfully restored.
11. User Tips
- For consistent and accurate results, always calibrate the meter before critical measurements or if it hasn't been used for a long period.
- When measuring, ensure the probe is fully immersed in the water and allow sufficient time for the reading to stabilize.
- Regularly check the electrode filling fluid level and replace the probe membrane sleeve if it appears worn or damaged.
12. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the contact information provided by your retailer or visit the official yieryi website. Keep your purchase receipt as proof of purchase.