1. Introduction
This manual provides essential information for the safe and efficient operation of your A-BF Laboratory DC Power Supply Unit. These units are designed for high precision, stability, and reliability, featuring a 4-digit display and memory functions for various applications.
Key Features:
- Complete operating protection functions (Overload, Over-temperature, Short Circuit).
- Highly efficient design with low ripple, small volume, and light weight.
- Double display for voltage and current with high precision (0.01V/0.01A resolution).
- Voltage stabilization and current limiting capabilities.
- Dual voltage input for convenient choice (110V/220V versions).
- Temperature-controlled fan for heat dissipation and low noise.
- Double potentiometers for more precise adjustments.
- Memory function (M1, M2, M3) for saving and recalling settings.
2. Product Overview
The A-BF DC Power Supply is a versatile instrument suitable for a wide range of electronic testing and maintenance tasks. It features an intuitive front panel with a clear 4-digit display and precise control knobs.
Front Panel (MP Series Example):

Image: Front and Rear Panel Diagram for MP Series DC Power Supply. This diagram illustrates the various controls and indicators on the front and rear panels, including voltage/current displays, control knobs, memory buttons, and power connections.
- Voltage Display (V): Shows the output voltage.
- Current Display (A): Shows the output current.
- Power Display (W): Shows the output power.
- CV/CC Indicators: Constant Voltage (CV) and Constant Current (CC) mode indicators.
- Key Lock Display: Indicates when buttons are locked.
- Voltage Regulation Knob: Adjusts the output voltage.
- Current Regulation Knob: Adjusts the output current.
- Output Switch Button (ON/OFF): Toggles power output.
- USB Interface (5V/2A): For charging or powering USB devices (Note: SS-3010MF/3010MP shipped from Russia may not have this feature).
- Output Terminals (+/-): Connect test leads here.
- M1-M3 Store/Recall Buttons: Save and recall frequently used voltage/current settings.
- OCP (Over Current Protection) Button: Sets overcurrent protection.
- OVP (Over Voltage Protection) Button: Sets overvoltage protection.
Rear Panel:
- Power Switch Button: Main power switch.
- Cooling Fan: Provides active cooling for continuous operation.
- AC Power Input Socket: Connects to mains power.
- Fuse Holder: Contains the protective fuse.
- SN: Serial Number of the unit.
3. Setup
- Unpacking: Carefully remove the power supply unit and all accessories from the packaging. Verify that all components listed in the 'Package Contents' section are present.
- Placement: Place the unit on a stable, level surface with adequate ventilation. Ensure the cooling fan at the rear is not obstructed.
- Power Connection: Connect the provided power cable to the AC Power Input Socket on the rear panel and then to a suitable mains outlet. Ensure the correct plug type (EU, UK, AU, US) is used for your region.

Image: Various plug types (EU, UK, AU, US) for regional compatibility.
- Test Lead Connection: Connect your test leads to the red (+) and black (-) output terminals on the front panel. Ensure a secure connection.
4. Operating Instructions
4.1 Basic Operation
- Power On: Flip the Power Switch button on the rear panel to the 'ON' position. The display will light up.
- Adjusting Voltage:
Press the 'U' button (Voltage) on the front panel. Rotate the Voltage Regulation knob to adjust the desired voltage. Press the knob to confirm the setting.Video: Demonstration of adjusting voltage on the A-BF DC Power Supply. Press 'U', rotate the knob, then press to confirm.
- Adjusting Current:
Press the 'I' button (Current) on the front panel. Rotate the Current Regulation knob to adjust the desired current limit. Press the knob to confirm the setting.Video: Demonstration of adjusting current on the A-BF DC Power Supply. Press 'I', rotate the knob, then press to confirm.
- Output ON/OFF: Press the 'ON/OFF' button to enable or disable the power output to your connected device. The button will illuminate when output is active.
4.2 Advanced Functions
- Memory Function (M1, M2, M3):
To save a setting: Adjust the voltage and current to your desired values. Long press one of the 'M1', 'M2', or 'M3' buttons until the display indicates the value is saved. The corresponding 'M' button will light up.
To recall a setting: Short press the 'M1', 'M2', or 'M3' button to load the saved voltage and current values.Video: Demonstration of using the memory function (M1, M2, M3) to save and recall voltage/current settings.
- Button Lock Function:
To lock the buttons: Long press the main adjustment knob. A lock icon will appear on the display, indicating that most buttons are locked. In this state, only the 'ON/OFF' button remains active to prevent accidental changes to settings.
To unlock: Long press the main adjustment knob again.Video: Demonstration of locking and unlocking the control buttons. When locked, only the output button is functional.
- Over Current Protection (OCP) and Over Voltage Protection (OVP):
These functions are designed to protect your connected device and the power supply itself. When OCP or OVP is activated and the set current or voltage limit is exceeded, the output will be blocked to prevent damage. Refer to the full user manual for detailed instructions on setting OCP and OVP limits.
5. Application Fields
The A-BF DC Power Supply is suitable for a variety of professional and educational environments:

Image: Visual representation of the power supply's diverse applications, including enterprise production lines, electrical repair and maintenance, product aging tests, circuit board repair, and laboratory research and teaching.
- Production line testing.
- Mobile phone, computer, and electrical appliance maintenance.
- Automotive electronics (Car DVD, car audio).
- Product aging tests (resistors, capacitors, relays, DC motors, LEDs).
- Battery charging.
- Laboratory research and teaching.
- General factory and maintenance power supply needs.
6. Maintenance
To ensure the longevity and optimal performance of your A-BF DC Power Supply, follow these maintenance guidelines:
- Cleaning: Regularly clean the exterior of the unit with a soft, dry cloth. Avoid using abrasive cleaners or solvents. Ensure no dust accumulates in the ventilation holes.
- Ventilation: Keep the cooling fan and ventilation holes clear of obstructions to prevent overheating. The unit is designed with four-sided cooling holes and an internal cooling fan for efficient heat dissipation.

Image: Detailed view of the internal components, cooling fan, and four-sided cooling holes, emphasizing the robust design and effective heat management.
- Fuse Replacement: If the unit fails to power on, check the fuse located in the fuse holder on the rear panel. Replace it with a fuse of the same type and rating if necessary. Always disconnect the power cable before attempting fuse replacement.
- Storage: When not in use for extended periods, store the power supply in a dry, dust-free environment within the specified storage temperature range (-20°C to 60°C, relative humidity ≤80%RH).
7. Troubleshooting
7.1 Unit Does Not Power On
- Check if the power cord is securely connected to both the unit and the mains outlet.
- Verify that the mains outlet has power by testing with another device.
- Inspect the power cord for any visible damage.
- Check the fuse in the fuse holder on the rear panel. Replace if blown (refer to Maintenance section).
- If there is voltage output from the power cord but the unit still doesn't turn on, carefully remove the power supply cover and check for any loose internal cables. Reinsert any loose cables and try again.
7.2 OCP/OVP Activation
If the output is blocked and the OCP or OVP indicator is on, it means the set overcurrent or overvoltage limit has been reached. These functions are designed to protect your circuit. To resume operation, reduce the load or adjust the voltage/current settings to be within the set limits, then reset the OCP/OVP if necessary.
8. Specifications
| Parameter | Value |
|---|---|
| Brand Name | A-BF |
| Model Numbers | SS-4020HP, SS-1605HP, SS-8010HP, SS-3010MF, SS-3010MP, SS-305MF, SS-305MP |
| Output Type | SINGLE |
| Output Power | 201 - 300W (varies by model) |
| Origin | Mainland China |
| Certification | CE, FCC |
| Voltage Load Regulation Rate | ≤0.1%+5mV |
| Voltage Power Regulation Rate | ≤0.01%+5mV |
| Voltage Setting Accuracy | ≤0.1%+1 digits |
| Voltage Readback Accuracy | ≤0.1%+1 digits |
| Voltage Ripple | 10mVrms |
| Current Load Regulation Rate | ≤0.2%+3mA |
| Current Power Regulation Rate | ≤0.2%+3mA |
| Current Readback Accuracy | ≤0.2%+3digits |
| Current Ripple | 8mVrms |
| OVP (Over Voltage Protection) | 0-Vmax±0.2%FS |
| Maximum Voltage | Vmax±0.2% |
| OCP (Over Current Protection) | 0-Cmax±0.2%FS |
| Maximum Current | Cmax±0.2% |
| Working Conditions | Temperature: 0~40℃, Relative humidity: ≤80%RH |
| Storage Temperature | Temperature: -20~60℃, Relative humidity: ≤80%RH |
| Operating Voltage | AC220V±10%50Hz (or 110V version) |
| Cooling Method | Air-cooled |
| Weight | 1.95KG |
| Size | 230*150*80mm |
Note: Specific voltage, current, and power ratings vary by model. Refer to the product label or the table below for your specific unit.

Image: Table detailing the output voltage, current, and power for various A-BF DC Power Supply models.
9. User Tips
- Memory Retention: The power supply retains the last set voltage and current values when turned off and on, provided the adjustment knobs are not moved.
- Protection Functions: OCP (Over Current Protection) and OVP (Over Voltage Protection) are crucial safety features. They are designed to block the output when predefined current or voltage limits are exceeded, protecting both the power supply and the connected load.
- RS485 Communication: Some models may feature an RS485 serial port for remote control and data logging. Details on the communication protocol can be found in the comprehensive user manual or by contacting customer support.
10. Package Contents
Upon opening your package, you should find the following items:
- 1 x A-BF DC Power Supply Unit
- 1 x Power Cable (appropriate for your region)
- 1 x Test Line (set of leads)
- 1 x Product Manual & Certificate

Image: Contents of the product package, including the DC power supply, test leads, power cord, user manual, and certificate.
11. Warranty and Support
Your A-BF DC Power Supply Unit comes with a 3-year warranty, ensuring quality and reliability. For detailed warranty terms, technical support, or further inquiries, please refer to the complete user manual included in your package or contact A-BF customer service.

Image: Certificates of Conformity (CE, FCC) for the A-BF DC Power Supply, indicating compliance with relevant safety and quality standards.