GVDA GD112 Series

Contents

GVDA GD112 Series Digital Clamp Meter User Manual

Models: GD112A, GD112B, GD112C

Brand: GVDA

1. Introduction

The GVDA GD112 Series Digital Clamp Meter is an intelligent true effective value multimeter designed for professional electricians, engineers, electronic enthusiasts, and general household use. It integrates intelligent identification and manual function modes, allowing for precise measurement of various electrical parameters.

This series includes models GD112A, GD112B, and GD112C, each offering a range of functionalities to suit different measurement needs. The meter can measure AC and DC voltage, AC and DC current, resistance, capacitance, continuity, diode, NCV (Non-Contact Voltage) detection, and more. Models GD112B and GD112C also feature temperature measurement and True RMS capabilities.

GVDA GD112B Digital Clamp Meter with test leads

Figure 1: GVDA GD112B Digital Clamp Meter with included test leads.

2. Safety Information

Please read and understand all safety warnings and operating instructions before using this instrument. Failure to follow these instructions could result in electric shock, fire, or serious injury.

  • Safety Rating: This device is rated EN61010-1, -2-030, EN61010-2-033, EN61326-1 CAT III 600V. Always adhere to these safety standards.
  • High Voltage/Current Warning: When measuring current greater than 3A or voltage greater than 80V, the LCD backlight will turn orange to warn the user of potentially dangerous conditions. Exercise extreme caution.
  • Current Measurement: When measuring AC current using the clamp, ensure you clamp only a single live line or a single neutral line. DO NOT clamp both lines combined together, as this will result in an inaccurate reading of zero.
Diagram showing correct and incorrect clamping for current measurement

Figure 2: Correct (right) and incorrect (left) clamping methods for current measurement. Always clamp a single wire.

  • Do not operate the meter if it appears damaged or if the test leads are damaged.
  • Ensure the battery cover is securely closed before operation.
  • Do not exceed the maximum input values for any function.
  • Always disconnect the test leads from the circuit before changing functions.

3. Product Features

  • Maximum Display: 6000 Counts digital display for high resolution readings.
  • Temperature Measurement: (Only for GD112B and GD112C) Supports K-type thermocouple probe for contact temperature measurement in the range of -20°C to 1000°C (-4°F to 1832°F).
  • NCV Detection: Non-Contact Voltage detection feature (Only for GD112B and GD112C) for identifying live wires without direct contact.
  • Large Jaw Design: 26mm jaw opening for clamping around larger conductors.
  • Safety Backlight: Orange backlight warning for voltages >80V or currents >3A.
  • Working Environment: Designed for operation between 0-40°C with relative humidity less than 80%.
  • True RMS Value: (Only for GD112B and GD112C) Provides accurate measurements for non-sinusoidal waveforms.
  • Smart Function: Automatic identification of current input for simplified operation.
  • Additional Features: Diode test, Bargraph display, Data Hold, Fuse Burning Prompt, Low battery indication, Auto Power Off, Flashlight, LowZ (Low Impedance), LPF (Low Pass Filter).
Image highlighting GD112A features: 6000 Counts, AC DC Voltage, AC Current, Resistance, Max/Min, Continuity, Auto Power Off, Data Hold, Flashlight, Double Backlight

Figure 3: Key features of the GD112A model.

Image showing NCV detection in use near a wall socket

Figure 4: NCV (Non-Contact Voltage) detection in action (GD112B/GD112C only).

4. Components and Controls

Familiarize yourself with the different parts of your clamp meter:

Diagram of GD112A clamp meter components

Figure 5: GD112A Components (Clamp, Flashlight, Date Hold, Knob Switch, MAX/MIN, Flashlight/Backlight, Function Selection, Display Screen, Measuring Input Jack, Trigger).

Diagram of GD112B/GD112C clamp meter components

Figure 6: GD112B/GD112C Components (NCV Probe, Clamp, Flashlight, Date Hold/NCV Key, Knob Switch, MAX/MIN, Flashlight/Backlight, Function Select/Zero, Display Screen, Measuring Input Jack, Trigger).

5. Setup

5.1. Battery Installation

The GVDA GD112 Series Clamp Meters are powered by two 1.5V AAA batteries (not included). To install or replace batteries:

  1. Ensure the meter is turned OFF and disconnect any test leads from the input jacks.
  2. Locate the battery compartment cover on the back of the meter.
  3. Use a screwdriver to loosen the screw(s) on the battery cover.
  4. Remove the battery cover.
  5. Insert two 1.5V AAA batteries, observing the correct polarity (+ and -) as indicated inside the compartment.
  6. Replace the battery cover and tighten the screw(s) securely.
Image showing the battery compartment of the clamp meter with two AAA batteries

Figure 7: Battery compartment for 2x 1.5V AAA batteries.

6. Operating Instructions

6.1. Power On/Off

Turn the rotary knob from the 'OFF' position to any desired function to power on the meter. To power off, turn the rotary knob back to the 'OFF' position.

6.2. Function Selection

Rotate the central knob to select the desired measurement function (e.g., ACV, DCV, ACA, DCA, Ohm, Capacitance, Hz, Temperature, Diode, Continuity). For some functions, the 'FUNC' button may be used to cycle through sub-functions (e.g., AC/DC voltage, Diode/Continuity).

6.3. AC/DC Voltage Measurement

  1. Turn the rotary knob to the 'V' (Voltage) position. Use the 'FUNC' button if necessary to select AC or DC voltage.
  2. Insert the red test lead into the 'V' input jack and the black test lead into the 'COM' input jack.
  3. Connect the test probes across the circuit or component to be measured.
  4. Read the voltage value on the display.

6.4. AC/DC Current Measurement (Clamp)

  1. Turn the rotary knob to the 'A' (Current) position. Use the 'FUNC' button if necessary to select AC or DC current (GD112C only for DC Current).
  2. Press the trigger to open the clamp jaws.
  3. Place only ONE conductor (live or neutral) within the clamp jaws. Ensure the jaws are fully closed.
  4. Read the current value on the display.

6.5. Resistance Measurement

  1. Turn the rotary knob to the 'Ω' (Resistance) position.
  2. Insert the red test lead into the 'V' input jack and the black test lead into the 'COM' input jack.
  3. Connect the test probes across the component to be measured. Ensure the circuit is de-energized.
  4. Read the resistance value on the display.

6.6. Capacitance Measurement (GD112B/GD112C only)

  1. Turn the rotary knob to the 'Capacitance' position.
  2. Insert the red test lead into the 'V' input jack and the black test lead into the 'COM' input jack.
  3. Connect the test probes across the capacitor. Ensure the capacitor is discharged before measurement.
  4. Read the capacitance value on the display.

6.7. Frequency and Duty Cycle Measurement (GD112B/GD112C only)

  1. Turn the rotary knob to the 'Hz%' (Frequency/Duty Cycle) position.
  2. Insert the red test lead into the 'V' input jack and the black test lead into the 'COM' input jack.
  3. Connect the test probes across the signal source.
  4. Read the frequency or duty cycle value on the display.

6.8. Temperature Measurement (GD112B/GD112C only)

  1. Turn the rotary knob to the '°C/°F' (Temperature) position.
  2. Connect a K-type thermocouple probe to the meter's input jacks (observing polarity).
  3. Place the tip of the thermocouple probe on the object or in the environment whose temperature is to be measured.
  4. Read the temperature value on the display.
Image showing temperature measurement using a K-type probe in a glass of water

Figure 8: Temperature measurement with a K-type probe (GD112B/GD112C only).

6.9. NCV (Non-Contact Voltage) Detection (GD112B/GD112C only)

  1. Turn the rotary knob to the 'NCV' position or press the 'NCV' button (if available).
  2. Move the NCV probe (top part of the clamp meter) close to the conductor or outlet.
  3. The meter will emit an audible alarm and the display will indicate the presence of AC voltage.

6.10. Diode Test and Continuity

  1. Turn the rotary knob to the 'Diode/Continuity' position. Use the 'FUNC' button to select between diode test and continuity.
  2. For Diode Test: Connect the red test lead to the anode and the black test lead to the cathode of the diode. The forward voltage drop will be displayed.
  3. For Continuity: Connect the test probes across the circuit or component. An audible beep indicates continuity (low resistance).

6.11. Data Hold, MAX/MIN, Flashlight

  • Data Hold: Press the 'HOLD' button to freeze the current reading on the display. Press again to release.
  • MAX/MIN: Press the 'MAX/MIN' button to record the maximum or minimum reading during a measurement session.
  • Flashlight: Press the 'Flashlight' button to turn on the built-in LED flashlight for illuminating dark work areas.
Image showing the clamp meter's LED flashlight illuminating wires in a dark electrical panel

Figure 9: Using the built-in LED flashlight.

7. Specifications

FunctionGD112AGD112BGD112C
Operating ModeManual RangeAuto RangeAuto Range
AC Voltage600V600mV/6V/60V/600V600mV/6V/60V/600V
DC Voltage600V6V/60V/600V6V/60V/600V
AC Current6A/60A/600A
DC CurrentXX6A/60A/600A
Resistance6KΩ/60KΩ600Ω/6KΩ/60KΩ/600KΩ/6MΩ/60MΩ
CapacitanceX10nF/100nF/1uF/10uF/100uF/1mF/10mF/100mF
FrequencyX10Hz/100Hz/1kHz/10kHz/100kHz/1mHz/10mHz
Duty CycleX1%-99%
TemperatureX-20~1000°C, -4~1832°F
Display Counts6000 Counts
Max/MinVVV
ContinuityVVV
Auto OffVVV
Data HoldVVV
Double BacklightV (>3A, >80V)
FlashlightVVV
Diode TestXVV
NCV DetectionXVV
True RMSXV T-RMSV T-RMS
LowZXVV
LPFXVV
Powered By2x 1.5V AAA Batteries (not included)
Product Size193mm x 73mm x 34mm
Jaw Size26mm
Safety RatingEN61010-1, -2-030, EN61010-2-033, EN61326-1 CAT III 600V

8. Maintenance

8.1. Cleaning

Wipe the meter with a dry, soft cloth. Do not use abrasives or solvents. Keep the input terminals free of dirt and moisture.

8.2. Battery Replacement

When the low battery indicator appears on the display, replace the batteries as described in the 'Battery Installation' section (5.1). Prompt battery replacement ensures accurate readings and proper device function.

8.3. Storage

If the meter is not to be used for an extended period, remove the batteries to prevent leakage and damage. Store the meter in a cool, dry place, away from direct sunlight and extreme temperatures.

9. Troubleshooting

  • Meter does not power on: Check if batteries are installed correctly and have sufficient charge. Replace batteries if necessary.
  • Inaccurate readings: Ensure test leads are properly connected and not damaged. Verify the correct function is selected for the measurement. For current, ensure only one conductor is clamped.
  • Display shows 'OL' (Overload): The measured value exceeds the selected range. Switch to a higher range if available, or ensure the input is within the meter's maximum specifications.
  • Auto Power Off: The meter features an auto power-off function to conserve battery life. If the meter turns off unexpectedly, simply turn the rotary knob to 'OFF' and then back to the desired function to restart.
  • Ammeter accuracy in DC mode: The GD112C model is designed to measure DC current accurately. If you suspect inaccuracies, ensure proper clamping technique and check battery levels.

10. User Tips

  • Safety First: Always be aware of the CAT III 600V safety rating and the orange backlight warning for high voltage/current. Prioritize safety in all electrical work.
  • DC Current Measurement: For models supporting DC current (GD112C), ensure the meter is set to the correct DC current range and that only a single conductor is clamped for accurate readings.
  • Auto-Off Feature: The auto power-off feature is designed to save battery. If you need continuous measurement, be prepared to reactivate the meter or check if your model has a disable feature for auto-off (refer to specific model manual if available).
  • Temperature Probe: For temperature measurements (GD112B/GD112C), ensure you are using a compatible K-type thermocouple probe.

11. Warranty and Support

11.1. Warranty

This product comes with a 24-month manufacturer's limited warranty for defective items from the date of receipt. Accessories typically come with a 3-month warranty. The warranty does not cover items damaged or misused after receipt. Defective items must be reported and returned within the warranty period, in their original packaging if possible, with a clear description of the defect and your order number.