1. Introduction
The JUNCTEK LC-200A is a high-precision digital capacitance and inductance meter designed for accurate measurements of small inductance and capacitance values. Utilizing the LC resonant principle and high-speed microcontroller computation, it offers a wide measurement range and high accuracy, capable of measuring inductance below 1uH and capacitance less than 1pF. Its compact size and light weight make it highly portable. The device features flexible online calibration to maintain measurement accuracy without the need for potentiometer adjustments, as calibration parameters are stored in FLASH memory, ensuring data retention even after power loss.
For a comprehensive guide, please refer to the official PDF user manual: LC200A_EN_manual.pdf
2. Package Contents
Upon receiving your LC-200A meter, please verify that all items listed below are included and in good condition. If any items are missing or damaged, please contact your supplier.

Image: LC-200A meter and its standard accessories.
- 1 x LC-200A Portable L/C Meter
- 2 x Test clips (Red/Black banana plugs to crocodile clip test leads)
- 1 x Mini USB cable
- 1 x User manual (PDF, referenced above)
- Optional: DC 5V adapter
- Optional: Chip testing clip (included with LC-200A-SMD package)
3. Instrument Overview
3.1 Structure Description
The LC-200A meter features a clear 1602 LCD display and intuitive controls. Below is a diagram illustrating the main components and their functions.

Image: Front and side views of the LC-200A with key components labeled.
- 1602 LCD Display: Shows measurement readings and status.
- Zero Button (Red): Used for calibration (zeroing) in both capacitance and inductance modes.
- Hi.C Button (Blue): Selects the large capacitance measurement range.
- Hi.L Button (Blue): Selects the large inductance measurement range.
- L/C Button (Blue): Selects the standard inductance or capacitance measurement range.
- Func Button (Blue): Extended function button, used to check the measurement frequency.
- Test Terminals (Lx/Cx): Red (+) and Black (-) terminals for connecting test clips.
- Power Switch: Turns the device on/off.
- Mini USB Interface: For external 5V power supply.
- 5.5 DC Socket: For external 5V DC adapter power supply.
- Cell Box: Houses 4 AA batteries for portable operation.
- Kickstand: Allows the meter to stand upright for easier viewing.
4. Setup
4.1 Powering the Device
The LC-200A can be powered using one of the following methods:
- Mini USB Interface: Connect a Mini USB cable to a 5V power source.
- 5V DC Adapter: Use a 5V DC adapter with a Φ5.5DC socket. Ensure the polarity is correct (positive in the hole, negative on the periphery).
- AA Batteries: Install 4 AA batteries into the cell box at the back of the device. The meter can operate for over 60 hours on batteries.
When powered by batteries, the meter will automatically turn off after 5 minutes of inactivity to conserve power. The backlight is off by default in battery mode; you can toggle it by simultaneously pressing the ZERO and FUNC buttons.
4.2 Initial Power On
After powering on the device using the power switch, the meter will display the company name and product model. It will then enter the capacitor test status, ready for measurement.
5. Operating Instructions
Before performing any measurement, it is crucial to select the appropriate measurement range and calibrate the device.
5.1 Range Selection
The LC-200A has four main measurement ranges, selected using the blue buttons:
- C range (L/C button): For capacitance measurements (0.01pF - 10uF).
- L range (L/C button): For inductance measurements (0.001uH - 100mH).
- Hi.L range (Hi.L button): For large inductance measurements (0.001mH - 100H).
- Hi.C range (Hi.C button): For large capacitance measurements (1uF - 100mF).
The Hi.C, Hi.L, and L/C buttons are self-locking. Pressing a button selects its corresponding range.
5.2 Calibration (Zeroing)
Calibration is essential for accurate readings. Always perform calibration before measuring components, even if the display shows '0'.
- For Capacitance Measurement (C and Hi.C ranges):
- Ensure the test clips are open (not touching anything).
- Long press the ZERO (red) button for 2-3 seconds.
- The display will show "CALCULATING..." then "CALCULATING...OK".
- Release the button. The display should show "0.00pF" (or "0.00uF" for Hi.C range). This indicates successful calibration.
Video: Demonstration of LC-200A operation, including calibration and measurement examples.
- For Inductance Measurement (L and Hi.L ranges):
- Short circuit the test clips (connect the red and black clips together).
- Long press the ZERO (red) button for 2-3 seconds.
- The display will show "CALCULATING..." then "CALCULATING...OK".
- Release the button. The display should show "0.000uH" (or "0.000mH" for Hi.L range). This indicates successful calibration.
Important: Do not perform calibration while a component is connected to the test terminals.
5.3 Measurement Operation
Once calibrated, you can proceed with measurements:
- Measuring Capacitance:
- Select the appropriate capacitance range (C or Hi.C).
- Crucially, ensure the capacitor is fully discharged before connecting it to the meter. Failure to do so may damage the device.
- Connect the capacitor to the test terminals using the red (+) and black (-) clips.
- Read the capacitance value displayed on the LCD.

Image: Measuring a capacitor with alligator clips.
- Measuring Inductance:
- Select the appropriate inductance range (L or Hi.L).
- Connect the inductor to the test terminals using the red (+) and black (-) clips.
- Read the inductance value displayed on the LCD.

Image: Measuring an inductor with alligator clips.
- Measuring SMD Components (with optional clip):

Image: Measuring an SMD component using the optional chip testing clip.
Note on Large Capacitance: Measuring large capacitances (above 10mF) may take longer than 1 second. For example, a 100mF capacitor may take approximately 7-8 seconds to measure.
5.4 Checking Measurement Frequency
To view the current measurement frequency while a component is connected and being measured, press the FUNC (blue) button. The display will show the corresponding frequency.
6. Technical Specifications
| Item | Parameter | ||
|---|---|---|---|
| Capacitance measurement (C) | Accuracy | 1pF~1uF | 1% |
| 1uF~10uF | 5% | ||
| Resolution | 0.01pF | ||
| Inductance measurement (L) | Accuracy | 1uH~100mH | 1% |
| Resolution | 0.001uH | ||
| Large inductance measurement (Hi.L) | Accuracy | 100mH~1H | 1% |
| 1H~100H | 5% | ||
| Resolution | 0.001mH | ||
| Large capacitance measurement (Hi.C) | Accuracy | 1uF~100mF | 5% |
| Resolution | 0.01uF | ||
| Test frequency | L, C | 500kHz | |
| Hi.L | 500Hz~50KHz | ||
| Inductance, capacitance and large inductance measurement methods | LC Resonance | ||
| Large capacitance measurement method | Charge-discharge | ||
| Display | 1602 LCD | ||
| Effective display digits | 4 digits | ||
| Interface to power supply | Mini USB & Φ5.5DC Socket, 4 of AA Batteries | ||
| Supply voltage | 5V | ||
| Product dimension (L * W * H) | 150*89*29mm | ||
| Product weight | 198g | ||
| Package dimension | 200*160*60mm | ||
| Package weight | 280g | ||
7. Environmental Requirements
To ensure optimal performance and longevity of your LC-200A meter, please adhere to the following environmental guidelines:
- Avoid placing the meter in environments with excessive dust, vibration, direct sunlight, or corrosive gases.
- Operating Temperature: 0℃ to 40℃
- Operating Humidity: ≤90%RH (at 40℃)
- Storage Temperature: -25℃ to 50℃. If the meter will not be used for an extended period, store it in its original packaging in a dry environment.
8. Cautions and Safety
- Always reset the meter to "0" before testing any capacitance or inductance. Failure to do so may result in inaccurate readings.
- Do not attempt to reset the meter while a component is being measured. If this occurs, immediately shut down the device, restart, and then perform the "0" reset.
- Critical Safety Warning: Never measure a capacitance that has not been fully discharged. Measuring an undischarged capacitor can severely damage the meter's internal circuitry.
- Be aware that measuring large capacitances (e.g., above 10mF) will take longer. For instance, a 100mF capacitor may require 7-8 seconds for a stable reading.
9. User Tips
- For precise measurements of small capacitances (1-30 pF), ensure proper calibration as described in section 5.2. The LC-200A is designed to measure down to 1pF with high accuracy.
- When measuring components, try to keep the test leads as short as possible to minimize stray capacitance and inductance, which can affect readings, especially for very small values.
- Always double-check the selected range before connecting a component to avoid potential damage or inaccurate readings.
10. Warranty and Support
JUNCTEK warrants to the original purchaser that this product will be free from defects in workmanship and materials for a period of one year from the date of purchase. During this period, JUNCTEK will, at its option, repair or replace defective products or component parts. A proof of purchase date is required for all warranty claims.