JUNCTEK MHS5200A

Contents

JUNCTEK MHS5200A Dual-channel DDS Signal Generator User Manual

Model: MHS5200A Series

1. Introduction

The JUNCTEK MHS-5200A series instruments are advanced dual-channel Direct Digital Synthesizer (DDS) signal generators. They integrate a function signal generator, arbitrary waveform generator, pulse generator, sweep function, counter, and frequency meter into a single device. Utilizing large-scale integrated circuits and a high-speed FPGA MCU microprocessor, the internal circuitry employs surface mount technology, significantly enhancing noise immunity and service life.

The display interface features an LC1602 LCD, divided into two lines. The top line shows the current frequency, while the bottom line displays additional parameters or function variables. Flexible flip-key settings greatly enhance operability. This instrument excels in signal generation, waveform scanning, and parameter measurement, making it an ideal tool for electronic engineers, laboratories, production lines, teaching, and research.

The MHS5200A series offers two completely independent channels, capable of outputting identical or different waveforms simultaneously. The output frequency and amplitude are digitally controlled and continuously adjustable. It also includes a frequency meter and counter function.

Product Overview Video

This video provides a visual introduction to the MHS5200A Dual-channel DDS Signal Generator, highlighting its main features and interfaces.

2. Key Features

  • Direct Digital Synthesis (DDS) technology, FPGA design, low power consumption.
  • Dual output channels, capable of synchronized phase adjustment.
  • Linear and logarithmic sweep function with up to 999 seconds sweep time.
  • Basic waveforms: Sine, Triangle, Square, Sawtooth (rise/fall), and variable duty cycle Pulse wave.
  • 16 sets of user-customized arbitrary waveforms (2048*12bits memory depth, 200MSa/s sampling rate).
  • 10 sets of parameter storage locations (M0-M9), with M0 automatically loaded on boot.
  • Output amplitude: up to 20Vp-p (for ≤12MHz) and 15Vp-p (for >12MHz).
  • Built-in precision -20dB attenuator for minimum amplitude resolution of 1mV.
  • DC bias function: 120% to +120%.
  • Pulse Duty precise adjustment to 0.1%.
  • Four variable phase difference TTL outputs.
  • Frequency measurement, period measurement, positive/negative pulse width measurement, duty cycle measurement, and counting function.
  • Four optional gate frequency measurement times for balancing speed and accuracy.
  • All parametric equalizer functions can be performed via internal calibration.
  • Open communication protocol for easy secondary development and PC control.
  • PC connectivity allows control of the instrument and editing/downloading of arbitrary waveforms.
  • Optional power module for signal output amplitude up to 30Vpp and maximum output current of 1A.
  • Digital signal output function, CMOS output range 0~10V.
  • High frequency accuracy (up to 10-6 magnitude) and resolution (0.01Hz).
  • Tracking function for frequency, amplitude, offset, duty cycle, and waveform.
  • Input over-voltage protection (DC5V supply, safe up to 5V).
  • Output short-circuit protection (all signal outputs can withstand 60s in short-circuit).

3. Product Components and Interfaces

Front Panel

Front view of the MHS5200A Signal Generator with LCD, channel outputs, and control buttons.

Figure 3-1: Front Panel Layout

Table 3-1: Front Panel Components
GradeDescription
LCD1602Main display for frequency, waveform, and parameters.
Status IndicatorIndicates operational status.
Operation buttonsButtons for various functions (CH1/2, SET, WAVE, AMPL, PgUp, PgDn, SHIFT, OK).
KnobAdjusts parameters and navigates menus.
Ext.IN input interfaceExternal input for AC signals.
CH1 output interfaceOutput for Channel 1.
CH2 output interfaceOutput for Channel 2.

Rear Panel

Rear view of the MHS5200A Signal Generator showing power, USB, and TTL interfaces.

Figure 3-2: Rear Panel Layout

Table 3-2: Rear Panel Components
GradeDescription
DV5V power inputDC 5V power input port.
USBUSB communication interface.
TTL input/output interfaceTTL input and output ports (TTL OUTPUT: 1,3,5,7; TTL INPUT: 9; GND: 2,4,6; Vcc(+5V): 8,10).
SwitchPower On/Off switch.

4. Setup and Unpacking

4.1 Unpacking and Initial Inspection

Upon receiving your MHS-5200A Series dual-channel DDS signal generator, please follow these steps to ensure the instrument is in proper condition:

  1. Check for Transportation Damage: Inspect the packaging (e.g., bubble bag cushions) for any serious damage. Retain all packaging materials until the machine and accessories have passed testing.
  2. Verify Package Contents: Compare the received items with the shipping list below. If any content does not match or if the instrument appears damaged, contact your dealer or manufacturer immediately.
  3. Inspect the Machine: Visually inspect the instrument for any physical damage. If it appears damaged, is not working properly, or fails performance tests, contact your dealer or the Company.

4.2 Shipping List

MHS5200A Signal Generator and its accessories: power adapter, TTL expansion board, USB cable, Q9 clip lines.

Figure 4-1: Included Accessories

  • MHS-5200A Series dual-channel DDS signal generator: 1 unit
  • Power Adapter (Input: 100~240V~50/60Hz, Output: 5.0V-0.2A): 1 unit
  • USB cable: 1 unit
  • Signal connection cable (Q9 Clip Line): 2 units
  • User Manual (PDF version): 1 (Reference PDF: User Manual PDF)

4.3 Powering On

  1. Connect Power: Connect the provided DC5V power adapter to the DC5V input port on the rear panel of the instrument. Ensure the power adapter is correctly connected.
  2. Power Switch: Flip the "Power" switch on the rear panel to the "On" position.
  3. Main Interface: The instrument will power on and display the main interface on the LCD.

5. Operating Instructions

The following instructions detail how to operate the instrument. Note that instructions for Channel 1 (CH1) also apply to Channel 2 (CH2) due to their symmetrical design.

5.1 Setting the CH1 Waveform

In the main interface, the LCD will show the current frequency and waveform type, for example:

*F00024.00000kHz
WAVE:SINE
  1. Select Waveform: Press the OK button. This will allow you to adjust the output waveform type. Available types include Sine, Square, Triangle, Rising Sawtooth, Descending Sawtooth, and 16 groups of Arbitrary waveforms.
  2. Navigate Waveforms: Rotate the ADJUST knob to cycle through the available waveform types.
  3. Quick Adjustment: To quickly adjust the output waveform, press SHIFT + WAVE PgUp.
  4. Confirm Selection: Press OK again to confirm your selection and return to the main interface.

5.2 Setting the CH1 Frequency

In the main interface, ensure the "*" cursor is on the first line (frequency display).

  1. Adjust Step Size: Use the CH1/2 or SET buttons to move the cursor and adjust the frequency step size.
  2. Adjust Frequency: Rotate the ADJUST knob to set the desired output frequency.

5.3 Setting the CH1 Amplitude

In the main interface:

  1. Enter Amplitude Adjustment: Press SHIFT + AMPL PgDn. A cursor will appear on the amplitude setting.
  2. Move Cursor: Use the CH1/2 or SET buttons to move the cursor position.
  3. Adjust Amplitude: Rotate the ADJUST knob to set the output waveform amplitude. The amplitude range is 5mVp-p to 20Vp-p (for frequencies ≤12MHz) or 5mVp-p to 15Vp-p (for frequencies >12MHz). Minimum step value is 0.01V (10mV).
  4. -20dB Attenuation: If the instrument is in the -20dB attenuation state, the maximum output signal is 2.000V, minimum 0.005V, and minimum step is 0.001V (1mV).

5.4 Setting CH1 DC Bias

In the main interface:

  1. Enter Bias Adjustment: Press SHIFT + WAVE PgUp or AMPL PgDn to access offset adjustment options.
  2. Move Cursor: Use the CH1/2 or SET buttons to move the cursor.
  3. Adjust Bias: Rotate the ADJUST knob to set the DC bias. The range is -120% to +120%.

5.5 Setting CH1 Duty Cycle

In the main interface:

  1. Enter Duty Cycle Adjustment: Press SHIFT + WAVE PgUp or AMPL PgDn to access duty cycle adjustment options.
  2. Move Cursor: Use the CH1/2 or SET buttons to move the cursor.
  3. Adjust Duty Cycle: Rotate the ADJUST knob to set the duty cycle. The precise adjustment range is 0.1% to 99.9%.

5.6 Setting Phase Difference Between Channels

In the main interface:

  1. Enter Phase Adjustment: Press SHIFT + WAVE PgUp or AMPL PgDn to access phase adjustment options.
  2. Move Cursor: Use the CH1/2 or SET buttons to move the cursor.
  3. Adjust Phase: Rotate the ADJUST knob to set the phase difference. This function is only meaningful when CH1 and CH2 have the same frequency.

5.7 Setting Frequency Display Unit

In the main interface:

  1. Enter Unit Adjustment: Press SHIFT + SET. The cursor "*" will move to the second row.
  2. Switch Unit: Press OK to cycle through frequency units: Hz, kHz, MHz.

5.8 Tracking Function

The tracking function synchronizes the frequency of CH2 to CH1. Users can also set amplitude tracking and duty cycle tracking.

  1. Enter Tracking Options: Press SHIFT + WAVE PgUp or AMPL PgDn to access tracking options.
  2. Toggle Tracking: Press OK to switch tracking ON or OFF.
  3. Frequency Tracking: When tracking is ON, CH2 frequency automatically tracks CH1 frequency.
  4. Amplitude/Duty Cycle Tracking: If amplitude or duty cycle tracking is enabled, CH1 and CH2 amplitude/duty cycle will also track each other.

5.9 External Signal Input Port Selection

To select the input source for frequency measurement:

  1. Enter Input Selection: Press WAVE PgUp or AMPL PgDn to navigate to the input port selection page.
  2. Switch Input: Press OK to toggle between Ext.IN (for AC signals) and TTL.IN (for digital signals).

5.10 Measurement Function

After selecting the input source, various parameters of the input signal can be measured:

  1. Enter Measurement Mode: Press WAVE PgUp or AMPL PgDn to navigate to the measurement function selection page.
  2. Select Measurement Type: Press OK to cycle through measurement objects: FREQ (Frequency), COUNTR (Counting function), POS-PW (Positive Pulse Width), NEG-PW (Negative Pulse Width), PERIOD (Cycle), DUTY (Duty cycle).
  3. Gate Time Selection: After selecting the measurement object, press AMPL PgDn to enter the gate time selection page. Press OK to select gate times: 10S, 1S, 0.1S, 0.01S. Different gate times affect measurement accuracy and speed.
  4. View Results: Once the gate time is set, the display will show the measurement results (frequency, duty cycle, pulse width, etc.).

5.11 Sweep Function

  1. Set Start Frequency: Navigate to the start frequency sweep function settings page. Adjust the start frequency (e.g., 5kHz).
  2. Set Stop Frequency: Press AMPL PgDn to enter the stop frequency sweep function settings page. Adjust the stop frequency (e.g., 10kHz).
  3. Set Sweep Time: Press AMPL PgDn to enter the sweep time setting page. Press SHIFT + SET to move the cursor to the second row. Rotate the ADJUST knob to set the sweep time (1-600S).
  4. Select Sweep Mode: Press AMPL PgDn to enter the sweep mode selection screen. Press OK to select between LINE (linear scanning) and LOG (logarithmic sweep).
  5. Toggle Sweep: After setting the scan mode, press AMPL PgDn to enter the sweep control page. Press OK to turn the sweep function ON or OFF.

5.12 Save / Load Function

The instrument can store 10 sets of parameters (M0-M9). M0 is automatically loaded on power-on.

  1. Enter Save/Load Page: In the main interface, press SHIFT + SET to move the cursor to the second row.
  2. Select Location: Rotate the ADJUST knob to select a storage location (M0-M9).
  3. Save Parameters: To save, select the desired location and press OK.
  4. Load Parameters: To load, select the desired location and press OK.

5.13 PC Software Instructions

The MHS5200A can be controlled via PC software. Follow these steps to install and connect:

  1. Install Runtime Software: Install visa540_runtime.exe.
  2. Install USB to Serial Drivers: Install CH341SER_SETUP.exe.
  3. Install Signal Generator Program: Install the main signal generator .exe program.
  4. Connect Online:
    1. Right-click on "My Computer" -> "Properties" -> "Device Manager".
    2. Under "Ports (COM & LPT)", identify the COM port assigned to "USB-SERIAL CH340".
    3. Open the DDS Signal Generator Control Panel software.
    4. Select the corresponding COM port from the dropdown menu.
    5. Click "Connect".

The PC software allows for standard output waveform control, external signal measurement, and arbitrary waveform generation/editing. For detailed instructions on using the PC software for arbitrary waveform generation (hand-painted, trapezoidal, amplitude modulation), refer to the full PDF manual.

6. Technical Specifications

Table 6-1: MHS-5200A Series Specifications
ProjectParameters
Model SelectionMHS-5206AMHS-5212AMHS-5220AMHS-5225A
Sine wave frequency range0~6MHz0~12MHz0~20MHz0~25MHz
Square wave frequency range0~6MHz
Pulse wave frequency range0~6MHz
TTL / COMS digital signal frequency range0~6MHz
Arbitrary / other waveform frequency range0~6MHz
Frequency Characteristics
Frequency minimum resolution10mHz
Frequency error±5x10-6
Frequency stability±1x10-6/5 hours
Amplitude Characteristics
Amplitude range (peak-to-peak value)5mVp-p~20Vp-p
Amplitude resolution1mVp-p (-20dB attenuation) 10mVp-p (No attenuation)
Amplitude stability±0.5% (Each 5 hours)
Amplitude error±1% + 10mV (Frequency 1KHz, 15Vp-p)
Offset range-120%~+120%
Offset resolution1%
Phase resolution1°
Waveform Characteristics
Waveform typeSine, Square, Pulse (adjustable duty cycle, precise adjustment of pulse width and period), Triangular wave, Partial sine wave, CMOS wave, DC level (set DC amplitude by adjusting offset), Half wave, Full Wave, Positive staircase wave, Anti-ladder wave, Noise wave, Exponential rise, Exponential drop, Symplectic pulse and Lorenz pulse, and 60 arbitrary waveforms.
Wave length2048 points
Waveform sampling rate200MSa/s
Waveform vertical resolution12 bits
Sine Wave
Harmonic Suppression≥40dBc (<1MHz); ≥35dBc (1MHz~25MHz)
Total harmonic distortion<0.8% (20Hz~20KHz)
Square Wave
Rise and fall time≤20ns
Overshoot≤10%
Duty cycle adjustment range0.1%-99.9%
TTL Signal
Output level≥3Vpp
Fan-out coefficient≥20TTL
Rise and fall time≤20ns
COMS Signal
Low level<0.3V
High level1V~10V
Rise and fall time≤20ns
Sawtooth Wave
Duty cycle >50%Sawtooth wave
Duty cycle <50%Sawtooth wave
Arbitrary Waveform
Quantity16 groups
Storage depth / group1KB / 16 groups
Waveform Output
Frequency measurement rangeGATE-TIME=10S0.1HZ-60MHZ
GATE-TIME=1S1HZ-60MHZ
GATE-TIME=0.1S10HZ-60MHZ
GATE-TIME=0.01S100HZ-60MHZ
Input voltage range0.5Vp-p~20Vp-p
Counting range0~4294967295
Counting methodManual
Positive and negative pulse width measurement10ns resolution, maximum measurement 10s
Period measurement20ns resolution, maximum measurement 20s
Duty cycle measurement0.1% resolution, measuring range 0.1% ~ 99.9%
Source selection1. EXT.IN input (AC signal), 2. TTL_IN input (digital signal)
Communication Characteristics
Interface methodUse USB to serial interface
Communication rate57600bps
ProtocolUsing the command line, the agreement is open
Other
Power supplyDC 5V±0.5V
Dimension180*190*72mm
Net weight550g (Host) 480g (Annex)
Gross weight1090g
Working environmentTemperature: -10℃~50℃ Humidity<80

7. Care and Maintenance

  1. Power Supply: Always ensure the DC5V power adapter is correctly connected and used.
  2. LCD Care: The LCD module is fragile. Avoid striking it or exposing it to corrosive chemicals. Clean dust and dirt with a soft cloth.
  3. Operating Environment: Operate the instrument within a temperature range of -10°C to 50°C and store it between -20°C and 70°C in a dry environment.
  4. Disassembly: Do not attempt to disassemble the equipment. Disassembly will void the warranty. The instrument contains no user-serviceable parts; repairs should only be performed by authorized service centers or the factory.
  5. Safety: Keep the instrument away from open flames (e.g., lighted candles), water, corrosive chemicals, and other unsafe items to prevent damage.
  6. Screen Protection: The display screen is susceptible to pollution and is a fragile component. Avoid touching it with bare hands, and protect it from external impacts. Keep children from playing with the instrument.
  7. Movement: Do not move the instrument while it is operating to prevent severe, irreparable damage to internal circuits.

8. Troubleshooting

If the instrument does not function correctly after re-powering it and you have ruled out common issues (e.g., incorrect power supply, loose connections), please contact your supplier for assistance. Do not attempt unauthorized repairs.

9. User Tips

  • PC Software Name: The PC software for controlling the MHS5200A is referred to as "DDS Signal Generator Control Panel" in the manual. Ensure you install all three required components: visa540_runtime.exe, CH341SER_SETUP.exe (USB to serial drivers), and the main signal generator .exe program.
  • Arbitrary Waveform Creation: When creating arbitrary waveforms using the PC software, remember that you can hand-draw, create trapezoidal waves, or amplitude-modulated waves. Always save your custom waveforms to one of the 16 available storage locations.
  • Channel Independence: Remember that CH1 and CH2 are completely independent. You can set different parameters for each, or use the tracking function to synchronize them.

10. Warranty and Support

10.1 Warranty

  • One-year guarantee from the date of receipt.
  • If the item is defective upon receipt, it should be returned within 7 days for exchange. The buyer must return all original and undamaged items in re-saleable condition.
  • The warranty does not cover products that are physically damaged or under abnormal operating conditions due to misuse or improper installation.

10.2 Contact Information

If you encounter any problems or require assistance with our products or this manual, please contact JUNTEK:

  • E-mail: junce@junteks.com
  • Website: www.junteks.com
  • Software and Communication Protocol Download: Refer to the official website or the provided PDF manual for download links.