ADC logo

Diagnostix ™
ADSTATION™
Modular Diagnostic System
5660T
Instructions for UseADC 5660T Diagnostix Adstation Modular Diagnostic System

Questions?
Call ADC toll-free: 1-800-232-2670

A SPECIAL THANK YOU

Congratulations on your purchase of an ADC ® Diagnostix ™ Adstation ™.
ADC professional diagnostic products are the instruments of choice where accuracy and dependability are critical.
With proper use and care, this physical exam and non-invasive diagnostic instruments will provide many years of dependable service.
Read this booklet thoroughly before using your new unit.

INTRODUCTION AND INTENDED USE

This manual is for the Diagnostix Adaptation. The wall unit described in these instructions is manufactured for use with various ADC instrument heads and modular components for non-invasive diagnostics and physical exams.
To learn more, visit our website at: www.adctoday.com.

ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 6WARNINGS AND PRECAUTIONS

Your Diagnostix Adaptation has been manufactured to the highest global standards and is subjected to rigorous quality control.
Read these instructions for use carefully before putting the unit into operation and keep them in a safe place.
If you have any questions, call our toll-free number or visit our website. Our contact information can be found on the last page of this booklet.
Please note that all instruments described in these instructions for use are only to be used by suitably trained personnel.
The performance and efficiency of this instrument are only guaranteed when original ADC parts and accessories are used.

WARNING: There may be a risk of gases igniting when the instrument is used in the presence of flammable or combustible gases. We recommend working in areas with good ventilation.

WARNING: Never attempt to take the instrument apart. There is a danger of life-threatening electrical shock.

WARNING: The device contains sensitive electronic components. Avoid strong electrical or electromagnetic fields in the direct vicinity of the device (e.g., mobile telephones, microwave ovens). This can lead to temporary impairment of the ADC Diagnostix Adstation devices.

WARNING: Do not use batteries or electrical cords other than those included with this product. Use only replacement parts supplied by the manufacturer.

CAUTION: Unplug the instrument before cleaning or disinfecting. Be careful that no liquid penetrates into the device.

ATTENTION: Some wall systems may include up to three additional extension modules. Take care that the connecting cable does not get caught behind the extension module. Push the connecting cable into the groove provided on the reverse side of the extension module.

SYMBOLS

Symbol Definition
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol Read operating instructions
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 2 Medical device
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 3 Type BF Equipment
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 4 Protection Class II Unit
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 5 Warning! The general warning sign indicates a possibly dangerous situation that can lead to serious injuries.
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 6 Caution! An important note in these instructions. The caution symbol indicates a potentially dangerous situation that can lead to minor or moderate injuries. It can also be used to warn of  unsafe practices
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 7 Direct current
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 8 Alternating current
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 9 Date of Manufacture
YYYY-MM-DD / (year-month-day)
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 10 Manufacturer
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 11 Manufacturer serial number
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 13 Lot number
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 14 Reference number
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 15 Temperature for transport and
storage conditions
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 16 Relative humidity for transport and
storage conditions
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 31 Air pressure for transport and storage
Air pressure for ambient operation
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 17 CE marking
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 18 Discard according to national regulations or EU directives
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 19 Non-ionizing radiation
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 21 Caution: Federal law restricts this device to sale by or on the order of a physician (licensed physician).
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 22 Not made with natural rubber latex
ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 23 Phthalate-free

ATTACHMENT

a.) Wall Mounting
If mounting to the wall, use an enclosed mounting template.

b.) Attaching the Wall Mounting Plates
If mounting to wall studs, use wood screws only. If mounting between studs, use the enclosed molly bolts.
Take the wall mounting plate and hold it onto the wall so that the screws can be pushed through the holes of the mounting plate into the molly bolts.
Screw in the screws with a screwdriver, as far as they will go.
If the optional wallboard is purchased, the mounting plates come pre-attached.

c.) Attaching the ADC Adaptation
When all screws have been screwed in tightly, take the Adaptation and guide the screw heads through the openings. Then slide downwards until it snaps into place. (See Figure 1.)

ADC 5660T Diagnostix Adstation Modular Diagnostic System - figure

d.) Attaching Optional Extension Modules (If Included) Additional extension modules must be electrically joined with the transformer station before mounting on the wall. 
Using the enclosed module connection cable, plug one side of the connector into the socket on the transformer module located on the right side transformer panel.
Plug the opposite end into the extension module contact, located on the left side panel.

WARNING: This unit is only compatible with an Adstation transformer whose serial number is 100 000 or higher. The extension module will be damaged when used with other devices.
Caution: Make sure the four contacts of the socket strip are positively connected to the four contacts of the pin strip (See Figure 2.)

ADC 5660T Diagnostix Adstation Modular Diagnostic System - figure 2

Tuck excess wiring into each extension to ensure a close side-by-side fit to each extension module.
Take the extension module and guide the screw heads through the openings as done with the main unit in step C above. Then slide the extension module downwards to snap  into place.

Attention:
Take care that the connecting cable does not get caught behind the extension module. Push the connecting cable into the groove provided on the reverse side of the extension module.
Repeat for any additional extension modules (up to three) you may have purchased.

OPERATION AND FUNCTION

Plug devise into an electrical outlet. Remove the handle from the cradle by sliding upwards. (See Figure 3.) Attach the desired instrument head by placing the two projecting guide cams onto the handle, pressing down lightly, and turning the handle in a clockwise direction until it stops. To remove the instrument head, turn the handle in a counter-clockwise direction.

ADC 5660T Diagnostix Adstation Modular Diagnostic System - figure 3

LOCKING INSTRUMENT HEADS

The enclosed Allen key can be used to lock the instrument head onto the power handle. To secure, insert the key into the slot on the head and tighten by turning clockwise. Do not over-tighten.  Keep the key in a convenient place in case the head must be removed for cleaning or service.

SWITCHING ON AND OFF

Turn on the wall system by using the rocker switch. (See figure 3.) The green control lamp in the rocker switch indicates that the instrument is ready to use. Each handle is automatically ready to operate at 100% light intensity as soon as it is taken out of the cradle and automatically switches off when replaced back in the cradle.

Light intensity modulation

The modulation of light intensity can be done with the handle; twist the switching ring in a clockwise or counter-clockwise direction and the light will get stronger or weaker.

ADC 5660T Diagnostix Adstation Modular Diagnostic System - Light

Attention: The handle switches off automatically when placed in the cradle, or after about two minutes. Make sure that no more than three handles are used at the same time, or the transformer may become overloaded and switch itself off.

CLEANING AND DISINFECTION

The Adstation can be cleaned externally with a moist cloth. The outside can also be disinfected.
Aldehydes (formaldehyde, glutaraldehyde, aldehyde derivatives),

surfactants, or alcohol may be used. Refer to disinfectant manufacturer’s instructions for use.
Always use a soft, lint-free cloth or cotton swabs to prevent damage to your Adaptation.

Sterilization
According to the Test Centre for Medical Devices in Tübingen, Germany, sterilization is only prescribed in the case of operative procedures.

TECHNICAL SPECIFICATIONS

Model:     5660TN ADC®  Diagnostix™  Adstation™  Transformer
Power Supply:  Input: 100V~240V AC / 50-60 Hz / 0.6 A
Output: 5V DC / 3A / 15W
Diagnostic Station:              Input: 5V DC / 3A / 15W
Output 1: 1 x 3.5V dc / 700 mA
Output 2: 2 x 5V dc / 2 x 1, 15 A
Extension Module:               Input:  5V DC / 3A / 15W
Output 1: 1 x 3.5V dc / 700 mA
Output 2: 2 x 5V dc / 2 x 1, 15 A
Classification:                      Application part type B
Operating Conditions:         32°F to 104°F (0°C to 40°C)
85 % relative humidity
Storage Temperature:          23°F to 122°F (-5°C to 50°C)
10% up to 85 % relative humidity
Air Pressure:                         700 bis 1050 hPa
Dimensions: 
Diagnostic Station:  7.88” x 7.12” x 2.95” (200 x 180.5 x 75 mm)
Extension Module:               7.88” x 3.94” x 2.95” (200 x 100 x 75 mm)
Weight Diagnostic Sta.: 1.80 lbs (800 g)
Weight Exten. Module:                1.10 lbs (500 g)
Switch-On Time:    ON: 1 Min / OFF: 5 Min

* only UL 60601-1 CAN/CSA C 22.2 No.601.01

WARRANTY

This Diagnostix Adstation transformer and extension modules are warranted for five years from date of purchase. Instrument heads are warranted for two years. The warranty does not apply to damage caused by improper handling, accidents, improper use, or alterations made to the instrument by third parties. The warranty is only valid after the product is registered online at www.adctoday.com/support/warranty-registration.

QUALITY STANDARDS

The device fulfills the stipulations of the International standard IEC 60601-1-2

 ELECTROMAGNETIC COMPATIBILITY

Guidance and manufacture’s declaration – Electromagnetic immunity The Diagnostic Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Diagnostic station should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level Compliance level

Electromagnetic environmental guidance

Electrostatic
discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Con: ± 8 kV
Air: ± 15 kV
5/50 ns, 100 kHz; ±2 kV
Con: ± 8 kV
Air: ± 15 kV
5/50 ns, 100 kHz; ±2 kV
Floors should be wood, concrete or ceramic tile. If the floor is covered with synthetic material, the relative humidity should be at least 30%. The quality of the supply voltage should be that of a typical business or hospital environment.
Surge
IEC 61000-4-5
1.2/50 (8/20) ps LtL: ± 1.0 kV LtG: ± 2.0 kV 1.2/50 (8/20) ps LtL: ± 1.0 kV
LtG: ± 2.0 kV
The quality of the supply voltage should be that of a typical business or hospital environment.
Voltage dips, short i  nterruptions and voltage variations on power supply input lines
IEC 61000-4-11
0% UT for 0.5 cycle (1 phase) 0% UT for 1 cycle
70% UT for 25/30 cycles (50/60 Hz)
0% UT for 0.5 cycle (1 phase)
0% UT for 1 cycle 70% UT for 25/30 cycles (50/60 Hz)
The quality of the supply voltage should be that of a typical business or hospital environment.
0% UT for 250/300 cycles (50/60 Hz) 0% UT for
250/300 cycles (50/60 Hz)
Power frequency (50Hz/60Hz) magnetic field
IEC 6000-4-8
30 A/m 50 Hz 30 A/m 50 Hz Mains frequency magnetic fields should be at a level characteristic of a typical location in a typical commercial hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test reel. v
Guidance and manufacture’s declaration – Electromagnetic emission
The Diagnostic Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Diagnostic station should assure that it is used in such an environment.
Emission Test Compliance Electromagnetic environment – guidance
RF-Emission  CISPR 11 Group 1 The Diagnostic Station uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF-Emission  CISPR 11 Class B The Diagnostic Station is intended for use in all establishments, including residential areas and those directly connected to a public supply network that also supplies buildings used for residential purposes.
Harmonic emissions IEC 61000-3-2 Pass
Voltage fluctuations/Flicker emissions IEC 61000-3-3 Pass
Guidance and manufacture’s declaration – Electromagnetic immunity
The Diagnostic Station is intended for use in the electromagnetic environment specified below. The customer or the user of the Diagnostic station should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Conducted RFIEC 61000-4-6 Radiated RF IEC 61000-4- 3 Proximity fields from RF  wireless communications  equipment Pass 3 V/m 80  MHz to 2,7  Hz  Pass P ass 3 V/m Pass Portable and mobile RF communications equipment should not be used closer to any part of the Diagnostic Station, including the cables, than the recommended distance, which is calculated using the equation applicable to the transmitter frequency. Recommended separation distance
d=1.2VP 150 kHz  to 80 MHz
d=1.2VP 80 kHz to 800 MHz
d=1.2VP 800 kHz to 2,7 MHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. b Interference may occur into the vicinity of  equipment  marked with the following symbol:
Where P is the maximum output power of the transmitter in watts (W) according to the transmitter manufacturer and the recommended distance is given in meters (m). Field strengths from fixed RF transmitters determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b. Interference may occur in the vicinity of devices marked with the following symbol.ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 19
NOTE 1: At 80 MHz and 800 MHz, the h higher frequency range applies. NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a.)Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios, amateur radio, AM and FM raadio broadcast, and N broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic sitesurvey should be considered. If the measured field strength in the location in which the Diagnostic Station is used exceeds the applicable RF compliance level above, the Diagnostic Station should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Diagnostic Station.
b.)Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the Diagnostic Station.
The Diagnostic Station is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the  Diagnostic Station can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment  (transmitters) and the Diagnostic Station as recommended below, according to the maximum output power of the communications equipment.
The rated maximum output power of the transmitter (W) Separation distance according to the frequency of the transmitter (m)
150 kHz to 80 80 kHz to 800 800 kHz to
MHz MHz 2.7 GHz
1=121/13 a=.1,21/F’ a=2,3413
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations, Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

HOW TO CONTACT US

To register your product and obtain further detailed user information about our products and services visit us at: www.adctoday.com and follow the links.
For questions, comments, or suggestions call us toll-free at: 1-800-232-2670

American Diagnostic Corporation
55 Commerce Drive, Hauppauge, New York 11788
Telephone: 631-273-9600 • Fax: 631-273-9659
Email: info@adctoday.com

ADC 5660T Diagnostix Adstation Modular Diagnostic System symbol 10ADC
55 Commerce Drive
Hauppauge, NY 11788
Made in Germany
Inspected and packaged in the U.S.A.
BliTZWOlF BW YPC110 Split Wi Fi Endoscope with 1200P HD Resolution - ICONMDI Europa GmbH
Langenhagen Str. 71
30855 Langenhagen
GERMANY
tel: 631-273-9600
toll-free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: info@adctoday.com

ADC logobook iconCE ICON