AandD LCC33-USB Series Mono Button Load Cell for Force Measurement Instruction Manual

LCC33-USB Series Mono Button Load Cell for Force Measurement

Specifications:

  • Model: LCC33-USB series
  • Product: USB Button Load Cell for Force
    Measurement
  • Rated Capacities: 5 N, 10 N, 20 N, 50 N
    (0.5099 kg, 1.020 kg, 2.039 kg, 5.099 kg)
  • Rated Output: 5.00000, 10.0000, 20.0000,
    50.0000
  • Nonlinearity: 1% of R.O.
  • Hysteresis: 1% of R.O.
  • Repeatability: 1% of R.O.
  • Power Supply Voltage: DC5 V (USB bus
    power)
  • Average Current Consumption: 60 mA or
    less
  • Compensated Temperature Range: Safe overload
    150% of R.C.
  • Website: LCC33-USB series
    website

Product Usage Instructions:

1. Outline:

The LCC33-USB series is a small compression-type load cell
calibrated for force measurement. It is connected to a computer via
a USB cable for measurement purposes.

2. Cautions:

  • Mount the load cell on a secure, rigid, and flat surface.
  • Clean the mounting surface to remove any dust or dirt before
    installation.
  • Avoid applying unbalanced load, lateral load, or bending
    moments to the load cell.

3. Names:

  • Load Cell Cable: 3 mm length, 0.9 m
  • USB Cable: 4 mm length, 1.2 m
  • Digital Filter: Select from None, 0.7, 1.0,
    1.4, 2.0, 2.8, 4.0, 5.6, 8.0, 11.0 Hz (Initial value 1.0 Hz)

Frequently Asked Questions (FAQ):

1. How do I mount the load cell securely?

To mount the load cell securely, ensure the surface is rigid and
flat. Clean the surface before mounting and avoid applying
unbalanced loads or bending moments.

2. What is the power supply voltage for the load cell?

The load cell operates on DC5V using USB bus power.

3. How should I handle the load cell when exposed to direct
sunlight?

If mounting the load cell in direct sunlight, use heat
insulating materials to prevent temperature gradients and avoid
damage to the load cell.

“`

LCC33-USB series

5. SPECIFICATIONS

USB Button Load Cell for Force Measurement

Model Item CD

A&D Company, Limited

Rated capacities 5 N

10 N

20 N

50 N

(0.5099 kg) (1.020 kg) (2.039 kg) (5.099 kg)

1WMPD4005366 LCC33-USB series website:
https://link.aandd.jp/Product_LCC33USB_EN
1. OUTLINE The LCC33-USB series is load cell of small type and compression type load cell. The load cell of the LCC33-USB series is calibrated by physical value of force, is connected to the computer using the USB cable and can measure it.

Rated output

5.00000

10.0000 20.0000 50.0000

±0.02500 [N] ±0.0500 [N] ±0.1000 [N] ±0.2500 [N]

Nonlinearity

1 % of R.O.

Hysteresis

1 % of R.O.

Repeatability

1 % of R.O.

Power supply voltage

DC5 V(USB bus power)

Average current consumption #1

60 mA or less

Zero balance

±10 % of R.O.

Temperature effect on zero

2.1 % of R.O. / 10 °C

Temperature effect on span

1.1 % of Load / 10 °C

Compensated temperature range

0 50 °C

Safe overload

150 % of R.C.

We recommend to use for simplified measurement of experiment and evaluation.

Cables

2. CAUTIONS Mount the load cell on a secure surface that is rigid and flat.
Clean the surface to ensure there is no residual dust or dirt before mounting the load cell.
Avoid applying unbalanced load, lateral load, bending moment to the load cell when mounting or applying load to the load cell.

Load cell material Conversion box
material Resonance frequency #2
Weight A/D conversion
rate

When mounting the load cell in a location that is exposed to direct sunlight or radiant heat, use heat insulating materials or take other measures to prevent temperature gradient.
To prevent malfunction, do not disassemble the conversion box.

3. NAMES

Load cell cable USB cable

Digital filter
Communication standard Baud rate Data
Communi bits cation Parity settings Stop bit Termin ator Code

Load cell cable: 3 mm length 0.9 m

USB cable: 4 mm length 1.2 m

A type connector

Aluminum

Stainless

Polycarbonate

2.3 kHz

116 kHz 115 kHz

0.5 kg

100 times/s

Select from None, 0.7, 1.0, 1.4, 2.0, 2.8, 4.0, 5.6, 8.0, 11.0 Hz (Initial value 1.0 Hz)

Conformed to USB Ver.2.0 Full Speed

38400 bps

8 bits Even 1 bit CR LF ASCII

#1: Reference value. #2: Load cell only.

Black area Mounting surface

Mesh area Don’t apply load.

Floating point type

measurement RFMV

RFMVXXXXXXXX

value reading

3 When using adhesive to secure the load cell, pay attention to the following points.

Floating point type

measurement value sequential

RCFM

RCFMXXXXXXXX

Use the cyanoacrylate adhesive.

reading

Push gently the load cell so as to keep the position. Paste the Fixed point type

adhesive to the place contacting the fixed surface and outer measurement RLMV

US,YYYYYYYYYZZZ

circumference of the load cell. Hold the load cell until maintaining the value reading

position. Do not apply excessive load and don’t defile with adhesive Fixed point type

on mesh area. Peel and shave adhesive using cutter and etc. when removing the load

measurement value sequential

RCLM

US,YYYYYYYYYZZZ

cell. Don’t hit and shock to the load cell when removing it. Note injury and reading

damage of the load cell in removing it. 4 Construct the grounding connection so that load cell and computer

Stop sequential reading

STOP

STOP

is the same voltage potential. If the load cell is charged static electricity, it may cause of malfunction.

XXXXXXXX : The floating point type measurement value, ZZZ : Unit YYYYYYYYY : The fixed point type measurement value

6.2. LOADING TO LOAD CELL 1 Load a vertical load to the load cell such as figure indicated
below.Avoid unbalanced load, lateral load and twist force to the load cell when loading to the load cell.
Direction of force

Response of command error

Items

Response command of load cell side

Format error

?

Setting value error V

9. LED DISPLAY

Orange …………. TX

(Sending)

Yellow …………… RX (Receiving)

Blue ……………… Power (Power supply)

7. PROCEDURE OF CONNECTING COMPUTER
1 Connect the USB cable to the computer. 2 Select the device manager at control panel in the computer. 3 Select “Ports (COM & LPT)”.

10. MAINTENENCE
1 Remove all dirt and dust from the load cell, and always use it in a clean environment.
2 When cleaning, use an air blower.

4

Confirm the displayed COM Port number. x of “USB Serial Port (COM x)” is COM Port number. If COM Port numbers are not confirmed and

11.

DIMENSIONS

are connected, identification of COM Port cannot recognize.

<LOAD CELL>

Therefore, confirm COM Port number each time when connecting it.

Additionally, when installation of driver software fails and COM Port

number isn’t displayed, refer to website of “Future Technology

Devices International Limited” and retry installation of driver software.

Refer to website of the LCC33-USB series for “USB load cell

computer connection communication manual” and “USB connection

manual”.

5 Select the “Port Settings” tab in property of USB Serial Port (COM x), then select “Advanced”.

6 In the “BM options”, set the “Latency Timer (msec)” under 10

Load cell

Conversion box

6. PROCEDURE OF INSTALLATION

(recommended value is 3). If it is not to set, a communication delay may result.

<CONVERSION BOX>

(20) 125

(30)

6.1. ATTACHING THE LOAD CELL 1 Attach the load cell to a rigid and flat base (the fixed surface). If there

8. COMMAND LIST

4- 5.4

(Attachment screw M4)

110

is slope or distortion on a part of the base, it affects the measurement In this document, only major commands are described. Concerning

accuracy.

of others, refer to “USB load cell computer connection

125

140

(155)

4. MEASUREMENT SOFTWARE The measurement data can be confirmed on the computer when using the measurement software “WinCT-DLC”. The “WinCT-DLC” can download from the LCC33-USB

2 The mounting surface is black area on the bottom of the load cell in the figure. Do not apply load to the concavity of the center of the bottom ( mesh area).

communication manual” from the LCC33-USB series website.

series website.

6 2

2 7

Documents / Resources

AandD LCC33-USB Series Mono Button Load Cell for Force Measurement [pdf] Instruction Manual
LCC33-USB Series, LCC33-USB Series Mono Button Load Cell for Force Measurement, Mono Button Load Cell for Force Measurement, Button Load Cell for Force Measurement, Load Cell for Force Measurement