GA logoS8-044-A1 GA-map Dysbiosis Test - Logo 1INSTALLATION GUIDE
for GA-map® Dysbiosis Test Lx v2 

OVERVIEW

This guide provides the information required to prepare a laboratory for performing the GA-map® Dysbiosis Test Lx v2 assay. Requirements for instruments, materials, and reagents are described.

INSTALLATION CHECK LIST

The following installation checklist summarizes the elements that must be completed during the installation phase. Mark each item as it is completed. Once these prerequisites have been met, please contact Genetic Analysis AS to schedule the training.

  • Instruments and consumable items for Step 1 Genomic DNA extraction (see tables on page 4-6) must be acquired
  • Instruments and consumable items for Step 2 Amplification of the bacterial 16S rRNA gene – Step 6 Hybridization and Signal Detection (see tables on page 6-8) must be acquired
  • All instruments, including the Luminex detection system, must be installed and ready for use
  • Operators must have been trained in use and maintenance of the Luminex detection system

RECOMMENDED SAFETY EQUIPMENT

warning - 1 Fecal samples should be treated as potentially infectious material until inactivation and require the use of BSL-2 grade laboratory equipment and precautions. This involves the use of appropriate PPE, biological safety cabinet, proper waste disposal and risk-minimizing routines for sample handling. Appropriate skin and eye protection should be worn during use of the DNA isolation kit (mag™ maxi, LGC genomics). For all other steps of the assay, suitable lab coat and disposable gloves should be used.

COMPUTER REQUIREMENTS

The GA-map® Analyzer is a cloud-based software compatible with most web browsers. See the GA-map® Analyzer Manual for details.

LABORATORY REQUIREMENTS

The GA-map® Dysbiosis Test Lx v2 requires a typical molecular laboratory set-up. It is recommended that the laboratory contains, at a minimum, dedicated pre-PCR, template-free, and PCR/post-PCR zones. The zones should be separated to prevent contamination.
The table below contains an overview of the different laboratory procedures that are performed in the individual zones, in addition to instruments and equipment needed in the different workstations. Please refer to the Equipment, materials, and reagents required section for equipment specifications. To equip the lab with appropriate pipettes, please refer to the IFU for volumes to be pipetted.

Zone Lab procedures Workstation
Pre-PCR Sample preparation and gDNA extraction.
Addition of template for the amplification step.
Biological safety cabinet equipped with:
• Vortex mixer
• Decapper for Lysing Matrix-E tubes
• Appropriate single- and 8-channel pipettes w/ wide orifice tips
Open bench equipped with:
• Bead beater
• Plate centrifuge for Lysing Matrix-E tubes
• Water bath
• DNA extraction robot
• Vortex mixer
• Microcentrifuge
• Dispenser pipette w/ tips
• Single- and 8-channel pipettes with tips
• Ice or cooling elements for other reagents and plate
Template- free Mastermix preparation for the amplification, clean-up, End-Labeling, and Hybridization steps. PCR cabinet with available equipment:
• Vortex mixer
• Microcentrifuge
• Dispenser pipette w/ tips
• Single channel pipettes w/ tips
• Freezing block for enzymes
• Ice or cooling elements for reagents and plate
PCR/Post-PCR Amplification reaction.
The quantification step.
Addition of template for the End-labeling step and the End-labeling reaction.
The Hybridization and signal detection step.
Open bench equipped with:
• Plate centrifuge for quick spin
• Thermal cycler
• DNA quantification system
• Ring magnet plate
• Luminex detection platform
• Vortex mixer
• Microcentrifuge
• Dispenser pipette w/ tips
• Single- and 8-channel pipettes w/ tips
• Ice or cooling elements for reagents and plate
• Freezing block for plates

EQUIPMENT, MATERIALS, AND REAGENTS REQUIRED

The lists below describe the required equipment, materials, and reagents for the GA-map® Dysbiosis Test Lx v2.
All items are required, unless otherwise specified. Genetic Analysis must be informed of any deviations from these requirements and a separate validation might be necessary.
The quantity listed for all equipment is based on the set-up described in the Laboratory Requirements section above and describes the minimum recommended quantity. It should be noted that different laboratory set-ups (e.g. with additional zones) might affect the recommended quantity.
FOR STEP 1 – GENOMIC DNA EXTRACTION
List of equipment

Equipment type Specifications/ requirement Alternative or recommended option Quantity
Decapper for Lysing Matrix-E tubes Capacity: 8-channel capper/decapper for matrix tubes 8-Channel Screw Cap Decapper (4105MAT), Thermo Scientific 1
AlteCap™ Switch (404000) with Cassette 12/96 – Internal thread Matrix screw caps (404014), AltemisLab
Bead beater FastPrep-96™ Homogenizer w/96-well plate insert (116010500), MP Biomedicals No validated alternatives 1
Plate centrifuge for Lysing

Matrix-E tubes

Capacity: Lysing Matrix-E 96 well rack with tubes (6 cm height) Force: 1300 rcf Eppendorf™ Centrifuge Centrifuge 5804/5810 w/ rotor for deepwell plates, Eppendorf 1
Water bath Capacity: deep well 96 plate Temperature: 65°C Any 1
DNA

extraction robot

KingFisher Flex – 96 deep well head (540 0630) w/magnetic micro-plate Separator, installed with KF Flex 96 KF heating block (2407 5420), Thermo Scientific No validated alternatives 1
Vortex mixer Speed: ~2800 rpm Any 2
Micro Centrifuge Capacity: 1.5ml/ 2ml tubes (spin down only) Any 1
Dispenser pipette w/tips Volumes to be dispensed: 20, 200, 250, 270 and 720 µl Multipette®E3/E3x (4987000371/4987000380),

Eppendorf

1
Pipettes single channel w/tips Volumes: see IFU
Wide orifice tips recommended for pipetting of fecal samples prior to gDNA extraction
Any Depends on volume range
Pipettes 8- channel w/tips Volumes: see IFU
Wide orifice tips recommended for pipetting of fecal samples prior to gDNA extraction
Any Depends on volume range
Ice or cooling blocks For keeping reagents/sample intermediates cold during handling Any NA

List of materials

Material Specifications/ requirement Alternative or recommended option
Lysing Matrix-E tubes Lysing Matrix E, 96-tube rack, barcoded tubes, 1 Rack (116984001B), (note: individual Matrix tubes with screw cap), MP Biomedicals No validated alternatives
Tube for Lysis mix Volume: 5, 15 or 50 ml Any
Deep well plate for DNA extraction robot KingFisher Deepwell 96 Plate, V- bottom, (95040450), Thermo Scientific 96 Well Plate SW 2ml with V- Bottom PP (BAKR43001-0504), VWR
Deep well Tip Combs for DNA extraction robot KingFisher 96 tip comb for DW magnets, (97002534), Thermo Scientific 96 Tip Comb PP (BAKR43001-0505), VWR
Elution plate for DNA extraction robot KingFisher 96 KF microplate

(200μL), (97002540), Thermo

Scientific

Well Plate 96 SW 0.2ml with V- Bottom PP (BAKR43001-0506), VWR
Adhesive PCR plate seal Suitable for deep well plate for DNA extraction robot Adhesive PCR Plate Seals (AB0558), Thermo Scientific
Microtiter sealing tape Suitable for plates for DNA extraction robot Adhesive Plate Seals (AB0580), Thermo Scientific
Microtiter plate w/seal for gDNA dilution Capacity: 96-well, ≥250µl Any
Reagent reservoir ≥25 ml Any

List of reagents 

Reagent Specifications/ requirement Alternative or recommended option
Extraction control positive (optional) Fecal sample of known quality Any
DNA extraction reagent kit mag™ maxi DNA purification kit, 288 tests (NAP40430), LGC Genomics
Note! Protease must be stored at -20°C upon reception
No validated alternatives
Ethanol (for extraction kit) 96-100% ethanol
≥500ml
Any
Acetone (for extraction kit) ≥99% acetone
≥350ml
Any
Water for dilution of gDNA and as PCR ctrl neg. Sterile, Nuclease-free water
≥500ml
Any

FOR STEP 2 – AMPLIFICATION OF THE BACTERIAL 16S RRNA GENE TO STEP 6 –HYBRIDIZATION AND SIGNAL DETECTION
List of equipment

Equipment type Specifications/ requirement Alternative or recommended option Used in assay steps Quantity
Plate centrifuge Capacity: 96-well plate (quick spin down only) Any Step 2
Step 3
Step 4
Step 5
Step 6
1
Thermal cycler VeritiPro™ 96-Well Thermal Cycler (A48141), Applied Biosystems Veriti™ 96-Well Thermal Cycler (4375786), Applied Biosystems Step 2
Step 4
Step 5
Step 6
1
T100™ Thermal Cycler (186- 1096), Bio-Rad
DNA

quantification system

FLUOstar OMEGA Microplate Reader with filter-based absorbance (415-103) w/ appropriate filters for the quantification assay, BMG LabTech Qubit® Fluorometer 3.0 (Q33216) or 4.0 (Q33238) or Flex (Q33327), Invitrogen Step 3 1
Ring Magnet Plate Capacity: 96-well Hybridization plate, 100 µL 96-Well Ring Magnet Plate (S380), Permagen Step 6 1
Luminex detection platform Luminex® 200™ system with xPONENT® version 4.2 or higher (LX200- XPON-IVD/RUO), Luminex* MAGPIX® system or NxTAG®‑Enabled MAGPIX® system with xPONENT® version 4.2 or higher (MAGPIX-XPON4.1-CEIVD), Luminex* Step 6 1
Vortex mixer Speed: ~2800 rpm Any Step 2
Step 3
Step 4
Step 5
Step 6
3
(including one of the two listed in Step 1)
Micro Centrifuge Capacity: 2 ml tubes
(spin down only)
Any Step 2
Step 3
Step 4
Step 5
Step 6
3

(including the one listed in Step 1)

Dispenser pipette w/tips Volumes to be dispensed: 20 µl and 40 µl Multipette®E3/E3x (4987000371/4987000380), Eppendorf Step 2
Step 5
Step 6
2
Pipettes single channel w/tips Volumes: see IFU Any Step 2
Step 3
Step 4
Step 5
Step 6
Depends on volume range
Pipettes 8- channel w/tips Volumes: see IFU Any Step 2
Step 3
Step 4
Step 5
Step 6
Depends on volume range
Ice or cooling blocks For keeping reagents/sample intermediates cold during handling Any Step 2
Step 3
Step 4
Step 5
Step 6
NA
Freezing block for tubes Capacity: 1.5/ 2 ml reagent tubes
For use during handling of enzymes
Benchtop cooler (5115- 0012), Thermo Scientific Step 2
Step 4
Step 5
1
Freezing block for plates Capacity: 96-well microtiter plate
For use during End- labeling
PCR-Cooler 0.2 mL (3881000031), Eppendorf Step 5 1

*Additional items for use and maintenance are required: kits for calibration and performance verification, sheath fluid or drive fluid, and an ultrasonic bath (40-60 kHz)
List of materials

Material Specifications/ requirement Alternative or recommended option Used in assay steps
Microcentrifuge tubes for mixing of reagents Volume: 1.5, 2 and 5 ml Safe-Lock tubes, Eppendorf Step 2
Step 4
Step 5
Step 6
Microtiter plate for PCR/End- labeling Capacity: 96-well PCR grade 96-Well Semi-skirted Flat Deck PCR Plates (AB1400), Thermo Scientific Step 2
Step 5
8-cap sealing strips Suitable for microtiter plate Flat PCR Caps, strips of 8 (AB0784), Thermo Scientific Step 2
Step 5
Microtiter sealing tape Suitable for microtiter plate Adhesive Plate Seals (AB0580), Thermo Scientific Step 2
Step 5
Tubes/plates for DNA quantification Suitable for the selected DNA quantification system For use with plate reader: Nunc™ 96-Well Microplate, Black (237108), Thermo Scientific Step 3
For use with Qubit® 3.0/4.0 Fluorometer: Qubit™ Assay Tubes (Q32856), Invitrogen
For use with Qubit® Flex Fluorometer: Qubit™ Flex Assay Tube Strips (Q33252), Invitrogen
Hybridization plate Thermowell™ 96-Well Polycarbonate PCR Microplates, Model P (6509), Corning 96-well Twin.tec™ PCR Plates, Unskirted, Divisible (0030133358), Eppendorf Step 6
Sealing film for Hybridization plate Microseal® ‘A’ (MSA5001), Bio-Rad No validated alternatives Step 6
Pierceable foil for Hybridization plate** Pierceable foil to cover 96-well plate Any Step 6
Reagent reservoir ≥25 ml Any Step 4
Step 6

**Only applicable if using the NxTAG®-enabled MAGPIX® detection platform
List of reagents

Reagent Specifications/ requirement Alternative or recommended option Used in assay steps
Water for PCR control neg. Sterile, Nuclease-free water, same as used for dilution of gDNA Any Step 2
Assay kit for DNA quantification For use with plate reader: Quant-iT™ 1X dsDNA HS Assay (Q33232), ThermoFisher Quant-iT™ PicoGreen™ dsDNA Assay Kit (P11496), ThermoFisher Step 3
For use with Qubit® Fluorometer: Qubit™ 1X dsDNA HS assay kit (Q33231), ThermoFisher Qubit™ dsDNA HS assay kit (Q32854), ThermoFisher

TRAINING PLAN

Training will be scheduled following system set-up. Genetic Analysis will train users in performing the process of GA-map® Dysbiosis Test Lx v2 from fecal gDNA extraction (or from PCR, if the extraction method differs from that described in the IFU) to generation of the reports. Basic instrument operation and use of the software will be included in the training. Those being trained will be required to have a basic knowledge of Microsoft Windows and use of general laboratory equipment and tools.
Operators must have been trained in use and maintenance of the Luminex system prior to the tech transfer.

CONTACT INFO

We are happy to help you with your inquiries.
Technical Support: support@genetic-analysis.com
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be considered unprotected by law. All trademarks and registered trademarks mentioned herein are the property of their respective owners.

S8-044-A1 GA-map Dysbiosis Test - Icon 2Genetic Analysis AS | Ulvenveien 80 | 0581 OSLO | Norway
www.genetic-analysis.com
Phone: +47 48 32 16 10
E-mail: info@genetic-analysis.com
© 2021-2023 Genetic Analysis, all rights reserved.