Introduction

The handheld optical multimeter is the integration of a power meter, optical light source, or visual fault locator, used to measure the optical power loss of the fiber link. It is
widely used in the fiber-optic line project construction, testing, and maintenance, such as digital data networks, telecommunication networks, and cable television. The figures
below display the product images of the two optical multimeters.
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - power meter

Accessories

FROM-201

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - AdapterFROM-103FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Adapter1

Installing

Inserting FC CablesFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - FC Cables

Install FC fiber cable.

Inserting ST Cables

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - SC connector

  1. Install ST connector.
    FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - ST connector
  2. Install ST fiber cable.FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - ST fiber cable

Inserting SC Cables

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - ST connector

1. Install SC connector.

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - SC fiber cable

2. Install SC fiber cable.

FROM-201

Function IntroductionsFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - MULTIMETER

Button

Description

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Power Power/Backlight Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Menu Menu Setting Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Up Button Up Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - REF Save/Query Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - REF
REF Setting Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - OPM OPM Reference Check/Set Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser On Off Button Laser On/Off Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Modulation Modulation Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - OLS OLS Wavelength/Down Button

Operation Instructions

  1. Power On/Off and Auto-off Function :
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Power button to turn on the instrument, then press for 2 seconds or longer to turn off. This multimeter has a power-saving function. If 10 minutes without any operation, the instrument will automatically shut down. If you need to shield this function and enable the multimeter to keep on working, only need to press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Power button and hold on when you boot the meter. After 2 seconds, the instrument will display FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - icon1which means permanent power on.
  2. Backlight Function:
    When the instrument is powered on, short press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Powerbutton to turn on/off backlight. The backlight function supports you to use the multimeter at night or on darker occasions, and you can set levels 0-9. If set as 0, backlight off, and no backlight even press button.
  3. Menu Setting
    Press the FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Menubutton to switch each setting interface, then press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Up Buttonor FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - OPMbutton to modify the specification. If all finished, press for 2 seconds to exit the menu setting. MENU
  4. OPM – Select Wavelength
    According to the project’s demand, we need to measure optical signals of different wavelengths. Then we need to select a corresponding wavelength to measure the
    optical power. If the measured wavelength is different from the wavelength we selected on OPM, it will lead to the measuring values meaningless. After turning it on, press theFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - OPM button to switch OPM wavelength from 850nm, 1300nm, 1310nm, 1490nm, 1550nm, 1625nm.
  5. OPM – Data Storage/Query
    On the OPM interface, long press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Savebutton, the display will show “SAVE” and storage number at the top left of the screen and will flash together 3 times. At the same time, the display will show the currently stored power value on the screen. After 2 seconds, the multimeter will restore the measuring interface, and save the current measurements in the instrument.
    The multimeter can store 1000 records. Each storage number automatically increases one by one, if, over the limit, the first data will be overwritten and so on.
    On the measuring interface, short press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Save button to the data-saving interface, you can RECORD -2s SAVE RECORD -2s SAVE

FROM-103

Function Introductions

Button

Description

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Power Power/Backlight Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser OPM Wavelength Shift/WAVE ID Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - VFL VFL On/Off Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Save1 OPM Data Save/Recall Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Unit OPM Unit Shift Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - REF OPM Reference Check/Set Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser Source On Laser Source On/Off Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser Laser Source Wavelength Shift Button
FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser Source Modulation Button Laser Source Modulation Button

Operation Instructions

  1. Power On/Off and Auto-off Function:
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Powerthe button to turn on the instrument with auto power off. If 10 minutes with any operation, it will automatically shut down.
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Powerthe button for 2 seconds when turning on the instrument, the auto-off function will be canceled, and the LCD will show “PERM”. Also, press it for 2 seconds to turn off the instrument.
  2. Backlight: 
    When the instrument power on, pressFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Power the button to turn on or turn off the backlight.
  3. OPM – Select Wavelength
    When testing, you must select the right wavelength. Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laserthe button for shifting the wavelength from 850nm, 1300nm, 1310nm, 1490nm, 1550nm, 1625nm.
  4. OPM – WAVE ID (optional)
    Long press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laserbutton to turn on the WAVE ID and display “id ” at the same time.
    Long press again to exit the WAVE ID.
  5. OPM – Select Unit:
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Unit button for shifting the unit from: dBm, nW/uW/mW, after pressingFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - REF button, it shifts to dB. After power off, the current unit will be saved. nW/uW/mW: 1mW = 1000uW, 1uW = 1000nW dBm: (dBm) = 10*log (mW) dB: (dB) = dBm – REF Unit REF
  6. OPM – Reference
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - REFbutton to check the reference value you set last time. And if keep pressing for 2 seconds, you can store the current dBm as a new reference value. Then it automatically shifts to dB. dB=dBm – REF REF You can store the REF value for each wavelength.
  7. OPM – Data Storage:
    Short press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Save1the button to check the data storage.
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laserthe button and FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Unitbutton to change the previous/next record. Long pressFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Save1 the button to save the current value.
  8. OLS On/Off
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser Source Onthe button to turn on/off the laser source output. Laser ON
  9. OLS – Select Wavelength/WAVE ID
    Short pressFS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser the button to shift the wavelength. Long press (>2s) the button to turn on the WAVE ID function, the wavelength glints on LCD. Long press it again to turn off the function.
  10. OLS – Modulation Select
    Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Laser Source Modulation Buttonthe button to shift from 270Hz, 1KHz, 2KHz. MOD
  11. Visual Fault Locator (optional)
    The instrument can be built in an optional VFL module. Press FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - VFLbutton to shift the conditions: on-> glint-> off. VFL
  12. Battery Energy Detect
    FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Sufficient power Sufficient power
    FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - More power More power
    FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Less power Less power
    FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Insufficient battery power Insufficient battery power
  13. Battery Charge
    When you use rechargeable batteries and the meter indicates insufficient power, it should be turned off and charged. Long time under voltage will cause the life of the
    battery to be shortened.
    When charging, the battery indicator on the LCD will flash. After charging fully, the indication will stop flashing and show full. Don’t charge for more than 48 hours.
    If charging while using the device, the time will be longer. Rechargeable batteries must be in the device when you use the AC/DC adaptor for charging.
    And do not charge the non-rechargeable batteries, or the device will be destroyed and also lose the guarantee.

Maintenance

  1. The interface is sensitive, please carefully plugin and pull out connectors.
  2. Keep using one type of optical adapter to avoid excess loss from different connectors.
  3. Please use a dust-proof cap for protection to avoid being scratched or contaminated when not in operation.
  4. It is important to keep all optical connectors and surfaces free from oil, dirt, or other contamination to ensure proper operation.
  5. In order to avoid the electric shock, please do not disassemble the components. Disobeying the standard instruction may lead to safety issues.
  6. Remove the batteries when the battery power becomes weak or when the instrument is not in use for extended periods. This will prevent damage to the multimeter from battery leakage at such time.

Online Resources

Download: https://www.fs.com/download.html
Help Center: https://www.fs.com/service/help_center.html
Contact Us: https://www.fs.com/contact_us.html

Product Warranty

FS ensures our customers that any damage or faulty items due to our workmanship, we will offer a free return within 30 days from the day you receive your goods.

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Warranty Warranty: All-Optical Multimeters enjoy a 1-year limited warranty against defects in materials or workmanship. For more details about the warranty, please check at https://www.fs.com/policies/warranty.html

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter - Return Return: If you want to return item(s), information on how to return can be  found at https://www.fs.com/policies/day_return_policy.html

Q.C. PASSED
Copyright © 2020 FS.COM All Rights Reserved.

Documents / Resources

FS FHOM-201 Power Meter + Laser Source Handheld Optical Multimeter [pdf] User Guide
FHOM-201, FHOM-103, Power Meter Laser Source Handheld Optical Multimeter

References

Leave a comment

Your email address will not be published. Required fields are marked *