F2 ZR Touch S Cordless Endodontic Motor

F2-ZR-Touch-S-Cordless-Endodontic-Motor-product

Specifications

  • Manufacturer: F2 Medical Supplies
  • Model: ZR Touch
  • Weight: Varies based on components
  • Speed range: 150-1000rpm
  • Torque range: 0.6-5.0Ncm
  • Operating conditions:
    • Temperature: +10 to +40 degrees Celsius
    • Humidity: 30% – 75%
    • Pressure: 70kPa – 106kPa
  • Storage conditions:
    • Temperature: -10 to +50 degrees Celsius
    • Humidity: 10% – 85%
    • Pressure: 70kPa – 106kPa

Product Usage Instructions

Warning

Read the manual carefully before use. Ensure the device is turned off before installing or removing components.

Product Introduction

The product includes a motor handpiece, a 1:1 contra angle, and an adapter.

Device Installation

Installation & Removal of Contra Angle

  • Installation: Connect the contra-angle at adjustable head positions by aligning the positioning pins with the slots and inserting until it clicks.
  • Removal: Pull out the contra-angle straight when removing.

Inserting and Removing the File

File Insertion: Press the top cover of the contra-angle, insert the file until it locks into place, then gently pull out to confirm locking.

Frequently Asked Questions

  • Q: Is the ZR Touch suitable for home use?
    • A: No, the ZR Touch is designed for professional use by qualified dentists in hospitals or dental clinics.
  • Q: How should I clean and sterilize the ZR Touch?
    • A: Refer to the “Cleaning, disinfection, and sterilization” section in the manual for detailed instructions on proper cleaning and sterilization procedures.

Please read the ins truction manual before operating

CONGRATULATIONS!

Thank you for your purchase. For optimum safety and performance, read this manual carefully and pay close attention to warnings and precautions before using the instrument.
Please keep this manual in a convenient place for quick and easy reference.

Equipment Introduction

The device is specially designed for dentists and used to drive root canal instruments to rotate and reciprocate. It is clinically used in root canal shaping and cleaning in root canal treatment

Scope of application

This equipment must be used in hospitals or dental clinic by qualified dentists

Contraindications

  • Disabled by doctors with pacemakers.
  • Patients with pacemakers (or other electrical equipment) and are warned not to use small appliances (such as electric shavers, hair dryers, etc.) are disabled.
  • Prohibited for patients with hemophilia.
  • Use with caution in patients with heart disease, pregnant women and young children.

Warning

  • Please read the instruction manual carefully before using this device for the first time.
  • This equipment must be used in hospitals or dental clinic by qualified dentists.
  • The patient shall not be an intended operator.
  • This device can only be used with its accompanying accessories, do not use any other power adapters, contra angle, lithium batteries, etc.。
  • There is an unacceptable risk of improper replacement lithium batteries.please use the lithium batteries provided by us and replace it according to the correct steps in the instructions.
  • Do not disassemble or repair the equipment by yourself. Any disassembly or repair by yourself may violate safety regulations and cause injury to the patient.
  • Any self-dismantling and repair will not be guaranteed by any promise.
  • Do not remove the contra angle and press the push button before the motor stops rotating,otherwise the contra angle may be damaged.
  • Do not remove the file before the motor stops rotating, otherwise the operator may be injured.
  • Before starting the motor,make sure that the file is in place and locked.
  • Please set the motor output speed and torque according to the parameters recommended by the file manufacturer.
  • Using the reciprocating mode for a long time may cause the motor handpiece to overheat. At this time, it should be left to cool down to reuse. If the motor handpiece frequently overheats, please contact your local dealer.
  • Do not put the equipment near flammable materials. Do not operate the device in the presence of a flammable anesthetic mixture of air, oxygen or nitrogen oxides.
  • Portable and mobile RF communication equipment may affect the performance of this equipment. Avoid strong electromagnetic interference when in use, such as close to mobile phones, microwave ovens, etc.
  • Do not place this equipment near a heat source. The device must be operated and stored in a reliable environment.
  • The equipment is usually operated at +10°C to +40°C, relative air humidity 30%-75%, and atmospheric pressure (70~106) kPa.
  • This equipment should be sterilized and disinfected for the first time and after the patient is used to avoid cross-infection.
  • The motor handpiece cannot be sterilized by high temperature and high pressure, only neutral disinfectant or alcohol can be used to wipe the surface

Product introductionF2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (2)F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (3)

Device installation

Installation & removal of contra angle

Installation

The contra-angle can be connected at З adjustable head positions, Align the positioning pins of the contra-angle with the positioning slots of the contra-angle and insert the head until it clicks.

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (4)

Removal

When removing the contra-angle, pull it straight out .

Warning

When install or removing the contra angle , make sure to turn off the device

Inserting and Removing the File

File insertion

Press the top cover of the contra – angle and insert the file until the file tail is loaded into the inner Iock slot of the contra – angle .Release the contra – angle top cover and gently pull the file out to confirm that the file is locked.

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (5)

File removal
Press the top cover of the contra-angle , and at the same time pull out the file at an angle perpendicular to the drive shaft of the hand piece.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (6)
Warning
When installing and removing files, the motor needs to be turned off to prevent damage to the operator. When installing and removing the file, press the top cover of the hand piece until the installation or removal action is completed. Otherwise, the locking needle part inside the contra angle will be damaged. Please use file that meets ISO standards.( ISO standards; Ф2.334-2.350mm)

Equipment function and use

Button definition and settingsF2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (7)

  • Power on and off
    Short press the main key“ F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (9)”to turn on the device. Press the key“F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (8) ”, then press the main key“F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (9) ”to turn off the device. The device will automatically shut down after 5 minutes without any operation.
  • Motor start and stop
    When the device is on, select the appropriate mode, press the main key“F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (9) ” to start the motor, press the main button“F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (9) ” again to stop the motor.
  • Terms and definitions
    When the device is on, select the appropriate mode, press the main key“ ” to start the motor, press the main button“ ” again to stop the motor.

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (10)

Mode introduction

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (11) F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (12)

Parameter settingsF2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (13)

Preset the file program selectionF2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (14)

Motor handpiece function settings

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (15)In order to satisfy the left-handed and right-handed operators, the viewing angle of the display screen can be adjusted by 180°. Press the “>” key without releasing it, and then press the main button”F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (9) “,to enter the adjustment interface. Then press the adjustment keys “<“”>” to select.the key” “to confirm the selection.

Protective function of automatic reverse

In the M1~M5 mode, the motor rotates in one direction. When the motor load torque reaches the preset value, the motor will automatically reverse; when the motor load returns to half of the preset torque value, the motor returns to the original direction of rotation.

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (16)

Warning

  1. The torque trigger automatic reverse protection function only works in M1~M5(one-way rotation) modes
  2. This function does not take effect in M6,M7,M8 modes.
  3. When the power of the device is less than 20%, it is not enough to support the maximum torque of the motor 4Ncm, please charge it in time.

Device charging

This product has a built-in rechargeable lithium battery. Insert the power adapter into the USB port at the bottom of the handle, and the device will enter the charging state.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (17)

After fully charged, the battery icon shows full and in a static state, please unplug the power adapter in time when fully charged.

Warning

Keep away from heat sources when charging.

  • Device calibration
  • Calibration method
  • Press the “<” button , then the main button to enter the calibration mode.

Warning

Please keep the battery above 50% to calibration.

Maintenance

Battery replacement

To replace the battery, use the lithium battery provided by the original manufacturer. Please contact your local dealer or manufacturer for replacement.

Replacement procedure:

  1. Ensure that the device is turned off
  2. Remove the silicone plug of the battery cover with tweezers, and then remove the fixing screws with a screwdriver
  3. Remove the battery cover and remove the old battery
  4. Put the new battery into the battery compartment and confirm that the battery is normal when you turn it on.
  5. Install the battery cover and lock it with screws, and then plug the silicone plug into the screw hole.

Warning

  • Do not disassemble parts that are not related to the replacement of the battery.
  • Do not replace the battery with wet hands, otherwise the battery may short-circuit and damage the device.
  • Do not use lithium batteries other than the original manufacturer, otherwise the equipment may be damaged.
  • If the device is not used for a long time, the lithium battery should be removed.

Battery replacement

To extend the use of the contra angle, it should be lubricated each time it is used or before sterilization.

  1. Rotate the oiler into the nozzle of the oiler bottle.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (18)
  2. Insert the oil injector into the tail of the contra angle and then fill for 2 to 3 seconds until the oil is flowing out of the contra angle.
  3. Place the tail of the contra-angle up right for more than 30 minutes and use gravity to drain excess lubricant.

Warning

The motor handpiece can not be filled with oil.

Note

When filling oil, hold the Contra-an-gle tightly to prevent the Contra-an-gle from leaving the oil injector due to the oil injection pressure.

Troubleshooting

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (21)

If you have not found the necessary information, you may consult the manufacturer via hotline 86-755-28540953,E-mail:endoservice@dental-perfect.com or address to the service department.

Cleaning, disinfection and sterilization

For the purpose of hygiene and safety, the components must be cleaned, disinfected and sterilized before the first use and after each use to prevent cross-infection.

Wipe disinfection

The device should be cleaned wiped and disinfected regularly. After use, use 75% medical alcohol to thoroughly wipe the contact surface of it to remove the residual substances or debris on the surface of the device.

Sterilization

The contra-angle must be cleaned, disinfected and sterilized for the first use and after each use.

Please seal the contra-angle in sterilized packaging that meets the EN-868-5 standard.

  • At 134°C, the sterilization time is at least 4 minutes.
  • At 121°C, the sterilization time is at least 20 minutes, and the drying time is at least 15 minutes.

Note

The contra-angle can be sterilized repeatedly for at least 250 cycles. The useful life is 1 year, and its lifespan is also different according to the frequency and time of clinical use, and the difficulty of the patient’s oral root canal during treatment.

Warning

  • Do not disinfect the device by steaming.
  • Do not heat and sterilize the device.

Disposal of the product

  • Do not throw the device into the household garbage system. Dispose of waste in accordance with local laws and regulation

Statement

ZR Touch S warranty for 12 months from the date of purchase by the user. The company does not provide technical data(such as circuit principles, component lists,etc.) to other organizations. If vou have any questions, please contact your local dealer or manufacturer.

The manufacturer is not responsible for:

  1. Use ZR Touch S for purposes and purposes that violate the specific provisions in this instruction manual.
  2. Use methods that are contrary to those stated in this manual for cleaning, disinfection and sterilization operations.
  3. Disassemble and repair by unauthorized personnel.

Symbol Description

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (23) F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (24)

EMC Declaration

Electromagnetic Compatibility

Attention

  1. ZR Touch endo motor meets the relevant requirements of IEC60601-1 standard electromagnetic compatibility;
  2. The user should install and use it according to the electromagnetic compatibility information provided by the accompanying document;
  3. Portable and mobile RF communication equipment may affect the performance of ZR Touch endo motor. Avoid strong electromagnetic interference when using it, such as close to mobile phones, microwave ovens, etc.;
  4. The guidelines and manufacturer’s declaration are detailed in the attachment

Warning

  1. ZR Touch endo motor should not be used close to or stacked with other equipment. If it must be used close or stacked, it should be observed and verified that it can operate normally under the configuration used;
  2. Except for the cables sold by the manufacturer of the ZR Touch as spare parts for internal components, the use of accessories and cables other than those specified may result in an increase in emission or a decrease in immunity of the ZR Touch S endo motor.

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (25)

Guidelines and manufacturer’s declaration-electromagnetic emissions

The model ZR Touch is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (26)

Guidance & Declaration — electromagnetic immunity

The model ZR Touch is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig 29

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig 30

NOTE1:At 80MHz & 800MHz,the higher frequency range applies.
NOTE2:The guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,objects and people.

  1. Stationary transmitters, such as base stations for wireless (cellular/cordless) telephones and terrestrial mobile radios, amateur radios, AM and FM radio broadcasts, and television broadcasts, whose field strength cannot be accurately predicted in theory. In order to assess the electromagnetic environment of fixed radio frequency transmitters, electromagnetic field surveys should be considered. If the measured field strength of the place where ZR Touch S is located is higher than the above applicable RF compliance level, ZR Touch S should be observed to verify its normal operation. If abnormal performance is observed, supplementary measures may be necessary, such as re-adjusting the direction or position of ZR Touch S.
  2. In the entire frequency range of 150KHz~80MHz, the field strength should be lower than 3 V/m.

Recommended isolation distance between portable and mobile RF communication quipment and ZR Touch S

ZR Touch is expected to be used in an electromagnetic environment where radio frequency radiation disturbance is controlled. According to the maximum rated output power of the communication equipment, the purchaser or user can prevent electromagnetic interference by maintaining the minimum distance between the portable and mobile radio frequency communication equipment (transmitter) and ZR Touch as recommended below.F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (27)

For the rated maximum output power of the transmitter not listed in the above table, the recommended isolation distance d, in meters (m), can be determined by the formula in the corresponding transmitter frequency column, where P is the emission provided by the transmitter manufacturer The maximum output rated power of the machine, in watts (W).

Note

  1. At 80MHz and 800MHz frequency points, the higher frequency band formula is used.
  2. The engineering medical frequency band between 150kHz and 80MHz refers to 6.765MHz~6.795MHz、13.553MHz~13.567MHz、 26.957MHz~27.283MHz和40.66MHz~40.70MHz.
  3. An additional factor of 10/3 is used to calculate the recommended isolation distance for transmitters in the 150kHz~80MHz engineering and medical frequency band and 80MHz~2.5GHz frequency range to reduce portable/- mobile  radio frequency communication equipment accidentally brought into the patient area The possibility of interference caused by time.
  4. These guidelines may not be suitable for all situations. Electromagnetic propagation is affected by absorption and reflection from buildings, objects and human bodies.

F2 Medical Supplies

Find Endodontic Files and more Endo Equipment Online

F2-ZR-Touch-S-Cordless-Endodontic-Motor-fig (28)

Shenzhen Perfect Medical Instruments Co., Ltd.

  • Address:Room 103, Building 3, No. 2, Weiqun Road, 4th Community, Henggang
  • Street, Longgang District, Shenzhen.
  • Post Code:518115
  • Tel:+86(755)28540953
  • Fax:+86(755)28540953
  • Website:www.dental-perfect.com
  • E-mail sales@dental-perfect.com
  • EC REP
  • SUNGO Europe B.V.
  • Olympisch Station 24, 1076DE Amsterdam, Netherlands
  • Tel:+31(0)202111106
  • E-mail: ec.rep@sungoglobal.com
  • Version:01
  • PF05S-IFUE-01
  • Issued:2023.5.27
  • Copyright: ShenzhenPerfect Medical Instruments Co., Ltd.

Documents / Resources

F2 ZR Touch S Cordless Endodontic Motor [pdf] Instruction Manual
ZR Touch S Cordless Endodontic Motor, ZR Touch S, Cordless Endodontic Motor, Endodontic Motor, Motor