IMPLEN DiluCell 10 Automatic Sample Dilution

Specifications
- Wavelength Range
- Optical Pathlength
- Volume Requirement
- Tolerance
- Dilution Factor
- Ordering Information
- Packaging
Product Overview
- The DiluCellTM 10 simplifies concentration measurements by reducing the optical pathlength, allowing direct reading of highly concentrated samples while staying within your photometer’s linear range. The Dilution factor is 1:10.
How DiluCellTM Works
- The Lambert-Beer-Law empirically relates the absorption of light to the properties of the sample. The built-in virtual dilution of DiluCellTM 10 enables a 1:10 dilution in accordance with the Lambert-Beer Law.
INTRODUCTION
Thank You for Choosing DiluCell™
- This advanced, disposable cuvette is designed to simplify workflows and enhance accuracy for absorbance measurements—especially when working with highly concentrated samples.
- DiluCell™ offers virtual dilution without the need to physically dilute your samples. The wide wavelength range of 340 nm – 950 nm enables the user to incorporate the DiluCellTM 10 cuvettes for all standard applications.
- The wide range of applications of our DiluCellTM 10 cuvettes include but is not limited to.
- Cell Density (OD600) – Monitoring bacterial/yeast/fungal growth
- Protein Assays – Bradford, BCA, Lowry, etc.
- Enzyme Kinetics
- Colorimetric Assays
- Turbidity/McFarland analyses
- Please read this manual carefully before use.
Product Overview
- DiluCell™ 10 simplifies concentration measurements by reducing the optical pathlength, allowing direct reading of highly concentrated samples while staying within your photometer’s linear range.
- Wavelength range: 340-950 nm
- Required sample volume: 300 µl
- Optical pathlength: 1 mm @ 8.5 mm center height
- Dilution factor: 1:10
How DiluCell™ Works
- The Lambert-Beer-Law, also known as Beer’s Law, empirically relates the absorption of light to the properties of the sample.
Lambert-Beer Law:
Where:
- I0 = incoming light
- I = outgoing light
- Ε = extinction coefficient
- C = concentration
- D = pathlength (1mm @ 8.5mm center height)

Built-in Virtual Dilution
- Traditional cuvettes require manual sample dilution and volumes of at least 1 ml.
- The innovative design of the DiluCell™ 10 reduces the light path from the standard 10mm to just 1mm, thereby enabling a built-in 1:10 virtual dilution in accordance with the Lambert-Beer Law.
- This allows you to work with smaller sample volumes (only 300 μl), while eliminating manual dilution steps, saving time, and minimizing the risk of dilution/pipetting errors and cross-contamination.

Operating Instructions
- Correct Insertion
- Place the DiluCell™ in the cuvette holder with the wide window facing the light beam.
- Implen devices such as the Implen OD600® and the *NanoPhotometer® product range feature directional arrows to assist with correct alignment.
- N80 and C40, cuvette-enabled devices.
- Important: DiluCell™ will not work in any other orientation.
- Do not mix cuvette types within the same measurement series (e.g., blanks).
- Sample Loading
- The geometry of DiluCell™ supports bubble-free filling.
- It is, however, still highly recommended to visually check your samples for air bubbles after loading.
- If bubbles are present, tap gently to eliminate them.
- Volume Requirement
- Use 300 µl of the sample for optimal performance.
- Data Handling
- Simply multiply your absorbance values by 10 post-measurement or set your instrument to apply a 10× dilution factor.
Important Usage Notes
Single Use Only
Reuse is not recommended due to:
- High cross-contamination risk
- Bubble formation during refilling
- Inability to properly decontaminate (especially for biological samples)
Do Not Autoclave
- Autoclaving may compromise the structure of the cuvette and result in inaccurate readings.
Solvent Compatibility
- For chemical resistance and solvent use, contact your local Implen representative or visit www.implen.de.
Specifications
- Parameter DiluCell™ 10
- Wavelength Range 340 nm – 950 nm
- Optical Pathlength 1 mm (at 8.5 mm center height)
- Volume Requirement 300 μl
- Tolerance ±10%
- Dilution Factor 1:10
- Ordering Information DC 10
- Packaging 96 disposable cuvettes per pack
Contact & Support
Europe, Asia, the South Pacific, North and South America Middle East, and Africa
- Phone: +49 89 72637180
- Fax: +49 89 726371854
- Email: info@implen.de
- Website: www.implen.de
- Implen GmbH Schatzbogen 52
- 81829 München Germany
North and South America
- Phone: +1 8187486400
- Fax: +1 8184490416
- Email: info@implen.com
- Website: www.implen.com
- Implen, Inc.
- Unit 104
- 31194 La Baya Drive
- Westlake Village, CA 91362
- USA
FAQ
Can DiluCellTM be reused?
No, DiluCellTM is designed for single-use only to avoid cross-contamination risks and maintain accuracy.
Can I autoclave DiluCellTM for sterilization?
No, autoclaving may compromise the cuvette's structure and lead to inaccurate readings.
Documents / Resources
![]() | DiluCell 10 Automatic Sample Dilution |
References
- mailto:info@implen.comimplen.com
- mailto:info@implen.deimplen.de
- Implen | The Most Innovative UV/Vis Spectrophotometerswww.implen.com
- Implen | The Most Innovative UV/Vis Spectrophotometerswww.implen.com
- Implen | The Most Innovative UV/Vis Spectrophotometerswww.implen.de
- Implen | The Most Innovative UV/Vis Spectrophotometerswww.implen.de
- User Manualmanual.tools
Published: January 15, 2026
Updated: January 16, 2026