Instruction Manual for novasina nLink models including: 8853 IP Analog Transmitter, 8853, IP Analog Transmitter, Analog Transmitter, Transmitter

Philippe Troesch

Operation Manual nLink AS IP & EC Transmitter

Novasina AG, Neuheimstrasse 12, 8853 Lachen, Switzerland Telephone 41 55 642 67 67 Fax 41 55 642 67 70, E-Mail: sales@novasina.ch, www.novasina.com


File Info : application/pdf, 17 Pages, 1.22MB

PDF preview unavailable. Download the PDF instead.

007523.01 Manual-nlink-AS-xx EN
Operation Manual nLink+ AS IP & EC
Transmitter
Novasina AG, Neuheimstrasse 12, 8853 Lachen, Switzerland Telephone +41 55 642 67 67 Fax +41 55 642 67 70, E-Mail: sales@novasina.ch, www.novasina.com

Operation Manual nlink+ AS IP & EC
Table of contents
1. General.............................................................................................................................3 2. Safety ...............................................................................................................................3 3. Product Description ........................................................................................................4
3.1. Technical Data nLink+ IP AS......................................................................................4 3.2. Technical Data nLink+ EC AS ....................................................................................5 3.3. Technical Data nLink+ dp IP.......................................................................................6 3.4. Technical Data nLink+ dp EC .....................................................................................7 3.5. Electrical Installation...................................................................................................8 3.6. Wiring Diagram ..........................................................................................................9 4. Initial Operation ............................................................................................................. 10 4.1. Windows PC Software Installation ............................................................................ 10 5. Configuration with nSoft-ACT-T ...................................................................................11 6. Calibration / verification of measuring values.............................................................15 6.1. Sensor cables and Configuration set ........................................................................ 15 7. Accessories nSens HT .................................................................................................. 16

July 2023 / Doc.No. 007523.01

2 / 17

Operation Manual nlink+ AS IP & EC
Scope of application: This manual is valid for all nlink+ transmitter systems with firmware version V01.00 or higher.
1. General
The nLink+ Transmitter has been designed for fixed mounting, either directly at or near the measuring location (IP version) or in the electrical cabinet (EC version=electrical cabinet). All nSens probes are compatible, either directly on the housing (IP version) or by the 3wire nSens cable or the nSens extension cable.
Configuration possible with the built in USB connection by Windows PC.
2. Safety
This instrument has left the factory in a faultless condition. No inappropriate modifications are allowed under the terms or the warranty. Please consider all notices and warning signs on the instrument and in this operating manual. Please also note:
· This instrument has been developed only for the measurement of clean air within the defined specifications, operate the instrument only for this purpose. In case of other applications, outside of these specified uses, the supplier accepts no responsibility for any damage caused.
· The installation work shall be only done by skilled personnel (electrician). · The instrument may only be operated under the specified operating conditions. · Any faults that may occur and cause damage to material and people, additional safety
precautions should be implemented. In case of any faults, the defined operating conditions have to be observed (e.g. limit switch etc.). · The instrument is not adequate for the installation in rooms with explosion hazard. · The installation has to be effected in accordance with the local electrical installation regulations as well as this operating manual. · The instrument contains ESD-sensitive parts. Please follow the indicated safety measures. · Use only original Novasina accessories and spare parts. · Without any written approval by Novasina no adaptations and modifications shall be undertaken on the instrument.

July 2023 / Doc.No. 007523.01

3 / 17

Operation Manual nlink+ AS IP & EC
3. Product Description
3.1. Technical Data nLink+ IP AS

Name Art-No. Power supply
Power consumption Display Analogue outputs (2 outputs)
Status LED Housing material Protection class Soldering material Working temp. Storage temperature CE-/EMC

nLink+ IP AS 2602078 24V DC, Permissible voltage range: 5 to 39V

<0.5W none, values on PC 2 scalable analogue outputs, current 0/4..20mA or voltage 0/2..10V

Accuracy <0.05% of span Linearity <0.05% of span Temperature effect 0.005% of span / °C

Load (I):

min. 0  / max. 500  or (Uin-2V)/Imax

Load resistance (U):

min. 10 k / max.  

LED for power On, LED for nSens connected

ABS

IP67

lead free (RoHS compliant)

0 to 50°C

-10 to 60°C (non-condensing)

Safety: IEC 61010-1:2020

EMC: IEC 61000-6-2:2016, EN 61000-6-2:2019

IEC 61000-6-3:2020, EN 61000-6-3:2007+A1:2011

July 2023 / Doc.No. 007523.01

4 / 17

Operation Manual nlink+ AS IP & EC
3.2. Technical Data nLink+ EC AS

Name Art.-No. Power supply
Power consumption Display Analogue outputs (2 outputs)
2* 0/4 - 20mA or 2* 0/2 - 10V
Status LED Housing material Protection class Soldering material Working temp. Storage temperature CE-/EMC

nLink+ EC AS 2602080 24V DC, Permissible voltage range: 5 to 39V

<0.5W none, values on PC 2 scalable analogue outputs current 0/4..20mA or voltage 0/2..10V

Accuracy <0.05% of span Linearity <0.05% of span Temperature effect 0.005% of span / °C

Load (I):

min. 0  / max. 500  or (Uin-2V)/Imax

Load resistance (U):

min. 10 k / max.  

LED for power On, LED for nSens connected

PA6.6 (UL94V0), mounting rail holder

none, installation in protected cabinet required

lead free (RoHS compliant)

0 to 50°C

-10 to 60°C (non-condensing)

Safety: IEC 61010-1:2020

EMC: IEC 61000-6-2:2016, EN 61000-6-2:2019

IEC 61000-6-3:2020, EN 61000-6-3:2007+A1:2011

July 2023 / Doc.No. 007523.01

5 / 17

Operation Manual nlink+ AS IP & EC
3.3. Technical Data nLink+ IP dp

nLink+ IP Art.-No. Measurement Range Accuracy at 20°C Temperature effect Max. permissible overpressure Response time T63 Max. Resolution Long term stability Flow rate Ambient pressure dependency Ambient pressure: Range Ambient pressure: Accuracy General Specification Power supply

AS A / AS AA

AS B / AS BB

2602079 / 2602141 2602172 / 2602173

-25 to +25 Pascal

-250 to +250 Pascal

Typical +/- 0.15%

Typical +/- 0.25%

Max. +/- 0.10Pa

Max. +/- 0.10Pa

2 bar (burst pressure 4 bar)

<1s typ.

0.1 Pa

+/-0.05% FSS (typ)

<200ul/min

Compensated with built in abs pressure sensor

700 ­ 1260 hPa / mBar

+/- 0.5 hPa

AS C / AS CC 2602174 / 2602175 -5000 to +5000 Pascal Typical +/- 1 % Max. +/- 0.20Pa 0.1 bar (burst pressure 0.3 bar)
None

24V DC, Permissible voltage range: 5 to 39V

Power consumption Display Analogue outputs
2* 0/4 - 20mA or 2* 0/2 - 10V

<0.5W None, values on PC 2 scalable analogue outputs, current 0/4..20mA or voltage 0/2..10V
Accuracy <0.05% of span Linearity <0.05% of span Temperature effect 0.005% of span / °C

Status LED Housing material Protection class Soldering material Working temp. CE-/EMC

Load (I):

min. 0  / max. 500  or (Uin-2V)/Imax

Load resistance (U):

min. 10 k / max.  

LED for power On, LED for nSens connected

ABS

IP67

lead free (RoHS compliant)

0 to 50°C

Safety: IEC 61010-1:2020

EMC: IEC 61000-6-2:2016, EN 61000-6-2:2019

IEC 61000-6-3:2020, EN 61000-6-3:2007+A1:2011

July 2023 / Doc.No. 007523.01

6 / 17

Operation Manual nlink+ AS IP & EC
3.4. Technical Data nLink+ EC dp

nLink+ EC Art.-No. Measurement Range Accuracy at 20°C Temperature effect Max. permissible overpressure Response time T63 Max. Resolution Long term stability Flow rate Ambient pressure dependency Ambient pressure: Range Ambient pressure: Accuracy General Specification Power supply

AS A / AS AA

AS B / AS BB

2602044 / 2602134 2602203 / 2602204

-25 to +25 Pascal

-250 to +250 Pascal

Typical +/- 0.1%

Typical +/- 0.25%

Max. +/- 0.10Pa

Max. +/- 0.10Pa

2 bar (burst pressure 4 bar)

<1s typ.

0.1 Pa

+/-0.05% FSS (typ)

<200ul/min

Compensated with built in abs pressure sensor

700 ­ 1260 hPa / mBar

+/- 0.5 hPa

AS C / AS CC 2602205 / 2602206 -5000 to +5000 Pascal Typical +/- 1% Max. +/- 0.20Pa 0.1 bar (burst pressure 0.3 bar)
None

24V DC, Permissible voltage range: 5 to 39V

Power consumption Display Analogue outputs

<0.5W None, values on PC 2 scalable analogue outputs, current 0/4..20mA or voltage 0/2..10V

2* 0/4 - 20mA or 2* 0/2 - 10V

Accuracy <0.05% of span Linearity <0.05% of span Temperature effect 0.005% of span / °C

Status LED Housing material Protection class Soldering material Working temp. CE-/EMC

Load (I):

min. 0  / max. 500  or (Uin-2V)/Imax

Load resistance (U):

min. 10 k / max.  

LED for power On, LED for nSens connected

PA6.6 (UL94V0), mounting rail holder

none, installation in protected cabinet required

lead free (RoHS compliant)

0 to 50°C

Safety: IEC 61010-1:2020

EMC: IEC 61000-6-2:2016, EN 61000-6-2:2019

IEC 61000-6-3:2020, EN 61000-6-3:2007+A1:2011

July 2023 / Doc.No. 007523.01

7 / 17

Operation Manual nlink+ AS IP & EC

3.5. Electrical Installation

Clamping range Wires:

Wire specification

0.13 - 1.5mm2 (Push-in Spring clip)

w. plastic collar ferrule DIN 46228/4: 0,25 - 0.75 mm²

w. wire end ferrule DIN 46228/1: 0,25 - 1.50 mm²

Solid, min. H05(07) V-U

0.2 - 1.50 mm²

Wire connection cross section AWG28 - 14

Remark: Cable specifications depend on the installation and have to be defined by the designer or installer. Heavy machinery and other instrumentation should not share the same power supply wiring. Use noise filters and surge protectors if required.

For EMC protection it is recommended to take the following measures:

· Wires emitting interference must be separated from measurement and analysis units · Parallel guidance of measurement cables and electrical power cables must be avoided, use different channels with separation (see European Standard EN50170 for detailed
information)

July 2023 / Doc.No. 007523.01

8 / 17

3.6. Wiring Diagram

Operation Manual nlink+ AS IP & EC

Remark: If more than 1 probe is connected, the single wires can be fixed in parallel in the same screw terminal.

July 2023 / Doc.No. 007523.01

9 / 17

Operation Manual nlink+ AS IP & EC
4. Initial Operation Before initial operation the power supply wiring and analogue output configuration should be checked (see wiring diagram).
Important remark: The power supply fuse protection has to be verified according to the local regulation. Voltages over 39VDC/27.6VAC result in a severe damage of the device!
4.1. Windows PC Software Installation
The software «nSoft-ACT-T» is a requirement for the configuration (signal outputs, sensor parameters etc). The software must be installed on a Windows PC with administrator rights and following minimum requirements:
Supported operating systems: · Windows XP with SP3 or higher (32+64bit) · Administrator-rights
Hardware: · CPU: mind. 1 GHz · USB- connection · min. 512 MB free memory · min. 4 GB free hardware memory · Software Adobe Acrobat Reader
The software can be obtained on the website www.novasina.ch or send an email to sales@novasina.ch. Software is license free (Freeware), the special USB configuration cable with adapter is necessary (see accessories)

July 2023 / Doc.No. 007523.01

10 / 17

Operation Manual nlink+ AS IP & EC

5. Configuration with nSoft-ACT-T
Connect the USB cable on the PC and the nlink+. Open the software by clicking on the Icon after installation.
The nlink+ does not require additional power supply. The USB connector is sufficient for configuration and sensor connection. Analog signals / outputs are only active with the additional power supply.

Startup

Connect configuration cable on the nlink+ and your PC. Choose the adapter in the dropdown menu.

Main Screen

1

2

3 4

6

5

7
July 2023 / Doc.No. 007523.01

11 / 17

Operation Manual nlink+ AS IP & EC
1: Activate password protection, requires a password set (2) 2: Set Password: the password is stored directly on the transmitter. 3: Read Data: Read configuration from the nlink+ 4: Write Data: Store the configuration on the nlink+
5: Details from the nlink+, such as serial numbers, measured values in real time etc.
6: Configuration-Register: · Measurement: Search connected sensors, choose channels and parameters · Analog Outputs: Configuration for analog outputs, low/high range etc · Manual Values: Set manual values for simulation and look control.
7: Action on new device. · Auto Read (default): Transmitter configuration is automatically transferred to the software and displayed. Prepared configuration in the software is overwritten. · Auto Write: The configuration prepared on the software is automatically written to the nlink+: Ideal to quickly configure multiple nlink+ with the same settings. · Show warning before execute (active by default): If active a warning message will appear before read or write. Deactivate if necessary.
Configuration procedure:
· Choose the adapter if the software does not automatically recognize
· Configure sensor (search sensor), measuring point and channels (6) · Configure analog output settings (6, register analog out) · After configuration activate «Write Data» (4) to store the settings on the transmitter.
Update required
If a firmware version on the nlink+ is not compatible with the installed nSoft-ACT-T this message «Update required» is shown.
Contact your local support for the newest firmware file and update instructions.

July 2023 / Doc.No. 007523.01

12 / 17

Operation Manual nlink+ AS IP & EC
New configuration Connect your nLink+ with the configuration cable to your PC. Connect the nSens as well.
Nlink+ is recognized. Register «i» displays additional information about the transmitter.
Click on «Search Sensor» Connected nSens probe will be recognized and displayed with the serial number.
The identified nSens must now be assigned to measuring points. If one nSens is connected it is automatically assigned to measuring point A.
« Measurement Channel» 1 and 2 correspond to analog out 1 and 2. Choose parameter and unit (depending on the type of nSens).
« Replacement Mode»: Automatic (default): If the nSens is replaced the nLink+ recognizes the different serial number and replace the removed nSens with the new nSens to the same measuring point. Configuration remains. Manual: A different nSens has to be configured with this software again.

July 2023 / Doc.No. 007523.01

13 / 17

Configuration analog outputs

Operation Manual nlink+ AS IP & EC
Register «Analog Outputs». Assign Signals (Loop mA oder V) and Range (Range low / Range high) according the specification from the automation system.
On Fail: In case an error is detected (e.g. nSens disconnected) this output is active. If power supply is interrupted signal output is zero.
Store configuration with Write Data:

Actual analog value on the output for each channel. Only active with additional power supply.
The configuration remains as long as no other nlink+ is connected and Read Data is activated. Choose "Auto Write" and connect another transmitter. The prepared settings will now be stored on the transmitter automatically.

« Manual Values»

Program fixed values for Simulation and loop checks. Choose Override mode and "Write Data" for activation. As long as manual values are active the sensor values are not displayed.
Beware: Deactivate override after completing the checks to receive real time data from the nSens again.

July 2023 / Doc.No. 007523.01

14 / 17

Operation Manual nlink+ AS IP & EC

6. Calibration / verification of measuring values
For humidity calibration and verification Novasina humidity standards are recommended. These humidity generators are easy to use and reusable multiple times. For the verification of the whole measurement range Novasina offers a set with 5 SAL-SC including carrying case and probe adapter rings. For an optimal calibration and/or verification the room temperature should be between 15...30°C and should not vary more than +/-0,2°C. In order to allow the standards to generate an accurate and stable air humidity the SAL-SC must be well sealed around the probe. For this purpose an adapter with O-ring can be used, which is put inside the SAL-SC hole and then put over the probe. The standards should be adapted to the climatic ambient conditions approx. 1 hour before they are used. If handled properly the SAL-SC generate very stable and accurate relative humidity and can be used as an alternative to humidity generators.

6.1. Sensor cables and Configuration set

nSens cable with end nSens cable extensions sleeves
Connects nSens with nlink+ Extension cable between any connector and nSens.

Configuration cable: nlink-USB&CA3
Configuration cable for nLink+ to Android or Windows PC. Software available for download

260 1080 nSens-cable 5m 260 1079 nSens-cable 10m 260 1078 nSens-cable 30m 260 1225 nSens-cable 60m 260 1226 nSens-cable 100m

260 1201 nSens Extension 2m 260 1136 nSens Extension 5m 260 1986 nSens Extension 10m 260 1987 nSens Extension 30m

260 1818 nlink-USB-CA3 (complete set)
Single items: 260 1755 CA-3 adapter 260 1075 nlink USB cable

July 2023 / Doc.No. 007523.01

15 / 17

Operation Manual nlink+ AS IP & EC

7. Accessories nSens HT

Art. No. 1110885 - SAL-SC 11 Art. No. 1110855 - SAL-SC 33 Art. No. 1110857 - SAL-SC 53 Art. No. 2600219 - SAL-SC 58 Art. No. 1110859 - SAL-SC 75 Art. No. 2518965 - SAL-SC 84 Art. No. 1110896 - SAL-SC 90 Art. No. 2518966 - SAL-SC 97

Sensor-Checks SAL-SC
(humidity standards) Reusable humidity standards based on saturated salt solutions in plastic cylinders with moisture permeable membranes. Each salt is delivered in a well-sealed box. Sensor Checks SC are obtainable for the following values (at 25°C):

11.3 % rh 32.8 % rh 52.9 % rh 57.6 % rh

75.3 % rh 84.3 % rh 90.1 % rh 97.3 % rh

Important: please consult the operation manual of your instrument to see which points can be calibrated. Other SAL-SC can be used for verification.

Humidity values in the temperature range 15°.... 30°C: 11.3 ..... 11.3% rh / 15....30°C 33.3 ..... 32.4% rh / 15....30°C 55.9 ..... 51.4% rh / 15....30°C 60.7 ..... 56.0% rh / 15....30°C 75.6 ..... 75.1% rh / 15....30°C 85.9 ..... 83.6% rh / 15....30°C 90.9 ..... 89.9% rh / 15....30°C 97.9 ..... 97.0% rh / 15....30°C

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

Weight :

90 g

Sensor-Checks SAL-SC with European certificate
Reusable humidity standards based on saturated salt solutions in plastic cylinders with moisture permeable membranes. Each salt is delivered in a well-sealed box. Sensor Checks SAL-SC are obtainable for the following values (at 25°C):

Internationally accredited laboratory

Art. No.1111044 - SAL-SC 11C Art. No.1111037 - SAL-SC 33C Art. No.1111040 - SAL-SC 53C Art. No.1111035 - SAL-SC 75C Art. No.2601272 - SAL-SC 84C Art. No.1111032 - SAL-SC 90C Art. No.2601275 - SAL-SC 97C

11.3 % rh 32.8 % rh 52.9 % rh 75.3 % rh 84.3 % rh 90.1 % rh 97.3 % rh

All Novasina humidity standards can also be supplied with an internationally recognised certificate from an accredited European laboratory (UKAS England).

Weight :

90 g

Art. No. 1117847 SAL-SC Sensor Check Set Art. No. 1117841 Empty case for SAL-SC

Set with 5 Humidity Generators SAL-SC
Reusable humidity standards SAL-SC in a case delivered incl. the needed adapters for Novasina probes and factory calibration certificates. Case set contains: - SAL-SC 11 - SAL-SC 33 - SAL-SC 53 - SAL-SC 75 - SAL-SC 90 - Factory calibration certificates of SAL-SC - 1 adapters for Novasina probes

Humidity values in the temperature range 15°... 30°C: 11.3 ..... 11.3% rh / 15....30°C 33.3 ..... 32.4% rh / 15....30°C 55.9 ..... 51.4% rh / 15....30°C 75.6 ..... 75.1% rh / 15....30°C 90.9 ..... 89.9% rh / 15....30°C

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

Weight :

900 g

July 2023 / Doc.No. 007523.01

16 / 17

Operation Manual nlink+ AS IP & EC

Art. No. 2601143 Adapter SAL-SC for nSens probes

Adapter SAL-SC for nSens probes
Plastic adapter for humidity standards. Used for diameter reduction and radial sealing around the nSens probe with diameter 13mm.
With integrated green coloured rubber sealing ring

Dimension : Out.diameter 30 mm Inner diam.13 mm
Weight : 5 g
Material : Thermoplastic resin

Thermal insulation styrofoam box
A styrofoam box providing optimal insulation and temperature stabilisation of a SAL-SC check during the calibration procedure. Consisting of two halfcovers for simple temporary mounting.

Dimension : 100 x 65 x 50 mm
Weight : 10 g
Material : thermal insulating styrofoam PPE

Art. No. 1111302 Styrofoam box
Technical information and other information subject to change without notice

July 2023 / Doc.No. 007523.01

17 / 17

