VOLTAGE/CURRENT STANDARDS 2422
Portable Calibrator

Yokogawa 2422 Portable Calibrator -

2422
257 X 182 X 40 mm 1.2 kg
(10-1/8 X 7-1/4 x 1-5/8″ 2.7 Ibs}
The 2422 is capable of generating DC V, DC mA and emf, and displays both generation and measurement values on its LCD display unit.

  • SIMULTANEOUS DISPLAYS OF OUTPUT SUPPLY AND INPUT MEASUREMENT
  • COMPACT AND LIGHTWEIGHT m AUTO-STEPPING OUTPUT & OUTPUT DIVIDING FUNC- TIONS
  • DIRECT THERMOMETER CALIBRATION
  • BUILT-IN Ni-Cd BATTERY/AC POWER OPERATION

AVAILABLE MODELS

ModelDescription
AC AdapterCorresponding TC emf
242211120 VANSI
242212220 to 240 VANSI
242221220 to 240 VDIN

SPECIFICATIONS

SettingTime: Approx. 1s
Divided Output: Output = set value X n/m, m (no, of divisions) $15, n=0,1,2..m
Set-Value Memory: 20 steps (10 steps X 2, excluding polarity signal) Auto Step: n/m and memory channel automatic feed, step time… 2,5, , 20s
Output Monitor: Load current monitor at 1, 10 and 30 V ranges (ac- aay? mA), load voltage monitor in 20 mA range (accuracy +0.2 V
Output Limiter: Activated under load of more than maximum output (manual return)
Output Setting: 4-digit UP/DOWN
Display: 4-1/2 digit LCD
RJC Sensor Measuring Range: —10 to +50°C, 14 to 122°F (accuracy in combination with 2422 … +0.5°C at 23£5°C, +1°C at —10 to 50°C)
Dielectric Strength: 500 V DC for one minute between input and output terminal
Power Source: AC adapter/internal Ni-Cd batteries
Optional Accessories: RJC sensor (B9638CR), terminal block (B9638BM)
Accessories Supplied: AC adapter, output and measuring cables (1 pc. each), carrying case

RangeGenerating and
Measuring ranges
Accuracy (23±5°C)*1ResolutionMax. OutputDescription
Generation Ranges30 V
10 V
1 V
100 mV
0 (0 =36.00 V Oto±12.000 V

0 to ±1200.0 mV 0 to ±120.00 mV

± (0.05% of rdg + 0.06% of range) ± (0.05% of rdg + 0.02% of range) ± (0.05% of rdg + 0.02% of range) ± (0.1% of rdg + 0.02% of range)10 mV 1 mV 100 AV 10 ILVApprox. 5 mA
Approx. 20 mA
Approx. 20 mA
Internal resistance
approx. 5 Ω
TCK-200.0 to +1370.0°C, -328.0 to 2502°F± (0.1% of rdg + 0.5°C), (0.1% of rdg +2°F)*20.1 °C, 1°FInternal resistance
approx. 5 Ω
E-200.0 to 700.0°C, -328.0 to 1292°F± (0.1% of rclg + 0.5°C), ±(0.1% of rdg + 1.0°F)*20.1°C, 0.2°F
J-200.0 to +1000.0°C, -328 to 1832°F
T-200.0 to +400.0°C, -328 to 752°F
R0 to +1600°C, 32 to 3216°F± (0.1% of rdg + 1.5°C), ± (0.1% of rdg +2.7°F)*21c C, 1e f
-40 to 0°C, -40 to 32°F± (0.1% of rdg +3°C), ± (0.1% of rdg +5.4°F)
20 mA0 to ±24.00 mA± (0.1% of rdg +0.1% of range)10 liAApprox. 12 V
Reference
resistance
100 0± 0.2%
Measuring Ranges30 V
10 V
1 V
± 36.00 V

± 11.999 V

± 1199.9 mV

± (0.1% of rdg + 2 digits)10 mV
1 mV
100µV
Input resistance
approx. 1 MΩ

*1  Voltage drop due to the load current through the attached lead is not included (lead resistance is about 0.08 9).
* 2 Accuracy of TC range does not include the error of RJC.